Introduction
Thank you for purchasing the HIOS torque meter. The HM series torque meters are carefully designed for exclusive use in automatic controlled systems. This series can measure torque without removing the screwdriver mounted on the system. Please read the following notes carefully before use.
Features
- Measure screwdriver torque as it is mounted on the automatic controlled system, without removing the bit.
- Compact detector allows easy measurement even in small spaces.
- Digital display provides easy and precise reading.
- Instrument saves and indicates peak values for precise measurement.
- Rechargeable and portable for easy carrying.
- Can use commercially available tools for data collection and analysis.
- Can measure the torque of a motor screwdriver for manual use (requires HIOS Fidaptor and a conversion plug).
- Analog outputs available for waveform observation, recording, or measurement result judgment (analog cord is optional).
! Caution
Read the notes below carefully for safe and proper use.
Before using this instrument
- Carefully read this operation manual and attached notes before use.
- The torque meter (display unit) and detector are calibrated with the same serial number; always use the proper combination.
- HIOS is not responsible for malfunctions caused by unauthorized modification, disassembly, or handling other than specified in the manual.
Cautions at work
- When measuring a rotational device, be aware of surroundings to prevent involvement in rotation.
- Stop operation immediately if anything abnormal is detected. Wear proper working clothes and close cuffs, buttons, and zippers.
- Do not wear gloves as they may be slippery.
- Fix the small detector firmly to avoid swinging.
- Do not throw or hit the detector, as this may cause malfunction.
- The detector's strain gauge is sensitive to the environment; set to [TRACK] mode at power-on, confirm zero is displayed, and adjust zero occasionally. (*1)
Notes for use
- Never apply torque exceeding the maximum displayed value, as it may break the instrument.
- Refer to page 5 for correct handling of Fidaptor.
- Do not use attachments other than Fidaptor and specified attachments.
- Do not use with devices that give repeated shocks, such as air or impact wrenches. (*2)
- When attaching Fidaptor to the socket, fix it with screws in four directions. (*3)
- Do not loosen socket fixing screws, as it may cause precision errors.
- Refer to page 12 for correct handling of the charger.
- Do not connect devices other than those specified to the data output connector.
- When inserting or removing connectors, hold the connector head and check the pin arrangement.
- Always turn off the power after use.
- Do not hit or apply load to the display plate.
- Do not change internal volume for calibration.
- Do not handle the detector roughly or drop it.
- Avoid using the detector in inappropriate places: where fluids may be scattered, where vibrations, dust, or heat exist, outdoors with electric noises, or in high temperature/humidity (15°C-35°C, 25%-65% humidity).
- Do not store the instrument where temperatures or humidity change significantly, as condensation may cause damage.
Important
- *1: Before zero adjustment, set to TRACK mode and check if "zero" is displayed.
- *2: For air or impact wrenches, use the HIT series; consult HIOS.
- *3: Fix Fidaptor to the socket at four points (see pages 6 and 16).
- *4: Turning the zero adjustment knob in PEAK mode disables the reset function. Switch to TRACK mode, confirm zero, then perform "Peak" measurement. (Refer to Section 7.8 on page 7.)
Parts Name
Diagram of the HIOS HM-100 Display Unit showing labeled parts: Printer Output connector, Display Panel, Power Switch (ON/OFF), Reset Button, Zero Adjustment Knob, Mode Switch (PEAK/TRACK), Socket, Analog Output Connector, Buzzer Set, and connection port for Charger. The back of the display unit shows the Serial No. sticker. The Detector is shown connected via a Cable (DPC-0506, 1.7m) with two connectors, one covered by a cap. The Unit Converter Switch is shown with labels for Lbf·in, kgf·cm, and N·m.
Mode Switch
- TRACK: The display value changes with the load to the detector. It disappears when no load is applied.
- PEAK: Captures the maximum load value. It disappears when the reset button is pressed.
Buzzer Set
The buzzer sounds when the torque reaches the specified torque.
