Test Fixture

Model: LCR-15A

Provided by: Test Equipment Depot - 800.517.8431 - TestEquipmentDepot.com

Quick Start Guide

GW INSTEK PART NO. 82CR-15A00M01

ISO-9001 CERTIFIED MANUFACTURER

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will Corporation.

The information in this manual was correct at the time of printing. However, Good Will continues to improve its products and therefore reserves the right to change the specifications, equipment, and maintenance procedures at any time without notice.

Manufacturer: Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

Components

Item Description Quantity
7 Low Electrode LOW side test pin 1
8 Stage Area for DUT placement 1
9 High Electrode HIGH side test pin 1
10 STD LOAD Standard resistor (100Ω) 10
11 L-Shape Bracket For stability, reducing BNC improper force (screws fixed by default) 1

Installation

1. Turn the both fixing handles of BNC terminals from the LCR-15A test fixture toward left side until the gap of the BNC external sleeve faces upward.

2. Align the gap with the salient point of the BNC terminal from the LCR test instrument and insert it firmly into place. Turn the both fixing handles of the BNC terminals from the LCR-15A test fixture toward right side until the external sleeve of the BNC is fixed stably into place.

Note: When the L-shape bracket is employed, take off the nut from the ground terminal prior to test fixture installation. Follow the steps above for proper installation followed by fastening the nut so that the test fixture can be held tightly when DUT is inserted.

Note: Refer to the diagrams below for details of L-shape bracket installation.

Diagram: Test Fixture Installation Illustrates the connection of the LCR-15A test fixture to an LCR instrument via BNC connectors. Shows the BNC connector handles being turned to secure the connection.
Diagram: L-shape Installation Shows the L-shape bracket being attached to the test fixture, likely for added stability or mounting.

Overview

The LCR-15A Test Fixture is designed for chip type components.

Specifications

Specification Value
Model LCR-15A
DUT Connector 2-Terminal
Measurement Frequency DC to 30MHz
Maximum Voltage ±45V Peak max. (AC+DC)
Application size 0201 ~ 1812, <5mm
Dimensions 105W x 47H x 69D mm
Weight 185g
Operating Environment 0~50°C, <70% RH

Additional Error Correction (OPEN/SHORT/LOAD ≥ 3MHz)

The following formulas apply for error correction:

Ze = {A + (Zs/Zx + Yo x Zx) x 100} %

Parameter Formula Unit
A Proportional error 0.375 x (f/10)² %
Yo Open repeatability 0.005 + 10 x (f/100) µS
Zs Short repeatability 20 + 30 x (f/10)

Where f is the frequency in MHz.

Fixture Overview

Diagram: Fixture Overview Shows a detailed view of the LCR-15A test fixture with numbered components.
Index Item Description Quantity
1 Test Fixture Main Body The main body of the test fixture. 1
2 Knob Pull or push to move the test pin of the HIGH side electrode. 1
3 Locking Screw Rotate clockwise [lock] or counterclockwise [unlock] the HIGH electrode. 2
4 BNC Measuring terminals that connect with instruments. 4
5 Fixing Handles Fastens or loosens terminals between BNC and instrument. Turn right [fix], turn left [loosen]. 2
6 Lever Place and rotate clockwise [fix] the LOW side test pin. Pull leftward [adjust] for DUT placement. Rotate counterclockwise [level] after test. 1

Measurement

WARNING: Before measurement, be sure to read the operating instructions to avoid danger.

1. Set measuring conditions with parameters, and install the LCR-15A test fixture.

2. Execute the CORRECTION mode. Set LCR-15A for the FIXTURE COMPENSATION item to execute parameter compensation (CABLE LENGTH item will be unavailable). If an extension cable is used, disable FIXTURE COMPENSATION (OFF) and select an appropriate CABLE LENGTH option.

3. Open Circuit Test:

Pull the LOW side test pin leftward [adjust]. Loosen the High side locking screw [unlock]. Move the HIGH side test pin rightward [adjust]. Loosen the LOW side test pin [adjust]. Ensure the test pins of both LOW and HIGH sides are contacted in a short circuit. Implement the open circuit test to display "OPEN" ON.

Diagram: Open Circuit Test Setup Shows the test fixture with LOW and HIGH side pins positioned to create an open circuit. The LCR-15A display is shown with "OPEN" selected and ON.

4. Short Circuit Test:

Pull the LOW side test pin leftward [adjust]. Loosen the High side locking screw [unlock]. Move the HIGH side test pin leftward [adjust] to the center of the test stage. Loosen the LOW side test pin [adjust] so the test pins of both LOW and HIGH sides contact each other in a short circuit. Implement the short circuit test to display "SHORT" ON.

