1 Micro USB cable. 3. Safety Precaution. Caution! Please refer to this manual. Improper use may damage the meter and its components.

Download

เครื่องวัด CO2/ฝุ่น PM2.5 PM10/อุณหภูมิ/ความชื้น TENMARS TM-380

10 feb 2023 — Contents: 1 Application .................................................................. 1. 2 Accessories.

PDF preview unavailable. Download the PDF instead.

TM-380
Multi IAQ Monitor
TM-380
User's Manual
HB2TM3800000

Contents:
1 Application .................................................................. 1 2 Accessories................................................................. 1 3 Safety Precaution........................................................ 1 4 Instrument Description ............................................... 2
4.1 Feature and function...................................... 2 4.2 Indication on the LCD display ....................... 3 5 Operation..................................................................... 4 5.1 Power on/off ................................................... 5 5.2 Auto Power Off............................................... 5 5.3 Backlight mode .............................................. 5

5.4

MAX/MIN,AVG and MAN MIN ................. 5

5.5

REC Manual record ................................ 5

5.6

LOG starts the auto-recording function 6

5.7

Read the data recorded in the memory. 6

5.8 °C or °F units .................................................. 6

5.9 Clear memory................................................. 6

5.10

Turn on / off the key sound .................... 7

5.11 Resume the factory settings.......................... 7

5.12 Function settings ........................................... 7

6 General Specifications.............................................. 12

7 General Specifications.............................................. 14

7.1 PM2.5 ............................................................ 14

7.2 PM10 ............................................................. 14

7.3 CO2................................................................ 14

7.4 Temperature ................................................. 15

7.5 Humidity ....................................................... 15

8 Maintenance or Repair .............................................. 16

9 Product disposal ....................................................... 16

TM-380
1 Application
This air quality meter can measure and display CO2, PM2.5, Humidity and Temperature measurements and is ideal for monitoring indoor air quality in a variety of environments.
2 Accessories
1 Meter (with a 3.7V Rechargeable lithium battery)
1 User's Manual 1 Micro USB cable
3 Safety Precaution
Caution! Please refer to this manual. Improper use may damage the meter and its components.
Complies with European Directive.
 Do not operate in environments with flammable gas or humid environments.
 Operating altitude: up to 2000M.  Operating environment: Indoor use; Pollution degree 2.  Clean with soft cloth when dirty, such as glasses cloth. Do not
clean with chemicals and other solvents.  EMC: EN61326-1:CISPR 11:Group 1, Class B  Class B ­ Equipment for use in all establishments other than
domestic.  Group 1 ­ RF energy generated is needed for internal
functioning.
1

4 Instrument Description
4.1 Feature and function

TM-380

1. LCD 2. Right-side button 3. Memory reading / Up button 4. MAX / MIN / AVG button 5. Power button 6. Single log and auto-record button/Down
button 7. Temperature and humidity sensing port 8. Air outlet for sensor 9. Air inlet for sensor 10. Micro USB jack
2

4.2 Indication on the LCD display

TM-380

1.MAX / MIN /AVG /MAN MIN
2.Manual recording 3.Auto-recording 4.Read record 5.Battery indicator 6.Auto-power off 7.Buzzer 8.Alarm

9.CO2 10. PM2.5 11. Temperature 12. MM minute,
SS second 13. PM10 14. Humidity

3

TM-380

5 Operation

Please hold the meter in the following way for the correct

measured value:

The placement follows the This is a suitable way to hold

electrical specifications

the meter, holding for a long

described in Section 7.

time may affect the accuracy.

This is not a proper way to hold the meter, it will block the temperature and humidity sensors.

This is not a proper way to hold the meter, it will block the PM2.5, PM10 and CO2 sensors.

4

TM-380

5.1 Power on/off

Press

to turn the power on or off.

5.2 Auto Power Off

At the power-on status, press and hold

for more than 2

seconds to disable or resume auto power off function. When

auto power off is activated, the indicator appearing, and it will automatically turn off after 30 minutes.

5.3 Backlight mode
Press "right-side button" to turn on or off the backlight.
(When the red backlight for alarm is on, this button is invalid)
The backlight mode is turned off automatically after 15 seconds.

5.4

MAX/MIN AVG and MAN MIN

Press

to enter this function, the MAX is shown on the

screen and it displays the maximum value.

Press

repeatedly to select the function of

MIN/AVG/MAX MIN/MAX and update the displayed value.

