1. Safety Precautions

1-1. DANGER

The connector for thermocouple is not insulated. Do not make measurement using a thermocouple when there is a risk of electrical shock hazard.

1-2. WARNING

KEW5515 is one of the portable laser applied instruments which are regulated by the consumer product safety law.

1-3. CAUTION

Please follow the precautions below for use to keep the instrument's functions proper and to enable accurate measurement.

1-4. Environmental Precautions

1-5. Maintenance

Enclosure

Clean it with a soft cloth. To remove heavy stains, wipe it with a cloth wrung out with diluted neutral detergent.

Cautions:

Do not use chemical agents such as thinner, benzin or alcohol since these may cause a crack or fog on the surface of the measuring window, and consequently it may change the infrared radiation transmittance of the plastic lens.

1-6. Relation of Distance and Measuring Diameter

The relation between the distance and measuring diameter is as follows. However, the ratio between distance and measuring diameter varies depending on a distance. The more the distance increases, the larger area is needed.

The distance to measuring diameter ratio (D:S) is 12:1.

2. Features

This is a contactless handheld infrared thermometer. Temperature measurement with a K-type thermocouple is also possible.

3. Specifications

Parameter Value
Measuring range -32 to 535°C
Accuracy ±3.0°C: -32 to -20 °C, ±2.0°C: -20 to +100°C, ±2%: 100 to 535°C
Infrared spectral band 5 to 14µm
Repeatability Within ±1°C
Resolution 0.1°C
Response 500ms
Emissivity Variable between 0.10 and 1.00 (by 0.01 steps): Before shipment: 0.95
Measuring diameter 1000mm/ø78mm (Distance/ Measuring dia.: 12:1)
Battery One 9 V Dry cell battery (006P)
Operating temperature & humidity 0 to 50°C/ 10 to 90% RH
Auto power off If no key is pressed for 6 seconds, the power is shut off automatically.
LCD display LCD with backlight (blinks in red when alarm function is activated)
Collimation Laser beam (630 to 670nm 1mW or less) specifies the center.
Display Max/ Min/ Average value display
Dual display Simultaneous display: Measured value and either of max, min, average or thermocouple value.
Thermocouple K-type
Measuring range of thermocouple -199 to 1372°C
Accuracy of thermocouple ±(1.5%+1°C): -40 to 1372°C
Dimension 180 x 130 x 40mm
Weight Approx. 195g (excluding battery)
Accessories Carrying case, One 9 V Dry cell battery (006P), Instruction manual
Approved standards EN61326, IEC61000-4-2, IEC61000-4-3, IEC61000-4-8, IEC60825-1, CE Mark
Environmental standards EU RoHS Directive compliant

4. Name of each part

4-1. Body

LASER RADIATION

DO NOT STARE INTO BEAM

CLASS 2 LASER PRODUCT

CAUTION

Max. 1.0mW, 630-670nm IEC60825-1: 2014

レーザー光をのぞきこまないこと

レーザー光を人に向けないこと

子供に使わせないこと

Components:

Spot Size/Distance:

Model: KEW 5515

4-2. Display Part (Buttons)

The display shows various indicators and values. The buttons allow for operation and setting adjustments.

5. Getting Started

5-1. Measurement Mode and Parameter Setting Mode

Pull the trigger on the instrument to take a measurement. Press the F button (Function button) to use other functions. The main display shows measured values and the sub display shows values obtained with each function. The following flow chart shows the sequence and corresponding action.

Measurement Mode
Parameter Setting Mode

* Auto power off function activates in 6 seconds. When the display becomes blank due to this power off function, pull the trigger.

5-2. Other Functions

6. Battery Replacement

When the battery is low, the battery mark ? appears on the display. Replace the battery with a new one. When nothing is displayed on the display if attempted, it may also be battery exhaustion. Replace the battery with new one.

Open the battery cover as shown in the figure, and insert a new battery with correct polarity, and then close the cover.

Caution: Always set the measurement unit to °C.

Enable/disable laser (Enable by default)

Dry cell battery 9 V (006P)

7. Setting Emissivity (Reference)

Quantity of infrared emitted radiation from objects depends on the objects' material, surface conditions, measuring temperatures and so on. The table below shows emissivity of some objects. Please note that these values are just for reference. Please also refer to the estimated emissivity using the separately sold black body tape MODEL8085.