Diagram: Short Circuit Test Setup Shows the test fixture with LOW and HIGH side pins touching each other to create a short circuit. The LCR-15A display is shown with "SHORT" selected and ON.

5. HF LOAD Calibration (Recommended for frequencies > 3MHz):

Pull the LOW side test pin leftward [adjust]. Loosen the High side locking screw [unlock]. Move the HIGH side test pin rightward [adjust] and place the STD-LOAD to the center of the test stage. Adjust the HIGH side test pin to contact the right side of the STD-LOAD, then fasten the locking screw [lock]. Loosen the LOW side test pin [adjust] so the test pins of both LOW and HIGH sides contact the STD-LOAD properly. Implement the HF LOAD test to display "HF LOAD" ON.

Diagram: HF LOAD Calibration Setup Shows the test fixture with the STD-LOAD component placed between the LOW and HIGH side pins. The LCR-15A display is shown with "HF LOAD" selected and ON.

6. DUT Measurement:

Pull the LOW side test pin leftward [adjust]. Loosen the High side locking screw [unlock]. Move the HIGH side test pin rightward [adjust] and place the DUT (Device Under Test) to the center of the test stage. Adjust the HIGH side test pin to contact the right side of the DUT, then fasten the locking screw [lock]. Loosen the LOW side test pin [adjust] so the test pins of both LOW and HIGH sides contact the DUT for measurement. The latest calibration status will be shown in the bottom of the LCD display (item and measuring cable length or test fixture model).

Diagram: DUT Measurement Setup Shows the test fixture with the DUT component placed between the LOW and HIGH side pins. The LCR-15A display shows measurement results including Ls (Inductance), Q (Quality Factor), R (Resistance), Rdc (DC Resistance), AC Voltage, Frequency, and Level. The CORR status is shown as "On, OPEN, SHORT, H/F, LCR-15A".

7. After completing the measurements, remove the component from the test fixture.

Note: It is required to place DUT into the test fixture in an exactly vertical way without any tilt, skew, or wobble. Do not fasten the adjusting screws overly in case of thread stripping and over compression on DUT. Also, do not overly loosen the screws in case of accidental drop out of screws.

Models: 82CR-15A00M01, LCR-15A Test Fixture, LCR-15A, Test Fixture, Fixture

File Info : application/pdf, 2 Pages, 1.56MB

PDF preview unavailable. Download the PDF instead.

instek lcr15a quickstart ecfc

References

Adobe Illustrator CS4 Adobe PDF library 9.00

Related Documents

Preview GW Instek GOM-804 & GOM-805 DC Milli-Ohm Meter User Manual
User manual for the GW Instek GOM-804 and GOM-805 DC Milli-Ohm Meters. This guide provides detailed information on features, operation, safety instructions, and specifications for precise low resistance measurements.
Preview GW Instek GPM-8213 Digital Power Meter User Manual
User manual for the GW Instek GPM-8213 Digital Power Meter, detailing its features, operation, safety, and technical specifications for precise electrical power measurement.
Preview GW INSTEK ASR-6000 Series Multi-phase Programmable AC/DC Power Source DeviceNet Manual
Comprehensive DeviceNet manual for the GW INSTEK ASR-6000 Series multi-phase programmable AC/DC power sources (models ASR-6450, ASR-6600). Covers safety, operation, DeviceNet communication protocols, and remote control commands.
Preview GW Instek GDS-1000B Series Digital Storage Oscilloscope Quick Start Guide
This Quick Start Guide provides essential information for setting up and operating the GW Instek GDS-1000B Series Digital Storage Oscilloscope. It includes safety instructions, panel overviews, setup procedures, main features, specifications, and package contents.
Preview GW Instek GPP-3610H/GPP-7250 Programmable DC Power Supply Quick Start Guide
Quick start guide for the GW Instek GPP-3610H and GPP-7250 programmable DC power supplies, detailing front and rear panel features, operational functions, and technical specifications.
Preview GW Instek PEL-2000B Series Programmable DC Electronic Load Quick Start Guide
Quick start guide for the GW Instek PEL-2000B Series Programmable DC Electronic Load, covering safety instructions, setup, features, specifications, and package contents.
Preview GW INSTEK GPD-3303 Series DC Power Supply User Manual
Comprehensive user manual for the GW INSTEK GPD-3303 Series DC Power Supply, covering safety instructions, operation, setup, remote control, command syntax, FAQ, and specifications.
Preview GW Instek GCT-9040 Ground Bond Tester User Manual
Comprehensive user manual for the GW Instek GCT-9040 Ground Bond Tester, covering safety instructions, setup, operation, external control, remote control, and FAQs. Learn about the features, specifications, and proper usage of this safety tester.