User Manual for Elimko models including: E-KC-102 Series Head Mounted Type Temperature Converter, E-KC-102 Series, Head Mounted Type Temperature Converter, Mounted Type Temperature Converter, Type Temperature Converter, Temperature Converter, Converter

KY KC102 0524 1

murat.aras

Elimko

23 ago 2024 — It s not used n med cal appl cat ons. KY-KC102-0524-1. E-KC-102 SERIES HEAD MOUNTED TYPE. TEMPERATURE CONVERTER USER MANUAL ...


File Info : application/pdf, 2 Pages, 240.37KB

PDF preview unavailable. Download the PDF instead.

KY KC102 0524 1
E-KC-102 SERIES HEAD MOUNTED TYPE TEMPERATURE CONVERTER USER MANUAL

Elimko

KY-KC102-0524-1
The E-KC-102 converter is designed for use in an industrial environment.
m The package of the E-KC-102 converter contans; the Converter, user manual and guarantee certfcate.
m After openng the package, please vsually check whether the type of the transmtter s sutable for the order, whether the above-mentoned parts are mssng and whether the transmtter has been damaged durng shpment.
m Before nstallng and operatng the controller, please read the user manual thoroughly. m The nstallaton and confguraton of the controller must only be performed by a person qualfed n nstrumentaton. m Keep the unt away from flamable gases, that could cause explotons.
m Do not use alcohol or other solvents to clean the transmtter. Use a clean cloth soaked n water tghtly squeezed to gently wpe the outer surface of the transmtter.
m It s not used n medcal applcatons.

1. DESCRIPTION
E-KC-102 series two wire transmitters are microprocessor based industrial instruments in plastic case. E-KC-102 transmitters are used to convert the thermocouple and resistance thermometer signals into standard 4-20 mA The configuration of the transmitter is done by PC, using a software programme supplied by Elimko. The RS 232 connection hardware is also available from Elimko.

2. TECHNICAL SPECIFICATIONS
2.1. OPERATING RANGE and MEASURING ACCURACY (at 24 V supply voltage and 25°C ± 3°C ambient temperature)

SENSOR

RTD Pt-100

B

E

J

K

T/C

N R

S

T

L

U

mV

STANDARD

LOWER LIMIT

UPPER MINIMUM LIMIT SPAN

ACCURACY

A/D

D/A

IEC 60751 -200°C 850°C 50°C ±0.25°C

100°C 1800°C 100°C ±2.00°C

-200°C 840°C 50°C ±0.50°C

-200°C 1120°C 50°C ±0.50°C

IEC 60584

-200°C 1360°C -200°C 1300°C -40°C 1760°C

50°C 50°C 100°C

±0.50°C ±0.50°C ±1.00°C

±0.1% Span

-40°C 1760°C 100°C ±1.00°C

-200°C 400°C 50°C ±0.50°C

DIN 43710 -200°C 900°C 50°C ±0.50°C -200°C 600°C 50°C ±0.50°C

-2000mV 500mV 25mV ±0.075mV

If the input type and scale is not specified while ordering, factory settings are; For KC-102-RT Input Type : Pt-100, and Scale : 0-200°C, For KC-102-TC Input Type : Type K, and Scale : 0-400°C.

KY-KC100-0224-1

2.2. General Specifications

Electrical: Supply Voltage Voltage Drop

10.0 - 30 V DC 10.0 V

Environmental Conditions: Operating Temperature
Protection Class

-10°C to +55°C
IP 00, IP 66 (DIN 43729 mounted on type B head)

Calbraton Temp. Mechancal: Dmensons Weght (approx.) Connecton Cables

25°C ±3°C
Ø 44.0 mm x 21.5 mm 40 g Max. 1.5 mm² (AWG 16)

Resstance Thermometer (RTD) / Resstance Input:

Sensor Connecton

3-wire or 4-wire (Configurable)

Maximum Wire Resistance 100 W

Error Signaling

Wire Break (Output current can be configured Up / Down.)

Thermocouple (TC) / mV Input:

Input Impedance

> 10 MW

Maximum Wire Resistance 100 W

Cold Junction Compensation (CJC)

Constant, Internal NTC, External Pt-100 (Configurable)

Error Sgnalng

Wre Break (Output current can be confgured Up / Down.)

Output: Output Signal Load Resistance
Operating Influences: Ambient Temperature CJC Error (For TC Inputs) EMC Immunity

4 - 20 mA or 20 - 4 mA [Vsupply-10]× 50 W (600 W at 22 V DC)
< ± 0.01% / °C < ± 1.0 °C < ± 0.5% Span

Electromagnetic Compatibility: The E-KC-102 meets the requirements of TS EN 61326-1.

E-KC-102 SERIES HEAD MOUNTED TYPE TEMPERATURE CONVERTER USER MANUAL

3. CONNECTION DIAGRAM

Measurement Unit

DC Power Supply

Measurement Unt

RL

RL

Elimko
KY-KC102-0524-1
DC Power Supply

Elimko Elimko

PC Connection Connector

102-RT 4W

PC Connecton Connector
102-TC

4 Wre RT
E-KC-102-RT Connecton Dagram

TC
E-KC-102-TC Connecton Dagram

m Before operating the converter, ensure that the converter is correctly configured.
Incorrect configuration could result in damage to the process being controlled.
m To minimize the pick-up of electrical noise, the wiring of low voltage lines, particularly the sensor
input should be routed away from the high-current power cables. Where it is not possible, use shielded cable and ground the shielded cable.
m The cables used for powering the transmitter and the power outputs must conform to the standarts
IEC 60245 and IEC 60227.

4. DIMENSIONS

Ld

Ld

1

2

ø = 7.2 mm 3

6

4

5

ø = 6.0 mm 33.0 mm
ø = 44.0 mm

KY-KC102-0524-1

21.5 mm

For thermocouple inputs, be sure to use the proper compensation cables and pay attention to the polarity of the connection.
Manufacturer / Technical Support : Elimko Elektronik malat ve Kontrol Ltd. ti. ASO 2. Organize Sanayi Bölgesi Alci OSB Mahallesi 2001. Cad. No:14 Temelli 06909 Ankara / TURKYE Tel: +90 312 212 64 50 (Pbx) · Fax: +90 312 212 41 43 E-mail: elimko@elimko.com.tr · www.elimko.com.tr

