User Manual for Dracal models including: PTH450, PTH450 Air Pressure Sensor, Air Pressure Sensor, Pressure Sensor, Sensor

User Manual

USB carbon dioxide (CO2) sensor, 0%-20%

PTH450

Comparaison de la série PTH - Dracal Technologies


File Info : application/pdf, 12 Pages, 1.78MB

PDF preview unavailable. Download the PDF instead.

Dracal PTH450 User Manual
User Manual
PTH450
Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

Model name

Part number Description

USB-PTH450

601081

USB-PTH450-CAL 608081

Robust, high-precision USB temperature, humidity and air pressure sensor
Robust, high-precision USB temperature, humidity and air pressure sensor ­ calibratable

VCP-PTH450

603081

VCP-PTH450-CAL 605081

HS code: 9025.19.80.80

Robust, high-precision USB temperature, humidity and air pressure sensor ­ with VCP mode
Robust, high-precision USB temperature, humidity and air pressure sensor ­ calibratable with VCP mode

Table of Contents
3 Compliance Certification 4 Overview 5 Product Overview and Specifications 6 Safety Guidelines 6 Installation and Setup 6 Usage Guidelines and Applications 9 Care, Storage, and Recalibration 9 Troubleshooting Guide and Support 10 Return Process and Warranty Information 10 Accessories and Peripherals 10 Disassembly and Environmental Considerations 11 Document Revision History
2 Dracal Technologies Inc.

Compliance Certification
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the manufacturer:
Dracal Technologies Inc. A204 - 7900 Taschereau Blvd Brossard (Quebec) J4X 1C2 Canada
for the following product(s): Model: PTH450 Type of Equipment: USB sensor CE Marking: Yes
Compliance with Directives
RoHS3 Directive (2011/65/EU, 2015/863) The USB Sensor Model PTH450 complies with the requirements of the RoHS Directive, restricting the use of hazardous substances in electrical and electronic equipment.
EMC Directive (2014/30/EU) The USB Sensor Model PTH450 complies with the essential requirements of the EMC Directive for electromagnetic compatibility.
Technical Documentation
Dracal Technologies holds the technical documentation required by the relevant directives at the following address: compliance@dracal.com.
It is available upon request to the relevant authorities.
Signed for and on behalf of: Dracal Technologies Inc., January 2024

Ariane Garon President Dracal Technologies Inc.

PTH450 ­ Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

3

Overview
At Dracal Technologies, we want to democratize access to precision data to engineers and scientists to ease their lives so they achieve their goals more quickly. We believe that precision data should not be compromised by ease of use. Dracal's measuring instruments connect through a USB port. Units with the VCP option allow users to integrate sensor data without using a 3rd party software. The USB units are compatible with DracalView, the graphical software, and command-ine interfaces. The calibratable units are equipped with a 3-point user-calibration mechanism that allows the handler to align multiple instruments together precisely or to adjust the accuracy of an instrument with a more precise reference. The PTH Series of instruments was designed for anyone interested in temperature, relative humidity (dew point) and atmospheric pressure. The PTH450: Robust, high-precision USB temperature, humidity and air pressure sensor is the tool to monitor ambient conditions in harsh environments.
4 Dracal Technologies Inc.

Product Overview and Specifications

Product Overview
The PTH450 features an enhanced precision temperature and humidity sensor. It pushes the boundaries, providing unparalleled accuracy in environmental data acquisition. Based on proven industry technologies, its digital sensor chips undergo individual factory calibration, linearization, and temperature compensation. The result? A sensor that stands at the forefront of performance. Engineered for environmental temperature, humidity, and atmospheric pressure data, the PTH450 adapts effortlessly to various conditions. Its unique construction allows it to thrive in a wide range of temperatures, making it a reliable companion for diverse scientific applications. With a built-in particle filter, the PTH450 shields against dust, soot, and contaminants, safeguarding its precision and longevity.
Key specifications

Temperature Probe operating range Typical accuracy from -40 °C to 125 °C Resolution Relative humidity Probe operating range Typical accuracy from 0 to 80 °C
Maximum accuracy at 25 °C
Resolution Atmospheric pressure Operating range for full accuracy Accuracy Resolution Digital resolution Altitude resolution Internal factory-calibration coefficients

