URGENT - Corrective Action for Medical Device

Philips PageWriter TC Electrocardiographs (TC20/30/50/70)

User Battery Maintenance

Dear Customer,

A problem has been detected in Philips PageWriter TC (TC20/30/50/70) electrocardiographs which, if it occurs again, could pose a risk to patients or users. This communication aims to inform you about:

This document contains important information for continuing to use the equipment safely and correctly. Please read the information below carefully and share its contents with all operational personnel in the department. It is essential to understand the implications of this communication.

Keep a copy to attach to the equipment's Instructions for Use.

Philips has received a report of overheating and fire involving the lithium-ion battery of a PageWriter TC electrocardiograph. The battery has exceeded its average lifespan and needs to be replaced. Battery replacement should occur upon reaching 300 charge/discharge cycles or when the remaining battery capacity falls below 80% of a new battery's capacity. Although the electrocardiographs can display battery status, the current labeling does not include complete instructions on when to replace batteries or the potential risks associated with not doing so. Therefore, Philips is distributing an addendum to the PageWriter TC (TC20/30/50/70) electrocardiograph service manual with the necessary information to assist users in battery management and replacement. The Service Manual Addendum is enclosed with this communication.

For information on identifying affected devices and the measures to be taken, please read the following pages carefully. Please also follow the instructions provided in the "Measures to be taken by the customer" section of the document. This communication has been forwarded to the relevant authority.

Philips apologizes for any inconvenience caused by this issue. Customer satisfaction with Philips products and our response in this situation are of paramount importance.

For questions or concerns regarding this corrective action, please contact Philips:

0800 80 3000

Sincerely,

Vivian Dai
Post Market Surveillance Manager

Affected Products

All PageWriter TC (TC20/30/50/70) electrocardiographs manufactured before November 20, 2018, that can be used with battery power and have currently installed lithium-ion batteries are affected. Specifically, the following PageWriter TC electrocardiographs with software versions up to and including A.07.05.22:

Product Description
860315 PageWriter TC70 Electrocardiograph
860310 PageWriter TC50 Electrocardiograph
860306 PageWriter TC30 Electrocardiograph
860332 PageWriter TC20 Electrocardiograph
860352 PageWriter TC70 with trolley Government Bundle
860353 PageWriter TC70 without trolley Government Bundle
860354 PageWriter TC30 with trolley Government Bundle
860355 PageWriter TC30 without trolley Government Bundle
860429 PageWriter TC50 Government Bundle

Note: The PageWriter TC10 (860392) electrocardiograph is not affected by this safety notice.

Description of the Problem

Philips has received a report of overheating and fire involving the lithium-ion battery of a Philips PageWriter TC electrocardiograph. The battery has exceeded its average lifespan and must be replaced upon exceeding 300 charge/discharge cycles or when the remaining battery capacity is less than 80% of a new battery's capacity. Although Philips PageWriter TC electrocardiographs allow for the display of battery health status (SOH) and charge/discharge cycles, the current labeling does not include complete instructions on how to use this information to determine when to replace the battery.

Associated Risks

Battery overheating can cause the device's outer casing to overheat, which in turn can lead to melting and/or fire of the device, potentially causing injury to patients and nearby users or damage to objects.

How to Identify Affected Products

It is possible to determine if the device in use is affected by this communication by identifying its software version. This can be done in the following ways:

  1. Locate and verify the PageWriter TC electrocardiograph part number on the cover of the Instructions for Use or on the electrocardiograph's rear label.
  2. Locate and verify the software version in the System Utility screen of the PageWriter TC electrocardiograph.

To determine if the device is battery-powered, check for the battery icon in the upper right corner of the electrocardiograph's display.

Image depicting four PageWriter TC models: TC20, TC30, TC50, and TC70.

Measures to be Taken by the Customer/User

Upon receiving this communication, please carefully read the enclosed PageWriter TC Service Manual Addendum.

Immediately determine the number of cycles and the health status (SOH) on each affected Philips PageWriter TC (TC20/30/50/70) electrocardiograph, as specified in the attached Service Manual Addendum.

If the cycle count is greater than 300 and/or the SOH is less than 80%, the battery has reached the end of its useful life and must be replaced. An approved replacement battery can be ordered using Philips' standard replacement procedures. Details on how to replace the battery can be found in the PageWriter TC electrocardiograph Service Manual Addendum. After determining that the battery does not require replacement, or after replacing the battery, the PageWriter TC electrocardiograph is safe and can resume operation.

