Resmed Vauto Manual

[PDF] VPAP Auto™ - Resmed

You can adjust the Exhalation rate so that the pressure drop is at a level that is most comfortable for the patient. Min EPAP and Max IPAP (VAuto mode). Max ...


File Info : application/pdf, 2 Pages, 162.32KB

PDF preview unavailable. Download the PDF instead.

268123 vpap-auto quick-setup-guide amer eng
VPAP AutoTM
Quick Setup Guide
English

Patient Menu

VAuto mode
SETTLE: 20min menu

S mode and CPAP mode
RAMP: 20min menu

EXHALATION: MED *

change

exit

USED:

0000hrs

000/000days exit

MASK: ULTRA

change

exit

SW: SX3650201 exit
* If enabled by a clinician

LCD screen Left key

Start/Stop and Mask fit key

Up key Right key Down key

To access the Patient Menu: Press the Left key. To move within a menu or between menus: Press the Up and Down keys. To enter a submenu: Press the Left key. To change a setting option: 1. Press the Left key to enter the menu. 2. Press the Up and Down keys until the desired setting
option appears. 3. Press the Left key to select the setting option. To exit without changing options: Press the Right key. To exit out of a menu or submenu: Press the Right key.
Note: You can return to the top level of the menu at any time by holding down the Right key for at least 3 seconds.

Pressure Support (VAuto mode)
Pressure
Pressure Support (Set)

Inspiratory P (auto-titrated)
(AutoSet P) (conceptual)

Time

Expiratory P (auto-titrated)

Pressure support allows you to set the difference between inspiratory and expiratory pressure to maximize patient comfort. The actual inspiratory and expiratory pressures are auto-titrated by the device.

Exhalation (VAuto mode)
Pressure

Medium
Fast Slow

To access the Efficacy data:
Hold down the Right and Left keys simultaneously for at least 3 seconds.
Choose Exit to return to the Patient menu.

To access the Mask Fit function:
Hold down the Start/Stop key for at least 3 seconds.
Press the Start/Stop key to return to the Patient menu.

Min EPAP and Max IPAP (VAuto mode)

Pressure

Max IPAP (Set)

Min EPAP (Set)
Time
These settings allow you to constrain the pressure range of the AutoSet algorithm. Note: The device will stay at Min EPAP pressure during the Settling time (if selected) at the beginning of a session.

Time
You can adjust the Exhalation rate so that the pressure drop is at a level that is most comfortable for the patient.

Clinical Menu

CLINICAL

menu

exit

SETTINGS

enter

exit

ADV SETTINGS

enter

exit

RESULTS

enter

exit

MODE: CPAP

change

exit

CPAP: 5.0

change

exit

SMARTSTOP: ON

change

exit

MASK: MIR FULL

change

exit

MODE: SPONT

change

exit

IPAP: 5.0

change

exit

EPAP: 3.0

change

exit

RISE TIME: 300ms

change

exit

Ti Max: 2.0s

change

exit

Ti Min: 0.1s

change

exit

SMARTSTOP: ON

change

exit

MASK: MIR FULL

change

exit

MODE: VAUTO

change

exit

PRESS SUP: 4.0

change

exit

Max IPAP: 18.0

change

exit

Min EPAP: 6.0

change

exit

Ti Max: 2.0s

change

exit

Ti Min: 0.1s

change

exit

EXHALATION: FAST

change

exit

SMARTSTOP: ON

change

exit

MASK: MIR FULL

change

exit

To access the Clinical Menu:
Hold down the Right and Down keys simultaneously for at least 3 seconds.

LEAK ALERT: OFF

change

exit

TUBE LENGTH: 2m

change

exit

MAX RAMP: 35min

change

exit

START CPAP: 4.0

change

exit

LEAK ALERT: OFF

change

exit

TRIGGER: MED

change

exit

CYCLE: MED

change

exit

TUBE LENGTH: 2m

change

exit

MAX RAMP: 35min

change

exit

START EPAP: 8.0

change

exit

LEAK ALERT: OFF

change

exit

TRIGGER: MED

change

exit

CYCLE: MED

change

exit

TUBE LENGTH: 2m

change

exit

MAX SETTLE: 45min

change

exit

EFFICACY DATA

enter

exit

LEAK: 11L/min

day...

exit

VT: 0-250

day...

exit

RESP RATE: 0-21

day...

exit

MV: 0.0-11.6

day...

exit

PRESS: 5.0

day...

exit

AHI: 15

day...

exit

AI: 7

day...

exit

LEAK: 12L/min

wk...

exit

VT: 0-245

wk...

exit

RESP RATE: 0-20

wk...

exit

MV: 0.0-11.7

wk...

exit

PRESS: 5.1

wk...

exit

AHI: 13

wk...

exit

AI: 9

wk...

exit

LEAK: 13L/min

mth...

exit

VT: 0-249

mth...

exit

RESP RATE: 0-19

mth...

exit

MV: 0.0-11.8

mth...

exit

PRESS: 5.2

mth...

exit

AHI: 12

mth...

exit

AI: 10

mth...

exit

LEAK: 14L/min

6mths...

exit

VT: 0-248

6mths...

exit

RESP RATE: 0-20

6mths...

exit

MV: 0.0-11.7

6mths...

exit

PRESS: 5.1

6mths...

exit

AHI: 14

6mths...

exit

AI: 11

6mths...

exit

LEAK: 13L/min yr...

exit

*

VT: 0-249 yr...

exit **

RESP RATE: 0-20

yr...

exit

**

MV: 0.0-11.8 yr...

exit **

PRESS: 5.0 yr...
AHI: 13 yr...

* 
exit exit

AI: 10

yr...

exit

USAGE DATA

enter

exit

OPTIONS enter
SERVICING enter

exit
¤
exit

REMINDERS

change

exit

PT ACCSS EXH: OFF

enter

exit

FACTORY DEFAULTS

reset

exit

ERASE DATA

erase

exit

DATE: 01 Jul 2007

change

exit

TIME: 15:54

change

exit

LANGUAGE:ENGLISH

change

exit

RUN HRS: 12345 exit

SN:123456789123

456789

exit

SW: S11601009 exit

USED: 12000hrs 102/106days exit

USAGE: 7:45hrs

day...

exit

MASK: 01Jul2007

change

exit

MASK RECUR: 6mth

change

exit

FILTER: 01Jul2007

change

exit

FILT RECUR: 6mth

change

exit

CALL: 01Jul2007

change

exit

CALL RECUR: OFF

change

exit

CARD: 01Jul2007

change

exit

CARD RECUR: OFF

change

exit

SERV: 01Jul2007

change

exit

SERV RECUR:24mth

change

exit

CUST: 01Jul2007

change

exit

CUST RECUR: OFF

change

exit

USAGE: 7:05hrs

wk...

exit

USAGE: 6.55hrs

mth...

exit

USAGE: 7:15hrs

6mths...

exit

USAGE: 7:20hrs

yr...

exit

* 95th percentile ** 5th and 95th percentile  CPAP: CPAP pressure
S: IPAP pressure VAuto: AutoSet prescription pressure ¤ VAuto Mode only
Note: All values for multiple sessions (ie, all except for day) are median values.

See the VPAP Auto User Guide and Clinical Guide for full details. ©2007 ResMed.

268123/1 07 11