AQ3STF Sensor - Sulfur Dioxide (SO2) Gas Sensor - Characterization Note

AQ3 Series - AQ3STF - Characterisation Note

Sulfur Dioxide (SO2) Gas Sensor - Characterization Note

DOCUMENT PURPOSE. The purpose of this document is to provide indicative, technical performance data for the AQ3STF gas sensor to assist in the.


File Info : application/pdf, 11 Pages, 635.21KB

PDF preview unavailable. Download the PDF instead.

sps-ast-aq3stf-char-note-002728-en ?download=false
CHARACTERIZATION NOTE
AQ3STF Sulfur Dioxide Gas Sensor

CONTENTS
Document Purpose The Gas Response Curve Linearity Temperature Characteristics
Variation of Sensitivity with Temperature Variation of SO2 Reading with Temperature after Compensation Variation of Baseline Offset with Temperature after Compensation Long-term Sensitivity Drift Repeatability Calibration Guidelines Recommended Gas Flow Rates Recommended Circuit Compensation Logics of AQ3 Series Sensors Cross Sensitivity Table

DOCUMENT PURPOSE
The purpose of this document is to provide indicative, technical performance data for the AQ3STF gas sensor to assist in the integration of the sensor into gas detection instrumentation. The sensor has been subjected to a testing program as part of the development process. Within this document, detailed information on the results of this program is presented. All data has been taken from equipment using a +5 Vdc power supply.
This document and the information contained within does not constitute a specification. The data is provided for informational purposes only and is not warranted by the manufacturer. It should be used in conjunction with the AQ3 Product Datasheet, Operating Principles(OP20) and the Product Safety Datasheet (PSDS 19).

, NOTICE
· Ensure the sensor is powered on for a minimum of 24 hours before use. · Sensor may experience higher failure risk when continuously exposed to
90 %RH/50ºC for > 168 hours. · All baseline tests are performed under clean dry air instead of ambient air.

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

THE GAS RESPONSE CURVE
The data in Figure 1 shows a typical response curve for the AQ3STF. Test data was taken from current production at the time of release of this document, and reflects the typical performance of a production batch at this time.
The data in Figures 2 and 3 shows typical response and recovery profiles based on the data above.

Concentration(ppb)

Figure 1. AQ3STF Gas Response and Recovery Profile
AQ3STF: EOL Test
500 Applied Concentration (ppb)
400
300
200
100

0

0

50

100

150

200

Time (s)

Figure 2. AQ3STF Gas Response Profile
AQ3STF: Response Curve
500

Concentration (ppb)

400

T90 : 5s

300

(Typical Value < 40 s)

200

100

0 0

20

40

60

80

100

Time (s)

Figure 3. AQ3STF Gas Recovery Profile
AQ3STF: Recovery Curve
500

400

Concentration (ppb)

300

200

100

0 0

20

40

60

80

100

Time (s)

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 2

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

LINEARITY
The data in Figure 4 shows the typical linearity performance of the AQ3STF gas sensor when subjected to differing Sulfur Dioxide concentrations which is 0 ppb to 1000 ppb.
The presented results reflect the performance of a typical production batch. Across typical measurement ranges for atmospheric monitoring, the sensor can often be considered linear.

Figure 4. Output Linearity from 0 ppb to 1000 ppb

Measured Concentration (ppb)

1000 900 800 700 600 500 400 300 200 100 0 0

AQ3STF: Linearity
100 200 300 400 500 600 700 800 Applied Concentration (ppb)

900 1000

Applied Concentration (ppb)

0

Measured concentration (ppb)

0

20

40

80

100

200

400

500

800 1000

18

39

79

99

201

400

499

799 1000

TEMPERATURE CHARACTERISTICS
Variation of Sensitivity with Temperature
The sensitivity of the AQ3STF Series gas sensor will vary as a function of ambient temperature. The data in Figure 5 shows the typical output performance across the operating temperature range and is presented normalized to the 20°C value with clean air.
For instruments that are expected to function across a wide range of ambient temperatures. Honeywell recommends that an electronic compensation algorithm is used to ensure maximum accuracy. The presented results reflect the performance.

Figure 5. Sensitivity vs. Temperature without Compensation
AQ3STF: Sensitivity vs Temperature
140

Sensitivity normalized at 20°C (%)

120

100

80

60

40

-20

-10

0

+3SD +2SD Mean -2SD -3SD

10

20

30

40

50

Temperature (°C)

Temperature (°C)

-20

-10

0

10

20

30

40

50

+3SD (equivalent to 99.7% confidence)

90

94

97

100

100

107

108

113

+2SD (equivalent to 95% confidence)

88

92

96

99

100

106

106

109

Mean (%)

82

88

93

96

100

102

102

101

-2SD (equivalent to -95% confidence)

76

84

90

94

100

99

99

92

-3SD (equivalent to -99.7% confidence)

74

83

88

93

100

97

97

88

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 3

SO2 Reading Variation

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

Variation of SO2 reading with Temperature after Compensation
To ensure maximum accuracy, an electronic compensation algorithm is being used for above result.
The Data in Figure 6 shows the AQ3STF reading performance across the operating temperature range after compensation.

