Disinfection guide

Clinical use only

This guide provides the reprocessing instructions for the AirTouch N30i, AirFit N30i and AirFit P30i masks, which are intended for multi-patient reuse in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed's validated procedures. Potable water with hardness ≤150 mg/L is recommended for all cleaning, rinsing and disinfection steps. Where local regulation impose water quality requirements for medical device reprocessing, these requirements shall be met.

Mask component¹Cleaning (Alconox)Cleaning (Neodisher Mediclean forte)Thermal disinfection (Manual)Thermal disinfection (Automated Washer-Disinfector)Validated number of cycles
Frame (AirTouch N30i)75°C for 30 minutes90 to 93°C for 1 to 10 minutes (5 to 10 minutes)15 or 30
Cushion (AirTouch N30i)75°C for 30 minutesA0 of 3000 to 11,971 (90 to 93°C for 5 to 10 minutes)30
Cushion (AirFit N30i)75°C for 30 minutesA0 of 3000 to 11,971 (90 to 93°C for 5 to 10 minutes)30
Cushion (AirFit P30i)75°C for 30 minutesA0 of 3000 to 11,971 (90 to 93°C for 5 to 10 minutes)30
Multi-hole vent75°C for 30 minutesA0 of 3000 to 11,971 (90 to 93°C for 5 to 10 minutes)30
Frame (SLM) (SLM AirFit N30i, SLM AirFit P30i)²75°C for 30 minutes90 to 93°C for 1 to 10 minutes (5 to 10 minutes)15 or 30
Frame (AirFit N30i, AirFit P30i)75°C for 30 minutes90 to 93°C for 1 to 10 minutes (5 to 10 minutes)15 or 30
Elbow body / elbow clip / elbow ring75°C for 30 minutesA0 of 3000 to 11,971 (90 to 93°C for 5 to 10 minutes)30
Headgear80°C for 10 minutesNot intended for multi-patient use30
QuietAir™ ventNot intended for multi-patient useNot intended for multi-patient use30

Mask Components

The following components are illustrated in the accompanying diagram:

Disassembly Steps

Refer to the following diagrams for disassembly:

Manual Thermal Disinfection Procedure

  1. Disassembly

    Disassemble the mask according to the instructions in the User Guide. Additionally:

    • Remove the elbow clip away from the elbow body by firmly levering one side of the elbow clip. (See disassembly step a.)
    • Pull the elbow ring out of the frame. (See disassembly step b.)
  2. Cleaning

    1. Make a solution of a mild alkaline, anionic detergent and water as per manufacturer's instructions. ResMed has validated: Alconox 1% concentration at 30-55°C.
    2. Soak the mask components for 5-10 minutes in the solution. Ensure there are no air bubbles.
    3. After the soaking time, while keeping the components in the solution, clean the inside and outside with a soft bristle brush (e.g., soft bristle toothbrush) as follows. The inside of the frame requires a soft bristle tube/bottle brush of an appropriate size to extend and contact all internal surfaces. Pay particular attention to the crevices and cavities.
      • Cushion (AirFit N30i, AirFit P30i) / elbow body: 30 seconds brushing
      • Elbow clip / elbow ring / multi-hole vent: 10 seconds brushing
      • Cushion (AirTouch N30i) / frame (SLM) / headgear: 1 minute brushing
      • Frame (AirTouch N30i): 2 minutes brushing, comprising 1 minute for the external surfaces and 1 minute for the internal surfaces
    4. Rinse the components as follows under running water at 20-60°C directly contacting all surfaces:
      • Cushion (AirFit N30i, AirFit P30i) / frame (SLM) / elbow body / elbow clip / elbow ring: 30 to 60 seconds rinsing
      • Multi-hole vent: 10 seconds rinsing
      • Cushion (AirTouch N30i) / headgear: 1 minute rinsing, while squeezing repeatedly
      • Frame (AirTouch N30i): 2 minutes rinsing, comprising of 1 minute for the external surfaces, while squeezing repeatedly, and 1 minute for the internal surfaces
  3. Inspection and drying

    1. Inspect all components to ensure they are visually clean. If not, repeat all steps in section [2. Cleaning] until all components are visually clean. If the sleeve on the frame becomes twisted, it can be repositioned when wet.
    2. Squeeze the fabric components to remove excess water.
    3. Allow the components to air dry out of direct sunlight. Drying is not required if thermal disinfection is continued immediately.
  4. Disinfection

    1. Fully immerse the mask components in a hot water bath using one of the following temperature-time combinations. Agitate the components in the water bath to ensure no air bubbles are trapped.
      • 75°C for 30 minutes, OR
      • 80°C for 10 minutes, OR
      • 90-93°C for 1 to 10 minutes
    2. Squeeze the fabric components to remove excess water.
    3. Allow the components to air dry out of direct sunlight.
  5. Inspection

    Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears, pilling etc.), the component should be discarded and replaced. Slight discolouration of the silicone and fabric components may occur and are acceptable. Disinfecting fabric components together may result in dye transfer stains and are acceptable.

