Instruction Manual
Tangential Flow Filtration System
Model No.
Tanfil 100
Video List
1. Important Safety Precautions
Before using this instrument, users must read this instruction manual thoroughly. This instrument must not be modified in any way. Any unauthorized modifications will void the warranty and may cause safety hazards. The company is not responsible for any machine damage or personal safety caused by unauthorized modifications.
- Connect to the correct power source according to the instrument's rated voltage.
- Before use, check the chemical compatibility of the flow path material and the medium to be used.
- Keep the instrument away from flammable materials. Do not place it near any flammable or explosive materials. Install and operate the instrument in a clean, dust-free, and well-ventilated environment.
- During operation, if the pressure is high or the viscous liquid causes a heavy load on the motor module, the motor may stop for self-protection after prolonged overload. In this case, turn off the power for five minutes before turning it back on to continue operation.
- After each use, clean the tubing and exterior according to the maintenance method.
- If you have any questions about product usage, please contact the maintenance personnel. Do not disassemble or replace parts improperly.
- Dispose of packaging materials in accordance with local regulations.
- For more detailed product information, please refer to the latest catalog and official website.
9. Operating Conditions
- Ambient Temperature: 5 ~ 50℃
- Relative Humidity: 80% RH Max.
- Power Supply: 100-240V~, 50/60Hz, 50W
- Fuse: T1.5A, 250V
- Altitude: Maximum 2000 meters
- Pollution Degree: II
- Indoor use only
Caution: The pressure inside the tubing must not exceed 40 psi, otherwise the silicone tubing may expand due to excessive pressure.
Caution: Before replacing the fuse, disconnect the power supply.
2. Pre-Assembly Inspection
Before unboxing this product, please ensure that the packaging box is not damaged. If there are any issues, please retain the serial number and packaging box, and contact your local distributor or contact us for service.
Item | Content | Item | Content |
---|---|---|---|
1 | Tanfil 100 Main Unit | 6 | Concentration Bottle (with PP filter cap assembly) |
2 | Support Base Assembly | 7 | Pressure Manifold Module (Upper and Lower, with pressure gauge) |
3 | Luer Connector Accessory Kit | 8 | Cassette Support Bracket |
4 | Power Cord | 9 | Support Base Mount x1 |
5 | Magnetic Stirrer | 10 | Instruction Manual |
3. Main Unit Introduction
(1) Description of Main Unit Components
A diagram shows the main unit with numbered components. The corresponding numbers and descriptions are listed below:
Item | Content | Item | Content |
---|---|---|---|
1 | Main Unit Body | 10 | Concentration Bottle |
2 | Operation Panel Area | 11 | Concentration Bottle Cap |
3 | Pump Head | 12 | Pressure Regulating Clamp |
4 | Thin Magnetic Stirrer | 13 | Return Pressure Gauge (Upper) |
5 | Inlet Pressure Gauge (Lower) | 14 | Support Base Mount |
6 | Support Rod | 15 | Support Base |
7 | Return Pressure Manifold (Upper) | 16 | SPDT Power Switch |
8 | Cassette Support Bracket | 17 | Power Socket with Fuse (Input) |
9 | Inlet Pressure Manifold (Lower) | 18 | Electromagnetic Stirrer Power Socket (Output) |
(2) Description of Accessories, Luer Connectors, and Silicone Tubing
Diagrams and descriptions of various accessories:
- Three-way Valve x4
- Male-to-Male Luer Connector x2
- Male Luer to 1/8" Barb Connector x12
- Female Luer to 1/8" Barb Connector x1
- Female Luer Cap x2
- Tubing Clamp x13
- Tubing ULTRA-C (1.5m) x1
- Return Pressure Manifold Module (Upper) x1
- Inlet Pressure Manifold Module (Lower) x1
- Support Base Mount x1
- Cassette Support Bracket x1
- Hex Wrench x1
- Pressure Regulating Clamp x1
(3) Material of Each Component
Name | Material | Name | Material |
---|---|---|---|
Concentration Bottle | PES | Pressure Gauge and Liquid Contact Parts | SS316 Stainless Steel |
Concentration Bottle Cap | PP | Pressure Manifold | PP |
Concentration Bottle O-ring | Silicone | Luer Connector | SS316 Stainless Steel, PP |
Magnet | PTFE Coated | Luer Connector O-ring | NBR |
Three-way Valve | PC, POM | Tubing | Silicone (Platinum Catalyzed Polymerization), TPE |