Display panel (unit)
Conversion of N·m: 1 N·m ≒ 10.2 kgf·cm
Model | N·m | N·cm | N·m | N·cm | kgf·cm |
---|---|---|---|---|---|
HM-100 | 0.00 | 0 | 10 | 1000 | 102 |
HM-10 | .000 | 0.0 | 1 | 100 | 10.2 |
9.81 | 981 | 100 | |||
0.981 | 98.1 | 10 |
Items included with the torque meter
Both the main unit and accessories are packed in a dedicated aluminum attaché case.
Model | Fidaptor | Fidaptor Screw for head replacement | Fidaptor spring | Charger | L-shaped wrench | Others (1 piece each) |
---|---|---|---|---|---|---|
HM-10 | P/N: TEM 26-Z Yellow spring (1 piece) |
5 pcs Phillips head pan screws, M2.6 x 6mm 2 pcs hollow set screws, M3 x 6mm |
- | AC100V P/N TCH-100N (1 piece) |
Opposite side dis.: 1.5mm | Cord for detector P/N: DPC-0506 Inspection report |
HM-100 | P/N: TEM40-Z Black spring (1 piece) |
5 pcs Phillips head pan screws, M4.0 x 8mm 2 pcs hollow set screws, M4.0 x 6mm |
Measurement range: 0.15-0.6 N·m Yellow spring (1 piece) |
Opposite side dis.: 1.5mm, 2.0mm |
Fidaptor list
Model | P/N | Measurement range (N·m) | Supporting bit No. | Head screw diameter |
---|---|---|---|---|
HM-10 | Standard accessory TFM 26-Z 0.15-0.6 |
Optional accessory (separately sold) TFM 20-Z 0.25 or smaller |
#+1 M2.6 (P/N: SPP26×060SUS) |
#+0 M2.6 (P/N: SPC26×060) |
HM-100 | Standard accessory TFM 40-Z 0.5-3 |
#+2 M4.0 (P/N: SPP40×080SUS) |
About Fidaptor (accessories)
Use Fidaptor for torque measurement of the screwdriver for an automatically controlled system.
Diagrams showing how to attach Fidaptor for #1 and #2 bits. Includes a note: "When the screw head on the threaded shaft wears, measurement results vary. Please replace it as soon as possible."
Replacement of the head screw (preferably replace it within 10 operation cycles.)
Use the attached L-wrench to loosen the hollow set screws (M3) before the head shape collapses.
Diagram showing the use of an L-wrench to lock the rotation of the hollow set screw and unfasten the head screw. Suitable wrench size: 1.5mm for HM-10, 2.0mm for HM-100. Spare screws are included.
Operation procedure
How to check torque using Fidaptor
- First, connect the torque meter (HM) to the detector. Then turn on the power and confirm if the battery has been charged.
- How to check the battery: Turn on the power. Set the measurement mode to TRACK. If the battery is low, "LOBAT" is displayed in the upper left corner of the display. Then, charge the battery (do not charge longer than eight hours) using the provided chargers. (Note) Do not take measurements while charging.
- Fix the detector. There are two points for fixing (see dimensional drawing on page 16).
- Connect the cord to the detector. Use one of the two connectors, considering workability. Ensure the connector shape is correct to prevent pin bending. (Note) Use one connector and cover the other with a cap.
- Use a handy screwdriver to loosen the threaded shaft ring of Fidaptor slightly. Loosen the shaft so the spring can be turned with fingers; the shaft should always return to the same position.
- Set Fidaptor to the socket.
- 5-1 Fix Fidaptor to the socket: Use the L-shaped wrench (1.5mm) to fasten the four hollow set screws (M3.0) on the side of the socket with even torque for stable measurement.
- 5-2. Set the bit end of the target screwdriver to the center of the head screw of Fidaptor.
Notes when fixing Fidaptor:
- Fix Fidaptor by fastening the four screws on the side evenly to ensure the bit is centered in the head screw.
- The bit tip should be engaged precisely in the head screw of Fidaptor for measurement, aligned with the thrust direction and screw center.
- When used in an automatically controlled system, it should be aligned with the center of the Z-axis.
