Fluke 374 FC/375 FC/376 FC/902 FC Clamp Meter Calibration Manual
February 2017 Rev. 1, 9/19
Introduction
⚠️ Warning: To prevent electric shock or personal injury, read the Safety Information before you use the Product.
This manual provides all the information necessary to perform basic maintenance and make calibration adjustments for the 374 FC/375 FC/376 FC and the 902 FC Clamp Meters (the Product). All figures show the 376 FC unless noted.
For complete operating instructions, refer to the 375 FC/375 FC/376 FC Users Manual and the 902 FC Users Manual.
Safety Information
A Warning identifies hazardous conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
⚠️ Warning: To prevent possible electrical shock, fire, or personal injury:
- Carefully read all instructions.
- Read all safety information before you use the Product.
- Use the Product only as specified, or the protection supplied by the Product can be compromised.
- Do not use the Product around explosive gas, vapor, or in damp or wet environments.
- Do not use the Product if it operates incorrectly.
- Do not use the Product if it is altered or damaged.
- Disable the product if it is damaged.
- Prior to use, ensure the Product is clean, dry, and in good repair. Do not use the Product if it has been exposed to water, cleaning solutions, battery electrolyte leakage, or other contaminants that may have entered the enclosure. These contaminants can reduce or eliminate the protection provided by the Product against electric shock or arc explosion. Have the product dried, cleaned, or repaired as necessary to ensure continued safe operation.
- Use only correct measurement category (CAT), voltage, and amperage rated probes, test leads, and adapters for the measurement.
- Do not exceed the Measurement Category (CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
- Comply with local and national safety codes. Use personal protective equipment (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed.
- Before each use, examine the Product. Look for cracks or missing pieces of the clamp housing or output cable insulation. Also look for loose or weakened components. Carefully examine the insulation around the jaws.
- Do not use test leads if they are damaged. Examine the test leads for damaged insulation and measure a known voltage.
- Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
- Do not measure current while the test leads are in the input jacks.
- Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.
- De-energize the circuit or wear personal protective equipment in compliance with local requirements before you apply or remove the Flexible Current Probe.
- Measure a known voltage first to make sure that the Product operates correctly.
- Limit operation to the specified measurement category, voltage, or amperage ratings.
- The battery door must be closed and locked before you operate the Product.
- Connect the common test lead before the live test lead and remove the live test lead before the common test lead.
- Remove all probes, test leads, and accessories before the battery door is opened.
- Keep fingers behind the finger guards on the probes.
- Hold the Product behind the tactile barrier.
- Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
⚠️ Warning: Do not use the HOLD function to measure unknown potentials. When HOLD is turned on, the display does not change when a different potential is measured.
Disconnect power and discharge all high-voltage capacitors before you measure resistance, continuity, capacitance, or a diode junction.
Remove the input signals before you clean the Product.
Use only specified replacement parts.
When batteries are changed, ensure that the calibration seal in the battery compartment is not damaged. If damaged, the Product may not be safe to use. Return the Product to Fluke for replacement of the seal.
Do not use in CAT III or CAT IV environments without the protective cap of test probe. The protective cap decreases the exposed probe metal to <4mm. This decreases the possibility of arc flash from short circuits.
Do not place magnet inside Category IV panel. Place it outside the panel instead.
For safe operation and maintenance of the Product:
- Repair the Product before use if the battery leaks.
- Have an approved technician repair the Product.
⚠️ Caution: To prevent possible damage to the Product or to equipment under test:
- Use the correct terminals, function, and range for measurements.
- Clean the case and accessories with a damp cloth and mild detergent only. Do not use abrasives or solvents.
Note: The Measurement Category (CAT) and voltage rating of any combination of test probe, test probe accessory, current clamp accessory, and the Product is the LOWEST rating of any individual component.
Symbols used on the Product and in this manual are explained in Table 1.