When select MAX MIN function, it displays the current

measured value.

5.5

REC Manual record

Press

to save one data log, the "REC" and the log

number appear on the LCD for one second. For an example

of the log number 10, it will increase by 1 when pressing

every time. When reaching the maximum log number

20,

appears.

5

TM-380

5.6

LOG starts the auto-recording function

Press and hold

for more than 2 seconds, " "

appears on the LCD, and auto-recording starts. If the power

is insufficient during auto-recording,

appearing,

auto-recording will be forcibly stopped. When reaching the

maximum log number 20,

appears.

Press and hold

again for more than 2 seconds to exit

the auto-recording function.

5.7

Read the data recorded in the memory

Press

to enter the memory-reading mode, while

and the log number appear on the LCD for one second, the

record is then displayed. Read the data recorded in the

memory with

or .

Press and hold

to exit this mode.

5.8 °C or °F units
At the shutdown status, press and hold , then press to turn on the meter, after the software version appears,
release . Repeat the above steps to switch the unit.

5.9 Clear memory

At the shutdown status, press and hold

and then press

to turn on the meter, the

and software version

appears successively on the LCD, release the the memory.

to clear

6

TM-380

5.10 Turn on / off the key sound

The sound is defaulted as on and the is displayed.

At the shutdown status, press and hold

and then press

to turn on the meter, after the software version appears,

release

and the

disappears, the key sound is off.

Repeat the above step to tun on the key sound.

5.11 Resume the factory settings

At the shutdown status, press and hold "right-side button"

and then press , the

and software version

appears successively on the LCD, release the "right-side

button" to complete the process of resuming to the factory

settings and clearing the memory.

 Auto-shutdown: OFF

 Key sound: On

 The calibration value resumed to 0

 CO2 Abc OFF

 Memory cleared

5.12 Function settings
Press "right-side button" for more than 2 seconds to enter the setting mode. Press the "right-side button" repeatedly to enter the setting function 1 ~ 8 in sequence:

7

SET 1. Time interval setting 1. The LCD displays:

TM-380

2. Press

or

to select the interval 5

seconds / 10 seconds / 5 minutes / 10 minutes /

20 minutes / 30 minutes / 60 minutes; the default

value is 5 seconds.

3. Press the "right-side button" again to enter set 2.

SET 2. CO2 alarm setting 1. The LCD displays:

2. Press

to move the digit, and the selected

digit will flash.

3. Press

or

to increase or decrease the

CO2 value.

4. Press the "right-side button" again to enter set 3.

8

SET 3. PM2.5 alarm setting 1. The LCD displays:

TM-380

2. Press

to move the digit, and the selected

digit will flash.

3. Press

or

to increase or decrease the

CO2 value.

4. Press the "right-side button" again to enter set 4.

SET 4. PM10 alarm setting 1. The LCD displays:

2. Press

to move the digit, and the selected

digit will flash.

3. Press

or

to increase or decrease the

CO2 value.

4. Press the "right-side button" again to enter set 5.

9

TM-380
SET 5. (Factory use only) CO2 Abc self-calibration mode 1. The LCD displays:

2. This is for factory calibration use only. Please do not change it.
3. Click the "right-side button" again to enter set 6.
SET 6. CO2 manual calibration 1. The LCD displays:

2. Directly increase or decrease the displayed value

of CO2, press the

to move the digit, and the

selected digit will flash.

3. Press

or

to modify the value.

4. Press the "right-side button" again to enter set 7.

10

SET 7. Temperature calibration 1. The LCD displays:

TM-380

2. Directly increase or decrease the temperature value, press the to move the digit, and the selected digit will flash.
3. Press or to modify the value. 4. Press the "right-side button" again to enter set 8.
SET 8. Humidity calibration 1. The LCD displays:

2. Directly increase or decrease the temperature

value, press the

to move the digit, and the

selected digit will flash.

3. Press or to modify the value. 4. Press the right-side button again to return to the
measurement mode.

11

TM-380

6 General Specifications
 Real-time measurement of CO2/PM2.5/PM10/temperature/humidity.
 Maximum / minimum / average value.  Auto-power on / off function.  Adjustment and calibration function
(CO2/temperature/humidity).  CO2/PM2.5/PM10 red backlight alarm.  Manual record and auto-record.  Key sound assists to ensure the operation works.