- Set the switch for measurement unit change to your desired unit (N·m, lbf·in, kgf·cm).
- Set the mode switch to TRACK and turn the zero adjustment knob for zero adjustment. Check if "000" is displayed. (Note) Zero adjustment is not available in PEAK mode.
- Set the mode switch to PEAK. After measurement, clear the value with the RESET button.
- Turn the screwdriver for measurement. When the clutch disconnects, the spring compresses, and the measurement is displayed and saved. This is the output torque.
- Let Fidaptor recover from contraction to return to constant height. Use a handy screwdriver to loosen the threaded shaft. (Note) Always let the spring recover height by loosening.
The target looseness is that the spring can be turned with fingers to always have a constant height.
- Cancel the value in the window by pushing the Reset button.
Note for resetting:
- Let Fidaptor return to constant height before pressing RESET.
- Ensure the socket section has no load; it cannot be reset if there is a load.
- If the measurement result is high torque, the value may remain. Press RESET again if necessary.
- Repeat steps 10 to 12 to determine average output torque. Adjust the screwdriver's nut for torque adjustment if needed.
- When finishing measurements, always loosen Fidaptor. Turn off the power and remove everything from the socket at the end of the day. (Note) Loosen the Fidaptor spring and store it properly.
- For aggregation of measurement data, turn the mode switch to PEAK. An input signal (torque = load) of 15 digits or larger is required. Pressing the Reset button outputs data and cancels the display value. (Refer to page 17 for data output tools.)
- About "digit": "digit" means the unit of a value displayed regardless of the decimal point. (Examples: 0.01 = 1 digit, 0.025 = 25 digits, 0.10 = 10 digits, 1.25 = 125 digits)
How to set the buzzer
Diagram showing the buzzer set knob and zero adjustment knob on the HIOS HM-100. Notes state the buzzer is set to 0.981 N·m (HM-10) and 9.81 N·m (HM-100) by default.
Setting procedure
- Turn on power.
- Set to TRACK mode.
- Use the zero adjustment knob to set the value.
- Turn the buzzer adjustment knob to set the desired buzzer start value. The buzzer sounds when the displayed value reaches this point. Stop turning when the buzzer is heard.
- Confirm the value.
- Set zero on display using the zero adjustment knob.
- Measure in PEAK mode and check the value after the buzzer.
Measurement after setting the buzzer
- Apply load to the detector.
- When the load reaches the preset value, the buzzer starts.
- In TRACK mode: Buzzer stops when load is released; "zero" is displayed.
- In PEAK mode: Display is retained with no load, but buzzer stops beeping.
If the buzzer setting is not used, turn the buzzer set knob to a value higher than the working value (default setting).
How to replace the spring (attached to HM-100)
HM-100 has two springs: black (strong) and yellow (weak). A black spring is attached by default. For measuring a smaller torque range, replace the black spring with the yellow one.
(Note) Set the screwdriver bit to the Fidaptor's head screw and turn it counterclockwise. The threaded shaft will then loosen, allowing spring replacement.
Diagram showing the springs and the process of replacing them.
Inspection of Fidaptor
Handling of Fidaptor
- Never apply load exceeding Fidaptor's measurement capability.
- Use Fidaptor appropriate for the screwdriver and torque.
- Always keep Fidaptor loose and remove from the socket after measurement.
- Keep screwdriver and Fidaptor upright during measurement, applying thrust load of 5kg or smaller (2kg or smaller for lower torque ranges).
- For continuous measuring, apply grease to Fidaptor component parts.
- Set measurement cycle to 5 seconds or longer to prevent rapid wear of component parts.
- Never leave or store Fidaptor in a fastened state; loosen the spring when not in use.
- Correctly engage the object for measurement with Fidaptor.
- Do not use a deformed or modified Fidaptor.
Maintenance and inspection of Fidaptor
- Apply grease (*) to components (1), (2), and (5) regularly.
- Before use, check components: (1), (2), and (5) must be greased; threaded shaft must not be bent or worn; threaded shaft must be free from foreign materials.
