Salt20 u
SALT20 POCKET SALINITY TESTER
USER MANUAL

sales@labcoscientific.com.au labcoscientific.com.au

1800 052 226 PO Box 5816, Brendale, QLD 4500

ABN 57 622 896 593 Version 1.1

CONTENTS BATTERY INSTALLATION
KEYPAD FUNCTIONS COMPLETE KIT CALIBRATION
SALINITY MEASUREMENT NOTES
PARAMETER SETTING TECHNICAL SPECIFICATIONS OTHER FUNCTIONS & PARAMETERS
WARRANTY

PAGE NUMBER 2 2 3 3 3 4 4 5 5 6

1

1. BATTERY INSTALLATION
Please install batteries according to the following steps. Please note polarity: "+" (anode) is upward; "-" (cathode) is downward.

Note: This salinity meter measures practical salinity converted from electrical conductivity (NaCl Concentration). It does not measure absolute salinity.
2. KEYPAD FUNCTIONS

Short press < 2 seconds

Long press > 2 seconds

1. Short press to turn on, long press to turn off; 2. When turned off, long press to enter setup; 3. In mode setting, short press to change parameter;

1. When turned on, long press to enter calibration mode. 2. In calibration mode, short press to confirm calibration; 3. In mode setting, short press to confirm parameter selection.

Diagram - 1

2

3. COMPLETE KIT

Diagram ­ 2

4. CALIBRATION

4.1 Rinse the probe in distilled water and dry it. Short press

to turn on the tester.

4.2 Long press

to enter calibration mode; Short press

to exit.

4.3 Dip the probe into the calibration solution. Stir gently, leave it to stand, LCD displays , short press

to complete

1-point calibration. The tester returns to measuring mode, and calibration icon

appears at the button left of LCD.

5. SALINITY MEASUREMENT

5.1 Short press

to turn on tester. Rinse probe in distilled water and dry it.

5.2 Stir probe in the sample solution gently, leave it to stand. Get readings after the smile icon comes up and stays.

3

6. NOTES
6.1 The tester has self-diagnosis functions:

6.2 The tester has already been calibrated after manufacture. Usually users can use the tester right out of the box or test it in the calibration solution to compare the reading with the standard value. When the error is large, calibrate it before using

7. PARAMETER SETTING
7.1 Parameter setting reference chart:

SYMBOL

PARAMETER SETTING CONTENT

P1

Select salinity unit

P2

Restore to factory default

CODE Ppt ­ mg/L
No - Yes

FACTORY DEFAULT
ppt No

7.2 How to Setup parameters:

When turned off, long press

to enter setup  short press

to switch P1-P2 Short press

, parameter

flickering  short press

to choose, short press

to confirm parameter selection  Long press

to switch off.

7.3 Parameter setting notes: a) Select salinity unit(P1): Optional units: ppt ­ mg/L, factory default is ppt.

b) Restore to factory default: Select Yes to restore the calibration to the theoretical values and parameter setting to original values. When meter's calibration or measurement performs abnormally, this function can be adopted so the meter goes back to factory default setting and then users can conduct calibration or take measurements again.

4

8. TECHNICAL SPECIFICATIONS

SALINITY TEMP.

RANGE RESOLUTION ACCURACY CALIBRATION POINTS AUTOMATIC TEMP. COMPENSATION
RANGE RESOLUTION ACCURACY

0 ­ 10.00 ppt 0.1 ppt ± 1%F.S
1 point auto. Calibration 0 ­ 50°C 0 ­ 50°C 0.1°C ±0.5°C

9. OTHER FUNCTIONS & PARAMETERS

CALIBRATED POINTS
INDICATION OF STABLE MEASUREMENTS SELF-DIAGNOSIS INFORMATION LOW POWER REMINDER
DIMENSION/WEIGHT

AUTO POWER-OFF WATERPROOF LEVEL

Power-off in 8 minutes if no operation
IP67, floats on water

Er1, Er2

POWER SUPPLY

AAA Batteries x4

Flashes to remind to replace batteries

BATTERY LIFE

1,000 Hours

Instrument: 40x31x178mm/107g; Carrying Case: 190x165x140mm/438g

5

10. WARRANTY
We warrant this instrument to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, and unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.