  6. Reassembly

    1. Push the elbow clip onto the elbow body until it clicks into place.
    2. Insert the elbow ring into the top of the frame.
    3. Reassemble the mask according to the instructions in the User Guide.
  7. Storage and transportation

    For storage and transportation details, check the User Guide.

Automated Washer-Disinfector (WD) Procedure

  1. Disassembly

    Disassemble the mask according to the instructions in the User Guide. Additionally:

    • Remove the elbow clip away from the elbow body by firmly levering one side of the elbow clip. (See disassembly step a.)
    • Pull the elbow ring out of the frame. (See disassembly step b.)
  2. Pre-wash

    1. Soak the mask components in a 5 litre rinse bath of cold water for 1 minute. Ensure there are no air bubbles on the components.
    2. Rinse the components under cold running water for 30 seconds. Ensure internal surfaces are flushed.
  3. WD

    1. Transfer the components into an ISO 15883 series compliant washer-disinfector. Components should be orientated to ensure effective drainage.
    2. Set the washer-disinfector using the following settings:
    Phase/StageRecirculation time (minutes)Temperature (°C)Detergent
    Pre-wash4Cold-
    Wash1043-602-10 ml/l (0.2-1%) Neodisher MediClean forte
    Neutralisation643-601-2 ml/l (0.1-0.2%) Neodisher Z
    Final Rinse343-60-
    Thermal Disinfection5-1090-93-
  4. Inspection and drying

    1. Inspect all components to ensure they are visually clean. If not, repeat all steps in section [3. WD] until all components are visually clean. If the sleeve on the frame becomes twisted, it can be repositioned when wet.
    2. Squeeze the fabric components to remove excess water.
    3. Allow the components to air dry out of direct sunlight.
  5. Inspection

    Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears, pilling etc.), the component should be discarded and replaced. Slight discolouration of the silicone and fabric components may occur and are acceptable. Disinfecting fabric components together may result in dye transfer stains and are acceptable.

  6. Reassembly

    1. Push the elbow clip onto the elbow body until it clicks into place.
    2. Insert the elbow ring into the top of the frame.
    3. Reassemble the mask according to the instructions in the User Guide.
  7. Storage and transportation

    For storage and transportation details, check the User Guide.

General Warnings and Cautions

Models: F20 Full Face Mask, F20, Full Face Mask, Face Mask, Mask

File Info : application/pdf, 4 Pages, 301.58KB

PDF preview unavailable. Download the PDF instead.

608525 airtouch-n30i-airfit-n30i-p30i disinfection-guide row eng

References

Adobe PDF Library 17.0 Adobe InDesign 19.2 (Windows)

Related Documents

Preview ResMed CPAP Prescription Form
A prescription form for ResMed CPAP machines and accessories, detailing patient information, diagnosis, prescribed therapy modes, mask interface, and accessories. Includes important disclaimers regarding medical advice and insurance.
Preview ResMed AirTouch N30i & AirFit N30i/P30i User Guide
Comprehensive user guide for ResMed's AirTouch N30i, AirFit N30i, and AirFit P30i CPAP masks. Learn about mask components, intended use, fitting, cleaning, reassembly, and technical specifications for optimal sleep therapy.
Preview ResMed AirFit F30i and AirFit X30i CPAP Mask Disinfection Guide
This guide provides essential reprocessing instructions for ResMed AirFit F30i and AirFit X30i CPAP masks intended for multi-patient use in clinical settings. It details cleaning, thermal disinfection, inspection, and reassembly procedures to ensure safe and effective mask reuse.
Preview ResMed AirTouch N30i / AirFit N30i / P30i Mask User Guide
Comprehensive user guide for ResMed AirTouch N30i, AirFit N30i, and AirFit P30i CPAP masks. Learn about mask components, fitting, cleaning, and technical specifications.
Preview ResMed AirFit N30i & P30i User Guide
Comprehensive user guide for ResMed AirFit N30i and P30i nasal and pillows CPAP masks, detailing intended use, fitting, cleaning, and technical specifications.
Preview ResMed AirFit N30i & P30i CPAP Mask Sizing Guide
Find the perfect fit for your ResMed AirFit N30i nasal mask or AirFit P30i nasal pillows mask with this easy-to-follow sizing guide. Learn how to use the printable template and frame fitting measurements for optimal comfort and therapy.
Preview ResMed AirFit N30i and P30i User Guide
Comprehensive user guide for ResMed AirFit N30i and P30i masks, detailing intended use, fitting, adjustment, cleaning, reassembly, and technical specifications for CPAP and bi-level therapy.
Preview ResMed AirFit N30i and P30i CPAP Mask User Guide
This user guide provides instructions for the ResMed AirFit N30i and P30i nasal CPAP masks, including fitting, cleaning, maintenance, and troubleshooting.