- Fidaptor is a consumable component; inspect and replace as needed.
Diagram of Fidaptor components (1) Threaded shaft, (2) Bearing set, (3) Bearing housing, (4) Spring, (5) Plate nut, with a replacement guide: Components (1) and (2) every 2,500 cycles; Component (1) if bent or worn; Component (4) with (1); Component (5) every 5,000 cycles. Grease is available from HIOS (P/N: TF-3G).
How to use the socket
This instrument can measure various torques by changing attachments. When using a customized attachment, observe the following:
- No play between socket and attachment.
- Attachment must have sufficient strength to prevent breakage.
- Vertical load to socket: 2kg or smaller for HM-10, 5kg or smaller for HM-100.
- No impact load to socket.
- No violation to torque measuring instrument principles.
For socket details, refer to the outline view on page 16.
Battery charge
Always use the dedicated charger and turn off the main unit power before charging. The first charge takes eight hours.
! Caution
Please read the following carefully:
- Do not charge the battery longer than eight hours.
- Use only NiMH batteries; no other types, even from HIOS Corporation.
- Do not use the instrument while charging.
- When "LOBAT" is displayed, stop measurement and charge the battery.
- Do not use the dedicated charger for other purposes.
- Do not place heavy items on the charger cord, or bend/bind it.
- Always turn off power when connecting or disconnecting the battery.
- Do not remove the battery from the main unit.
! Danger
Charging the battery for more than eight hours may cause heating, explosion, deterioration, or fire.
How to charge the battery
- Turn off the power switch and connect the charger's plug to the connector.
- When charging is complete, disconnect the plug and turn on the power switch to check the display.
Customer service
Repair
- Service charges apply for repairs under the following circumstances:
- Malfunctioning or damage from incorrect use, disassembly, or attempted repair.
- Oil added to socket, switches, or inside the instrument.
- Damage during shipping, dropping, etc.
- Damage from fire, gas exposure, earthquake, water, irregular power supply, or other disaster.
- Calibration, inspection, or parts replacement for Fidaptor, etc.
- No charge for inspection/calibration of the same part within three months of previous service (except for conditions 1-4 above).
Shipping and handling charges for repair service must be paid by the customer. Contact HIOS Corporation or your HIOS dealer for customer service questions.
Attention
The product contains a rechargeable battery, which is recyclable. Disposal into municipal waste streams may be illegal. Check with local solid waste officials for recycling options or proper disposal.
Recycling symbol and "Ni-MH" text.
Troubleshooting
Before deciding the instrument has failed, refer to the table below. If problems persist, contact your dealer or HIOS.
Symptom | Possible cause | Action to take |
---|---|---|
"LOBAT" in the display | Insufficient charging of the battery | Charge the battery (Refer to P. 12). If no display after charging, contact HIOS. |
Nothing appears on the display | Instrument not used for a long time, or battery died. | Turn off power, charge for 10 minutes, then turn on. If display activates, charge for ordinary use (8 hours). If problem persists, contact HIOS. |
Cannot set it to zero | Reset does not work if display value exceeds 10 digits without load. Mode switch is set to TRACK. Zero adjustment not done. |
Set Mode switch to TRACK and check for zero display (perform zero adjustment in TRACK mode occasionally). Set it to PEAK. Use the zero adjustment knob for adjustment. |
The value cannot be held. | Connection cable wrongly connected or broken. | Replace the cable. |
The instrument cannot be charged | Charger plug not inserted correctly. Plug connected to wrong terminal. |
Check if plug is correctly inserted. Connect to the right connector. If problem persists, contact HIOS. |
After charging, still displays insufficient charging. | Battery died. Still insufficient charging. |
Repair is required. Charge battery again, not exceeding eight hours. |
An irrelevant value is displayed | Noises cause wrong value display (when power is on or in PEAK mode). | Press the Reset button to delete the value. |
Specifications
Model | HM-10 | HM-100 |
---|---|---|
Peak measuring range (N·m) | 0.015-1.000 | 0.15-10.00 |
Peak measuring range (lbf·in) | 0.15-9.00 | 1.5-90.0 |
Peak measuring range (N·cm) | 1.5-100.0 | 15-1000 |
Accuracy | Within ±0.5% (F.S.) | |
Battery pack | 6V NiMH | |
Charging time | 8 hours or less | |
Weight (kg) | Display unit: 1.0kg Detector: 0.35kg |
Display unit: 1.0kg Detector: 0.35kg |
Continuous operating time on a full charge | 8 hours | |
Battery life | 300 cycles of charging | |
Detector cord | 1.7m (6P cord), P/N: DPC-0506 | |
Exclusive battery charger | Input: AC100V, 120V, 220-240V Output: DC7.2V 120mA (P/N: TCH-100N) |
Never apply load more than the maximum allowable load.
