15W High Reliable Green Medical Encapsulated Type

MPM-15 series
User's Manual

EN 60601-1 IEC 60601-1
8
BS EN/EN60335-1 ANSI/AAMI ES60601-1 BS EN/EN60601-1 IEC60601-1 TPTC004
 Features
2.06"x1.07" compact size Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1 Suitable for BF application with appropriate system consideration No load power consumption<0.1W Extremely low leakage current Wide operating temp. range -40 ~ +85 EMI class B for class  configuration Protections: Short circuit / Overload / Over voltage / Over temperature No minimum load required 3 years warranty

++
 Applications
Portable medical device Mobile clinical workstation Medical computer monitor Medical examination instrument
 GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx

 Description
MPM-15 is a 15W high density and small size (52.4*27.2*24mm) AC/DC module type medical power supply series offered in pin type. It features the operation for 80~264VAC, a low no load power consumption less than 0.1W, a high efficiency up to 87%, Class  (no FG) double insulation, outstanding dissipation and high lifespan thanks to the interior potting, 2G anti-vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1version three with 2xMOPP level and ultra-low leakage current (<80A). It is very suitable for BF (patient contact) type medical device or relevant equipment.

 Model Encoding MPM - 15 - 5
Output voltage Rated wattage Series name
Downloaded from Arrow.com.

File Name:MPM-15-SPEC 2024-04-15

M P M - 1 5 15W High Reliable Green Medical Encapsulated Type

series

SPECIFICATION

MODEL

MPM-15-3.3

MPM-15-5

MPM-15-12

MPM-15-15

MPM-15-24

DC VOLTAGE

3.3V

5V

12V

15V

24V

RATED CURRENT

3.5A

3A

1.25A

1A

0.63A

OUTPUT

CURRENT RANGE Note.2 0 ~ 3.5A

PEAK CURRENT

3.85A

RATED POWER

11.6W

PEAK LOAD(10sec.) Note.3 12.7W

RIPPLE & NOISE (max.) Note.4 150mVp-p

VOLTAGE TOLERANCE Note.5 ±2.0%

LINE REGULATION

±0.5%

0 ~ 3A 3.3A 15W 16.5W 150mVp-p ±2.0% ±0.5%

0 ~ 1.25A 1.38A 15W 16.6W 150mVp-p ±2.0% ±0.3%

0 ~ 1A 1.1A 15W 16.5W 180mVp-p ±2.0% ±0.3%

0 ~ 0.63A 0.69A 15.1W 16.6W 180mVp-p ±2.0% ±0.3%

LOAD REGULATION

±0.5%

±0.5%

±0.5%

±0.5%

±0.5%

SETUP, RISE TIME

1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at full load

HOLD UP TIME (Typ.)

40ms/230VAC 10ms/115VAC at full load

VOLTAGE RANGE Note.6 80 ~ 264VAC

113V~370VDC

FREQUENCY RANGE

47 ~ 440Hz

EFFICIENCY (Typ.)

83.5%

85.5%

86.5%

87%

INPUT

AC CURRENT (Typ.)

0.6A/115VAC 0.3A/230VAC

INRUSH CURRENT (Typ.)

COLD START 20A/115VAC

45A/230VAC

LEAKAGE CURRENT (max.) Note.7 Touch current <80A/264VAC

86.5%

OVERLOAD

110% ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed

PROTECTION OVER VOLTAGE

3.8 ~ 5V

5.8~ 6.8V

13.8 ~ 16.2V

Protection type : Shut off o/p voltage, clamping by zener diode

17.3 ~ 20.3V

27.6 ~ 32.4V

ENVIRONMENT

OVER TEMPERATURE WORKING TEMP. WORKING HUMIDITY STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT SOLDERING TEMPERATURE

Protection type : Shut down o/p voltage, recovers automatically after temperature goes down -40 ~ +85 (Refer to "Derating Curve") 20 ~ 90% RH non-condensing -40 ~ +85, 10 ~ 95% RH non-condensing ±0.03%/ (0 ~ 60) Wave soldering: 265,5s (max.); Manual soldering: 390,3s (max.)

