Page 6/36 (39/22) Page 7/36 (39/22) • Before mounting or using the stand up and transfer aid Vita-Lift carefully read all instructions of this user manual, as these have been drawn up for your safety


File Info : application/pdf, 36 Pages, 4.47MB

PDF preview unavailable. Download the PDF instead.

um-vita-lift-en
User manual
Stand up, stand and transfer aid
Hoists
edition 48/21

Dear Customer!
We would like to thank you for your trust extended to us and for purchasing the VitaLift stand aid.
The stand up and transfer aid Vita-Lift distinguishes itself by its quality, security, easy operation and modern design.
For optimal use of this aid, we ask you and possible other users to carefully read this operating manual.

Foreword

Please read through these instructions carefully before first use and handle equipment appropriately!
Page 3/36 (48/21)

Table of contents

2 Table of contents

Topic

Page

1 Foreword ............................................. 3

2 Table of contents .................................... 4

3 Safety instructions ................................. 5

4 Warranty and product liability .................. 9

5 CE sign and position .............................. 9

6 Indications and prophylaxis ..................... 10

7 Purpose of use ....................................... 10

8 Product overview .................................... 11 8.1 Basic model .................................... 11 8.2 Accessories / options ........................... 12

9 Initial operation .................................... 13

10 Assembly and adjustment ........................ 14 10.1 Basic model .................................... 14 10.2 Accessories / options ........................... 21

11 Types of models .................................... 22 11.1 Basic model .................................... 22 11.2 Accessories / options ........................... 23

12 Technical details .................................... 24

13 Cleaning and disinfection ........................ 26

14 Maintenance and overhaul ........................ 28

15 Operating life and Re-use ........................ 30

16 Faults, repairs, spare parts ........................ 31

17 Storage ................................................ 32

Page 4/36 (48/21)

Safety instructions

3 Safety instructions

The stand up and transfer aid Vita-Lift is CE-marked and fulfils the requirements of the lastest directives. It was developed in line with most recent findings. Production is subject to an on-going quality control.
Please read this user manual carefully before operating the stand up and transfer aid.
In case of loss, request a new copy or download directly from our download area on our homepage www.ato-form.com.
Please keep the user manual in a safe place, accessible to care and maintenance staff.
The user manual does not replace the experience of supervisory staff with the experience and knowledge of operators, which they gained from (for example) training courses.

In the event of specific or difficult applications, the user manual represents a guide to the most important handling.
Ensure all users of this aid understand its handling and the meaning of the symbols attached to the device.
Accidents can be prevented when technical instructions according to the EWG guidelines are followed.
Protective coverings, labels and warning notices may not be removed.
The maximum load-bearing capacity of the Vita-Lift is 150 kg.

Regular monitoring at your specialist dealer (at least once a year) lengthens the life time of the Stand up and transfer aid and gives you security. Please also see section: maintenance and overhaul.
Page 5/36 (48/21)

Safety instructions

3 Safety instructions

· Before mounting or using the stand up and transfer aid Vita-Lift carefully read all instructions of this user manual, as these have been drawn up for your safety and as a guideline for the correct use and maintenance of this product. This operating manual is an indispensable component of the delivery and should be stored carefully for further use.
· The operating manual is intended for retailers and users of the stand up and transfer aid Vita-Lift, (or their carers). It contains the required information to secure the correct use of the product, it cannot, however replace the necessary experience of the user or assistant in handling the device.
· The reseller has the duty to deliver the device ready for operation and to explain precisely the functions and technical data of the device, as well as provide the necessary support and briefing needed for a correct and efficient use of the device.
· The given instructions of the operating manual, refers to the point in time of the device being placed on the market. The manufacturer reserves the right to introduce any modifications which prove necessary to the device and associated operating manual, without being bound to change previous devices and associated operating manuals.
· Use of the rehabilitation aid and its accessories, must be administered under medical supervision.

· The stand up and transfer aid Vita-Lift must only be operated under supervision of a qualified person, or respectively a family member of the user.
· Never leave the user of the aid unattended.
· The Vita-Lift may only be operated by persons who have familiarized themselves with the standards and rules of the present operating manual.
· The Vita-Lift may only be used by persons of a suitable stature and physical aptitude.
· Never place a patient in an aid, which is not coordinated with his or hers body height and weight.
· Adjust settings only in the unloaded condition.
· Do not place any hot beverages or objects onto the aid you have purchased, as this would seriously damage the aid.
· Do not load any heavy objects onto the aid, as this would impair the structure, the centre of gravity and consequently impair the balance.
· Always ensure there are no obstacles in operating range of the rehab aid, such as carpets, stairs or objects on the floor, which could lead to the stand up and transfer aid Vita-Lift overturning.
· Move all objects out of the way, which could be hit or fall whilst operating the aid.
· Never use the aid on rough terrain, steeply sloping ground or stairs.

