RD0370C-Manual
Scigiene Reference Thermometer Part# RD0370C
Instruction Manual
1

Table of Contents
Features.................................................................................................................................................................3 Specifications........................................................................................................................................................3 Instrument Description....................................................................................................................................4-5 Operating Instructions......................................................................................................................................6-7 Temperature Measurement................................................................................................................................8 Battery Replacement............................................................................................................................................8 Product Maintenance...........................................................................................................................................9
2

Features
Scigiene RD0370C is a high accuracy certified RTD thermometer accurate to better than +-0.4C. It includes a full 3-point N.I.S.T traceable certificate at 3 points over -20C to 140C.
· Includes 3 point N.I.S.T traceable certificate · IP67 waterproof housing · Platinum resistor temperature sensor requires less temperature compensation than
thermocouple units · 4-digit LCD readout with LED backlight · °C/°F switchable · Min/Max/Avg functions · Relative mode and Data Hold · Low battery indication with auto power-OFF · Includes a pointed 4.5" Stainless Steel RTD probe
Specifications
· Temperature Range: -100°C to +300°C (-148°F to +572°F) · Resolution: 0.1°F/C · Accuracy: +-0.4C over -20 to 100C · Sensor Types: Platinum resistance for Pt-100, Pt-500, Pt-1000 (4 wires) selectable · Power Supply: 3 x 1.5V AAA alkaline batteries · Battery Life: Approximately 100 hours
3

Instrument Description
1 -- Pt type temperature sensor connector 2 -- LCD display 3 -- Power ON/OFF button 4 -- HOLD button 5 -- MAX/MIN/Average control button 6 -- Back light button 7 -- Relative readout button
4

Display Symbol definitions and button locations

C t xxx HD MA
R

This

ha h inus

is s ens ed

C l ius and

la inu T p

This

ha h displa da a is b in

Th axi u alu is b in

Th ini u alu is b in

Th

a alu is b in

Th

is und la i

Th

po l l

This

u o o Off is

5

Operating Instructions
Turning the meter on Press the button to turn the thermometer ON or OFF.
Connecting the temperature probe For measurement, plug the temperature probe into the input connectors.
Selecting the Temperature Scale When the meter is first turned on, the default setting is set to the Celsius (°C) scale. It can be changed to Fahrenheit (°F) by pressing "°C/°F"button and vice versa back to Celsius.
Data-Hold Operation To hold the present reading and keep it on the display, press the "HOLD" button. When the hold data is no longer needed, release the data-hold operation by pressing "HOLD" button again.
When the meter is under Data Hold operation, the " REL", AVG MAX MIN and "°C/°F" buttons are disabled.
Back Light Operation Press the "Back Light" button to turn the back light on, press it once again to turn it off. The meter will turn the back light off automatically if the "Back Light" button is not pressed after 10 seconds.
Relative Operation When the " REL" button is pressed, the meter will memorize the present reading and the difference between the new reading and the memorized reading and the result will be shown on the display. Press the " REL" button again to exit the relative operation.
MAX/MIN/AVG Operation When the AVG MAX MIN button is pressed, the meter will enter the MAX/MIN/AVG mode. Under this mode the maximum value, minimum value and average value will appear simultaneously. The average value is calculated using the last 8 readings from the memory, it will update with every new reading.
When the MAX symbol is displayed, the Maximum value is shown on the display.
Press AVG MAX MIN again and the NIN symbol is on the display and also the minimum reading.
Press AVG MAX MIN again and the AVG symbol is on the display as well as the average reading.
6

Press AVG MAX MIN again and the MAX, MIN and AVG will blink together. This means that all these readings are updated in the memory and the reading is the current temperature.
Press AVG MAX MIN to circulate the display mode among these options.
When the meter is under AVG MAX MIN operation, the " REL" and "°C/°F" functions are disabled.
To exit the MAX/MIN mode, press and hold AVG MAX MIN for two seconds.
Auto Power Off By default, the meter is in auto power off mode. The meter will power itself off after 30 minutes without operation.
To disable the auto power off, press and hold the "HOLD" button and turn the meter on. There will be two successive beeps to indicate that the auto power off has been disabled.
Low Battery This meter indicates the battery power level directly on the display. Use this reference guide to establish battery power level.

a po take
a po o b

l l is F LL you an s ill
l is LO h ou an s ill

a po o
an no lon

l is E (siz
a

TY h . V x ).

For complete instructions on how to replace the batteries see page 8.

7

Temperature Measurement
Correct Measurement Method The temperature sensor is located at the end of the metal sheath of the sheath type temperature probe. To accurately test internal temperatures, insert the probe to a distance of at least 15 times the diameter of the sheath directly into the item you want to measure.
Connector Configuration
.

N

Battery Replacement
1) Remove all of the back cover screws to be able to remove the cover. 2) Verify the polarity of the batteries and install the new LR03 (AAA size) alkaline batteries
into the instrument. 3) Replace cover properly onto the meter and tighten screws.

Th p op l
h

ba o is fi d i h

in s. f

h

ha h

in s e

b fo ins allin h ba o .

of h

in s

h uni s

and

sul in

8

Product Maintenance

In order to ensure the accuracy of the thermometer for a long period of time it should be calibrated once a year. For service (repairs or calibration) on this or any other products, contact Scigiene at 416-261-4865 or visit our website at: www.scigiene.com

Clean the device and the window of the display with a clean, lint-free, antistatic and dry-cleaning cloth.

.

Do no us l anin

ha on ain a bon o

al ohol o an hin si ila o l an h

as h s

h su fa of h

Do no us ools i h sha p

an hin si ila o l an h

Temperature Probe
Piercing type temperature probe

Piercing type temperature probe Specification

Sensor Type
Accuracy Measurement Range Temp. Sensor Dim. Temp. Sensor Length Cable Length Water-resistant

Platinum resistance thermometer sensor Pt 100 (4 wires)
IEC751, class A
+/- 0.15°C +/- 0.002 measurement temperature
-100 to 400°C Approx. 3.2mm (0.125") Approx. 120mm (4.72") Approx. 1100mm (43.3") EN60529:1991, IP67

9