Figure 6. SO2 Reading vs Temperature after Compensation
AQ3STF: Reading Aer Compensaon
550 33%
500 23%

SO2 Reading After Compensation (ppb)

450

13%

400

350

300

250

-20

-10

0

10

20

30

Temperature (°C)

3%

-8%

+3SD +2SD Mean (ppb) -2SD -3SD

40

50

-18% -28% -38%

Temperature (°C)

-20

-10

0

10

20

30

40

50

+3SD (equivalent to 99.7% confidence)

449

427

424

418

411

425

434

450

+2SD (equivalent to 95% confidence)

429

417

415

410

407

417

424

433

Mean (%)

387

397

398

393

399

401

403

399

-2SD (equivalent to -95% confidence)

346

376

382

377

391

385

382

365

-3SD (equivalent to -99.7% confidence) 325

366

373

368

387

377

372

348

Baseline Offset with Temperature

Figure 7. Baseline vs Temperature after Compensation

after Compensation

The electrical output in the absence of 80
target gas

(baseline offset) of the AQ3STF will

60

AQ3STF: Baseline Reading@T

Reading after compensation(ppb)

vary as a function of the ambient

40

temperature . The data on the right shows

20

typical AQ3STF performance across 0

the operating temperature range, for

sensors calibrated at 20°C with clean air.

-20

Although the variation is relatively small,

-40

Honeywell recommends the use of offset

-60

correction factors so as to minimize inaccuracies in the span measurement. The presented results are being

-80

-30

-20

-10

0

10

20

30

40

50

60

Temperature (°C)

compensated with correction factors.

You may find the correction factors in the

note on page 9. The presented results

reflect the typical performance of a

production batch.

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 4

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

LONG-TERM SENSITIVITY DRIFT
The typical long term sensitivity of the AQ3STF is represented in Figure 8, which reflects the performance of a typical production batch The sensor batches under test were stored and tested in ambient conditions.

Figure 8. Long-Term Sensitivity Drift
AQ3STF: Long-term Stability Performance
30%
20%
10%
0%

Sensitivity drift (%)

-10%

-20%

-30% 0

30

60

90

120

150

180

210

Time (day)

REPEATABILITY
The data in Figure 9 show the repeatability performance of the AQ3STF sensor when exposed repeatedly to SO2. The presented results reflect the performance of a typical production batch.

Figure 9. Repeatability of AQ3STF Sensor response to 400 ppb SO2

Response (ppb SO2)

450 400 350 300 250 200 150 100
50 0
-50 0

AQ3STF: Repeatability when Exposed to 400 ppb SO2
100 200 300 400 500 600 700 800 900 1000 1100 Time (s)

1st

2nd

3rd

4th

5th

application

Mean response (ppb)

412

408

406

405

404

Standard deviation (ppb)

4

3

5

6

5

6th
402 4

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 5

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

CALIBRATION GUIDELINES
Honeywell AQ3 Series gas sensors provide very stable signals over time and for many applications, instruments containing AQ3 gas sensors only require periodic recalibration. The time interval required between initial calibration and subsequent recalibrations is dependent on various factors. In strenuous applications involving extremes of operation, or for sensors used in safety applications, frequent instrument calibration may be required. Electrochemical gas sensors need a certain amount of oxygen to function. Generally, a few thousand ppm oxygen is sufficient. However, as calibration normally involves exposing the sensing face of the AQ3 gas sensor to gas for a relative short period of time, a calibration gas need not contain oxygen-sufficient oxygen is supplied from the ambient air, for a limited time. In most cases, a five minute exposure time is sufficient to achieve a stable calibration signal.
Safety Note: Many AQ3 gas sensors are designed to be used in safety critical applications. To ensure that the sensor and/or instrument in which it is used, is operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas (bump check) before each use of the sensor and/or instrument. Failure to carry out such tests may jeopardize the safety of people and property. Please be aware the recommended flow rate below is for a batch (6 sensor).