 Overload indication .

 Battery: 3.7V 2000mAh rechargeable lithium battery

 Transformer for charging: 1A/5V

 Battery life: continuous use for approximately 6 hours

 5V micro USB interface for external power supply

 When battery-low indicator

shows as the following

figure, the power will be turned off after 5 seconds.

 Charging indication: The battery indicator will

sequentially vary as   from right to left

during charging and keep

when fully charged.

 °C or °F unit selection.

 When the external power supply is connected, the

auto-power off is disabled, and the white backlight is

always on.

12

TM-380

 CO2 /PM2.5/PM10 red backlight alarm setting function

Setting

CO2

PM2.5

PM10

value 1000PPM

35µg/m³

75µg/m³

When the measured value reaches the setting value for

alarm, the LCD backlight shines in red and related

measurement items flash.

 20 logs for manual and automatic recording values.

 Auto-recording time interval setting: 5 seconds, 10

seconds, 5 minutes, 10 minutes, 20 minutes,

30 minutes, 60 minutes.

 Sampling time: 2.5 seconds (5 types, 0.5 seconds for

each).

 Dimensions: 150x65x22mm (L x W x H).

 Weight: 190 g.

 Operating temperature and humidity: 0°C ~ +50°C,

<80%RH (no condensation).

 Storage temperature and humidity: 0°C ~ +60°C,

<70%RH (no condensation).

 Display size: 52mm (L) x 40mm (W).

Caution If to use the meter in a high-humidity environment, please place the meter in a low-humidity environment for 24 hours after use.

13

TM-380

7 General Specifications
Ambient temperature range for accurate measure: 18°C (64°F) ~ 28°C (82°F) 7.1 PM2.5

Sensor Measuring range
Accuracy
Resolution Reaction time

Laser particulate matter sensor
0-500 g/m3
±8g :<=50µg/m3 ±10% of reading value:>50µg/m3
1
<10seconds

7.2 PM10 Sensor Measuring range
Accuracy
Resolution Reaction time
7.3 CO2 Sensor Measuring range
Accuracy
Resolution Reaction time

Laser particulate matter sensor 0-500 g/m3
±10g :<=50µg/m3 ±15% of reading value:>50µg/m3
1 <10 seconds
Non-Dispersion Infrared(NDIR) 400 to 9999ppm
±70ppm ±3% of readings. (400-2000ppm) ±1ppm
90% during about 2 minutes
14

TM-380

7.4 Temperature

Sensor Effective range
Accuracy
Resolution
Reaction time

Bandgap temperature sensor -10°C~50°C / 14°F~122°F
±1.0°C/±1.8°F (Non-charging and 30 minutes after
charging) 0.1°C / 0.1F (0~35°C / 32°F~95°F) ;
1°C / 1°F (Above 100°C /°F or below -10°C /°F)
About 1 second

7.5 Humidity Sensor Effective range
Accuracy
Resolution Reaction time

Capacitive humidity sensor 1-99%
±5.0%RH(10~90%); ±8.0%RH(<10%,>90%) no
condensation, at 25°C 0.1%
About 4 seconds

15

TM-380

8 Maintenance or Repair

1.For the battery-low indicator:

indicates the battery

full;

indicates the battery low, please charge it in

time.

2.If dirty, please wipe it with a soft cloth, such as a glasses cloth. Do not wipe it with chemical solvents.
3.If fault on the meter, it can only be sent to the authorized service suppliers or sent back to the original factory for maintenance.

Caution When the LCD shows the battery-low indicator , please charge it in time.

9 Product disposal
Caution: This symbol indicates that this product and its accessories shall be subject to a separate collection and correct disposal, please not throw it away directly to avoid environmental pollution.

16

Professional Electrical and Environment Test & Measurement Instruments:
LED light meter, Temperature & Humidity meter,Infrared Thermometer, Sound level meter, Light meter, EMF meter, UV Light meter, RF meter, Hot wire Anemometer, Co meter,Anemometer, Lan cable tester, Co2 meter, Solar power meter, Radiation meter, Clamp meter, Multimeter, Phase Rotation
test, Digital Insulation tester
Our products of high quality are selling well all over the world
TENMARS ELECTRONICS CO., LTD. 6F, NO.586 Ruiguang Rd, Neihu Dist.
Taipei City, Taiwan E-mail: service@tenmars.com
http://www.tenmars.com

