User Guide for mantracourt models including: PSDS-HSBK Portable Sensor Display, PSDS-HSBK, Portable Sensor Display, Sensor Display, Display

PSD-HSBK Quick Start Guide

SensorSpace Cloud Platform

PSD-HSBK, PSD, portable, sensor, display, system, measurement, strain

Mantracourt

PSD-HSBK Quick Start Guide

Quick Start guide: (psds quick start guide )

PSDS kannettava näyttölaite voima-antureille - KLINGER Finland Oy


File Info : application/pdf, 14 Pages, 948.28KB

PDF preview unavailable. Download the PDF instead.

psds quick start guide
PSDS-HSBK Quick Start Guide
mantracourt.com
PSDS-HSBK
Portable Sensor Display ­ Strain Bridge

Introduction/overview .........................................................................................................................................2 Getting started......................................................................................................................................................3
Hardware overview.....................................................................................................................................................................................3 Front ............................................................................................................................................................................................................3 Back .............................................................................................................................................................................................................. 4 Top ...............................................................................................................................................................................................................5
Setting up without the Toolkit ............................................................................................................................6 Calibration .....................................................................................................................................................................................................6 Change units.................................................................................................................................................................................................8 Change decimal places .............................................................................................................................................................................8 Overload/underload alarm......................................................................................................................................................................8 Using TEDS ................................................................................................................................................................................................. 10
Toolkit ................................................................................................................................................................ 12 Toolkit overview ....................................................................................................................................................................................... 12

Mantracourt Electronics Limited

1

PSDS HSBK Quick Start Guide

Introduction/overview
The PSDS-HSBK allows simple display of strain bridge based measurements such as load cells and pressure gauges with sensitivity up to +/-480 mV/V. Up to six calibration ranges are available allowing for different loading modes (tension and compression) or different sensors. Each calibration range will remember settings that contribute to the user experience such as selected units and tare values. TEDS devices using templates 33, 40 and 41 can be connected and will update the viewed calibration. The last twenty TEDS devices connected will be remembered and recognized when connected again reverting to the last user experience settings for that device. TEDS can be disabled and the internal calibration ranges become available again. There are also up to six display modes available. These determine what is visible on the LCD display and what actions are available from the set of three soft keys. Full configuration is available with a free PC based toolkit. Some simple configuration such as two point calibration is available from within the handheld using the menu system.

Mantracourt Electronics Limited

2

PSDS HSBK Quick Start Guide

Getting started
This document is designed to give a very quick overview of the PSDS-HSBK and its general, standard functionality. For more details, see the full manual.
Hardware overview
Front

Mantracourt Electronics Limited

3

PSDS HSBK Quick Start Guide

Back

Mantracourt Electronics Limited

4

PSDS HSBK Quick Start Guide

Top

The Load Cell connector fitted to PSD Standard Product is BINDER 770-8. The cable fitted mating connector is a BINDER 771 8-way male connector. There are three versions available with different field cable diameters.

Cable pair
1 2 3 4 Screen

View from solder connector side of the connector

Suggested Colour White Black Green Black Red Black Blue Black Grey

Connector Pin 1 2 3 4 5 6 7 8

Function
Loadcell Reference +ve Loadcell Reference -ve Loadcell Signal +ve Loadcell Signal -ve Loadcell Excitation +ve Loadcell Excitation -ve TEDS Ground Cable screen should only be connected to chassis of the sensor. If this cannot be achieved, then it should be connected to Excitation -ve.

Mantracourt Electronics Limited

5

PSDS HSBK Quick Start Guide

Setting up without the Toolkit
Quick, limited setup is possible using just the handheld and no toolkit. To access all the configuration features, you will need to use the toolkit.
Calibration
Please note that this must be done in the correct order for the calibration to work correctly. Changing the Sensitivity after inputting mV/V readings will cause the calibration to be void. Always make sure you finish with Apply

'CALIBRATION'

'USER CALIBRATION'

'ENTER PASSWORD'

(if set)

Scrolls through available characters

Moves between digits

'LIVE CALIBRATION'

'SENSITIVITY'

Select required input sensitivity steps)

(Note, this MUST be selected before the following

Return to previous menu

'UNITS'