Sensor AQ3CO AQ3STF AQ3OZ AQ3ND

Warm-up time
24 hours
24 hours
24 hours
24 hours

Calibration Gas Concentration

Gas Path Saturation

1 ppm CO

1 ppm 2 L/min for 10 minutes

400 ppb SO2 400 ppb O3 400 ppb NO2

5 ppm 2 L/min for 5 minutes 2 ppm 2 L/min for 40 minutes 5 ppm 100 mL/min for 30 minutes

EOL Step
Ambient Air 1 min à Clean Air 3 mins à 1 ppm CO 3 mins à Clean Air 2 mins
Ambient Air 2 mins à Clean Air 3 mins à 400 ppb SO2 2 mins à Clean Air 3 mins Ambient Air 1 min à Clean Air 40 mins à 400 ppb O3 5 mins à Clean Air 3 mins Ambient Air 1 min à Clean Air 40 mins à 400 ppb NO2 5 mins à Clean Air 5 mins

Figure 10. Calibration Schematic

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 6

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

RECOMMENDED GAS FLOW RATES
A suitable flow rate is required to ensure accurate calibration ­ it also means that the response from an AQ3 gas sensor is equivalent in configurations where gas is flowing over the sensor and those where the sample is allowed to diffuse into the sensor. The minimum flow rate which is required will be different depending on the gas sensor type ­ these are shown in the table Please be aware the recommended flow rate is for a single sensor.

Gas Carbon Monoxide, CO Nitrogen Dioxide, NO2 Ozone, O3 Sulfur Dioxide, SO2

Sensor Type AQ3CO AQ3ND AQ3OZ AQ3STF

Flow Rate (ml/min) 500 500 500 500

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 7

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

RECOMMENDED CIRCUIT
The recommended circuit for un-biased 4-electrode AQ3 Series gas sensor is shown below. The description below can be applied to all AQ3 Sensors (unless otherwise noted). The circuit is essentially divided into two parts. The first part is a potentiostat circuit designed to keep the sensor sensing -reference voltage constant. An op amp is used to compare voltage at the reference pin to a stable bias voltage. Any movement of the reference pin voltage is compensated for by adjusting voltage on the counter pin. The circuit should draw no current from the sensor reference pin or the sensor output will be unstable. Further, the offset voltage of this op amp needs to be very low (typically 60 uV to 100 uV) or be nulled out, as offset in the circuit will appear as baseline offset in the sensor / instrument. The second part of the circuit, applied to both sensing and auxiliary pins, is a trans-impedance amplifier (TIA). The job of this circuit is to amplify the small current generated at the sensing pin when gas is detected and convert to a much larger voltage output which can be easily measured by instrumentation. Input is via a small load resistor (R106, R121 in the example circuit), whose value is specified on the sensor datasheet. Gain of this circuit is controlled by the feedback resistor, which is typically 20k to 1M (R107, R122 in the example). The circuit gain should be high enough to give a readable output, but not so high as to saturate the op amp at its highest (overload) output. The circuit below is based on operation from a single rail 5 V op-amp circuit with rail to rail output and a virtual ground reference for the sensor of 2.5 V. The output from the circuit will be positive with respect to virtual ground for sensors measuring oxidizing gases (CO and SO2) while the op amp output voltage ranges from 2.5 V to 4.5 V (2 V margin). Output will be negative with respect to virtual ground for sensors measuring reducing gases (NO2 and O3) while the op amp voltage ranges from 2.5 V to 0 V (2.5 V margin). Please be aware that signals for NO2 and O3 sensors are negative when using below circuit, remember to reverse the output signal when you use this circuit in real applications.

U100 ­ This LDO (LP5907 or similar, with low noise and low IQ) is to provide a stable voltage for the circuit. Please refer to chosen LDO datasheet for more detail.
U104 ­ This amplifier act as a trans-impedance amplifier (TIA) - current to voltage converter only.
U102 ­ This dual op amp amplifier should have either a low offset (<100 uV typical) or have its offset nulled out. This amplifier should also have a low power consumption. A suitable op amp is the OPA2336E or similar. This amplifier is used both as potentiostat and a current to voltage converter (trans-impedance amplifier).
U101 ­ The Zener diode circuit is to provide an accurate and stable reference voltage (2.5 V) to serve as virtual ground.

Figure 11. Recommended Circuitry for 4-electrode AQ3 Series Gas Sensors

, NOTE
Other op amp configurations may be used, including single- and dual-supply rails. In this case the reference voltage (bias voltage) will need to change to suit the circuit output range and sensor output, avoiding saturation of the op amp at limits of operation and ensuring a stable bias/virtual ground reference voltage.
AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 8

