Instruction Manual for BeaverLAB models including: DDL-MX Intelligent Microscope, DDL-MX, Intelligent Microscope, Microscope

English Manual

MANUAL – BEAVERLAB

BeaverLAB Intelligent Microscope MX

14 sept. 2023 — Please read this user manual carefully! Better reading experience! Please ... Read the pictures in the microscope memory with the mobile phone or copy them to the ...


File Info : application/pdf, 22 Pages, 2.98MB

PDF preview unavailable. Download the PDF instead.

1040691
BeaverLAB Intelligent Microscope MX
DDL MX Instruction Manual
Please read this user manual carefully! Better reading experience! Please keep the manual properly!

Catalogue

1. Product Overview

01

2. Instructions For Use

02

3. Instructions For Using Microscope

03

4. Instructions For Operate Steps of the microscope

06

5. User Guide For Interface

11

6. Function Description Of External Equipment

15

7. Accessories And Tools

26

8. Troubleshooting Of Microscope

19

9. Trademarks And Legal Statement

20

10. Product Warranty Instructions

20

Overall Parameters Of Microscope

1 Product Overview
 Thank you for purchasing BeaverLAB intelligent microscope, hereinafter referred to as microscope.  Please read this manual carefully and keep it properly before use. Do not use the microscope without any
instructions.
01

2 Instructions For Use
Please read this manual carefully and use it correctly. Pay attention to protect the lens. Be sure to close the lens cover after using, avoid entering dust or foreign matters. Pay attention to protect the display screen. Please close the screen after using, avoid damaging the display screen. The observation table of this microscope is a lifting table. Please keep the table clean and do not press it heavily. When placing specimens or other observation objects, pay attention to the placement position and place them in the center accurately below the lens as far as possible. Avoid being unable to see objects and bring a bad experience. Import of external files is not supported (external files such as videos, pictures, and documents do not support import and reading, which may cause the device to malfunction.)
02

3 Instructions For Microscope
Microscope Composition
Screen
Host Head Objective Lense
Lifting Table
03

Upright column Base

Knob Function Description
Objective Lens Lighting Adjusting
Knob
Turn it on clockwise & increase the brightness Turn the brightness down
Objective Lens Switching Knob
100x / 600X / 1200X switch by rotation
Platform Move Back And Forth Button
Rotate counterclockwise to move the left platform backward
Rotate the platform clockwise and move it forward

Platform Bottom Light Adjustment Knob
Turn it on clockwise, turn it up, turn it down
Focusing Knob (Fine Adjustment)
Zoom out clockwise / zoom in counterclockwise
Platform Lifting Adjustment Knob
Decrease counterclockwise / increase clockwise
Platform Left And Right Movement Knob
Rotate counterclockwise and move the platform to the left Rotate the platform clockwise and move it to the right
04

Interface Usage

USB
It can be operated by connecting the mouse
Use Environment

Charging / Data Transmission Port
DC5V Charging / Using Adapter
You can connect your mobile phone to export data Connect windows computer and operate
with computer software

Long press on /

Please use it at10  ~ 45 

Please prevent the microscope from getting wet by water, beverages and corrosive lampliquids

Please keep the microscope away from
heat source, open
explosive gas (liquid)

Please avoid dust enteringthe microscope lens
and inside

Please avoid the impact and
violent vibration of the microscope

·At low temperature, the available capacity of the battery will decay to varying degrees.
·In case of any abnormality, please contact the after-sales department for maintenance. Disassembling the microscope without permission will not enjoy the warranty policy and may lead to irreparable damage.

05

Charging Instructions

The red light is always on during charging and the green light is always on after charging

Can use daily mobile phone charger, computer The power bank charges the microscope.

12 h

Do not charge for more than service life of the battery.