Symbol | Description |
---|---|
ℹ | Consult user documentation. |
⚠ | WARNING. RISK OF DANGER. |
⚡ | WARNING. HAZARDOUS VOLTAGE. Risk of electric shock. |
↘︎ | Do not apply around or remove from uninsulated hazardous live conductors without taking additional protective measures. |
↗︎ | Application around and removal from uninsulated hazardous live conductors is permitted. |
⌂ | Earth |
□ | Double Insulated |
~ | AC (Alternating Current) |
– | DC (Direct Current) |
Ⓐ | Conforms to relevant Australian EMC standards. |
Ⓟ | Certified by CSA Group to North American safety standards. |
CE | Conforms to European Union directives. |
CAT II | Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation. |
CAT III | Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation. |
CAT IV | Measurement Category IV is applicable to test and measuring circuits connected at the source of the building's low-voltage MAINS installation. |
♻ | This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. |
The Product
374 FC/375 FC/376 FC and 902 FC Clamp Meters
Description of Figure 1: A Fluke 374/375/376 FC clamp meter is shown, highlighting the clamp jaws, display, rotary dial, buttons (HOLD, MIN/MAX, LOG), and the tactile barrier for finger safety. The model shown is the 376 FC.
⚠️ Warning: To prevent possible electrical shock, fire, or personal injury:
- When measuring current with the Jaw, keep fingers behind the Tactile Barrier.
Electrical Specifications
374 FC/375 FC/376 FC Electrical Specifications
AC Current via Jaw
Range | 374 FC and 375 FC | 376 FC | Resolution | Accuracy | Crest Factor (50 Hz/60 Hz) |
---|---|---|---|---|---|
Range | 600.0 A | 999.9 A | 0.1 A | 2% ±5 digits (10 Hz to 100 Hz) 2.5% ±5 digits (100 to 500 Hz) |
3 @ 500 A (376 FC) 2.5 @ 600 A (376 FC) 1.42 @ 1000 A (376 FC) |
374 FC and 375 FC Crest Factor | 2.5 @ 350 A 1.42 @ 600 A |
Note: Add 2% for C.F. > 2 |
AC Current via Flexible Current Probe
Range | Resolution | Accuracy | Crest Factor (50 Hz/60 Hz) |
---|---|---|---|
2500 A | 0.1 A (≤999.9A) 1 A (≤ 2500 A) |
3% ±5 digits (5 Hz to 500 Hz) | 3.0 at 1100 A (375 FC and 376 FC only) 2.5 at 1400 A 1.42 at 2500 A Note: Add 2% for C.F. >2 |
Description of Figure 2: Position Sensitivity diagram showing a cross-section of the clamp with the conductor positioned at different distances (A, B, C) from the center, indicating the corresponding error percentages for i2500-10 Flex and i2500-18 Flex probes.
Distance from Optimum | i2500-10 Flex | i2500-18 Flex | Error |
---|---|---|---|
A: 0.5 in (12.7 mm) | 1.4 in (35.6 mm) | ±0.5% | |
B: 0.8 in (20.3 mm) | 2.0 in (50.8 mm) | ±1.0% | |
C: 1.4 in (35.6 mm) | 2.5 in (63.5 mm) | ±2.0% |
Measurement uncertainty assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature range.
DC Current
Range | 374 FC and 375 FC | 376 FC | Resolution | Accuracy |
---|---|---|---|---|
Range | 600.0 A | 999.9 A | 0.1 A | 2% ±5 digits |
AC Voltage
Range | Resolution | Accuracy |
---|---|---|
1000 V | 0.1 V (≤600.0 V) 1 V (≤1000 V) |
1.5% ±5 digits (20 Hz to 500 Hz) |
DC Voltage
Range | Resolution | Accuracy |
---|---|---|
1000 V | 0.1 V (≤600.0 V) 1 V (≤1000 V) |
1% ±5 digits |
mV dc (375 FC and 376 FC)
Range | Resolution | Accuracy |
---|---|---|
500.0 mV | 0.1 mV | 1% ±5 digits |
Frequency via Jaw
Range | Resolution | Accuracy | Trigger Level |
---|---|---|---|
375 FC and 376 FC: 5.0 Hz to 500.0 Hz | 0.1 Hz | 0.5% ±5 digits | 5 Hz to 10 Hz, ≥10 A 10 Hz to 100 Hz, ≥5 A 100 Hz to 500 Hz, ≥10 A |
Frequency via Flexible Current Probe