-40 °C * to 125 °C ±0.3 °C, Maximum ±0.6 °C 0.015 °C (typical)
0 %RH to 100 %RH (non-condensing) ±2 %RH from 0 %RH to 100 %RH ±2 %RH from 0 to 80 %RH; ±3 %RH from 80 %RH to 100 %RH 0.01 %RH (typical)
45 to 110 kPa ±0.15 kPa at 25 °C 0.0065 kPa 24 bits Approx. 10 cm

Miscellaneous
· Probe cable material: silicone, · Probe first filter material: punctured anodized aluminum. · Supply: Powered by a USB port or a powered USB HUB. · Operating temperature range (USB cable and housing): -20 °C to 70 °C. · Supported by the SensGate DAQ unit. · Computes the dew point, heat index, humidex, and altitude. · Simultaneous operation of multiple Dracal sensors is supported.**
* Only if the cable is not moved/flexed while the temperature is below 0 °C.
** Only limited by allowable USB topologies and OS capacity. Fifty simultaneous USB sensors have been tested under Windows 8.

PTH450 ­ Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

5

Safety Guidelines
WARNING: Electromagnetic interference (EMI) may decrease the sensor's accuracy. Avoid using this device near EMI sources such as motors, high-voltage transformers, and fluorescent tubes. WARNING: This product should not be used in applications where its failure may cause personal injury.
Installation and Setup
TIP: As for any precision measurement equipment, powering the unit at least 15 minutes before using it is advised.
Installation
1 Download the DracalView* software at https://www.dracal.com/en/software/
2 Follow the installer's instructions. The whole process should take less than a minute. The setup is now completed.
3 Insert the unit into a free USB port.
* Includes a set of command-line tools.
Usage Guidelines and Applications
Operation ­ DracalView
We recommend launching DracalView to check if the instrument is connected correctly and works as expected. Here are the main steps of the Getting Started with DracalView guide.
1 Connect an instrument. 2 Launch DracalView. 3 Verify the instrument appears in the Sources tab. 4 Modify units and decimals in the Configuration tab (Optional). 5 Customize the names of channels to record in the Sources tab
(Optional). 6 Visualize data on GraphView.
6 Dracal Technologies Inc.

Please refer to the PTH450 page for the complete guide, specifically the "Software & Documentation" tab.
Of note, to log data with DracalView, these are the remaining main steps of the Getting Started guide:
7 Initiate the logging task in the Logging tab. 8 Keep an eye on data with BigView. Watch an introductory video for a quick overview of setting up and logging data in less than 3 minutes.

Operation ­ Command-line interface
The command-line tools are available with the download of DracalView. In the Getting Started guide, locate them and call the -h (help) argument, whether on a Windows, MacOS X, or Linux-based computer.
For complete documentation on how to use dracal-usb-get or dracal-sensgate-get, code examples, and integrating sensor data into LabVIEWTM, click on the relevant link in the same Getting Started guide.
Briefly, to log data to a file, here is the command line:
dracal-usb-get -s [serial number] -i [chosen channels] -L [location]
To redirect the output, give the argument -L the parameter "- " to have a stdout output.

PTH450 ­ Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

7

Operation ­ COM Protocol
"VCP-" prefix: If present, the product is equipped with the virtual COM port (VCP) option, which allows it to be freely converted between USB and VCP modes. It is delivered in USB mode. To convert the product from USB to VCP mode,
1 Open a command-line terminal. 2 Call "dracal-usb-set -f set_protocol VCP" and disconnect the product. To reverse the product back to USB mode,
Send the "PROTOCOL USB" command and disconnect the product. Please note that under Windows, once an instrument is converted to VCP, it is not recognizable by DracalView and the command-line tools. For additional information, please click the relevant link in the Software & Documentation section of the PTH450 page.
Calibration
"-CAL" suffix: If present, the product is equipped with a 3-point user-calibration mechanism. The Getting Started guide for calibration is summarized as follows:
1 Determine the calibration points. 2 Perform the calibration using DracalView or dracal-usb-cal. A comprehensive video tutorial on the best practices for calibrating instruments using DracalView can be found here.
The Getting Started guide for successful calibration, mentioned in this section, provides many tips and complementary documentation.
8 Dracal Technologies Inc.

