G7 GETTING STARTED GUIDE
HARDWARE DETAILS
Cartridge (standard, single-gas or multi-gas)
LCD screen Up button OK OK button Power button Microphone (G7c only)
POWERING ON
Press and hold the power button for two seconds.

Top lights Operational and low urgency High urgency notification LiveResponse
Connectivity light

CHARGING
1. Insert the micro USB cable into the removable charging clip
2. Slide the charging clip onto the charging port at the bottom of your G7
3. A solid red light at the bottom of the device confirms your G7 is charging

Speaker

Down button Latch pull Latch push button Charging light

Charging port Removable charging clip Micro USB cable

MENU
To use G7's menu:
1. Press OK once to wake up G7's screen 2. Press OK again to enter the main menu 3. Use the up and down arrow keys to
navigate to the selection you would like 4. Press OK to select a menu item 5. Repeat as necessary

SOS ALERT

What does the green light mean?
Flashing A flashing connectivity light indicates that your G7 is storing data. It is not currently connected to the network and will send data when the light is solid.
Solid A solid connectivity light indicates that G7 is actively transmitting data and is connected to the Blackline Safety Network.
©2023 BLACKLINE SAFETY CORP Rev. 11

OK button

Up and down buttons

What is an SOS alert?
If you are in need of assistance, you can pull down the red latch to signal for help. This feature must be enabled by your safety supervisor. They can contact Blackline Customer Care for setup assistance.

G7 GETTING STARTED GUIDE | 1 2023-05-09

G7 GETTING STARTED GUIDE

GAS DETECTION
Which cartridge do I have?

Standard (no gas)

Single-gas (1 gas)

Quad-gas (2-5 gases)

Some G7 devices are equipped with a Single or Multi-gas cartridge capable of detecting gases in your environment.
If you have a Standard cartridge, you can ignore the GAS DETECTION section of the guide.

How to manually bump or calibrate:
1. Attach calibration cap to one end of the tubing
2. Attach the other end of the tubing to a fixed flow regulator on your calibration gas cylinder
3. Attach the calibration cap to the cartridge 4. Press OK to access the main menu 5. Using the up and down arrows,
navigate to gas options 6. Select OK to enter the gas options menu 7. Use the OK button to select bump test
or calibration 8. Follow the instructions on G7's screen to
complete the calibration or bump test
Tubing
Calibration cap

Single-gas G7

Multi-gas G7

G7 Dock
Bump tests and calibrations can also be completed with G7 Dock, where this process is automated for you.
To bump test or calibrate with Dock:
1. Place G7 in Dock at a 30 degree angle 2. Lower the top of G7 into Dock,
pressing until you hear a click 3. Close G7 Dock's lid 4. Select bump test or calibration from
the Dock menu on G7 5. Allow G7 Dock to run and remove G7 when the
screen prompts you

YELLOW WARNING ALARM

RED ALERT

LEARN MORE
Visit support.blacklinesafety.com for technical support and access to the following materials.  Technical User Guide  Video training tutorials  Datasheets  Certifications  Accessory products

Low warning alarm for gas If a sensor detects a low concentration gas level.
UL (under limit) If a sensor's baseline shifts and is unable to obtain an accurate gas reading.
Sensor error If a sensor stops working for any reason.
Calibration and bump test If a sensor is due for a calibration or bump test.
©2023 BLACKLINE SAFETY CORP Rev. 11

High alert for gas If a sensor detects gas levels above the maximum gas concentration.
STEL (short term exposure limit) If a sensor detects you have reached the short-term exposure limit configured.
TWA (time weighted average) If you have reached the average gas amount of repeated exposure in an eight-hour period.
OL (over limit) If excessive amounts of gas are detected and sensors can't provide an accurate gas reading.

CUSTOMER CARE
Please contact us for warranty activation, technical support, advance replacement and account setup assistance. North America (24 hours) Toll Free: 1-877-869-7212 | support@blacklinesafety.com Europe/United Kingdom (8am-5pm GMT) +44 1787 222684 | eusupport@blacklinesafety.com International (24 hours) +1-403-451-0327 | support@blacklinesafety.com