Range | Resolution | Accuracy | Trigger Level |
---|---|---|---|
375 FC and 376 FC: 5.0 Hz to 500.0 Hz | 0.1 Hz | 0.5% ±5 digits | 5 Hz to 20 Hz, ≥25 A 20 Hz to 100 Hz, ≥20 A 100 Hz to 500 Hz, ≥25 A |
Resistance
Range | 374 FC | 375 FC and 376 FC | Resolution | Accuracy |
---|---|---|---|---|
Range | 6000 Ω | 60 kΩ | 374 FC: 0.1 Ω (≤600.0 Ω), 1 Ω (≤6000 Ω) 375 FC and 376 FC: 0.1 Ω (≤600.0 Ω), 1 Ω (≤6000 Ω), 10 Ω (≤60.00 kΩ) |
1% ±5 digits |
Capacitance
Range | Resolution | Accuracy |
---|---|---|
1000 μF | 0.1 μF (≤100 μF) 1 μF (≤1000 μF) |
1% ±4 digits |
902 FC Electrical Specifications
Function | Range | Resolution | Accuracy |
---|---|---|---|
Voltage DC | 600.0 V | 0.1 V | 1.0% ±5 counts |
Voltage AC (True-rms) | 600.0 V | 0.1 V | 1.5% ±5 counts (45 Hz to 400 Hz) 2.0% ±5 counts, (45 Hz to 65 Hz) 2.5% ±5 counts, (65 Hz to 400 Hz) |
Current AC (True-rms) | 600.0 A | 0.1 A | Max Crest Factor (50 Hz/60 Hz): 3 @ 180 A, 2.5 @ 220 A, 1.42 @ 600 A Note: Add 2% for C.F. >2 |
Current DC (True-rms) | 200.0 μA | 0.1 μA | 1.0% ±5 counts |
Resistance | 600.0 Ω 6000 Ω 60.00 kΩ |
0.1 Ω 1 Ω 0.01 kΩ |
1.0% ±5 counts |
Continuity | <30 Ω | -- | -- |
Temperature | -40.0 °C to 400.0 °C | 0.1 °C | 1.0% ±8 counts |
Capacitance | 100.0 μF 1000 μF |
0.1 μF 1 μF |
1.0% ±4 counts |
Mechanical Specifications
374 FC/375 FC/376 FC | 902 FC | |
---|---|---|
Size (L x W x H) | 249 mm x 85 mm x 45 mm | 230 mm x 83.7 mm x 45.4 mm |
Weight | 410 g | 382 g |
Environmental Specifications
Parameter | Specification |
---|---|
Operating Temperature | -10 °C to +50 °C |
Storage Temp | 374 FC/375 FC/376 FC: -40 °C to +60 °C 902 FC: -30 °C to +60 °C |
Operating Humidity | Non-condensing (<10 °C) ≤90% RH (at 10 °C to 30 °C) ≤75% RH (at 30 °C to 40 °C) ≤45% RH (at 40 °C to 50 °C) |
Operating Altitude | 2000 meters |
Storage Altitude | 12 000 meters |
Safety | 374 FC/375 FC/376 FC: IEC 61010-1, Pollution Degree 2; IEC 61010-2-032: CAT III 1000V / CAT IV 600V; IEC 61010-2-033: CAT III 1000V / CAT IV 600V 902 FC: IEC 61010-1, Pollution Degree 2; IEC 61010-2-032: CAT III 600 V / CAT IV 300 V; IEC 61010-2-033: CAT III 600 V / CAT IV 300 V |
Intrusion Protection (IP) Rating | IEC 60529: IP30 |
Radio Frequency Certification | FCC ID:T68-FBLE IC:6627A-FBLE |
Wireless Radio Frequency Range | 2412 MHz to 2462 MHz |
Output Power | <100 mW |
Electromagnetic Compatibility (EMC) | International: IEC 61326-1: Portable, Electromagnetic Environment, IEC 61326-2-2, CISPR 11: Group 1, Class A Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself. Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances. Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments. Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of this standard when test leads and/or test probes are connected. |
Temperature Coefficients | Add 0.1 x specified accuracy for each °C above 28 °C or below 18 °C |
Performance Tests
Performance tests verify that the Product operates within the published specifications. These tests should be performed periodically and after service or repair. If the Product fails any part of the verification test, repair, and/or calibration adjustments are required. Refer to the Calibration Adjustment section.
In this section, the Product is referred to as the UUT (unit under test). Use these verification performance tests to ensure the values indicated on the UUT correspond, as closely as possible, with the actual measurement values.
For each test:
- Set the UUT to the appropriate function.
- Connect the UUT to the Calibrator.
- Set the Calibrator output.