During charging, if the microscope shell and battery have slight heating, it is normal. Please rest assured to use it

Storage Environment
Store in a cool and dry place for daily use and avoid direct sunlight. Avoid storing the microscope in a place where there is a risk of falling, which may cause lens deviation inside the microscope or damage to other components, as well as other irreparable defects damage problem.
06

4 Instructions for using the microscope

1. Place Stably

2. Open the screen

long time to start the machine
07

4. Move the platform to a suitable

5. Remove the lens cover and put in the observation items (note that the specimen should be facing the eyepiece)

6. Select the multiple to be observed (note that the specimen should be facing the eyepiece)

7. Raise the platform to a suitable height

8. Adjust the intensity of bottom lighting and top lighting

08

9. Focus through the focusing knob and observe the display screen to a clear position

10. The observation position can be moved by adjusting the moving knob under the platform

The Pithy Formula For Microscope Use: large multiples should be close and small multiples should be far
Use Of Specimen Holder
Specimen Fixing Clip
Fixed specimen or observation target
09

Matters Needing Attention: the base. adsorption of other metal substances.
Use Of Observation Panel

Center Indicator Arrow

Observation

Matters Needing Attention: The observation panel shall be wiped regularly to avoid oil stain or other stains affecting the observation experience.
Description of light adjustment function of observation window
pay attention to the adjustment and use.
10

5. Interface User Guide
Software interface function introduction

Gallery Take A Picture
Record Video Dimensional calibration
Measure the internal dimensions of the calibration system
Connect to the computer
through settings switching
Gallery Observe, browse and edit the photographed Pictures and videos.
11

9

9

Click the screen to call out / Hide the

function option interface

Take A Picture
Take pictures of the currently observed objects.

Dot Straight Line Circular Casual Painting Angle Rectangle With Rounded Ellipse Delete Undo / Return
Record Video Take a video of the currently observed object

Dimensional calibration
Measure the internal dimensions of the calibration system

Set Up
Language: Chinese / English Time :time/ date Sleep: never / 3 min / 5 min / 10 min / 15 min Timed Shutdown: never / 5 min/ 10 min / 30 min / 1 hour Memory Card Switching: in / out Camera Switching: in / out Color Selection: blue, green, red, yellow, purple

Shut Down

Grid
locate the center point of the image

Straight Line
Measure the length of the observation object

Circular
Measure the S (area) of the selected range C (perimeter),D (diameter)

Casual Painting Tagging

Angle
Measure the angle of the observation object

Rectangle with rounded corners Measure s (area) c (perimeter) of the selected range

Ellipse
Measure s (area) and d(diameter) of the selected range

Delete
Delete all dimensions on the interface

Undo / Return Undo the previous action
12

6 Function Description Of External Equipment
Connect To Mobile Phone Purpose: Read the pictures in the microscope memory with the mobile phone or copy them to the mobile phone. Operation method:  Plug the data cable into the DC5V charging port of MX microscope, plug the adapter into the mobile phone end, and connect the data cable with the adapter.  Open the MX microscope and click: Setting  Memory card switching  Cut out in sequence. At this time, the screen will darken.
13

 Open the mobile phone  File manager. At this time, you can see a USB storage device. After opening, you can see all the pictures in the memory of MX microscope. Computer Device Connection Visit www.dangdangli.com Download and install the software installation package Purpose: Directly connect to the computer, use the computer software to observe and operate, and the screen is larger. Operation Method:  Plug the data cable into the DC5V charging port of MX microscope, and connect one end to the USB interface of computer.  Open BeaverLAB computer software (only windows computers are supported).  Open the MX microscope and click: Setting  Camera Switching  Cut Out, and the screen darkens at this time.
14

 In the computer software, click: connect device button  select device (select the recognized camera device)  open
 At this time, the software on the computer side is directly connected to the camera, so you can use the computer to observe directly.
15

Click Connect Device Click Select Device
Click Confirm

7 Accessories And Tools

Prepared Slides x10

Adapter

Data Line

Magnetic specimen holder x2

8 Troubleshooting Of Microscope

Problem Phenomenon
Lifting is not smooth / Unable to lift
Lifting is not smooth / Unable to lift
The light cannot be adjusted
The device cannot be charged

Reason

Processing Method

During charging, the current is large and the calorific value of the battery is high, which is a normal phenomenon

