SHORT INSTRUCTION MANUAL - poolcomet.com

SHORT INSTRUCTION MANUAL General description The CP11 is an accurate hand-held datalogger that displays & records relative humid-ity, temperature, CO2 and calculated parameters such as dew point and wet bulb. These short instructions are limited to a description of the main functions and installa

Medidor CO2 Rotronic CP-11 | PoolComet


File Info : application/pdf, 1 Pages, 283.42KB

PDF preview unavailable. Download the PDF instead.

manual-de-instrucciones-tecnicas-reducido
SHORT INSTRUCTION MANUAL
Congratulations on your purchase of the new portable humidity, temperature and CO2 instrument. Please read these short instruction carefully before installing the device.
General description The CP11 is an accurate hand-held datalogger that displays & records relative humidity, temperature, CO2 and calculated parameters such as dew point and wet bulb. These short instructions are limited to a description of the main functions and installation of the device. The detailed instruction manual can be found on the internet at www.rotronic-humidity.com
Programming
Most of the settings, such as memory clear, CO2 alarm, units( C/ F), sampling rate, pressure compensation and real time clock can be changed using the function keys.

Integrated probe
Built-in CO2 sensor w/ calibration cap

5VDC adaptor port PC connect port

Tripod hole
Battery compartment

Power Supply The CP11 uses 4pcs AA alkaline batteries or 5VDC adaptor (optional accessory). To insert (replace)the batteries, use the screw driver enclosed to loose 2 screws and put in the batteries in correct polarity with good contact.
Connection to PC or Laptop Connect the CP11 via mini USB cable to a PC or Laptop. Recorded data download can be easily downloaded with the software enclosed.

Functions Key

ON/OFF/SETUP -Press to turn on or turn off the instrument -In normal mode, press longer to enter setup mode. -Press with HOLD key to disable sleep mode
HOLD/UP -Freezes the current readings -Cancels data hold function -Selects unit or increases value in setup

START/ESC -Start automatic logging mode -Exits setup/recall/logging

MODE/RECALL -Press to switch displayed parameter -Press longer to enter memory recall

MEM/DOWN

MIN/MAX/AVG/ENTER

-Press to manually record the reading

-Activates MIN,MAX,AVG function

-Selects unit or decreases value in setup -Saves and finishes settings

Display

CO2

TEMPDPWBT UpperLCD Humidity/TEMP/CO2....reading LowerLCD Real time clock display

HOLDMAXMINAVGCAL

CO2 HOLD

Carbon dioxide reading Readings are hold unchanged

CO MIN/MAX Minimum/Maximum readings

FO

AVG

Average reading Low battery indicator

%rh DP

Dew point temperature

TEMP Temperature

WBT

Wet bulb temperature

AM

REC

%rh oC/oF

Unit of relative humidity Celsius/Fahrenheit of temperature

PM

Re call CAL

In calibration status

REC

In manual/automatic logging status

RECALL In manual records recall mode

Operation

POWER ON/OFF

Press

ON/OFF SET

to turn the meter on. At power up,

it emits a short beep and per-forms 28 sec.

countdown (Fig.1) for meter warm up, then

enters normal mode with real time clock

displayed (Fig.2). Date and time are

displayed in turns.

To power off, press

ON/OFF SET

again to turn off the

meter.

Fig.1

CO2
AM Fig.2

DATE/TIME SETUP

ON/OFF
When meter is powered on , hold down SET for more than 2 seconds until entering

meter setup mode. Press HOLD or MEM key to select the program P80 (rtc) and press

MIN MAX/AVG

key to enter.

First, choose the time format as 12hour or 24hour by pressing HOLD or MEM and then

press

MIN MAX/AVG

to confirm. Second, choose the date format as Month-Day-Year or Day-

Month-Year by pressing HOLD

or MEM

and

then,

press

MIN MAX/AVG

key to confirm. Now, start to

input the real time clock value from year, month, day, hour, minute to second. Press

HOLD

or

MEM

to select and then press

MIN MAX/AVG

to confirm. Press START to return to normal.

ESC

UNIT SETUP

ON/OFF
When meter is powered on , hold down SET for more than 2 seconds until entering

meter setup mode. Press HOLD or MEM key to select the program P50 (unit) and press

MIN MAX/AVG

key to enter.

Press

HOLD

or MEM

to select oC or

oF and then press

MIN MAX/AVG

to

confirm.

Press

START ESC

to return

to normal.

TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change, it takes 30 seconds to respond for CO2 sensor. NOTE: Do not hold the meter close to faces in case exhalation affects CO2

Humidity

Press MODE RECALL

to switch to humidity display.(Fig.3).The

lower display shows the real time clock.

TEMP,DP,WBT

Press

MODE RECALL

to switch to temperatures display(Fig.4,5,6).

The lower display shows the real time clock.

%rh Fig.3

CO2 (Carbon Dioxide)

Press

MODE RECALL

to switch the mode to CO2.User can get the

CO2 reading in ppm unit on main display.(Fig.7). The

lower display shows the real time clock.