Additionally, review this information with all staff members responsible for device management for the Philips PageWriter TC electrocardiographs.

Keep the following Service Manual Addendum together with the Philips PageWriter TC Electrocardiograph Service Manual.

Complete and return the provided response form.

PHILIPS INTERVENTION PROGRAM

Philips plans to release a customer-installable software update for the PageWriter TC (TC20/30/50/70) electrocardiograph that will provide warning messages to support users in managing the battery replacement cycle. Customers will be notified once the user-installable software update is available.

FURTHER INFORMATION AND ASSISTANCE

For further clarification and assistance, please contact your local Philips organization:

0800 80 3000

Customer Response Form

Subject: Safety Notice Philips Healthcare
Response from customer for FSN86000263A
Electrocardiographs PageWriter TC (TC20/30/50/70)
Addendum to the service manual

Please complete and send via email to customercare.ch@philips.com

Contact Name
Phone Number
Email Address
Facility Name
Street: City, Postal Code

Send the completed form via email to the address provided above.

CUSTOMER CONFIRMATION:

The PageWriter TC (TC20/30/50/70) electrocardiograph Service Manual Addendum has been attached to the first page of the section related to battery maintenance, so that it can be correctly filed and kept together with the Service Manual for consultation.

Confirm that all batteries of PageWriter TC (TC20/30/50/70) electrocardiographs with a cycle count greater than 300 and/or a health status (SOH) ≤ 80% have been replaced with approved replacement batteries, as indicated in the Service Manual Addendum.

CUSTOMER NAME (in block letters) QUALIFICATION
CUSTOMER SIGNATURE DATE

Send the completed response form via email to customercare.ch@philips.com. In case of problems with the instructions in this communication, please contact your local Philips organization.

PageWriter TC Cardiograph Service Manual Addendum - Battery Maintenance

About the Battery

The rechargeable lithium-ion battery used in the PageWriter TC cardiographs is a smart battery with built-in circuitry that communicates battery status information to the cardiograph.

To properly maintain the battery and prevent damage to the cardiograph, observe these guidelines:

Supported Batteries

Supported Batteries Design Capacity PN
Lithium-ion battery ME202EK 7800 mAh 989803194541
Lithium-ion battery ME202C Rev D 7200 mAh 989803170371 (China only)

Note: If your cardiograph contains batteries with PN 989803160981, Philips strongly recommends that you replace these batteries as soon as possible with the supported batteries listed above.

When operating a PageWriter TC cardiograph with two batteries installed, both batteries must have the same part number. If a cardiograph is operated with two batteries with different part numbers, the cardiograph will display an error message and will not operate.

WARNING
Batteries other than those listed here are not supported. Failure to follow these instructions can lead to undesired consequences (battery overheating, shortened battery life, etc.).

Copyright ©2018 Koninklijke Philips N.V. All rights are reserved.

Manufacturer: Philips Medical Systems, 3000 Minuteman Road, Andover, MA 01810 USA, Tel: 1 (978) 687-1501

453564844701 Revision A, December 2018

TC20 Cardiograph

TC30 Cardiograph

TC50/TC70 Cardiograph

TC50 Cardiograph One Battery Operation

Viewing Battery Information

As a battery ages, its capacity decreases, and the battery status indicator becomes increasingly less accurate as the total number of charges and discharges increase. You can view information about the battery on the About the Cardiograph screen. To access this screen, see “Using the About the Cardiograph Screen” in Chapter 4 of the Service Manual. Ensure that the monitor is connected to AC power before attempting to review battery information.

Note: If only one battery is installed, information will not be displayed for the empty battery compartment.

Battery information displayed on the About the Cardiograph screen includes:

Image showing the 'About the Cardiograph' screen with various battery status details.

If the battery cycle count exceeds the recommended limit of 300 cycles, or the battery's state of health (SOH) has decreased to 80%, the battery should be replaced. A battery's state of health can be determined using the following formula:

SOH = Full capacity/Design capacity*

* See page 1 for design capacity.

Installing or Replacing the Batteries

WARNING
Properly dispose of or recycle depleted batteries according to local regulations. Do not disassemble, puncture, or incinerate the depleted batteries.