VIBRATION

10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

OPERATING ALTITUDE Note.8 5000 meters

SAFETY STANDARDS

IEC 60601-1:2005+A1,TUV BS EN/ EN 60601-1:2006+A1+A12+A2,ANSI/AAMI ES60601-1:2005/A2:2021, CAN/CSA C22.2 No. 60601-1:2014+A2,IEC 60335-1:2010/AMD2:2016, Dekra BS EN/ EN 60335-1:2012+A11+A13+A1+A14+A2+A15 ,EAC TP TC 004 approved

ISOLATION LEVEL

Primary-Secondary: 2xMOPP

WITHSTAND VOLTAGE ISOLATION RESISTANCE

I/P-O/P:4KVAC

I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH

Parameter

Standard

Test Level / Note

Conducted emission

BS EN/EN55011 (CISPR11)

Class B

EMC EMISSION
SAFETY & EMC (Note.9)

Radiated emission Harmonic current Voltage flicker BS EN/EN55035, BS EN/EN60601-1-2 Parameter ESD

BS EN/EN55011 (CISPR11) BS EN/EN61000-3-2 BS EN/EN61000-3-3
Standard BS EN/EN61000-4-2

Class B Class A -----
Test Level / Note Level 4, 15KV air ; Level 4, 8KV contact

EMC IMMUNITY

RF field susceptibility
EFT bursts Surge susceptibility Conducted susceptibility Magnetic field immunity

BS EN/EN61000-4-3
BS EN/EN61000-4-4 BS EN/EN61000-4-5 BS EN/EN61000-4-6 BS EN/EN61000-4-8

Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz ) Level 3, 2KV Level 3, 1KV/Line-Line Level 3, 10V Level 4, 30A/m

Voltage dip, interruption

BS EN/EN61000-4-11

100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods

MTBF OTHERS DIMENSION

7319.8K hrs min. Telcordia SR-332 (Bellcore) ; 1210.0K hrs min. MIL-HDBK-217F (25) 52.4*27.2*24mm (L*W*H) or 2.06*1.07*0.94 inch

NOTE

PACKING

0.056Kg; 240pcs/14.4Kg/0.94CUFT

1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.

2. No minimum load required.

3. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.

4. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1F & 47F parallel capacitor.

5. Tolerance : includes set up tolerance, line regulation and load regulation.

6. Derating may be needed under low input voltages. Please check the derating curve for more details.

7. Touch current was measured from primary input to DC output.

8. The ambient temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan models for operating altitude higher than 2000m(6500ft).

9. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still

meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.

(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

 Product Liability DisclaimerFor detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

File Name:MPM-15-SPEC 2024-04-15

Downloaded from Arrow.com.

M P M - 1 5 15W High Reliable Green Medical Encapsulated Type

series

Block Diagram

I/P

EMI FILTER

RECTIFIERS &

POWER SWITCHING

FILTER

RECTIFIERS &
FILTER

fosc : 100KHz
+V -V

O.L.P.

PWM CONTROL

DETECTION CIRCUIT

Derating Curve
100 80 70 50 30

3.3V,5V

12V~24V

LOAD (%)

-40 -30 -20 -10 0 10 20 30 40 50 55 60 70 80 85
AMBIENT TEMPERATURE ()

(HORIZONTAL)

Output Derating VS Input Voltage
100 90 80 70 60 50 40
80 95 100 115 120 140 160 180 200 220 240 264
INPUT VOLTAGE (VAC) 60Hz

LOAD (%)

Downloaded from Arrow.com.

File Name:MPM-15-SPEC 2024-04-15

M P M - 1 5 15W High Reliable Green Medical Encapsulated Type

series

Mechanical Specification

Case No.219A Unit:(mm)

3.2

52.4

3.4

45

AC/N

-V

+V

AC/L BOTTOM VIEW

3.2 8

1.04

24 3.5±1mm
P/N diameter:1.04
SIDE VIEW

20.8

27.2

Installation Manual
Please refer to : http://www.meanwell.com/manual.html

Downloaded from Arrow.com.

File Name:MPM-15-SPEC 2024-04-15



References

Acrobat Distiller 9.0.0 (Windows); modified using iText 5.5.6 ©2000-2015 iText Group NV (AGPL-version)