Select required calibration units

Mantracourt Electronics Limited

6

PSDS HSBK Quick Start Guide

Return to previous menu

'LOW/HIGH CALIBRATION'

'SET OUTPUT LOW'

Manually enter the low output required using the value

to select digits and

to change

changes the sign of the entered number

moves the decimal place to the left

moves the decimal to the right

to save changes (long press cancels changes and returns to previous view)

'SET OUTPUT HIGH' Manually enter the high output required (as above)

to save changes (long press cancels changes)

'CAPTURE mV/V LOW' Apply low load, let load settle then: Confirm or edit low mV/V input
(You can manually edit the mV/V to a value from a calibration certificate here)

to save changes (long press cancels manual changes)

'CAPTURE mV/V HIGH' Apply high load, let load settle then: Confirm or edit high mV/V input
(You can manually edit the mV/V value to a value from a calibration certificate here)

to save changes (long press cancels manual changes) If everything is as planned

Return to previous menu

Mantracourt Electronics Limited

7

PSDS HSBK Quick Start Guide

'APPLY' Return to previous selected display mode.
Change units

In all standard display modes,

scrolls through the available units for the selected calibration.

Change decimal places

In all standard display modes, a 2 second press

shifts the decimal place position for the selected units.

Overload/underload alarm
Please note, the values saved for the thresholds are applied in the calibrated units of the currently selected range. This means that different ranges will trigger at different loads if the calibrated unit is different.

'ALARM'

'ALARM SOURCE'

Select `GROSS' or `NET'

Return to previous menu

'ALARM MODE'

Select `NORMAL' or `LATCHED' (When the alarm is latched but no longer being triggered, cancel the alarm by:

'ALARM'

Mantracourt Electronics Limited

8

PSDS HSBK Quick Start Guide

'**ALARM CANCEL**'

)

'ALARM THRESHOLDS'

'ALARM LOW'

Manually enter the low threshold required using

to select digits and

to change the value

changes the sign of the entered number

moves the decimal place to the left

moves the decimal to the right

to save changes (long press cancels changes and returns to display mode)

'ALARM HIGH' Manually enter the high threshold required as above.

to save changes (long press cancels changes and returns to display mode)

Return to previous menu

'ALARM TRIGGER'

Select from: Disabled Outside limits (<low, >high) Inside limits (>low, <high) Above high (>high) Below high (<high) Above low (>low) Below low (<low)

Mantracourt Electronics Limited

9

PSDS HSBK Quick Start Guide

Return to previous menu
'ALARM ACTION' Select `NONE', `BEEP', `FLASH' or `BOTH' Return to previous menu
Return to previous selected display mode.
Using TEDS
Plug in a TEDS enabled load cell.
Message: `NEW TEDS DEVICE USE SESSION DEFAULTS' TEDS table(s) will be automatically loaded.
Change selected TEDS calibration
Change displayed units To disable TEDS and use the PSD-HSBK's internal calibration:

'CALIBRATION'

'USER CALIBRATION'

'ENTER PASSWORD'

(if set)

Scrolls through available characters

Mantracourt Electronics Limited

10

PSDS HSBK Quick Start Guide

Moves between digits 'TEDS'
'TEDS MODE' Select from:
`ENABLED (TEDS)' `DISABLED (PSD)'
Return to previous menu Return to previous selected display mode. The next time that you connect that TEDS device you will get the message: `KNOWN TEDS DEVICE SESSION RESTORED'

Mantracourt Electronics Limited

11

PSDS HSBK Quick Start Guide

Toolkit
Toolkit overview
More complex setup is possible using the supplied toolkit software. Install the toolkit onto your windows PC and connect the USB to the PSDS-HSBK. Then open the toolkit:
Navigation is achieved using the icons along the top of the toolkit.

Mantracourt Electronics Limited

12

PSDS HSBK Quick Start Guide

For further information, please see the PSDS-HSBK User Manual.

Document Title: Applies To: Part Number: Issue Number: Dated:

PSDS-HSBK Quick Start PSD (Portable Sensor Display) 517-952 01.00 23rd May 2022

In the interests of continued product development, Mantracourt Electronics Limited reserves the right to alter product specifications without prior notice.

www.mantracourt.com

Mantracourt Electronics Limited

13

PSDS HSBK Quick Start Guide



References

Microsoft Word for Microsoft 365