WARNING
Carefully follow the instructions for replacing the batteries. Only use batteries with Philips part number 989803194541 or 989803170371 (available only in China).

WARNING
Incorrect replacement of lithium batteries or fuel cells or replacement by inadequately trained personnel could result in an unacceptable risk (e.g., excessive temperatures, fire, explosion).

Caution

Note: If the status Recommend Calibration appears in the Battery Status window, calibrate the battery(ies) per the calibration procedure on page 7. If the status Recommend Replacement appears, the battery(ies) have reached the end of their useful life and need to be replaced. Note that the Recommend Replacement notice only appears after you have calibrated the battery(ies).

Notes about Battery Installation

To install the batteries:

  1. Open the battery door.
  2. Locate the two gold pull tabs inside of the battery compartment. Pull the tabs straight out of the battery compartment and lay flat.
  3. Insert the battery with the external connector facing the bottom rear of the compartment.
    Note: If operating the TC70, TC50 or TC30 cardiograph with one battery installed, the battery may be inserted into either battery compartment.
  4. Push in the battery and ensure that the battery is fully inserted into the slot. The pull tab will insert along with the battery. Insert the second battery following the same procedure.
  5. Replace the battery door.
  6. Connect the AC power cord to the cardiograph. Charge the batteries for five hours before operating the cardiograph on battery power only.

Battery Calibration

Battery calibration may be necessary in order to enhance the accuracy of the battery level indicator that displays on the Status Bar. If the accuracy of this indicator is affected, the other battery indicators included on the Battery Status window (that is opened by tapping the battery icon on the Status Bar) may also be less accurate.

Figure 2-5 Battery Power Indicator on the Status Bar

Image showing the Status Bar of the PageWriter TC cardiograph with battery indicators.

The recommended intervals for battery calibration are dependent upon factors in your clinical use model. When the battery power indicators are not functioning so that they are useful in your daily work environment, calibrate the batteries as described here.

The battery calibration procedure requires that the cardiograph be taken out of active use for up to 35 hours.

Note: If the status Recommend Calibration appears in the Battery Status window, calibrate the battery(ies) per the following procedure. If the status Recommend Replacement appears, the battery(ies) have reached the end of their useful life and need to be replaced. Note that the Recommend Replacement notice only appears after you have calibrated the batteries.

To calibrate the batteries on the cardiograph:

  1. Attach the AC power cord to the cardiograph. Ensure that the AC power supply is connected to a grounded electrical outlet and that the cardiograph is receiving AC power. Check that the AC power indicator light (located next to the power button) is lit.
  2. Fully charge the batteries. To ensure that the batteries are fully charged, view the Charge Current field in the Service Utility. Touch the Setup button on the tool bar. Select the Service Utility from the Configuration Setup and Service Utilities main menu.
    Note: Accessing the Service Utility may require entering a password. If a password is lost and cannot be retrieved, contact Philips Customer Service for assistance.
  3. From the Service Utility screen, ensure that the About the Cardiograph button is selected (top of screen). A selected button is highlighted in blue.
  4. Underneath the Battery Status column (middle of screen), check that the Charge Current field for both batteries display 0 mA, ensuring that both batteries are fully charged.
    Note: For devices with battery PN 989803194541 or 989803170371, the battery should remain in fully charged status for 5.25 hours or longer.
  5. Touch the Print button (top of screen) to print out a report of the cardiograph settings displayed on this screen.
  6. Touch the Exit button (lower right hand corner of screen). Touch the Exit button again on the Setup main menu.
  7. After the Main screen appears, touch the ID button to open the ID entry screen. Ensure that the ID screen remains open.
  8. Unplug the cardiograph from AC power. Ensure that the AC power indicator light (located next to the power button) is not lit.
  9. Keep the ID screen displayed and allow the batteries to deplete of all battery power. This process will take approximately 8 hours to complete. When the batteries are depleted of all power, the screen is black and the cardiograph cannot be returned to active use by touching the power button.
  10. Once the batteries are fully depleted, reconnect the cardiograph to AC power. Press the On/Standby button to power on the cardiograph in order to confirm successful calibration. Afterward, charge the batteries fully before returning it to active use.
    Note: For devices with battery PN 989803194541 or 989803170371, the battery should remain in fully depleted status for 5.25 hours or longer.
  11. On the Main screen, touch the Setup button.
  12. Select the Service Utility from the Configuration Setup and Service Utilities main menu.
  13. From the Service Utility screen, ensure that the About the Cardiograph button is selected (top of screen). A selected button is highlighted in blue.
  14. Underneath the Battery Status column, check the Expected Max Error (%) and Full Capacity (mAh) values as viewed on the screen are different than the values printed on the report generated from the Service Utility screen. If the values are different, the battery calibration procedure is complete.
    Note: If the Expected Max Error (%) field has not been reset to 2%, another calibration procedure may be necessary

