TSI VelociCalc Air Velocity Meter

Model 9545/9545-A

Operation and Service Manual

P/N 1980564, Revision G, November 2024

Description: The TSI VelociCalc Air Velocity Meter is a handheld instrument used for measuring air velocity, temperature, and humidity. It features a digital display, a telescoping probe, and connectivity for data download.

Website: www.tsi.com

Benefits of Registration

TSI encourages users to register their instrument purchase. Registration allows TSI to send important information regarding software updates, product enhancements, and new products. Registration can be completed at http://register.tsi.com. The registration process also includes an opportunity to provide feedback on TSI products and services.

Copyright, Warranty, and Service Information

Copyright © TSI Incorporated / 2007-2024 / All rights reserved.

Address: TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA

Fax: (651) 490-3824

Limitation of Warranty and Liability (effective May 2024)

(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)

TSI warrants its goods (excluding software) against defects in workmanship and material under normal use for a period determined by the longer of 24 months or the time specified in the operator's manual from the date of shipment. This warranty is subject to specific exclusions and exceptions, including:

This warranty is in lieu of all other warranties. No other express or implied warranty of fitness for a particular purpose or merchantability is made. TSI's liability for infringement is limited to direct infringement.

The buyer's exclusive remedy is the return of the purchase price (discounted for wear) or, at TSI's option, replacement with non-infringing goods.

TSI's total liability for any losses, injuries, or damages is limited to the return of goods and refund of the purchase price, or repair/replacement of the goods. For software, defective software will be repaired/replaced, or the purchase price refunded. TSI is not liable for lost profits, business interruption, or incidental/consequential damages. Installation/dismantling costs are not covered. Actions must be brought within 12 months of accrual.

Acceptance of the product constitutes acceptance of this Limitation of Warranty and Liability. Amendments require written consent from a TSI Officer.

Service Policy

TSI is committed to prompt attention for any instrument problems. If a malfunction is discovered, contact your nearest sales office or representative, or call Customer Service at (800) 680-1220 (USA) or (1) 651-490-2860 (International).

Trademarks

TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country's trademark registrations.

Chapter 1: Unpacking and Parts Identification

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI Incorporated immediately.

  1. Carrying case
  2. Instrument
  3. USB cable

Chapter 2: Setting-up

Supplying Power

The TSI Model 9545/9545-A VelociCalc Air Velocity Meter is powered by four size AA batteries.

Installing the Batteries

Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The instrument operates with either alkaline or NiMH rechargeable batteries, but it will not recharge NiMH batteries. Battery life is shorter with NiMH batteries. Carbon-zinc batteries are not recommended due to the risk of acid leakage.

Using the Optional AC Adapter

When using the AC adapter, the batteries (if installed) are bypassed. Ensure the correct voltage and frequency, as marked on the adapter, are used. The AC adapter does not recharge the batteries.

Using the Telescoping Probe

The telescoping probe contains the velocity, temperature, and humidity sensors. Ensure the sensor window is fully exposed and the orientation dimple is facing upstream during use.

NOTICE: For temperature measurements, ensure at least 3 inches (7.5 cm) of the probe is in the flow to allow the temperature sensor to be in the air stream.

Extending the Probe

To extend the probe, hold the handle in one hand and pull the probe tip with the other. Do not hold the cable while extending, as this can prevent proper extension.

Retracting the Probe

To retract the probe, hold the handle in one hand and gently pull the probe cable with the other.

Connecting to a Computer

Use the provided Computer Interface USB Cable to connect the Model 9545/9545-A to a computer for downloading stored data. Refer to Chapter 3, "LogDat2 Downloading Software," for detailed instructions.

CAUTION: A symbol (indicating a prohibition for network connection) signifies that the data port of the Model 9545/9545-A is NOT intended for connection to a public telecommunications network. Connect the USB data port only to another USB port.

