Model: LX1330B
Instruction Manual

Table of Contents
I. Instruction..........................................................1 II. Features............................................................1 III. Specification.......................................................1 IV. Name of Parts & Positions.....................................3 V. Operating Instructions...........................................4 VI. Battery Check & Replacement................................5 VII. Spectral Sensitivity Characteristic.........................6 VIII. Maintenance.....................................................6 IX. Recommended Illumination. ..................................7

. Instruction: The Digital Illuminance Meter is a precision instrument used to measure illuminance in the special field such as construction, inspection, photography, greenhouse gardening and etc. It is fully cosine corrected for the angular incidence of light.
The illuminance meter is compact, tough and easy to handle because of its construction. The light sensitive component uses a very stable and long life silicon diode to ensure stability with spectral response filter and fully corrected for the angular incidence of light.
 Features: z Light-measuring levels ranging from 0.1Lux to
200,000Lux, 0.01FC to 20,000FC repeatedly. z Highly accurate with precise and rapid response even in
high ambient light. z Data-Hold function for holding measuring values. z Units and signs display are arranged logically and
clearly for ease and comfort of reading results z Automatic zeroing.

z Meter corrected for Luminous Efficiency function. z Correction factor need to be manually calculated for
nonstandard light sources. z Short rise and fall times. z Accessories:
Carry case, instruction manual, battery.

 Specifications: z Display:
3-1/2 digit LCD. z Measuring Range:
200, 2,000, 20,000 and 200,000Lux (20,000Lux range reading×10, 200,000Lux range reading×100) And/or 20, 200, 2,000, 20,000FC (20,000FC range reading×10) *1FC = 10.76Lux z Out of range display: Highest digit of "1" is displayed. z Accuracy: ±(3%rdg +10dgt) 20,000Lux/2,000FC ±(5%rdg+10dgt) >20,000Lux/2,000FC z Repeatability: ±2% z Temperature Characteristic: ±0.1%/. z Measuring Rate: Approximately 2.0 time/sec.

z Photo detector: One silicon photo diode with filter.
z Operating Temperature and Humidity: 0 to 40 (32 to 104) 0 to 80% RH
z Storage Temperature and Humidity: -10to 60(14 to 140) 0 to 80% RH
z Power Source: One 9Volt Battery (NEDA 1604 or JLS 006P or IEC6F22)
z Battery life (typical): 200hours (Alkaline Battery).
z Photo Detector Lead Length: 150cm (approx).
z Photo Detector Dimensions: 100mm(H) ×60mm(w) ×28mm(D).
z Dimensions: 149mm(H) ×71mm(w) ×41mm(D).
z Weight: 250g (5.8oz).

 Name of Parts & Positions:

1. LCD Display: 3-1/2 Digits with a maximum reading of 1999.
2. Power Switch: The power switch key turns the digital illuminance meter ON or OFF.
3. Data-Hold Switch: Pressing the HOLD key selects HOLD mode. When the HOLD mode is selected, the digital illuminance meter stops all further measurements. Pressing the HOLD key again cancels the HOLD mode, causing the digital illuminance meter to resume taking measurements.
4. Data-Peak Switch: Pressing the PEAK key again to clear the peak recording mode.
5. LUX/FC Unit Switch: Pressing the LUX/FC button to choose LUX or FC unit.
6. Range Switch: Pressing the range key changes 200LUX/20FC, 2,000LUX/200FC, 20,000LUX/2,000FC, 200,000LUX/20,000FC ranges, circularly.
7. Photo Detector.

 Operating Instructions: 1. Power-up: Press the power key to turn the meter ON or
OFF. 2. Selecting the LUX/FC scale: Set the range selection
switch to desired LUX/FC range. 3. Open the photo detector cap and face it to light source in
a horizontal position. 4. Read the illuminance result from the LCD display. 5. Out of Range: If the instrument only display one "1" in
the M.S.D., that means the input signal is too strong, so a higher range should be selected. 6. Data-Hold mode: Press the HOLD key to select Hold mode. When the mode is selected, the illuminance meter stops all further measurements. Press the HOLD key again to cancel HOLD mode. Then it resumes normal operation and you can do the measurement again. 7. Data-Peak mode: Press the PEAK key to select PEAK mode. When PEAK mode is selected, the illuminance meter stops all further measurements. Press the HOLD

key again to cancel HOLD mode. Then it resumes normal operation. 8. When the measurement is completed, place the photo detector cap back and turn the power selector OFF.

 Battery Check-up & Replacement:

1. When the battery power is not sufficient, LCD will

display "

"; therefore a new battery type 9V is

required.

2. After turning off the meter, press the battery cover and

push in the direction of the arrow to open.

3. Disconnect the battery from the instrument and replace

it with a standard 9-volt transistor battery and go for the

cover.

 Spectral Sensitivity Characteristic:

z To the detector, the applied photo diode with filters

makes the spectral sensitivity characteristic almost

meet C.I.E.

(INTERNATIONAL COMMISSION ON ILLUMINTION)

photopic curve V as the following chart described.

S pectral S ensi tivity

100

V( )

80

...

60

Rela tive S ensitiv it y( %)

40

20

0

400

500

600

700

W avelength (nm)

 Maintenance: 1. The white plastic disc on the top of the detector should be cleaned with a damp cloth when necessary. 2. Do not store the instrument where temperature or
humidity is excessively high. 3. The reference level, as marker on the face plate, is the
tip of the photo detector globe. 4. The calibration interval for the photo detector will vary
according to operational conditions, but generally the sensitivity decreases in direct proportion to the product of luminous intensity by the operational time. In order to maintain the basic accuracy of the instrument, periodic calibration is recommended.

IX Recommended Illumination:

Conference, reception room

Office

Clerical work

Typing drafting

Packing work, entrance passage

Factory

Visual work at production line Inspection work

Electronic parts assembly line

Hotel

Public room, cloakroom Reception, cashier

Indoors stairs corridor

Store

Show window, packing table

Forefront of show window

Sickroom, warehouse

Hospital

Medical examination room Operation room, emergency

treatment

Auditorium, indoor gymnasium

School

Class room

Laboratory, library, drafting room

200750Lux 7001,500Lux 1,0002,000Lux 150300Lux 300750Lux 7501,500Lux 1,5003,000Lux 100200Lux 2201,000Lux 150200Lux 7501,500 1,5003,000 100200Lux 300750Lux 7501,500Lux
100300Lux 200750Lux 5001,500Lux

