Maintenance
[Warning] Before disassembly of any meter or accessory component, ALL INTERNAL PRESSURES MUST BE RELIEVED AND ALL LIQUID DRAINED FROM THE SYSTEM IN ACCORDANCE WITH ALL APPLICABLE PROCEDURES. Pressure must be 0 (zero) psi. Close all liquid and vapor lines between the meter and liquid source. Failure to follow this warning could result in property damage, personal injury, or death from fire and/or explosion, or other hazards that may be associated with this type of equipment.
Tools Required:
- Flat blade screwdriver
- 14mm box end or open end wrench
Disassembling
The optical air eliminator consists of a housing, optical sensor, and control solenoid valve. Only the solenoid valve is serviceable. If any part of the solenoid is damaged, a new solenoid assembly (Part Number 502011) must be ordered. The optical sensor has no serviceable parts as its internal components are potted; if it fails, the complete assembly (Part Number 81947) must be replaced.
- Step 1: Loosen the thin hex nut holding the solenoid in place using a 14mm wrench. Remove the nut and coil from the armature guide post.
- Step 2: Remove the plastic bonnet from the armature guide post. This should be easy to remove without tools.
- Step 3: Using a screwdriver, loosen the upper left and lower right screws of the valve body. These two screws hold the valve body in place. The upper right and lower left screws fasten the armature guide post and valve body together. Remove the valve body from the optical air eliminator housing.
- Step 4: Place the valve body on a flat surface. Using a flat blade screwdriver, remove the two screws holding the armature guide post and valve body together. Lift the armature guide post off the valve body. The internal components are a plunger and a spring. Inspect the spring for damage.
- Step 5: The armature guide post is composed of four components: two O-rings, the guide post, and the flange. Inspect these components for damage. Image shows an armature guide post with O-rings highlighted.
- Step 6: The valve body has two O-rings on the face that mates with the housing. These O-rings are identical. Inspect them for damage. Inspect the ports for blockage. Image shows the valve body face with O-rings highlighted.
Reassembling
- Step 1: Place the spring inside the plunger and insert the plunger, spring end first, into the armature guide post. Place the armature guide post assembly onto the valve body. Fasten the armature guide post to the valve body using the two screws removed earlier. Ensure screws are inserted into the threaded holes. Note the number stamped on the valve body; this is crucial for proper orientation.
- Step 2: Align the solenoid valve body so the stamped number faces the housing assembly. The valve body can be fastened in two orientations, but only one is correct.
[Important] If the stamped numbers face away from the housing, the optical air eliminator will not function properly, the port will be blocked, and the air eliminator will fail.
With the correct orientation, fasten the valve body to the housing using the two screws. Tighten to a torque of 15 to 18 in-lbs (1.7 to 2.0 Nm). - Step 3: Place the plastic bonnet over the valve body and snap it into place.
- Step 4: Place the coil over the armature guide post and fasten with the thin hex nut. Tighten the nut with a 14mm wrench to a torque of 4.5 in-lbs (0.5 Nm).
Optical Sensor Replacement
If the optical sensor needs replacement, use a 1" open end wrench to remove it from the housing. When installing a new sensor, do not exceed a torque of 75 in-lbs (8.5 Nm) to avoid damage.
A light coating of grease or anti-seize lubricant should be applied to the sensor's threads before assembly.
Troubleshooting
PROBLEM: Excessive liquid flowing out of vent to spit tank.
Scenario 1: Solenoid not closing.
- Check S3 solenoid wiring.
- Measure resistance across S3 solenoid. Value should be approximately 15 Ω. If not, replace S3 solenoid.
- Inspect S3 solenoid for blockage. Refer to disassembly instructions.
- LectroCount CPU failure. Replace CPU board.
Scenario 2: Optical Sensor not functioning.
- Check optical sensor wiring.
- Measure resistance between the RED and WHITE wires. Value should be approximately 10kΩ. If not, replace optical sensor.
- LectroCount CPU failure. Replace CPU board.
PROBLEM: No liquid flowing through meter during delivery.
Scenario 1: Liquid level not rising in optical air eliminator.
- Check S3 solenoid. It may not be opening to allow air/vapor to vent.
- Check S3 solenoid wiring.
- Measure resistance across S3 solenoid. Value should be approximately 15 Ω. If not, replace S3 solenoid.
- Inspect S3 solenoid for blockage. Refer to disassembly instructions.
- LectroCount CPU failure. Replace CPU board.
Scenario 2: Meter outlet valve not opening.
- Check wiring of the outlet valve S1 solenoid.
- Measure resistance across S1 solenoid. Value should be approximately 15 Ω. If not, replace S1 solenoid.
- Inspect S1 solenoid for blockage. Refer to the manual which accompanies the valve.
- LectroCount CPU failure. Replace CPU board.