- Compare the results to Table 2. (Where X = not applicable.)
The test passes if the UUT reads between the upper and lower limit.
Table 2. Performance Tests
Function | Calibrator Output | 374 FC | 375 FC | 376 FC | 902 FC | Meter Reading Limit (Lower Limit) | Meter Reading Limit (Upper Limit) |
---|---|---|---|---|---|---|---|
VAC | 20 V, 20 Hz | ✔️ | ✔️ | ✔️ | ❌ | 19.2 | 20.8 |
10 V, 50 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 9.4 | 10.7 | |
500 V, 50 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 492.0 | 508.0 | |
1000 V, 50 Hz | ✔️ | ✔️ | ❌ | ✔️ | 980 | 1020 | |
10 V, 400 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 9.4 | 10.7 | |
500 V, 400 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 492.0 | 508.0 | |
-1000 | ✔️ | ✔️ | ✔️ | ❌ | -1015 | -985 | |
-10 | ✔️ | ✔️ | ✔️ | ✔️ | -10.6 | -9.4 | |
10 | ✔️ | ✔️ | ✔️ | ✔️ | 9.4 | 10.6 | |
500 | ✔️ | ✔️ | ✔️ | ✔️ | 494.5 | 505.5 | |
1000 | ✔️ | ✔️ | ❌ | ✔️ | 985 | 1015 | |
VDC | 50 mV | ❌ | ✔️ | ✔️ | ❌ | 49.0 mV | 51.0 mV |
500 mV | ❌ | ✔️ | ❌ | ✔️ | 494.5 mV | 505.5 mV | |
-500 mV | ❌ | ✔️ | ✔️ | ❌ | -505.5 mV | -494.5 mV | |
Ω | 0 Ω | ✔️ | ✔️ | ✔️ | ✔️ | -0.5 Ω | 0.5 Ω |
10 Ω | ✔️ | ✔️ | ✔️ | ✔️ | 9.4 Ω | 10.6 Ω | |
300 Ω | ✔️ | ✔️ | ✔️ | ✔️ | 296.5 Ω | 303.5 Ω | |
3000 Ω | ✔️ | ✔️ | ✔️ | ✔️ | 2965 Ω | 3035 Ω | |
30 kΩ | ❌ | ✔️ | ✔️ | ✔️ | 29.65 kΩ | 30.35 kΩ | |
Capacitance | 5 μF | ✔️ | ✔️ | ✔️ | ✔️ | 4.6 μF | 5.5 μF |
90 μF | ✔️ | ✔️ | ✔️ | ✔️ | 88.7 μF | 91.3 μF | |
500 μF | ✔️ | ✔️ | ✔️ | ✔️ | 491 mF | 509 μF | |
AAC | 2 A, 50 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 97.5 | 102.5 |
with 50 turns coil | 12 A, 50 Hz | ✔️ | ✔️ | ✔️ | ✔️ | 587.5 | 612.5 |
19.5 A, 50 Hz | ❌ | ❌ | ✔️ | ❌ | 955 | 995 |
Table 2. Performance Tests (cont.)
Function | Calibrator Output | 374 FC | 375 FC | 376 FC | 902 FC | Meter Reading Limit (Lower Limit) | Meter Reading Limit (Upper Limit) |
---|---|---|---|---|---|---|---|
A | -19.5 A | ❌ | ❌ | ✔️ | ❌ | -995 A | -955 A |
-12 A | ✔️ | ✔️ | ✔️ | ❌ | -612.5 A | -587.5 A | |
-0.2 A | ✔️ | ✔️ | ✔️ | ❌ | -10.7 A | -9.3 A | |
0.2 A | ✔️ | ✔️ | ✔️ | ❌ | 9.3 A | 10.7 A | |
ADC | 2 A | ✔️ | ✔️ | ✔️ | ❌ | 97.5 A | 102.5 A |
with 50 turns coil | 12 A | ✔️ | ✔️ | ✔️ | ❌ | 587.5 A | 612.5 A |
19.5 A | ❌ | ❌ | ✔️ | ❌ | 955.0 A | 995.0 A | |
iFlex Current Probe (with simulation) | 3 mV, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 96.5 A | 103.5 A |
27 mV, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 872.5 A | 927.5 A | |
60 mV, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 1935 A | 2065 A | |
75 mV, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 2420 A | 2580 A | |
750 mV, 500 Hz | ✔️ | ✔️ | ✔️ | ❌ | 2420 A | 2580 A | |
iFlex Current Probe (with 50 turns r-coil) | 0.2 A, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 9.2 A | 10.8 A |
10 A, 50 Hz | ✔️ | ✔️ | ❌ | ✔️ | 484.5 A | 515.5 A | |
18 A, 50 Hz | ✔️ | ✔️ | ✔️ | ❌ | 872.5 A | 927.5 A | |
6 A, 440 Hz | ✔️ | ✔️ | ✔️ | ❌ | 290.5 A | 309.5 A | |
μA | 0 μA | ❌ | ❌ | ❌ | ✔️ | -0.5 μA | 0.5 μA |
10 μA | ❌ | ❌ | ❌ | ✔️ | 9.4 μA | 10.6 μA | |
200 μA | ❌ | ❌ | ❌ | ✔️ | 197.5 μA | 202.5 μA | |
MADC | -200 μA | ❌ | ❌ | ❌ | ✔️ | -202.5 μA | -197.5 μA |