Page 6/36 (48/21)

Safety instructions

· Do not use the rehab aid before ensuring a complete and correct fitting of the safety belts, if available.
· Check all adjustable parts are locked, before every use of the aid.
· The brakes must always be locked during patient transfer.
· Always lock brakes before use, when setting, adjusting functions or accessories.
· Please note the individual and overall safety relevant information when using several rehabilitation aids simultaneously.
· All aspects of spatial influences (uneven, wet and smooth floors, steps or similar) must be taken into consideration by the carer or user.
· Do not use the Vita-Lift before checking the wheels of the aid.
· Wet and dirtied wheels can decrease the braking effect.
· Check regularly if all screws and handles etc. are firmly tightened. Worn and defect parts are to be repaired or replaced immediately. This is especially important to ensure a safe use of the rehab aid! Carefully check the clamping jigs and the like are securely fixed.
· Do not use the aid, if parts of it are damaged or missing. Missing parts are only to be replaced with original spare parts.
· Do not carry out any interferences or measurements if you are unsure of their effects; for your own safety contact the nearest service centre or contact the dealer directly.

· In case of accidents or malfunctions, contact the dealer or manufacturer from whom you purchased the device.
· The manufacturer is not liable for damage to persons, objects or animals resulting from false usage of the equipment by the user or third parties, through incorrect mounting of parts and accessories, by neglect or poor maintenance, neglect, through unauthorized alterations and modifications, usage of non-original spare parts or accessories, through unpredictable and exceptional events resulting from non-compliance of the regulations in the presented operating manual.
· Clean the stand up and transfer aid VitaLift in accordance with the instructions given in the chapter ,,cleaning".
· Please ensure before use the product is cleaned according to the cleaning instructions and if indicated also disinfected.
· Any serious incidents occurring in connection with the product must be reported to the manufacturer (ATO FORM GmbH) and the competent authority of the country in which the user is resident.
· Make sure that the battery is charged regularly. Complete or total discharge of the battery will destroy the battery cells.
The Vita-Lift is equipped with automatic shut-off and an emergency-stop switch.

Page 7/36 (48/21)

Safety instructions

Page 8/36 (48/21)

We constantly develop new aids and improve existing products. We are happy to receive your suggestions of every description.

Warranty and product liability CE-sign and position

4 Warranty and product liability

ATO FORM grants the required by law warranty of your country against defects in workmanship and materials.
The warranty would be adversely affected if the customer's responsibility of servicing and/or daily maintenance is not carried out according to the guidelines and intervals prescribed by the supplier and/ or stated in the manual.
For further information, we refer to the ATO FORM homepage/download www. ato-form.com.
The warranty can only be sustained if the ATO FORM product is in use in the same country where it was purchased and if the product can be identified by the serial number. The warranty does not cover accidental damage, including damage caused by misuse or neglect.
The manufacturer`s obligations transfer then to the person carrying out the modifications.

The warranty does not extend to nondurable parts, which are subject to normal wear and tear and need periodic replacement.
The warranty is null and void if non-original ATO FORM parts/accessories are used, or if the product is repaired or altered by anyone other than an authorized ATO FORM representative or by trained personnel officially recognized by ATO FORM for repair and maintenance of ATO FORM products.
ATO FORM reserves the right to inspect the product being claimed for and the relevant documentation before agreeing to the warranty claim and to decide upon whether to replace or repair the defective product. It is the customer's responsibility to return the item being claimed for under warranty to the address of purchase.
The warranty is given by ATO FORM or, subsequently, an ATO FORM dealer.

5 CE-sign and position
The CE label with details of the manufacturer, article number, serial number and production date is located on the right side of the frame.

Page 9/36 (48/21)

Indications and prophylaxis Purpose of use
6 Indications and prophylaxis
Minimal mobility left e. g.: · Paralysis of lower extremities · Muscular diseases · And others
Contraindications
· Limb loss of the lower extremities · Open wounds (at the belt points)

The list of contraindications does not grant guarantee completeness. The patient`s general condition is an important part of the risk assessment. In individual cases, the decision must be made by the attending physician, who must weigh the risks versus the expected benefits.