Chapter 3: Operation

Keypad Functions

Common Terms

Menus

DISPLAY SETUP

Configure desired parameters for the running screen. Use ON/OFF soft keys to show/hide parameters. The PRIMARY soft key displays one parameter in a larger format. One primary and up to two secondary parameters can be selected.

SETTINGS

Set general parameters including Language, Beeper, Units, Time Constant, Contrast, Time, Date, Time Format, Date Format, Number Format, Backlight, and Auto Off. Use ⬅️ or ➡️ soft keys to navigate options and ← to accept.

FLOW SET UP

Select from four flow types: Round Duct, Rectangle Duct, Duct Area, and Horn. Use ⬅️ or ➡️ soft keys to cycle through types and ← to accept. Highlight "Enter Settings" and press ← to change values.

NOTICE: Horn numbers correspond to horn models (e.g., AM 100). Compatible horn models include AM 100, AM 300, AM 600, and AM 1200. Selecting a horn model enables the instrument to use a preprogrammed curve for flow rate calculation from velocity.

ACTUAL/STANDARD SETUP

Choose between Actual/Standard measurements. Select Standard Temperature, Standard Pressure, and the source for actual temperature. Barometric pressure must be entered for converting air velocity and volume measurements to actual conditions.

DATA LOGGING

Measurements

Logged measurements are independent of displayed ones and must be selected under DATA LOGGING → Measurements.

Log Mode/Log Settings

Log Mode options: Manual, Auto-save, or Cont-key.

Auto-save and Cont-Key modes offer additional Log Settings:

ModeLog Settings
Auto-saveLog Interval
Cont-keyLog Interval

Pressing ⬆️⬇️ simultaneously locks the keypad. Pressing them again unlocks it.

Delete Data

Use to delete all data, tests, or samples.

% Memory

Displays available memory. "Delete All" under Delete Data clears memory.

LogDat2 Downloading Software

Download LogDat2 software from the TSI Software and Firmware Wizard page. Install by running the application file and following on-screen instructions.

To download data, connect the Model 9545/9545-A to a computer via USB cable. Run the LogDat2 software and select tests for download or double-click a test to open it.

Chapter 4: Maintenance

The Model 9545/9545-A requires minimal maintenance.

Recalibration

For optimal accuracy, TSI recommends annual recalibration. Contact TSI offices or a local distributor for service arrangements and a "Service Request" number. An online "Service Request" form is available at tsi.com/service.

TSI Contact Information:

TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 680-1220
(651) 490-2860
Fax: (651) 490-3824

Field recalibration is possible via the CALIBRATION menu for minor adjustments. For complete, multiple-point calibration, the instrument must be returned to the factory.

Cases

Clean the instrument or storage case with a soft cloth and isopropyl alcohol or mild detergent. Do not immerse the instrument. If the enclosure or AC adapter is broken, it must be replaced immediately to prevent exposure to hazardous voltage.

Storage

Remove batteries when storing the unit for over one month to prevent battery leakage damage.

Chapter 5: Troubleshooting

Table 5-1 outlines common symptoms, causes, and solutions for the Model 9545/9545-A. For unlisted symptoms or unresolved issues, contact TSI Incorporated.

Table 5-1: Troubleshooting the Model 9545/9545-A

SymptomPossible CausesCorrective Action
No DisplayUnit not turned on
Low or dead batteries
Switch unit on.
Replace batteries or plug in AC adapter.
Velocity reading fluctuates / unstableDirty battery contactsClean the battery contacts.
Fluctuating flowReposition probe in less-turbulent section or use longer time constant.
Instrument Error message appearsMemory is fullDownload data if desired, then DELETE ALL memory.
Fault in instrumentFactory service required on instrument.

WARNING: Remove the probe from excessive temperature immediately, as excessive heat can damage the sensor. Operating temperature limits are detailed in Appendix A, Specifications.