DATA HOLD

AM

In normal display mode, press HOLD to freeze the

readings, "HOLD" icon is flashed on the left top of LCD

(Fig.8). All current readings are kept unchanged.

Press HOLD again to cancel the hold function.

NOTE:
The hold function is disabled in Min/Max/Avg mode. AM

BACKLIGHT

The backlight will be activated for 10 seconds by pressing any key.

MIN,MAX,AVG,CURRENT REVIEW

This meter allows you to check the minimun, maximum,

average, current value from the moment you press the AM

MIN MAX/AVG

key. Under normal mode, each press of

MIN MAX/AVG

displays

MIN, MAX, AVG and current value in sequence.

CO2

(Fig.9~11). To return to normal mode, press START to ESC

escape(Fig.12)

In this max/min/avg/current modes, it shows the

corresponding readings on main display and the

accumulated time of how long you are in this mode. The maximum time interval for this function is 18 hours. AM

TEMP

CO Fig.4

DP
CO Fig.5

WBT
CO Fig.6

Fig.7

Press MODE RECALL

can switch the displayed parameters. (Fig.13)

HOLD
Distinguishing the normal display and current review by checking the time display format. In normal display, date & time are displayed in turns. In current review, the lower display is the accumulated time of how long you are in max/min/avg mode.

%rh Fig.8

NOTE: "HOLD" & "MANUALLY RECORD" & "RECALL" functions are disabled in Min/ Max/Avg mode.

CO2

CO2

CO2

CO2

MAX

AVG

MAX

Fig.9

Fig.10

Fig.11

AM
Fig.12

%rh
Fig.13

ALARM
The meter features audible alarm to give warnings when CO2 concentration exceeds the limit. (See "Meter Setup" section for setting alarm threshold). It emits beeps (Abt. 80dB) when CO2 level goes over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.

AUTO POWER OFF

The meter turns off automatically after 20 minutes of inactivity.

To override the function, first, the meter is in power off status.

Hold down ON/OFF SET

and HOLD for 2 seconds to turn on the meter until

"n" appears.(Fig.14)

NOTE: Auto sleep function will be disabled during calibration & auto recording mode. Auto sleep function WILL NOT be disabled automatically via using adaptor.

Fig.14

MANUALLY RECORDING

MEM
The meter features a 99 point memory. In the normal or hold mode, press key to start recording, the REC icon and the main display will be flashed 3 times. The memory sequential number will show on the main display with 99 points in maximum. The data in the memory contains all parameters information and not only the displayed parameter. When all 99 points memory are occupied, a "FULL" message will be displayed on the screen. If you think the reading is changing too fast, you can press
HOLD key to freeze current reading first then manually record the data. This recording function will be disabled in the Min/Max mode.

99 MEMORIES RECALL

In normal or hold mode, press

MODE RECALL

key

for

more

than

2

seconds

until

the Re ca ll icon

flashes. Press HOLD or MEM to scroll the memories. The memory sequential number

displays on main LCD first and reading comes after. (Fig16&17). Press

MODE RECALL

key to

switch the display parameter.(Fig.18).The time displayed in memory recall mode is

the recording time of this memory.

CO2

%rh

AM

REC

Fig.15

DATA LOGGING

Rec al l
Fig.16

Rec al l
Fig.17

Rec al l
Fig.18

The meter can automatically record Humidity/TEMP./CO2 for long term environment monitor. The memory capacity is 6000 points for each parameter. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds. The factory preset rate is 30 second.

To setup the sampling rate, when meter is powered on , hold down

ON/OFF SET

for more than

2 seconds until entering meter setup mode. PressHOLD orMEM key to select the program

P60

(rAtE)

and

press

MIN MAX/AVG

key to enter.

Press HOLD or MEM to select the sample rate from hour, minute to second and then press

MIN MAX/AVG

to confirm. Press START

to return to normal.

ESC

After sampling rate is selected, press

START ESC

for 2 seconds under normal mode to start

logging. The REC icon flashes to indicate the logging status and LCD main display

shows the real time value. Lower displays are real time clock.(Fig.19)

Repeat above to start another run of logging. While the logging CO2 space is full, it shows "FULL". To terminate data logging, press

START ESC

for 2 seconds,

REC icon stops flashing.

Press MODE RECALL

key to switch to the mode you want to see. Except the

mode switching function, auto power off, min/max, manual

record, hold and recall functions are all disabled during logging. AM

REC

Meter Setup

Fig.19

ON/OFF
When meter is powered on , hold down SET for more than 2 seconds until entering meter setup mode. To exit setup, press START to return to normal mode.
ESC

Press HOLD

or MEM

key

to

select

the

program

and

press

MIN MAX/AVG

key to select.

The programmable setting flashes on display, press HOLD or MEM key to select and

press MIN MAX/AVG

key to confirm. To leave without saving, press START key to return.

ESC

P10, 99 points MEMORY clear P11, choose yes or no

Note:
" "is displayed while memory is cleared.

