User Guide for AtriCure models including: ACM1 CryoICE Box, ACM1, CryoICE Box, Box

cryoICE BOX Quick Start Guide

cryoICE BOX Quick Start Guide - AtriCure

13 gen 2023 — Refer to ACM Cryo Console ‒ Technical Reference. Guide for troubleshooting tips for connecting probe. 3. Verify tank indicator. Note: Flashing tank indicates ...

Product Labeling | AtriCure


File Info : application/pdf, 4 Pages, 8.90MB

PDF preview unavailable. Download the PDF instead.

IFU-0043.B
TECHNICAL REFERENCE GUIDE
cryoICE® BOX Quick Start Guide

Scan QR code to view AtriCure's cryoICE BOX Quick Start Guide Video

cryoICE BOX Set-Up Sequence

1. Open tank
· Open valve "all the way open" · Active tank should have orange heater band
on it. Verify valve assembly is attached to tank with orange heater band.

C002021.D C002022.C

CAUTION: Federal (U.S.A.) Law restricts this device to sale by or on order of a physician.

Mason, OH 45040 USA 1.866.349.2342 www.AtriCure.com

ACM1 115 (100 -120) V~ 50/60 Hz 4.0 A
2x T 4A H 250V

2. Turn on machine
· Turn on power switch located on back of unit
· cryoICE BOX should be set up at least 15-minutes prior to the procedure

Note: Be sure to route the N2O vent tubing to a safe area prior to use.
2022-12 | IFU-0043.B

TECHNICAL REFERENCE GUIDE CRYOICE BOX QUICK START GUIDE
cryoICE BOX Set-Up Sequence Continued
3. Verify tank indicator Note: Flashing tank indicates the tank is closed. Open tank valve to resolve.

A BC
Note: Be sure to route the N2O vent tubing to a safe area prior to use.
cryoICE BOX Operation

4. Connect probe
· Connect probes to color matching connections
A. Thermocouple connectors: Red to red and black to black
B. Orange exhaust connector: push back locking collar prior to connecting
C. Blue inlet connector: pushes straight in
Note: Pull back on orange and blue connectors after connecting to verify good connection.
Refer to ACM Cryo Console  Technical Reference Guide for troubleshooting tips for connecting probe.

A

B

C

1. Start a freeze cycle
A. Press "Start/Stop" button to initiate the freeze cycle (button will turn blue)
B. Timer starts when temperature reaches ­40°C
C. Temperature display
Refer to ACM Cryo Console  Technical Reference Guide for troubleshooting tips

2. Freeze completed, auto-defrost cycle begins

A. When timer hits "0", machine will go into defrost (orange) mode and begin to warm probe to 10°C before returning to green ready status

Note: Wait untill "Start/Stop" button turns green

A

before beginning a new freeze cycle

3. Aborting a freeze cycle
A. Freeze can be stopped at any time by pressing "Start/Stop" button once · Machine will initiate a defrost cycle (orange) and return to ready status (green)
A

2022-12 | IFU-0043.B

TECHNICAL REFERENCE GUIDE CRYOICE BOX QUICK START GUIDE

cryoICE BOX Purge and Power Down Sequence
Note: Leave unit powered ON until purge sequence is completed

AB

1. Disconnect probe
A. Disconnect all connections and discard probe (single use device)
B. Push locking collars of orange and blue connectors towards cryoICE BOX to disconnect

2. Close tank · Close valve completely (turn to right)

C002021.D C002022.C

CAUTION: Federal (U.S.A.) Law restricts this device to sale by or on order of a physician.

Mason, OH 45040 USA 1.866.349.2342 www.AtriCure.com

A

C002021.D C002022.C

CAUTION: Federal (U.S.A.) Law restricts this device to sale by or on order of a physician.

Mason, OH 45040 USA 1.866.349.2342 www.AtriCure.com

3. Purge gas from machine

ACM1 115 (100 -120)V~ 50/60 Hz 4.0 A400
2x T 4A H 250V 300
200

500 600 700
800

A. Press and hold vent button until gas gauge stops moving · Gas gauge is mounted on cart frame

ACM1 115 (100 -120) V~ 50/60 Hz 4.0 A
2x T 4A H 250V

4. Turn off machine

See AtriCure Cryo Module (ACM) User's Manual for a complete list of safety related instructions and instructions for use.
2022-12 | IFU-0043.B

TECHNICAL REFERENCE GUIDE CRYOICE BOX QUICK START GUIDE
This Page Intentionally Left Blank
2022-12 | IFU-0043.B



References

Adobe PDF Library 17.0 Adobe InDesign 18.0 (Windows)