Battery life may vary depending on usage pattern.
Instruments for foreign countries are available (input: AC120V, 220-240V).
Outline dimensional drawing (and detailed socket dimensions)
Outline dimensional drawings for the "Main unit" and "Detector" of the HIOS torque meter, showing dimensions in millimeters (mm). The detector drawing includes detailed socket dimensions and notes that the drawing is not to full scale.
Specifications for analog output
Output is about 0.72 V at maximum torque (9.81N·m for HM-100; 0.981N·m for HM-10).
For waveform observation, devices like oscilloscopes, voltage meters, or data loggers may be needed. A dedicated analog data output cord (P/N: HP-8060, 1.5m) is also required. Use devices after reading their manuals.
Data output
To transfer data to a PC, use the input tool of Mitutoyo. A connection cable (sold separately) is required.
Types of input tools
- USB keyboard conversion type, P/N: IT-020U
- Connection cable: 06AGF590, 5 pins, 2m
Contact Mitutoyo or your dealer for input tools and printer connection cables. Contact HIOS for import of measurement data.
Specifications for serial output
1. Connector pin arrangement: Mitutoyo MQ65-5P
- ① GND: Ground
- ② DATA: Data output format specified below.
- ③ CK: Clock
- ④ RD: Request for data
- ⑤ REQ: Request of data output from the outside
- ① to ④: Open drain; -0.3 to +7V (400µA max.)
- ⑤: Pulled up to VDD (1.55V).
2. Data output format
13 digits are output in the sequence: d₁ d₂ d₃ d₄ d₅ d₆ d₇ d₈ d₉ d₁₀ d₁₁ d₁₂ d₁₃ (Data type, Sign, Entry number, Decimal point position, Measured value). Each digit is expressed with a 4-bit binary, output from LSB (2⁰→2¹→2²→2³).
3. Timing chart
MIN | MAX | UNIT | |
---|---|---|---|
T₀ | 2 | - | sec |
T₁ | 0.2 | 0.4 | sec |
T₂ | 0.2 | 0.4 | mS |
T₃ | 0.5 | 1 | mS |
T₄ | 0.2 | 0.4 | mS |
CHINA RoHS2
The following table is for CHINA RoHS2. If asked by China Customs, please show this table.
部件名称 (Component Name) | 有害物质 (Hazardous Substances) | |||||
---|---|---|---|---|---|---|
铅(pb) | 汞(Hg) | 镉(Cd) | 六价铬 (CR (VI)) | 多溴联苯 (PBB) | 多溴二苯醚 (PBDE) | |
充电池 (Battery) | ○ | ○ | ○ | ○ | ○ | ○ |
外壳 (Housing) | ○ | ○ | ○ | ○ | ○ | ○ |
电源适配器 (Power Adapter) | X | ○ | ○ | ○ | ○ | ○ |
This table is prepared according to SJ/T 11364. ○: Hazardous substance content in all homogeneous materials of the component is within the limit specified by GB/T 26572. ×: Hazardous substance content in at least one homogeneous material of the component exceeds the limit specified by GB/T 26572.
In addition, China RoHS marks are required on the product and product box. If the mark cannot be found, please ask your distributor. In case of emergency, cut the mark below and stick it on the product and product box.
China RoHS mark (two circular logos with '10' inside).