P20, auto LOG data clear

P21, choose yes or no

P30, CO2 ALARM setup P31, choose from 100 to 9900
P40, CO2 ABC setup
P41, Choose on or off

P50, Temp. UNIT setup

P51, choose oC or oF

oC

P60, logging sampling RATE

P61, choose from 04:59:59 to 00:00:01
P70, PRESSURE compensation

P71, choose from 700 to 1990hpa

P80, REAL TIME CLOCK setup

P81, choose 12 or 24 hour format

choose DATE format

Input Year/ Month/ Day
Input Hour/ Min./Sec.

year month date Hour Minute Second

oF
or

Note: " "is displayed while memory is cleared.
Note: -Factory preset 1400ppm. -Adjustable scale is every 100ppm
Note: -Factory preset ABC
function as ON
Note: -Factory preset oC
Note: -Factory preset at 30 sec. -The format is
hour:minute:second
Note: -Factory preset at 1013hpa -The adjustable scale is
every 1hpa.
Note: -Factory preset at 12H & dd-mm-year format.

Battery low indicator

A battery low icon " " appears when the supplied power is too low to maintain a normal operation. Please replace the batteries for accurate measuring. Suggest to use adaptor (optional part: AC1214) for long term measuring & datalogging.

Humidity Calibration

This meter can be calibrated either via 35% & 80% salt bottle or humidity generator.

The ambient condition is recommended as stable 25oC. While the meter is powered

off,

plug

the

sensor

probe

into

35%

salt.

Press

ON/OFF SET

,

START ESC

,

MEM

simultaneously for 3

seconds to turn on the meter and enter calibration mode. Press HOLD or MEM to select

35.0% calibration and press

MIN MAX/AVG

to start. "CAL" and calibrating value are blinking on

the LCD.

CAUTION: Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact ROTRONIC for calibration salt or services.
Single point calibration will cause error code E11 or non accurate reading. Always do dual point calibration to complete a process. Suggest to calibrate low humidity first.

CO2 Calibration

The sensor has an automatic calibration function to secure long-term accuracy. If a

calibration

is

desired,

press

ON/OFF SET

START ESC

MEM

simultaneously for 3 seconds to turn on the

meter and enter calibration mode. Press HOLD or MEM key to select 400 ppm CO2

calibration (Outside Air calibration). Press

MIN MAX/AVG

to

start

and

see"CAL"

and

CO2

value

blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is

completed and back to normal mode. While choose 0 ppm CO2 calibration, remove

the cap from meter rear side and plug the CO2 CALIBRATOR pipe into the sensor

hole beneath. Wait about 10 min until the blinking stops and the calibration is

completed and back to normal mode.

CAUTION: Do not calibrate the meter in the air with unknown CO2 concentration. Otherwise, it will be calibrated as 0 or 400ppm by default and leads to inaccurate measurements. Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration. When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.

Trouble Shooting

Error E01 E33 E02 E03 E04
E11 E31 E32 E33

Messages CO2 sensor is out of order CO2 sensor is out of order measured value is under range measured value is over range Original data error results in this error (DP, WB) RH calibration error Temp. sensor or AD damaged Memory IC damaged RH sensor or circuit damaged

Solution Turn off meter and re-start again Retry CO2 calibration Put meter in normal condition Put meter in normal condition Solve the original data error
Retry humidity calibration Return for repair Return for repair Return for repair

Technical Data Humidity/accuracy:
Temperature/accuracy:

0.1% to 99.9%/ ±3.0%(10~95%@25 °C), ±5% (other) -20...60oC/+0.3oC@5~40oC

CO2 /accuracy:

0...9999ppm/+(30ppm+5% of reading)@0~5000ppm

Storage and transit:

-20...60oC/10...90%rh, non condensing.

Operating limit at electronics:0...50oC for CO2, -20...60oC for rest/ non condensing.

Memory:

99 records manually, 18000 records auto logging.

Dimensions:

234(L)x77(W)x42(H),mm

Weight:

200gr.

Battery type:

4pcs AA alkaline

DC power type:

5VDC, 250mA. Optional available.

PC connection:

mini USB port

Optional Accessory
5VDC universal adaptor: CO2 calibrator:

Part Number AC1214
CO2 CALIBRATOR

ROTRONIC AG, CH-8303 Bassersdorf Tel. +41 44 838 11 44, www.rotronic.com ROTRONIC Messgeräte GmbH, D-76275 Ettlingen Tel. +49 7243 383 250, www.rotronic.de ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg Tél. +33 1 60 95 07 10, www.rotronic.fr ROTRONIC Italia srl, I- 20157 Milano Tel. +39 2 39 00 71 90, www.rotronic.it ROTRONIC Instruments (UK) Ltd, West Sussex RH10 9EE Phone +44 1293 571000, www.rotronic.co.uk ROTRONIC Instrument Corp, NY 11788, USA Phone +1 631 427-3898, www.rotronic-usa.com ROTRONIC Instruments Pte Ltd, Singapore 159836 Phone +65 6376 2107, www.rotronic.com.sg ROTRONIC Shanghai Rep. Offi ce, Shanghai 200233, China Phone +86 40 08162018, www.rotronic.cn

