Technical Compliance Statement
EMC Test Report
Product Information
Product | Switching Power Supply |
---|---|
Model Names | (1)FSP250-60MPC, (2)FSP220-60MPC, (3)SPI250F4BMB, (4)SPI220F4BMB |
Brand Name | (1)~(2)FSP, (3)~(4)SPI |
Applicant | FSP Group Inc. |
Ref. File No. | C1M2305071 |
Original Safety Certificate Type | EMC/Medical Class (EMC/醫療類) |
Applicable Standards
Emission Standards
- EN 60601-1-2:2015 +A1:2021 (IEC 60601-1-2:2014 +AMD:2020)(Edition 4.1)
- BS EN 60601-1-2-2015 +A1:2021
- CISPR 11:2015 +A1:2016 +A2:2019, Class B
- EN 55011:2016+A11:2020
- EN IEC 61000-3-2:2019+A1:2021 (IEC 61000-3-2:2018 +A1:2020), BS EN IEC 61000-3-2:2019+A1:2021
- EN 61000-3-3:2013 +A1:2019 (IEC 61000-3-3:2013 +A1:2017), EN 61000-3-3:2013 +A1:2019 +A2:2021 (IEC 61000-3-3:2013 +A1:2019 +A2:2021)
- BS EN 61000-3-3:2013 +A1:2019, BS EN 61000-3-3:2013 +A1:2019 +A2:2021
Immunity Standards
- IEC 61000-4-2:2008
- IEC 61000-4-3:2020
- IEC 61000-4-4:2012
- IEC 61000-4-5:2014+A1:2017
- IEC 61000-4-6:2013
- IEC 61000-4-8:2009
- IEC 61000-4-11:2020
- IEC 61000-4-39:2017
Certification Statement
We hereby certify that the above product has been tested by us with the listed standards and found in compliance with the Regulation (EU) 2017/745 concerning medical device. The test data and results are included in the test report EM-E160240B.
Test Laboratory Information
Signature: Tony Lee/Deputy Manager
Date: 2023. 05. 24
Test Laboratory: Audix Technology Corporation, EMC Department
TAF Accreditation No.: 1724
Web Site: www.audixtech.com
The statement is based on a single evaluation of one sample of the above-mentioned products. It does not imply an assessment of the whole production and does not permit the use of the test lab logo.
FSP Company Information
Company: FSP (全漢企業股份有限公司)
Address: No. 22, Jianguo E. Rd., Taoyuan Dist., Taoyuan City 330, Taiwan(R.O.C)
Contact: TEL: +886-3-375-9888 FAX: +886-3-375-6966