BAX Q7 Software v4213 installation instruction guide
Guide | BAX® System Q7 Software Upgrade
BAX® System Q7 Version 4.21.3.9366 Features and Instructions

BAX System Q7 software release v4.21 contains several user interface features, new target options, and error logging improvements.
We strongly recommended that you always install the most recent version as soon as possible for optimal performance. Please read all instructions thoroughly and ensure all required prerequisites are confirmed prior to prior to beginning installation of the software update.
If you currently have a WIN7 computer system, please contact diagnostics.support@hygiena.com or your local Hygiena® representative for upgrade support. This release is NOT applicable for WIN7 systems.

Software Updates and Features for Release v4.21.3.9366 for WIN10 ONLY  Install Package  Single executable install with reduced size (2 GB)  Reorganization of software versioning:  Package: 4.21.3.9366  BAX System: 4.21.3.9316  Bridge: 4.21.3.9359  SDS: 1.3.1035.1  Analysis: 4.21.3.8313
 SDS Equipment Driver Improvements:  Auto-recovery of instrument communication disconnection with notifications
 User Interface Enhancements:  Option to export results as a .csv file  About Dialog updates  Removed ability to initialize a run without instrument connection  Feature to preserve calibration file during installation
 Application Insights:  Improvements with more detailed logs for troubleshooting
 Certified Algorithms  Additional targets:  STEC Screener eae ONLY  STEC Screener stx ONLY  Real-Time E. coli O157:H7 S-Target ONLY
 Display messages and confidence levels for:  Real-Time E. coli O157:H7 Exact  Real-Time Salmonella
www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade
Software Upgrade Instructions:
1. Required Prerequisites:  Regional Settings are in English  Read/Write Permissions for the below file locations for all users  C:\Program Files\MATLAB  C:\Program Files (x86)\BAX  C:\ProgramData\BAX Q7 System  All Power and Sleep Settings are set to Never
2. Calibration storage and removal of current software:  Back up the system's calibration file  Select Operations > BAX Maintenance > Export Calibration from the menu bar  Navigate to a new folder in which to store the backup file and click Save  Name the calibration file (example: Q7Cal_date.bqc)
 Uninstall the previous version of the BAX Q7 application and ANY VERSION of MATLAB (REQUIRED)  Navigate to Windows Settings and select Apps

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade  Click on BAX System Q7, select Uninstall, and agree (click Yes) on next window to initiate removal
 Click on ANY version of the MATLAB application, select Uninstall, and click Agree on next window (this may take several minutes)
Restart Computer BEFORE continuing to step 3

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade
Software Installation Instructions for step 3 and 4
3. Download 4.21.3.9366 software by clicking this link or the button from the BAX Q7 Page  Transfer download to a memory/USB stick if computer doesn't have internet access  Click on the Q74.21.3.9366 icon and the program will automatically start execution of install.
 A windows protect screen may pop-up. Select More Info, then select Run Anyway.  On the next pop-up, select Yes to allow install to begin
4. MATLAB updates will install first. Follow prompts below and select the default responses:

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade
Install error: VCRUNTIME140.dll was not found. (shown below)
 The install will end after you click on OK. To resolve this issue, please download and install the patch vc_redist.x64.exe from https://www.microsoft.com/en-us/download/details.aspx?id=52685. After installing this patch, go back and retry installing the updated BAX System Q7 software.

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade

5. Following MATLAB installation, the BAX Q7 application install will initiate. Read the prompts carefully. Choose the default responses.
 Welcome Window ­ click Next

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade  Calibration File Warning ­ ENSURE YOU HAVE BACKED UP THE CALIBRATION FILE  Check box and click Next
 License Agreement  Check Accept Terms box and click Next

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade  Custom Setup  Click Next
 Password Protection  BAX default is disabled password protection. Choose an option for your lab, then click Next

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade  Ready to Install Program  Click Install
 Installing BAX System Q7 Software  During installation, an Error Log Reporting window will pop-up. We recommend NOT checking the box so the system can collect software and/or hardware errors in the Log Files to assist Diagnostics Support in troubleshooting. However, please review and decide what is appropriate for your lab.  Select Next to complete install

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade  InstallShield Wizard Complete  Select Finish
 Click on the desktop icon to initiate program. New open screen will have software version 4.21.3.9366
 Enter Instrument Name and Serial Number  This will appear on reports and should be utilized

www.hygiena.com

DIAG-INS2049-REVA

Guide | BAX® System Q7 Software Upgrade
6. CHECK system's calibration (NEW FEATURE)  From the menu bar, select Diagnostics > Calibration to open the Q7 Calibration Diagnostics window  Click the Check Calibration button to determine the status:  If calibration is available, no further actions are needed  If calibration is NOT optimal or NOT found: Import the backed-up calibration file by selecting Operations > BAX Maintenance > Import Calibration from the menu bar and selecting the backed-up calibration file from step 2

7. Ready to perform BAX System Analysis
 Define proper targets that are used in your lab by selecting View > Select Targets > then in window, highlight targets being used, and then use arrow buttons to move targets from Define in System to Available for Selection window