Capillary Tube | PTFE |
4. Operation Panel / Buttons / Symbols / Screen Description
(1) Operation Panel Description
A diagram of the operation panel with numbered controls:
Item | Content |
---|---|
1 | Screen |
2 | Function Switching Button |
3 | Up/Down Buttons |
4 | Start/Stop Button |
(2) Button Description
Button Symbol | Name | Description |
---|---|---|
▲ | Up Button | Increase time or speed (hold for fast forward) |
▼ | Down Button | Decrease time or speed (hold for fast forward) |
MODE | Function Switching Button | Switch between speed, time, or direction settings |
START/STOP | Start/Stop Button | Start/End operation (Hold to enter fast cleaning mode) |
(3) Indicator Symbol Description
Indicator Symbol | Name | Description |
---|---|---|
(Arrow indicating direction) | Pump Head Direction Indicator | Clockwise/Counter-clockwise pump head operation |
(Arrows indicating direction) | Count Down/Count Up Operation | Timing countdown/count up |
(Arrows indicating direction) | Fast Cleaning Mode | Fast cleaning, direction change |
(4) Screen Description
Diagrams of the operation screen showing different modes:
- Screen 1: Shows speed (100rpm), time (23min), and direction indicator.
- Screen 2: Shows speed (600rpm) and time (23sec).
Item | Name | Item | Name |
---|---|---|---|
1 | Speed Setting (rpm) | 4 | Count Up/Down Timer |
2 | Pump Head Direction | 5 | Fast Cleaning Mode |
3 | Time Setting (sec/min) |
5. Installation Instructions
(1) Support Base Assembly
Install the support base onto the main unit casing. Use the hex wrench to remove the two internal hex screws, then secure the support base. Ensure the support rod is oriented correctly, with the machined surface facing backward.
A diagram illustrates the installation of the support base.
(2) Pressure Manifold Module Installation
a. Insert the inlet pressure manifold module (lower) into the support bracket at the lowest position. Ensure the gauge faces forward and tighten the hand-tightening screws at the rear.
b. After installing the support base mount into the cassette support bracket, slide it to the middle height of the support rod and tighten the hand-tightening screws at the rear.
c. Insert the return pressure manifold module (upper) into the upper side of the support rod and tighten the hand-tightening screws at the rear.
Diagrams show the installation steps for the pressure manifold modules.
(3) Magnetic Stirrer and Concentration Bottle Installation
Place the magnetic stirrer into the recess on the top cover of the main unit. The power plug can be connected to the electromagnetic stirrer power socket (output) on the rear of the main unit. Install the disc filter onto the concentration bottle cap first, and then place it above the magnetic stirrer.
Diagrams show the placement of the magnetic stirrer and concentration bottle.
(4) Three-way Valve Installation
Refer to Section 3, Main Unit Introduction, items 7, 9, and 10. Install the three-way valves at the bottom of the concentration bottle, in front of the inlet pressure manifold (lower), and to the left of the return pressure manifold (upper). Remove the male Luer caps before installation.
Diagrams illustrate the three-way valve installation.
(5) Tubing Installation
a. Cut the silicone tubing to different lengths and install Luer connectors as shown in the diagram and table below.
Name | Silicone Tubing Model/Color | Length | Connector |
---|---|---|---|
Peristaltic Pump Tubing | Pharmed/White | 50 cm | Double Male Luer to 1/8" Barb |
Return Tubing | Ultra-C/Transparent | 15 cm | Double Male Luer to 1/8" Barb |
Concentration Tubing | Ultra-C/Transparent | 15 cm | Single Male Luer to 1/8" Barb |
Collection Tubing | Ultra-C/Transparent | 20 cm | Single Male Luer to 1/8" Barb |
Drain Tubing | Ultra-C/Transparent | 20 cm | Single Male Luer to 1/8" Barb |
Filtration Tubing (Note) | Ultra-C/Transparent | 20 cm | Single Female Luer to 1/8" Barb |
Cleaning Tubing | Ultra-C/Transparent | 15 cm | Double Male Luer to 1/8" Barb |
Rinsing Tubing | Ultra-C/Transparent | 20 cm | Single Male Luer to 1/8" Barb |
Notes:
- The return tubing and filtration tubing can be adjusted with Luer connectors according to the actual cassette specifications.
