Saevo Dental Practice Unit Pre-Installation Manual
Property Facilities Specifications for Dental Practice Units
Property Facilities Specifications
Understanding the property facility specifications for dental practice units is crucial for their performance and durability. This manual outlines the necessary requirements before installing your equipment. Saevo reserves the right to make changes to products and this manual without prior notification. Installation must be performed by an authorized Saevo technician to maintain the warranty. Property facility design, sizing, and execution must be handled by qualified professionals in compliance with current regulations.
Dental Practice Unit Positioning
Optimal positioning of the dental practice unit enhances ergonomics, ambience, and productivity. Installations should adhere to International Organization for Standardization (ISO) and International Dental Federation (FDI) recommendations. The room can be visualized as a clock face with the patient's mouth at the center (12 o'clock position). Concentric circles define zones:
- Circle A (0.5m radius): Transfer Zone for instruments and manual parts, and seating for the dentist and assistant.
- Circle B (1.0m radius): Working Area (maximum reach) for auxiliary tables and the main equipment body.
- Circle C (1.5m radius): Total Area for fixed cabinets and sinks, ensuring drawers remain within Circle B when opened.
Compressed Air
Compressed air installations must be free of particles, liquids, and oil. Maintain pressure between 80-120 PSI. Install the air compressor close to the supply point to minimize losses and avoid sanitary facilities to prevent air pollution. An air filter with a pressure regulator is required for each unit, using 1/4" x 10mm elbow at the inlet and 1/4" x 6mm elbow at the outlet. Pipes should slope 5%-10% towards airflow with an automatic drain valve at the lowest point. Use 10mm Polyurethane tubes preferably.
Electricity
Use appropriately sized circuit breakers. Separate electrical circuits for the dental unit, compressor, vacuum pump, X-ray, and autoclave. Ensure all electrical power points are properly grounded.
Water
Use only clean, potable water. Employ PVC pipes for installations. Install a water tap for each dental practice unit and a water register for the vacuum pump.
Sewer
Drainage pipes require a 2%-5% slope towards the drain flow. Separate drainage for sinks, the dental chair, and the vacuum pump. Connect drain sections only in siphoned boxes, preferably with a blind lid or lattice with closure. Use 45° fittings to prevent blockages and ensure correct drain direction. Install ventilation branches in the drainage network to prevent negative pressure. Use rigid PVC pipes.
High Power Suction
High power suction sewer pipes should be rigid PVC, 25mm diameter, with 45° joints and inclined in the flow direction. Electrical command installations for high power suction should be parallel when used in multiple dental units.
Technical Specifications for Sizing Installations
Equipment | Voltage (V) | Power (VA) | Air Consum. (l/min) | Water Consum. (l/min) |
---|---|---|---|---|
Dental chair | - | - | - | - |
Venturi | - | - | 25 | - |
Vac Plus | - | - | 35 | - |
Micro motor | - | - | 50 | 0,09 |
High rotation | - | - | 35 | 0,09 |
Triple syringe | - | - | 19 | 0,13 |
X-rays | 127 or 220 | 1200 | - | - |
1/2 HP Vacuum Pump | Bi-volt (key) | 1220 | - | 0,4 |
1 HP Vacuum Pump | Bi-volt (key) | 1690 | - | 0,4 |
12 L Autoclave | Bi-volt (key) | 1600 | - | (see manual) |
21 L Autoclave | Bi-volt (key) | 1700 | - | (see manual) |
40L Compressor | 127 or 220 | 830/1100 | - | - |
65L Compressor | 127 or 220 | 1660/2200 | - | - |
150L Compressor | 127 or 220 | 2948 | - | - |
Installation Points Diagram
The following diagram illustrates the recommended installation points for various services relative to the central axis of the dental chair. Distances are critical for proper setup.
Diagram Description:
A central axis of the dental chair is shown with concentric circles labeled A (0.5m), B (1.0m), and C (1.5m) indicating different functional zones. Several installation points are marked:
- POINT R: Reference point, 125mm from the central axis.
- POINT P: Head support work point.
- POINT E: High Power Sucker.
- POINT D: Independent Sewer Water Unit.
- POINT B: Electrical connection.
- POINT F: Electric command for Vacuum Pump.
- POINT A: Compressed Air connection.
- POINT C: Water connection.
Detailed connection diagrams are provided for:
- Point A (Compressed Air): Shows a 50mm x 500mm space with air supply via a 10mm Polyurethane Tube (02) within a corrugated flexible conduit (01).
- Point B (Electricity): Shows a 50mm x 400mm space with 3 cables of 2.5mm² (03) within a corrugated flexible conduit (04).
- Point C (Water): Shows water inlet connection Ø1/2" (05) with a weldable nipple with brass cap Ø20mm x 1/2" (06), a weldable 90° elbow Ø20mm (07), and a weldable rigid PVC tube Ø20mm (08).
- Point D (Independent Sewer for Water Unit): Requires a minimum 2% slope, using a siphoned box (14) and rigid PVC tubes (13, 18).
- Point E: Suction interconnection with the Vacuum Pump.
- Point F: Electrical control of Vacuum Pump suckers.
Note: Dimensions are in millimeters unless otherwise specified. The diagram also shows the recommended layout for the dental set positioning within the room.
Spare Connection Box
This section details the installation for a separate connection box, which is an optional kit. The standard installation uses an integrated connection box at the base of the dental chair. If using the separate kit, follow this pattern.
Diagram Description:
A layout showing the dental chair and the recommended distances for the spare connection box. Outputs are labeled:
- Pedal Output
- Flex Equipment Output (Air)
- Flex Equipment (Air) and Water Unit Output
- Water Unit Output
- Cart Equipment Output
The diagram also reiterates the installation points (A, B, C, D, E, F) relative to the central axis of the dental chair.
ATTENTION: For installations with a standard connection box separated from the base, it is necessary to acquire the kit for loose connection box (sold separately).