7 Purpose of use

The Vita-Lift is a standing and transfer aid for home care, acute care and residential care. Due to its manoeuvrability and ease of use, it can be operated by one person.
It has become an indispensable aid for patients in transfer, standing and walking.
The Vita-Lift may be used to assist individuals from sitting to standing, transporting ans walking. All of the above should only be done following a risk assessment of the person being moved to asses their ability and safety using the equipment.
The Vita-Lift may only be used for the applications described above. For all other applications and any resulting damages in connection therewith, the manufacturer accepts no liability.
You also have the opportunity to use the extensive system accessories available from the ATO FORM range of products.

The device must be prescribed by a medical specialist, who has tested beforehand and determined if the Vita-Lift is suitable for the individual therapeutic goal of the patient. Furthermore, the medical specialist has determined the aid to be suitable according to measurements, weight and willingness of the patient to use the aid.
The aid is delivered via an authorised technician in compliance with the requirements of the country where it is being used and in strict accordance with the information and instructions in this manual for adjustments of the aid according to the measurements of the user. Only compliance of the regulations and instructions can lead to positive ergonomically effects.
We wish you every success when using the stand up and transfer aid Vita-Lift.

Page 10/36 (48/21)

8 Product overview 8.1 Basic model

1 Hooks to secure the belt loops

2 Lifting arm

1

3 Actuators with emergency lowering

4 Battery box

5 Braked wheels

6 hand control

7 Knee pad unit

8 Frame

9 Footplate

10 Legs (width adjustable) 8 with step protection

11 Control unit incl.

7

Emergency stop button

12 Connections for actuators, hand control and charging socket

13 Electrical width adjustment

3

9

10

5

Product overview
2 6
4 11 13 12 Page 11/36 (48/21)

Product overview

8.2 Accessories / options

15 Cross-plate

16 Handles

16

17 Cross-plate belt

18 Cross-plate belt incl. suede inside cover

15

17

18

19 Standard belt

19

20 Standard seat belt

21 Standard belt incl. abdominal belt 22 Standard belt incl. abdominal belt and 20
suede inside cover

23 Coat role

21

22

23 Page 12/36 (48/21)

For a complete overview of our extensive range of accessories for the Vita-Lift see item 11.2 Types and accessories on page 23.

Initial operation

9 Initial operation

The reseller is to unpack the unit, ensure that the components and accessories listed in the scope of delivery are all present and undamaged, and give the buyer the initial instructions on use of the aid. Above all the reseller should read through the complete present user manual carefully, in order to then explain it to the customer in detail.
This way, with help of the reseller, the user as well as the attendant must become familiar with the correct operation of the aid.
Once the demonstration has been completed, the reseller must ensure the user and/or

the assistant have fully understood the basic functions of the equipment. This is the only way to prevent the user from dangerous situations.
The above mentioned requirements must be observed for first use and also for reuse.
The seller must give instruction for use, as well as explain the basic functions of the device according to MPG (medical products law) to qualified staff.

Page 13/36 (48/21)

Assembly and adjustment

10 Assembly and adjustment
10.1 Basic model
Preparation: The Stand aid has been delivered to you as a complete hoist unit, so no assembly is required. Unpack the product and remove the packing materials. Ensure that all supplied components are present and undamaged. Please check before using, that all plug connections are plugged in correctly and secured. All sockets are located on the underside of the control unit case (Image 1 ­ 1 ).
Before using the VITA LIFT please note following: The battery (Image 1 ­ 2 ) must be loaded before the first use. For loading insert the plug in the power cord, ensure that the Emergency stop is in the out position and allow the battery to recharge for 2 hours prior to use. Use only the originally supplied power cord. If the Emergency stop is pressed in, then hold and rotate the red button clockwise and the button will pop out and the unit will function. If the control unit is connected to the power plug on power grid, the two LED displays (ON/Charge) flash on. As soon as full capacity is reached, the orange LED display ,,Charge" goes out.
Page 14/36 (48/21)

2

Image 1

1

The operation of the Vita-Lift is not available while loading! Therefore we recommend to charge the battery after using evry day.