- After the tubing installation is complete, install the pressure regulating clamp onto the concentration tubing.
b. Location of each tubing connection:
Diagrams show the tubing connections for cleaning/rinsing and operating states.
6. Tubing Operation Instructions
This system has multiple operating stages. Please refer to the following instructions for setting up the tubing. For cassette-related cleaning steps and usage of solutions, please refer to the operation manuals of each brand.
(1) Cleaning
A diagram illustrates the tubing setup for cleaning.
a. For initial operation, the entire system must be cleaned. Connect the cleaning tube between the return pressure manifold (upper) and the inlet pressure manifold (lower), as shown in Figure 1.
b. Open the concentration bottle cap, pour in the cleaning solution, and close the bottle cap.
c. Refer to Figure 1 for the three-way valve and flow path settings.
d. Start the peristaltic pump and slowly increase the speed to 50 rpm.
- Check the tubing and connectors for leaks. If leaks are found, tighten the connectors.
e. Allow the system to circulate for 15-30 minutes, then stop the peristaltic pump.
(2) Drainage and Sample Collection
Diagrams illustrate the tubing configurations for drainage and sample collection.
a. With the peristaltic pump stopped, set the three-way valve direction as shown in Figure 2.
b. Open one of the airtight caps on the concentration bottle. The liquid in the concentration bottle will be collected into a container via the collection tubing due to gravity.
c. If the system tubing needs to be drained after the experiment, stop the peristaltic pump, adjust the three-way valve direction, remove the filtration cassette, and connect the cleaning tube, as shown in Figure 3.
d. As shown in Figure 4, adjust the three-way valve direction and start the peristaltic pump to collect all the solution in the system tubing into the concentration bottle, then stop the peristaltic pump.
e. As shown in Figure 5, adjust the three-way valve at the bottom of the concentration bottle, open one of the airtight caps on the concentration bottle, and the solution will naturally drain out through the collection tubing by gravity.
f. If gravity drainage is insufficient, adjust the three-way valve and flow path as shown in Figure 6, start the peristaltic pump, and drain the solution from the drain tubing using the pump's power.
- After initial cleaning, the tubing can be removed for high-temperature sterilization as needed.
(3) Cassette Rinsing
A diagram shows the tubing setup for cassette rinsing.
a. Adjust the three-way valve direction and flow path as shown in Figure 7. Pour the rinsing solution into the concentration bottle.
b. Start the peristaltic pump and slowly increase the speed until the inlet pressure reaches the recommended pressure for the cassette.
Note: Check the tubing for leaks. If leaks are found, tighten the connectors.
c. Continuously circulate the rinsing solution. Refer to the operation manual of each cassette brand for the recommended rinsing solution volume.
II. Cassette and System Rinsing
a. Adjust the three-way valve direction and flow path as shown in Figure 8.
b. Start the peristaltic pump and slowly increase the speed until the inlet pressure reaches the recommended pressure for the cassette.
Note: Check the tubing for leaks. If leaks are found, tighten the connectors.
c. Adjust the pressure regulating clamp and peristaltic pump speed until the return and inlet pressures reach the recommended pressures for the cassette.
d. Continue circulating until the concentration reaches the desired level. Stop the peristaltic pump and refer to (2) Drainage Steps to drain the solution from the tubing into a container.
- Before rinsing the cassette, rinse the cassette first. Refer to the operation manual of each cassette brand for rinsing methods.
(4) Pre-treatment and Concentration
I. Pre-treatment
a. Adjust the three-way valve direction and flow path as shown in Figure 9.
b. Pour the buffer solution into the concentration bottle, start the peristaltic pump, and slowly increase the speed. Refer to the recommended cassette values for inlet pressure, and ensure the tubing is filled with solution to avoid air bubbles.