Care, Storage, and Recalibration
Care
Use within the specifications available in the PTH450 datasheet.
NOTE: Note that this product is not waterproof and requires protection if contact with water is possible.
Storage
Store within the housing temperature and humidity ranges specified in the PTH450 datasheet.
Recalibration
Dracal's measuring instruments are factory-calibrated. It is possible to purchase a calibration certificate traceable to the SI along with the instrument and receive a calibrated instrument. An ISO17025-accredited laboratory performs this calibration.
If the instrument has a "-CAL" suffix, it can be adjusted for accuracy during the calibration.
The long-term drift of the PTH450 is less than 0.03 °C per year. Therefore, depending on the user's application, an adjustment might be required more frequently.
Troubleshooting Guide and Support
Troubleshooting
TIP: Avoid installing the sensor in a location where strong vibration is likely to occur. Strong vibrations may cause slight inaccuracies in the reading.
Feel free to browse the website's Resources section for any supplementary information. Here are interesting articles:
· Most frequent questions about Dracal · Top challenges and their solutions · FAQs about calibration

Support
Dracal Technologies has a dedicated email address for support: support@dracal.com.
As customer satisfaction is a priority, someone (an actual human!) will respond within 24 hours.

PTH450 ­ Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

9

Return Process and Warranty Information
Return Process
If a customer experiences issues with a Dracal sensor, they can contact support@dracal.com and explain their problem. In most cases, there will be an easy solution.
In the rare event of a faulty instrument under warranty, Dracal Technologies will proceed to a replacement upon evaluating the situation. The procedure will be explained through email once the customer advises support@dracal.com.
Warranty
The complete warranty can be found on Dracal Technologies' website here. Briefly, Dracal Technologies only warrants to the original product purchaser for 36 months (3 years) from the delivery date.
Accessories and Peripherals
To use Dracal Technologies' complete DAQ solution, the user must download the software on a computer with a USB port. Aside from sensors, Dracal Technologies offers a remote data logging station: the SensGate. Both a Wi-Fi/Ethernet gateway for Dracal's sensors and a data logger, it fulfills the needs of customers requiring a remote solution to precision data acquisition.
Disassembly and Environmental Considerations
Disassembly Instructions
1 Turn Off Power: Before disassembling the USB sensor, ensure it is disconnected from any USB port.
2 Remove External Connections: Unplug any cables connected to the USB sensor.
3 Identify Screws and Fasteners: Locate and identify any screws or fasteners securing the casing of the USB sensor.
10 Dracal Technologies Inc.

4 Separate Casing: Gently separate the casing; be careful about damaging internal components. If resistance is encountered, recheck for any overlooked screws.
5 Document Component Positions: Before further disassembly, document the positions of internal components for ease of reassembly.
6 Handle Components with Care: Handle internal components carefully to avoid electrostatic discharge and damage. Use an ESD (Electrostatic Discharge) mat or wrist strap if available.
7 Follow Reverse Order for Reassembly: When reassembling the USB sensor, follow the disassembly steps in reverse order, ensuring components are correctly positioned and secured.
Environmental Considerations
1 Recycling Recommendations: Dispose of components and packaging materials in accordance with local recycling regulations. Check with local authorities for proper e-waste disposal facilities.
2 Reuse Opportunities: Consider reusing components or donating functional parts to minimize environmental impact.
3 Minimize Energy Consumption: Power down or disconnect the USB sensor when not in use to minimize energy consumption.
RoHS Compliance
Our USB sensors adhere to RoHS (Restriction of Hazardous Substances) directives, ensuring they are free from harmful substances such as lead, mercury, and cadmium.
CE and REACH Compliance
These USB sensors comply with CE (Conformité Européenne) and REACH (Registration, Evaluation, Authorization, and Restriction of Chemicals) regulations, meeting European safety and environmental impact standards.
Document Revision History
Version 1.0 2024-02-08 Creation of the User Manual.

PTH450 ­ Robust, High-Precision USB Temperature, Humidity and Air Pressure Sensor

11

info@dracal.com dracal.com
+1 450.812.8612
Dracal Technologies Inc. A204 - 7900 Taschereau Blvd Brossard (Quebec) J4X 1C2 Canada



References