Operation of the electrical parts
Check the emergency stop button (red button) located on the control box is in the on position (out position).
If the button is pressed in, then hold and rotate the red button clockwise an the button will pop out and the unit will function.

Assembly and adjustment

Fit the battery pack to the control box making sure it clicks into the position in the top of the bracket, as shown.

Image 2

Fit the hand control unit to the hand control socket located on the base of the control unit.
Note: The hand control lead can only be fitted in one orientation into its socket. Dot not use force to insert the hand control into its socket. The plug when correctly inserted should be pushed firmly home.

Image 3

Image 4

Page 15/36 (48/21)

Assembly and adjustment

Operation of the electric actuators
The raising and lowering of the lifting arm and the spreading of the suspension will be achieved by pressing the corresponding button on the control.
Before you use the stand-up aid, please insure, that the emergency-stop-button (Image 6 ­ 3 ) is not pushed. if applicable solve the button by turning the knob.
In dangerous situations push the red emergency-stop-button on the control unit immediately.

LED
Raises the lifting arm
Lowers the lifting arm
Opens the legs on the base unit
Closes the legs on the base unit

Image 5

During use:

If the battery is discharged while using the

VITA-LIFT the client can be moved to the

3

lower position by making use of the emer-

gency lowering (on the control unit). The

battery must be loaded immediately.

After using the Vita-Lift please always push the emergency-stop-button, otherwise the battery is going to unload.
Page 16/36 (48/21)

Image 6

Operation the emergency lowering
In the case of a manual operation fault, the electrical functions can also be operated via the two micro push-buttons of the emergency control on the control unit.
The electrical emergency lowering (Image 7 ­ 1 ) allows the lifting arm to be lowered to the lowest position.
To do this, press with a ballpoint pen or other pointed object int the right hole (micro button) below the redt button on the control unit.
You have the option of emergency operation (lifting the support arm) via the left micro button (Image 7 ­ 2 ).

Assembly and adjustment

2

1

Image 7

Additionally you can use the manual emergency lowering facilitated (Image 8
­ 2 ) by pulling the red lever on the lift actuator upwards until the support arm is in the desired position.

Image 8 Page 17/36 (48/21)

Assembly and adjustment
Operation the mechanical parts Knee pads For knee pad adjustment loosen the two set screws (Image 9, 10 ­ A for height, Image 9, 10 ­ B for depth).
B

A

1

Image 9

Page 18/36 (48/21)

B A
Image 10

Footplate model variant VI-1000-L
The footplate (Image 9 ­ 1 , Images 11 & 12) can be folded upwards. However, make sure that it locks into place so that it does not fold down unintentionally. To fold it down, press it down again.

Assembly and adjustment

Image 11

Image 12

Footplate model variant VI-5000
For patients who can still walk with support, the footplate (Image 13) can be removed. However, care must be taken to ensure that the footrest is securely attached in the correct position below the knee pad unit.

Image 13

Page 19/36 (48/21)

Assembly and adjustment

Operation ­ general procedures

Firstly ensure you select the correct sling size for the person being transferred.

The sling should be fitted around the persons stomach / lower back and not around the chest. If necessary, a sling with an additional seat belt has to be used.

Push the VITA-LIFT approach the patient

and lock the brakes of the wheels. Extend

the carry bar towards the person and attach Image 14

1

the relevant slings loops to the hooks

(Image 14). Tension the carry bar and lock

into position using the two screws (Image

15, 16 ­ 1 ).

Please make sure to straddle the legs (Image

16 ­ 2 ) by pressing the knob on the remote control before you start lifting.

Use the up function on the remote control to raise the person into the standing position.

1

Abb. 15
During use: If the battery is discharged while using the VITA-LIFT the client can be moved to the lower position by making use of the emergency lowering (on the control unit). The battery must be loaded immediately.

2 Image 16
Parts to insert such as loops, chains or rings must always be fixed correctly in the appropriate hooks (Image 15, 16 ­ 1 ).

Page 20/36 (48/21)

10.2Accessories / options

Use of cross plate belt

1

The cross plate belt can only be used in conjunction with the cross plate.

To adjust the suspension loops, loosen the star grip (Image 17 ­ 1 ) as shown in the illustration.

Make sure that the loops of the sling are securely hooked in at the same length and in the correct position.