Note: Check the tubing for leaks. If leaks are found, tighten the connectors.
c. When the liquid level in the concentration bottle reaches the bottom, stop the peristaltic pump. Remove the filtration tubing, pour out the liquid in the tubing, and reinstall the cassette to complete the pre-treatment before the concentration experiment.
II. Concentration
a. Pour the sample into the concentration bottle, start the peristaltic pump, and refer to the recommended cassette values for inlet pressure.
b. Adjust the pressure regulating clamp and peristaltic pump speed until the inlet and return pressures reach the recommended cassette values.
Note: Ensure the inlet pressure does not exceed the recommended cassette value, and the return pressure does not exceed 40 psi.
c. Continue circulating until the desired concentration is reached, then stop the peristaltic pump.
d. Refer to (2) Drainage Steps to collect the sample from the system into a container.
(5) Continuous Sampling
If the sample volume to be filtered exceeds the concentration bottle capacity of 500 mL, the excess sample can be placed in another beaker. Use a silicone tube, with one end inserted into the septum on the concentration bottle cap and the other end placed in the beaker. As shown in Figure 10. During the filtration process, the concentration bottle will gradually become negative pressure, drawing the sample from the beaker into the concentration bottle for continuous sampling.
7. Cassette Usage and Maintenance Steps
This section usesModels: Tanfil 100, 184100-01 02, 184100-20, 184100-40, 184100-62, 203100-01, 203100-10
File Info : application/pdf, 28 Pages, 1.50MB
DocumentDocumentReferences
適用於 Microsoft 365 的 Microsoft WordRelated Documents
![]() |
ROCKER Tanfil 100 TFF Tangential Flow Filtration System - Product Overview Discover the ROCKER Tanfil 100, a versatile Tangential Flow Filtration (TFF) system designed for efficient concentration, diafiltration, separation, and purification of biomolecules and particles like proteins, exosomes, and antibodies. Key features include reduced membrane clogging, easy scale-up, and simple setup, backed by CE and RoHS certifications. |
![]() |
Rocker 2023-2024 Product Guide Explore the comprehensive 2023-2024 Product Guide from Rocker, featuring laboratory pumps, filtration systems, and solutions for food analysis and water analysis. Discover innovative products designed to make filtration easy and enhance laboratory efficiency. |
![]() |
Rocker STREAM Continuous Filtration System: Features, Specifications, and Ordering Detailed information on the Rocker STREAM Continuous Filtration System, including its features for effortless and user-friendly filtration, chemical resistance, applications in solvent purification and large volume filtration, technical specifications for 300 ml and 1000 ml models, and ordering details for components and compatible accessories from Rocker Scientific. |
![]() |
Rocker Laboratory Filtration Accessories: Bottles, Adaptors, Holders & Filters Explore Rocker's comprehensive range of laboratory filtration accessories, including PES/PC vacuum bottles, multilayer filtration adaptors, hoop holders, membrane filters, and GL45 adaptors and suction caps. Featuring products from Pall and Whatman. |
![]() |
Rocker MultiVac Manifold and Filtration System Instruction Manual This instruction manual from Rocker Scientific provides detailed guidance for the MultiVac Manifold and Filtration System. It covers installation, operation, maintenance, and troubleshooting for various models including MultiVac 310/610-MS, MultiVac 300/600-MS, MultiVac 300/301/600/601-MB, and their T/A variants, often used with Alligator 200 or Rocker 400 vacuum pumps. Essential for laboratory use, it ensures proper setup and usage of these filtration tools. |
![]() |
SolarVac Flexible Filtration Systems by ROCKER Explore ROCKER's SolarVac flexible filtration systems, featuring rotary bases for various laboratory filtration needs. Compatible with multiple filter cup and holder brands. |
![]() |
Rocker BioWasher 100 Gravity-fed Microplate Washer Instruction Manual This instruction manual provides comprehensive guidance on the installation, operation, maintenance, and troubleshooting of the Rocker BioWasher 100 and BioWasher 100 - T gravity-fed microplate washers, designed for laboratory use. |
![]() |
Rocker Oil Free Vacuum Pump Guide: Features, Specifications, and Performance Comprehensive guide to Rocker oil-free vacuum pumps, detailing features like compact design, overheat protection, and overflow protection. Includes specifications and performance graphs for models like Rocker 300, 400, 900 series. |