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

COMPENSATION LOGICS OF AQ3 SERIES SENSORS
This compensation logics provides customers with insights into our air-quality AQ3 Series sensors.
AQ3 Series sensors, including AQ3CO, AQ3STF, AQ3ND and AQ3OZ, are specially designed for high-precision and high-accuracy detection of ppb-level CO, SO2, NO2 and O3 in ambient environment. As electrochemical sensors, AQ3 Series are sensitive to the temperature variation, resulting in zero-background current change and sensitivity change that are well-known for long years. Compensation for these changes is indispensable for real-time and high-accuracy monitoring of ppb-level air pollutants. Therefore, in this Characterization Note we are mainly focusing on wellsimulated algorithms to make up for these changes, thus enhancing the accuracy and precision of result in the field application.
There are four electrodes in AQ3 Series sensors, i.e. sensing electrode, auxiliary electrode, reference electrode and counter electrode. For customers, the signals of sensing electrode and auxiliary electrode are most useful to compensate for zerobackground current and sensitivity in practical application. Based on our experiment results, we provide the best-fit compensation algorithms for different kinds of AQ3 Series sensors. Some key parameters are explained as below:

WE
WET
WE20 AE AE20 ST S20 CF RNO2 RO3 r1 r2 r3

real-time signal of sensing electrode, nA signal of sensing electrode at specific temperature in highly pure and dry air, nA signal of sensing electrode at 20°C in highly pure and dry air, nA real-time signal of auxiliary electrode, nA signal of auxiliary electrode at 20°C in highly pure and dry air, nA sensitivity of sensing electrode at specific temperature, nA/ppm sensitivity of sensing electrode at 20°C, nA/ppm cross-factor, different sensors have different CF, no unit AQ3ND concentration reading, ppb AQ3OZ concentration reading, ppb a function of temperature, usually r1=a1*T2+b1*T+c1, T/°C a function of temperature, usually r2=a2*T2+b2*T+c2, T/°C a function of temperature, usually r3=a3*T2+b3*T+c3, T/°C

The table below lists the compensation formula for different sensors, and the table above gives the parameters for different sensors.

Sensor Type
AQ3STF
AQ3OZ AQ3CO AQ3ND

Compensation Formula
Concentration (ppb) = [(WE-WE20) - r1*(AE-AE20)-r2CF1*RNO2*r3*S20-CF2*RO3*r3*S20] /(r3*S20) *1000 Concentration (ppb) = [(WE-WE20-r2-CF1*RNO2*r3*S20)/(r3*S20)]*1000
Concentration (ppb) = (WE­WE20­ r2)/(r3*S20)*1000

Note: r1 and r2 compensates for the zero-background current change due to temperature variation
r3 compensates for the sensitivity change due to temperature variation;
r1, r2 and r3 are usually function of temperature /°C

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 9

AQ3STF

Sulfur

Dioxide

(SO ) 2

Gas

Characterization

Note

CROSS SENSITIVITY TABLE
Whilst AQ3 Series gas sensors are designed to be highly specific to the gas they are intended to measure, they will still respond to some degree to various other gases. The table below is not exclusive and other gases not included in the table may still cause a sensor to react.
IMPORTANT NOTE: The cross sensitivity data shown below does not form part of the product specification and is supplied for guidance only. Values quoted are based on tests conducted on a small number of sensors and any batch may show significant variation. For the most accurate measurements, an instrument should be calibrated using the gas under investigation.

Gas Carbon Monoxide, CO Nitric Oxide, NO Nitrogen Dioxide, NO2 Hydrogen Sulfide, H2S Ozone, O3 Isobutylene, C4H8

Gas Concentration 5 ppm 5 ppm 0.4 ppm 5 ppm 0.4 ppm 5 ppm

Cross Interference ~None ~None -110%<x%<0% ~None -50%<x%<0% ~None

AQ3STF Sulfur Dioxide Gas Sensor Characterization Note | sps.honeywell.com/ast | 10

WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship during the applicable warranty period. Honeywell's standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgment or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items that Honeywell, in its sole discretion, finds defective. The foregoing is buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.
While Honeywell may provide application assistance personally, through our literature and the Honeywell web site, it is buyer's sole responsibility to determine the suitability of the product in the application.
Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this writing. However, Honeywell assumes no responsibility for its use.

m WARNING
MISUSE OF DOCUMENTATION
· The information presented in this characterization note is for reference only. Do not use this document as a product installation guide.
· Complete installation, operation, and maintenance information is provided in the instructions supplied with each product.
Failure to comply with these instructions could result in death or serious injury.

FOR MORE INFORMATION Honeywell Advanced Sensing Technologies services its customers through a worldwide network of sales offices and distributors. For application assistance, current specifications, pricing or the nearest Authorized Distributor, visit our website or call:

USA/Canada Latin America Europe Japan Singapore Greater China

+1 302 613 4491 +1 305 805 8188 +44 1344 238258 +81 (0) 3-6730-7152 +65 6355 2828 +86 4006396841

Honeywell Advanced Sensing Technologies 830 East Arapaho Road Richardson, TX 75081 sps.honeywell.com/ast

AQ3STF Characterization Note ECN 5062 | 002728-1-EN | 1 | 05/21 © 2021 Honeywell International Inc. All rights reserved.