Image 17

top

bottom

Image 18

Assembly and adjustment

For a complete overview of our extensive range of accessories for the Vita-Lift see item 11.2 Types and accessories on page 23.
Page 21/36 (48/21)

Types and accessories

11 Types and accessories 11.1 Basic model

Item No. VI-5000 VI-1000-L

Description Stand up and transfer aid Vita-Lift, colour silver, footplate fold away Stand up and transfer aid Vita-Lift, colour silver, footplate removable

Page 22/36 (48/21)

Types and accessories

11.2 Accessories / options

Item No.

Description

VG-0111

Standard belt, size 1

VG-0112

Standard belt, size 2

VG-0211

Standard seat belt, size 1

VG-0212

Standard seat belt, size 2

VG-0411

Standard seat belt, incl. abdominal belt, size 1

VG-0412

Standard seat belt, incl. abdominal belt, size 2

VG-0413

Standard seat belt, incl. abdominal belt, size 3

VG-0411-WL Standard seat belt, incl. abdominal belt with suede inside cover, size 1

VG-0412-WL Standard seat belt, incl. abdominal belt with suede inside cover, size 2

VG-0413-WL Standard seat belt, incl. abdominal belt with suede inside cover, size 3

VG-0300

Coat role

VG-0311

Cross-plate belt, size 1

VG-0312

Cross-plate belt, size 2

VG-0313

Cross-plate belt, size 3

VG-0311-WL Cross-plate belt, incl. suede inside cover, size 1

VG-0312-WL Cross-plate belt, incl. suede inside cover, size 2

VG-0313-WL Cross-plate belt, incl. suede inside cover, size 3

VI-3050

Cross-plate

VI-3065

Foot plate, elevated

VI-3070

Handles

VI-3090

Calf belt incl. belt cushion*

VI-3095

Calf belt fitting

*always add to VI-3095 Calf belt fitting order

The most up-to-date overview of our extensive range of accessories for VitaLift and their description as well as much more useful information can be found on our homepage www.ato-form.com.

For special problems, or if it seems the accessories you require are not demonstrated nor described, please contact us.
We may already have the right solution for you.

Page 23/36 (48/21)

Technical details

12 Technical details
Vita-Lift L Oberall length B Width with open legs (outside) B1 Width with closed legs (outside) H Overall height H1 Chassis height H2 Lifting arm height (extended)
Elektric drive (Akku) Supply Safety IP protection class Product weight Maximal load Service life Emergency switch Mechanical and electrical emergency lowering

cm cm cm cm cm cm Volt Volt Ampere
kg kg

94.8 58.0 up to 96.2 42.0 up to 80.2
100.4 16.0 89.0 bis 154.0 24 24 10 IPX4 45.5 165.0 10%, Max. 2 min./18 min.
Ja Ja

Technical changes reserved!

Page 24/36 (48/21)

Technical details
L

H H2

H1
B1 B

Construction:
The stand-up aid is consists of:
a powder-coated tubular steel frame, chrome-plated adjustable metal components, foam padded supporting elements with comfortable and easy-toclean leatherette upholstery, 4 polyamide double safety castors, each with full brake.
Hand control:
· Anti-slip coating on the back to prevent dropping,
· Essential information about battery state, overload occurrence and service always at hand,
· Button type: Easy touch tactile switches

Control Box:
· Service indicator, informing that it is time
for a service visit
· Internal charger · Diode display (LED indicators)

Operating conditions

Temperature: +5 °C to +40 °C

Relative humidity: 20% to 90% ­ noncondensing

Air pressure:

800 to 1060 hPA (Operation in an estimated height  2000 m)

Page 25/36 (48/21)

Cleaning and disinfection

13 Cleaning and disinfection

Leftovers, dirt and dust are always to be removed. Moving parts work easier, if you apply one drop of oil after cleaning.

Disinfection: For disinfecting, use household disinfectants and consider the following indications:

Cleaning the metal parts: The painted and chromed parts of the stand up and transfer aid should only be cleaned using a soft cloth and mild and alcohol based customary cleaning agents.
Cleaning the rollers/wheels and joints: Keep these clean at all times; regularly remove dust, hairs and other accumulating dirt from rollers.
Cleaning of textiles and imitation leather: These are best cleaned using water and a mild detergent.
When using the cleaning agents the instructions of the manufacturer must be observed.

The assistant must wear suitable protective clothing (for example gloves) in order to avoid infections. When using disinfecting agents always follow the instructions of the manufacturer.
Warning Do not use strong solvents, which could damage the surface of the device. The use of a high-pressure cleaner will void the warranty.
Ensure that the upholstery and wooden surfaces are kept clean. Urine and sweat can damage the product.