Normal state

Internal lifting gear damaged

Send for repair

Dimmer circuit board damaged / Lamp board damaged / Connecting wire open circuit
 The charging interface is not firmly plugged in  No power connection  Equipment failure

Send for repair Unplug the interface Connect the power For repair

16

The device cannot be used normally
The device cannot be connected to the computer software
Lens display image blur
There are foreign bodies in the specimen and the observation is not clear
The shell is broken and cracked due to falling and impact
The product cannot be turned off or the key fails

Internal damage of equipment due to Send for repair *Man made damage is not water spray, collision and other reasons covered by warranty

The interface is not plugged tightly, the data line is damaged, and the connecting line with data transmission function is not used
The ambient temperature changes sharply, causing the lens to fog

Reseat the interface, replace the data cable, replace the data cable with data transmission function, and then connect it for use
Wait for the temperature to be constant and use it after the fog disappears for a period of time

The surface of the specimen is dirty, and the panel is dirty

Wipe the surface of the specimen and the surface of the observation panel

Damage caused by external use

Man made damage is not covered by the warranty

Crash

Please reset and restart

9 Trademark And Legal Declaration

"

"" BeaverLAB " is a trademark applied for or registered in Chinese Mainland by Beaver

Technology Co., Ltd. and is used on this product use. Without the permission of thetrademark owner, no

person or institution may use the above trademark logo on unapproved goods without authorization.

This manual is made and copyrighted by Beaver culture and innovation technology (Shenzhen) Co.,

Ltd. no organization or individual may copy or distribute all or any part of this manual without permission.

Due to the continuous improvement of product functions, design changes and other reasons, this manual may be

inconsistent with the products you purchased. Please take the actual products as the standard. Product executive

standard: GB 4943.1-2011 GB / T 9254-2008

17

10 Product Warranty Description
1.Warranty Period  Within 7 days from the day after you sign for it, if the product has any performance failure listed in the performance failure table of BeaverLAB intelligent microscope products, you can enjoy the return or replacement service free of charge after being tested and determined by BeaverLAB after-sales service center;  Within 8-15 days from the next day after you sign for it, if this product has any performance failure listed in the performance failure table of BeaverLAB intelligent microscope products, it can enjoy the replacement or maintenance service free of charge after being tested and determined by BeaverLAB after-sales service center;  Within 6 months from the day after you sign for it, this product has the performance faults listed in the performance fault table of BeaverLAB intelligent microscope products. After being tested and determined by BeaverLAB after-sales service center, you can enjoy the maintenance service free of charge. Friendly Tip: because the packing box needs to be used during transportation to ensure the safety of product transportation, it is recommended that you keep the packing box for at least 30 days from the date of signing.
2.Warranty Content The host is guaranteed for 6 months.
3.Non Warranty Regulations
incorrect use methods, or tearing, altering labels and anti-counterfeiting marks;
18

 The warranty period has expired;
 Damage caused by force majeure;
 It does not comply with the performance faults listed in the performance fault table of BeaverLAB intelligent microscope products;
 Due to human factors, this product and its accessories have the performance faults listed in the performance fault table of BeaverLAB intelligent microscope products.
Overall Parameters Of Microscope

Product Brand: Product Name: Beav Product Model: Product Color: Battery Capacity: Texture Of Material:

BeaverLAB BeaverLAB Smart Microscope MX DDL-MX Gray/ White/ Yellow 3500mAh ABS+PC

19

Intelligent Microscope MX

Product Name: Intelligent microscope MX Product Model: DDL-MX Product Size: 214x224x324mm

Input Voltage: DC5V 1A Battery Capacity: 3500mAh Display Size: 9 inches

Brand Name: BeaverLAB Executive StandardCE FCC

Brand Company: Beaver Technology (Shenzhen) Co., Ltd.

Service Mailbox: support@beaverlabtech.com www.beaverlabtech.com / www.dangdangli.com



References

Adobe Illustrator 26.0 (Windows) GPL Ghostscript 9.52