Temperature | -40 °C | ❌ | ❌ | ❌ | ✔️ | -41.2 °C | -38.8 °C |
0 °C | ❌ | ❌ | ❌ | ✔️ | -0.8 °C | 0.8 °C | |
50 °C | ❌ | ❌ | ❌ | ✔️ | 48.7 °C | 51.3 °C | |
400 °C | ❌ | ❌ | ❌ | ✔️ | 395.2 °C | 404.8 °C |
Calibration Adjustment
Use the calibration procedures to adjust the Product so that the values shown on the Product correspond as closely as possible with the actual measured values. Table 3 lists the equipment required for the calibration adjustment.
Table 3. Required Equipment
Equipment | Required Characteristics | Recommended Model |
---|---|---|
Calibrator | 4.5-digit resolution | Fluke 55xxA Calibrator |
Wired coil | 50 turns | 5500A/COIL |
Test Lead for iFlex | -- | PN 666602 |
Test Lead for other | -- | PN 2070140 |
Power Supply | +3.0 V | Common power supply or a 2 x AA or AAA battery container |
Static Awareness
Semiconductors and integrated circuits can be damaged by electrostatic discharge during handling. This notice explains how to minimize damage to these components.
- Understand the problem.
- Learn the guidelines for proper handling.
- Use the proper procedures, packaging, and bench techniques.
Follow these practices to minimize damage to static sensitive parts.
⚠️ Warning: To prevent electric shock or personal injury. De-energize the product and all active circuits before opening a product enclosure, touching or handling any PCBs or components.
|
|
|
Calibration Setup
To set up the Product for calibration:
- Turn the Product over and use a flat-head screwdriver to remove the battery compartment screw. See Figure 3.
- Remove the battery door.
- Remove the batteries.
Description of Figure 3: Illustrates the steps to remove batteries: unscrewing the battery compartment, removing the door, and taking out the batteries.
- Remove the calibration sticker.
- Connect the Power Supply to the Product battery terminals.
- Turn on the Product.
- Use a small jumper to short the two pads together under the calibration sticker. See Figure 4.
Description of Figure 4: Shows the calibration activation process, depicting the bottom of the clamp meter with two pads that need to be shorted using a jumper.
To setup the current calibration see Figure 5.
Description of Figure 5: Illustrates the "Current Calibration Setup", showing the clamp meter connected to a Fluke 5522A Calibrator via test leads, with the calibrator displaying various functions and outputs.
For each calibration adjustment:
- Turn the rotary knob to select the function.
- Apply the required output from the source to the Product.
- Wait until each applied output stabilizes.
On the 37x FC Products, push [confirm] to confirm the value and move to the next step in the Adjustment Procedure. When you have completed all the steps, push [HOLD] to save the data to NVRAM storage.
On the 902 FC Products, push [confirm] to confirm the value and move to the next step in the Adjustment Procedure. When you have completed all the steps, push [HOLD] to save the data to NVRAM storage.
To view the target calibration point, push [MIN/MAX].
When calibration is complete:
- Remove the power supply.
- Replace the batteries.
- Replace the battery compartment door and tighten the battery compartment screw.