Ordering a Replacement Battery

To order a replacement lithium ion battery, contact your Philips Customer Care center. For information on how to contact Philips Customer Care for your country, go to http://www.healthcare.philips.com. Select your country and language, then navigate to the Customer Care page.

Product Description Battery PN* Battery PN**
860332 TC20 Cardiograph 989803194541
(11.1V 7800 mAh)
989803170371
(11.1V 7200 mAh)
860306 TC30 Cardiograph
860310 TC50 Cardiograph
860315 TC70 Cardiograph
860354 TC30 w/trolley government bundle
860355 TC30 w/o trolley government bundle
860429 TC50 government bundle
860352 TC70 w/trolley government bundle
860353 TC70 w/o trolley government bundle

* World-wide, except China
** China only

Note: Battery PN 989803160981 is no longer available for purchase. If your cardiograph is using batteries with PN 989803160981, you will need to replace both batteries as soon as possible.

PDF preview unavailable. Download the PDF instead.

Vk 20190123 09-IT-1 PDF Writer - bioPDF / http://www.biopdf.com / FPG / Philips IT Infrastructure

Related Documents

Preview Philips Device Management Dashboard: Instrucciones de Uso (Versión A.02)
Guía de instrucciones para el Philips Device Management Dashboard (Versión A.02), software de gestión y monitorización remota para electrocardiógrafos y estaciones de trabajo cardiacas. Cubre instalación, configuración, gestión de dispositivos, solución de problemas y recursos adicionales.
Preview Philips 12-Lead Chest Lead Set Instructions for Use
Comprehensive instructions for the Philips 12-Lead Chest Lead Set, covering product description, intended use, technical specifications, compatible devices, safety precautions, and handling guidelines for ECG monitoring.
Preview Philips Avent Video Baby Monitor Recall and Replacement Information
Philips is voluntarily recalling and replacing certain Philips Avent Video Baby Monitors (SCD620, SCD625, SCD630, SCD831, SCD833, SCD835, SCD841, SCD843, SCD845) manufactured between January 2016 and January 2020 due to a potential overheating risk of the battery. Learn how to check if your product is affected and how to get a free replacement.
Preview Philips BG5470/15 Series 5000 Body Groomer: Features and Specifications
Discover the Philips BG5470/15 Series 5000 Body Groomer, featuring the Triple Protect-Rasiersystem for a comfortable, close shave. Learn about its waterproof design, long battery life, and versatile attachments for full body grooming.
Preview Philips SFL2000RH/10 Rechargeable LED Headlamp - 330 Lumens, IPX4
Discover the Philips SFL2000RH/10 rechargeable LED headlamp. Featuring 330 lumens, 9-hour runtime, IPX4 water resistance, and a 120-meter beam distance for versatile lighting needs.
Preview Philips Shaver Series 9000 Prestige SP9841/32: Close and Smooth Shaving with SkinIQ Technology
Discover the Philips Shaver Series 9000 Prestige SP9841/32 for a close, smooth shave. Featuring SkinIQ technology, Dual Precision NanoTech blades, Hydro SkinGlide coating, and Ultraflex suspension for ultimate comfort and efficiency. Enjoy wet & dry shaving, a 60-minute cordless run time, and a 5-year warranty.
Preview Philips Fyce 9290033648 Outdoor LED Garden Light User Manual & Installation
User and installation guide for the Philips Fyce outdoor LED garden light (model 9290033648). Covers assembly, IP44 rating, battery disposal, and optimal solar placement. Features multilingual instructions.
Preview Philips S1000 Electric Shaver User Manual and Safety Guide
Comprehensive user manual and safety guide for the Philips S1000 electric shaver, covering operation, cleaning, maintenance, and recycling information.