Material Emissivity Material Emissivity
Water, Ice 0.98 Cloth, Fabric (colored) 0.95
Soil 0.92 to 0.96 Leather, Fur 0.96
Concrete (wet) 0.96 to 0.98 Human skin 0.99
Concrete (dry) 0.91 to 0.95 Vegetable, Fruit 0.98
Ceramics 0.85 to 0.95 Dough 0.98
Stone, Asbestos 0.92 Meat 0.98
Plastics 0.90 to 0.95 Copper oxide 0.50 to 0.60
Rubber (black) 0.95 Ferric oxide 0.70 to 0.80
Wood 0.98 Painted surface 0.80
Paper 0.92 Tile 0.80

Estimated emissivity using the black body tape

If a measuring object accepts sticking of adhesive tape, stick a piece of the black body tape (emissivity: 0.94) to the object, and measure the temperature after setting the emissivity to 0.94.

* Black body is a substance which absorbs all incident radiation: no transmission and no reflection, and would have an emissivity value of 1.

* Infrared thermometers can easily measure temperature from a distance with safe. They can be used for electrical or equipment maintenance, building control and process monitoring and other applications.

Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.

KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.

2-5-20, Nakane, Meguro-ku, Tokyo, 152-0031 Japan

Phone: +81-3-3723-0131

Fax: +81-3-3723-0152

Factory: Ehime, Japan

www.kew-ltd.co.jp

Models: KEW 5515 Hand-Held Infrared Thermometer, KEW 5515, Hand-Held Infrared Thermometer

File Info : application/pdf, 14 Pages, 2.00MB

PDF preview unavailable. Download the PDF instead.

Adobe InDesign CS6 (Windows) Adobe PDF Library 10.0.1

Related Documents

Preview KYORITSU KEW 1011 & KEW 1012 Digital Multimeter User Manual
Comprehensive user manual for the KYORITSU KEW 1011 and KEW 1012 digital multimeters, covering safety instructions, technical specifications, measurement procedures, and maintenance.
Preview KEW 6305 Digital Power Meter Instruction Manual
Comprehensive instruction manual for the KYORITSU KEW 6305 Digital Power Meter, detailing its features, operation, safety precautions, and specifications for accurate power measurements.
Preview Kyoritsu KEW 8031/8031F Phase Indicator Instruction Manual
Instruction manual for the Kyoritsu KEW 8031 and KEW 8031F Phase Indicators. Covers safety warnings, instrument layout, specifications, and operating instructions.
Preview Kyoritsu KEW SNAP 6 / KEW SNAP 2608A AC Clamp Meter Instruction Manual
Comprehensive user manual for the Kyoritsu KEW SNAP 6 and KEW SNAP 2608A AC Clamp Meters. Covers safety warnings, features, detailed specifications, measurement procedures for AC/DC voltage, AC current, and resistance, battery/fuse replacement, optional accessories, and after-sales service.
Preview KYORITSU Test and Measuring Instruments General Catalogue 2020-2021
Explore the comprehensive KYORITSU General Catalogue 2020-2021, featuring a wide range of test and measuring instruments including multimeters, clamp meters, insulation testers, earth testers, and more. Discover new products and detailed specifications for electrical testing and measurement needs.
Preview KYORITSU KEW 8602 EVSE Adapter: Installation, Testing, and Troubleshooting Guide
Comprehensive guide to the KYORITSU KEW 8602 EVSE Adapter, detailing its features for installation, inspection, maintenance, and troubleshooting of Mode 3 AC EVSE. Includes specifications, test capabilities, and accessories.
Preview KYORITSU Digital Earth Clamp Tester MODEL 4200/KEW 4202 Instruction Manual
Instruction manual for the KYORITSU Digital Earth Clamp Tester, models 4200 and KEW 4202. Covers safety warnings, features, specifications, measurement methods, and troubleshooting.
Preview Kyoritsu Digital Earth Clamp Tester Model 4200/KEW 4202 Instruction Manual
Instruction manual for Kyoritsu Digital Earth Clamp Testers (Model 4200, KEW 4202). Covers safety, features, specifications, measurement methods for earth resistance and AC current, and Bluetooth connectivity for KEW 4202.