VAC Adjustment Procedure
Step | LCD Display | 374 FC | 375 FC | 376 FC | 902 FC | Calibrator Output |
---|---|---|---|---|---|---|
1 | C-00 | ✔️ | ✔️ | ✔️ | ❌ | 0 V, 0 Hz |
2 | C-01 | ✔️ | ✔️ | ✔️ | ✔️ | 600 V, 50 Hz |
3 | C-02 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 50 Hz |
4 | C-03 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 100 Hz |
5 | C-04 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 200 Hz |
6 | C-05 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 300 Hz |
7 | C-06 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 400 Hz |
8 | C-07 | ✔️ | ✔️ | ✔️ | ✔️ | 300 V, 500 Hz |
9 | Save | ✔️ | ✔️ | ✔️ | ✔️ | STBY |
VDC/mVDC Adjustment Procedure
Step | LCD Display | 374 FC | 375 FC | 376 FC | 902 FC | Calibrator Output |
---|---|---|---|---|---|---|
1 | C-08 | ✔️ | ✔️ | ✔️ | ❌ | 0 V |
2 | C-09 | ✔️ | ✔️ | ✔️ | ✔️ | 600 V |
3 | C-10 | ✔️ | ✔️ | ✔️ | ✔️ | 0 V |
4 | C-11 | ✔️ | ✔️ | ✔️ | ✔️ | 0.5 V |
5 | Save | ✔️ | ✔️ | ✔️ | ✔️ | STBY |
Ohm/Cap Adjustment Procedure
Step | LCD Display | 374 FC | 375 FC | 376 FC | 902 FC | Calibrator Output |
---|---|---|---|---|---|---|
1 | C-12 | ✔️ | ✔️ | ✔️ | ✔️ | 0 Ω |
2 | C-13 | ✔️ | ✔️ | ✔️ | ✔️ | 600 Ω |
3 | C-14 | ✔️ | ✔️ | ✔️ | ✔️ | 660 Ω |
4 | C-15 | ✔️ | ✔️ | ✔️ | ✔️ | 6000 Ω |
5 | C-16 | ✔️ | ✔️ | ✔️ | ✔️ | 6600 Ω |
6 | C-17 | ✔️ | ✔️ | ✔️ | ✔️ | 60 000 Ω |
7 | C-18 | ✔️ | ✔️ | ✔️ | ✔️ | 0.1 μF |
8 | C-19 | ✔️ | ✔️ | ✔️ | ✔️ | 0.5 μF |
9 | C-20 | ✔️ | ✔️ | ✔️ | ✔️ | 1.5 μF |
10 | C-21 | ✔️ | ✔️ | ✔️ | ✔️ | 110 μF |
11 | C-22 | ✔️ | ✔️ | ✔️ | ✔️ | 500 μF |
12 | C-23 | ✔️ | ✔️ | ✔️ | ✔️ | 1000 μF |
13 | Save | ✔️ | ✔️ | ✔️ | ✔️ | STBY |
AAC Adjustment Procedure (374 FC, 375 FC, 376 FC)
Step | LCD Display | 374 FC | 375 FC | 376 FC | Calibrator Output |
---|---|---|---|---|---|
1 | C-24 | ✔️ | ✔️ | ✔️ | 0 A, 0 Hz |
2 | C-25 | ✔️ | ✔️ | ✔️ | 8 A, 50 Hz |
3 | C-26 | ✔️ | ✔️ | ✔️ | 3 A, 50 Hz |
4 | C-27 | ✔️ | ✔️ | ✔️ | 3 A, 100 Hz |
5 | C-28 | ✔️ | ✔️ | ✔️ | 3 A, 200 Hz |
6 | C-29 | ✔️ | ✔️ | ✔️ | 3 A, 300 Hz |
7 | C-30 | ✔️ | ✔️ | ✔️ | 3 A, 400 Hz |
8 | C-31 | ✔️ | ✔️ | ✔️ | 3 A, 440 Hz |
9 | Save | ✔️ | ✔️ | ✔️ | STBY |
AAC Adjustment Procedure (902 FC)
Step | LCD Display | Calibrator Output |
---|---|---|
1 | C-24 | 0 A, 0 Hz |
2 | C-25 | 3 A, 50 Hz |
3 | C-26 | 8 A, 50 Hz |
4 | C-27 | 3 A, 50 Hz |
5 | C-28 | 3 A, 100 Hz |
6 | C-29 | 3 A, 200 Hz |
7 | C-30 | 3 A, 300 Hz |
8 | C-31 | 3 A, 400 Hz |
9 | C-32 | 3 A, 440 Hz |
10 | Save | STBY |
ADC Adjustment Procedure (374 FC, 375 FC, 376 FC)
Step | LCD Display | 374 FC | 375 FC | 376 FC | Calibrator Output |
---|---|---|---|---|---|
1 | C-32 | ✔️ | ✔️ | ✔️ | 0 A |
2 | C-33 | ✔️ | ✔️ | ✔️ | 10 A |
3 | Save | ✔️ | ✔️ | ✔️ | STBY |