Appendix A: Specifications

Velocity Probe

Duct Size

Volumetric Flow Rate

Temperature from Velocity Probe

Relative Humidity from Velocity Probe

Wet Bulb Temperature from Velocity Probe

Instrument Temperature Range

Instrument Operating Conditions (Optional AC Adapter)

Data Storage Capabilities

Logging Interval

Time Constant

User selectable

Response Time

External Meter Dimensions

3.3 in. × 7.0 in. × 1.8 in. (8.4 cm x 17.8 cm x 4.4 cm)

Meter Probe Dimensions

Articulating Probe Dimensions

Meter Weight

Power Requirements

Four AA-size batteries (included) or AC adapter (optional) 9 VDC, 300 mA min.

1 Temperature compensated over an air temperature range of 40 to 150°F (5 to 65°C).

2 Accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is greater, begins at 30 ft/min through 6000 ft/min (0.15 m/s through 30 m/s).

3 Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F (0.03°C/°C) for change in instrument temperature.

4 Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for change in probe temperature. Includes 1% hysteresis.

Notes and Contact Information

TSI Contact Information:

Visit www.tsi.com for more information.

P/N 1980564 Rev G ©2024 TSI Incorporated Printed in U.S.A.

Models: 9545, 9545A, 9545 Air Velocity Meter, 9545, Air Velocity Meter, Velocity Meter, Meter

File Info : application/pdf, 24 Pages, 317.54KB

PDF preview unavailable. Download the PDF instead.

9545A-VelociCalc-US 1980564-web

References

Microsoft Word 2019

Related Documents

Preview TSI VelociCalc Air Velocity Meter Model 9545/9545-A Operation and Service Manual
Comprehensive operation and service manual for the TSI VelociCalc Air Velocity Meter, Model 9545/9545-A. Covers setup, operation, maintenance, troubleshooting, and detailed specifications for accurate airflow measurements.
Preview HVAC Assessment Handbook: A Practical Guide to Performance Measurements
A comprehensive guide to understanding and measuring the performance of Heating, Ventilation, and Air Conditioning (HVAC) systems, covering key parameters like air velocity, ventilation, thermal comfort, and air quality. Includes information on system design, common issues, and relevant measurement instruments.
Preview TSI Air Velocity Transducer Model 8455/8465/8475 Operation and Service Manual
Comprehensive operation and service manual for the TSI Model 8455, 8465, and 8475 Air Velocity Transducers, covering specifications, setup, troubleshooting, warranty, and maintenance information.
Preview TSI VelociCalc™ 9600 系列:高精度HVAC通風測量儀器
TSI VelociCalc™ 9600 系列多功能通風儀提供專業人士專用的引導式測量流程,確保精確可靠的HVAC測試、平衡和IAQ調查。探索9600、9630、9650和9650-NB型號。
Preview Alnor AVM430/AVM430A and Airflow TA430/TA430A Air Velocity Meter Operation and Service Manual
Comprehensive operation and service manual for the Alnor AVM430/AVM430A and Airflow TA430/TA430A Air Velocity Meters, covering setup, operation, maintenance, troubleshooting, and specifications. Includes warranty information and contact details.
Preview TSI Gas Flow Multi-Meter Models 5300/5310/5320/5330/5303 Operation and Service Manual
This document provides comprehensive operational and service guidance for the TSI Gas Flow Multi-Meter series, including models 5300, 5310, 5320, 5330, and 5303. It details setup procedures, measurement capabilities, user interface navigation, data management, maintenance, and troubleshooting for these advanced gas flow measurement instruments.
Preview TSI Model 4199 Flow Meter Battery Pack and Stand Operation and Service Manual
Operation and service manual for the TSI Model 4199 Battery Pack and Stand, detailing installation, removal, and compatibility with TSI Series 4000 and 4100 flow meters. Includes component list and usage notes.
Preview TSI Model 8640 SUREFLOW Face Velocity Controller User's Guide
Comprehensive user's guide for the TSI Model 8640 SUREFLOW Face Velocity Controller, detailing its operation, calibration procedures, maintenance, troubleshooting steps, technical specifications, theory of operation, and wiring diagrams.