Page 26/36 (48/21)

Page 27/36 (48/21)

Maintenance and overhaul

14 Maintenance and overhaul

For the protection of the warranty and safety, it is especially important that the stand up and transfer aid is serviced regularly once a year from the date of purchase.
It is recommended to make service arrangements. The manufacture will gladly provide the maintenance check list to your specialist dealer on request.
Please contact your specialist dealer to carry out corrective maintenance works for your Vita-Lift.
After 3 years at the latest, the unit must be inspected by the manufacturer or its commissioner.

Maintenance Checklist
Castors
· No damage or wear · Firm seat
Brakes
· Free from grease and dirt · Full brake function · No signs of damage or wear
Frame
· Metal parts; no signs of damage or wear,
especially for adustable parts
· Welding seams are intact · Fasteners, bolts, nuts and springs are
secure
· Replacement of parts if there is any sign
of wear
· No signs of wear of damage of the safety
components

Upolstered padding
· No signs of wear

Page 28/36 (48/21)

Date

Service performed

Maintenance and overhaul

Name

Sign

Page 29/36 (48/21)

Operating life and Re-use

15 Operating life and Re-use

The Vita-Lift is made from high quality materials and is subject to a continuous quality control in our house. Because of this and the solid construction, the expected operating life of our product is up to 5 years, depending on the usage intensity and number of re-uses, under normal and compliance with its intended use (according to the instructions given in our user manual).
The expected operating life does not apply to wear parts.
The product Vita-Lift was designed so it could be reused after the application of various measures.

Following measures should be followed:
· Thorough cleaning and disinfecting using appropriate cleaning agents.
· Replacement of wear parts
· Maintenance, following the maintenance schedule of the manufacturer or authorized specialist dealer.
· Preparing accompanying documents and release for re-use.

Page 30/36 (48/21)

Faults, repairs, spare parts

16 Faults, repairs, spare parts

If your stand up and transfer aid lift Vita-Lift cannot be operated?
Please ensure.....
· The plug is plugged in the socket and is supplied with power.
· the main switch is activated and there are batteries in the remote control.
· the emergency-stop-button on the battery housing is not depressed.
· All plug connections are plugged in correctly and secured.
Try and shut down the fault.
Failing that, contact your specialist dealer / reseller.

Only original accessories and original spare parts may be used. Non-compliance with these instructions and repairs carried out or modifications made to the Vita-Lift or its parts by unauthorized persons, voids any warranties.
In these cases, any product liability granted by the manufacturer (ATO FORM GmbH, Zur Lauterhecke 34, 63877 Sailauf, Germany) is voided in advance.
For all questions about spare parts and accessories contact your specialist dealer or the ATO FORM-Service 0049 6093-944-0.

Spare parts and accessories must not be removed or replaced while the patient is using the stand up and transfer aid!

Page 31/36 (48/21)

Storage

17 Storage

Keep this operating manual in a safe place, where you or care and maintenance staff have access to it.

The current version of these instructions for use can be found on our homepage in the download area.
You can find more information about the product on our homepage at www.ato-form. com or under this QR code.

Page 32/36 (48/21)

Page 33/36 (48/21)

Page 34/36 (48/21)

Disposal of battery operated devices
The battery pack must not be disposed of together with normal household waste at the end of its service life. Please observe the legal requirements in your area. Your municipal government and locally based waste disposal companies will be happy to provide you with information in this regard.
In environmental protection matters:
All packaging parts are re-useable. The cardboard packaging is to be recycled as waste paper. We therefore use paper as filler material. Possible used plastic film of the packaging are made out of PE and are therefore low in pollutant and can either be recycled or burnt.

2021-48/St.I-2021-18

Complete Range:  Head and body protection  Walking aids  Gait trainer  Standing aids  Seating aids  Hoists  Customized solutions
Thank you for your interest in our products. For questions please contact ATO FORM GmbH or your specialist dealer. Your ATO FORM Team
ATO FORM GmbH
Zur Lauterhecke 34, 63877 Sailauf, Germany Tel. +49 6093 944-0, Fax +49 6093 944-444 E-Mail: info@ato-form.com
www.ato-form.com
Your specialist dealer



References

Adobe PDF Library 16.0.3 Adobe InDesign 17.0 (Macintosh)