uADC Adjustment Procedure (902 FC)
Step | LCD Display | 902 FC | Calibrator Output |
---|---|---|---|
1 | C-32 | ✔️ | 0 A |
2 | C-33 | ✔️ | 200 μA |
3 | Save | ✔️ | STBY |
ROGO Adjustment Procedure (374 FC, 375 FC, 376 FC)
Step | LCD Display | 374 FC | 375 FC | 376 FC | Calibrator Output |
---|---|---|---|---|---|
1 | C-34 | ✔️ | ✔️ | ✔️ | 0 V, 0 Hz |
2 | C-35 | ✔️ | ✔️ | ✔️ | 60 mV, 50 Hz |
3 | C-36 | ✔️ | ✔️ | ✔️ | 30 mV, 50 Hz |
4 | C-37 | ✔️ | ✔️ | ✔️ | 60 mV, 100 Hz |
5 | C-38 | ✔️ | ✔️ | ✔️ | 120 mV, 200 Hz |
6 | C-39 | ✔️ | ✔️ | ✔️ | 180 mV, 300 Hz |
7 | C-40 | ✔️ | ✔️ | ✔️ | 240 mV, 400 Hz |
8 | C-41 | ✔️ | ✔️ | ✔️ | 300 mV, 500 Hz |
9 | Save | ✔️ | ✔️ | ✔️ | STBY |
Temperature Adjustment Procedure (902 FC)
Step | LCD Display | 902 FC | Calibrator Output |
---|---|---|---|
1 | C-35 | ✔️ | 0 V, 0 Hz |
2 | C-36 | ✔️ | 0.02 V, 0 Hz |
3 | Save | ✔️ | STBY |
902 FC Temperature Zero Procedure
After you do the temperature adjustment (See Temperature Adjustment Procedure (902 FC)), you must zero the temperature.
- Insert the K-type thermocouple connector to attach the 5522A and the Product.
- Turn off the Product and wait 20 minutes.
- Turn on the Product.
- Use a small probe to push the calibration button.
- Turn the control knob to [select function].
- Push [AC/DC] twice, until the LCD shows C-37.
Step | LCD Display | 902 FC | Calibrator Output |
---|---|---|---|
1 | C-37 | ✔️ | 0 °C |
2 | Save | ✔️ | STBY |
Maintenance
This section contains basic maintenance procedures.
Clean the Product
⚠️ Caution: To prevent possible damage to the Product or to equipment under test, do not use abrasive cleaners. They will damage the case.
To clean the Product, use a cloth with a mild cleaning solution.
Battery Replacement
⚠️ Warning: To prevent possible electrical shock, fire, or personal injury, and to prevent incorrect measurements, replace the batteries when the low battery indicator shows.
Replacement Parts
Replacement parts and accessories are listed in Tables 4. To order parts and accessories, see How to Contact Fluke.
Table 4. Replacement Parts and Accessories
Fluke Part Number | Description | Quantity |
---|---|---|
4696918 | Battery Door Assembly | 1 |
3752958 | Soft Case (374 FC/375 FC/376 FC) | 1 |
1997276 | Soft Case(902 FC) | 1 |
4705494 | User Manual (374 FC/375 FC/376 FC) | 1 |
4748982 | User Manual (902 FC) | 1 |
376756 | Battery (AA 1.5 V) | 2 |
855742 | TL75 test leads | 1 |
3798105 | Fluke i2500-18 Rogowski coil | 1 |
1997234 | Thermocouple Assembly, K-Type, Beaded, Molded Dual Banana Plug (80BK) | 1 |
4744076 | FLUKE-374 FC, DECAL CELL | 1 |
4744083 | FLUKE-375 FC, DECAL CELL | 1 |
4698317 | FLUKE-376 FC, DECAL CELL | 1 |