User Manual for Graviron models including: Wide Band PD Series O E Converter, Wide Band PD Series, O E Converter, Converter

User's Manual for SPD-3&4

GRAVITON INC. M.NAGATOMO

SPD-3 Wide Band PD Series 3.0[dB electrical] SPD-3 3.0[dB electrical] SPD-4

0.4 0.6 0.8 1 1.2 1.4 1.6 300 400 500 600 700 800 900 1000 Set up 1. First of all, connect the O/E converter to input connector of an instrument. Conversion connector may be required, if the instrument has no BNC jack.

SPD-3&4 Manual Eng           

Download-Graviton

User's Manual for SPD-3&4


File Info : application/pdf, 2 Pages, 386.38KB

PDF preview unavailable. Download the PDF instead.

SPD-3&4 Manual Eng
User's Manual
Wide Band PD Series
O/E CONVERTER
Please read this manual carefully before use

Introduction
Thank you for purchasing PD Series O/E Converter. PD Series is Optical to Electric signal converter and has following feature.
SPD-3500mV/mW SPD-4300mV/mW @ 850nm Wave Length, 50/125 GI Fiber Wide Frequency Range
DC2000MHz Flatness 0.5 ­3.0[dB electrical] SPD-3 DC3000MHz Flatness0.5 ­3.0[dB electrical] SPD-4 Standard FC optical connector Optional SC, F05, SMA, and ST connector are also available. Standard BNC plug
Able to connect directly to an oscilloscope, a spectrum analyzer etc. Able to get DC power from instruments through included Power Cable Assembly.
It is not guaranteed that power cable corresponds to all instruments. Able to use for instrumentation of lightwave products, optical links or
other various applications.
Please read this user's manual carefully and use it appropriately according to this manual, will make this product useful for development of light wave equipments, lightwave communication converter or other various applications.

Model

SPD-3

SPD-4

Graviton INC https://www.graviton.co.jp/index_En.html
Page 1
Contents of this package
The contents of this package are listed below. Please contact immediately the shop you purchased this product, if there is any missing item s found.

Do not re-produce or re-publishing a part or all this manual without written permission from Graviton Inc. This manual is subject to change without notice. Please follow exporting regulations/rules in individual countries when
exporting this product to other countries.
Page 2
Front and rear panel view









 O/E converter  Power Cable Assembly  Auxiliary Power Connector

 User's Manual(this manual)

Please keep original packaging, in case of re-shipping.

Page 3

 Power switch This switch is for controlling power. Please on/off this switch after
connecting power cable assembly to power connector() also keep this switch off when not in use.
 Power input connector The power can be supplied from probe power connector that equipped with an
instrument through the Power Cable Assembly. Or, alternately, external regulated power supply (not included) can be used through the Auxiliary Power Connector. Power supply voltage for PD series is±15[V]. Applying over and/or wrong polarity may cause permanent damage. Please refer the specification in page 7 for current consumption and page 5 for the detail of the power connection.
 Optical signal input connector Light signal supplied through this FC connector is up to -1[dBm] for SPD-3
and up to +1[dBm] for SPD-4. Please do not apply 10dBm or higher light signal. It may cause permanent damage.
 Electrical signal output connector 50[] input impedance with wide frequency range instruments is suitable to
observe light signal. Recommended to use wider frequency range instruments to withdraw full spec.DC2000MHz SPD-3, DC3000MHz SPD-4
Page 4

Set up

Connecting External Power Supply
.An external power supply is necessary, if an instrument going to use has no probe power output. The requirements for the power supply are regulated±15[V] ±5[%]and capable of supplying more than 150[mA] current with low ripple and noise. And connect as following figure.
Wrong power supply connection may cause permanent damage on the O/E converter. Please make sure the connection as RED Cable for Positive terminal, Black Cable for Ground Terminal and Blue Cable for Negative Terminal.

. First of all, connect the O/E converter to input connector of an instrument. Conversion connector may be required, if the instrument has no BNC jack.
. Confirm power switch on the O/E converter is off then connect Power Cable Assembly supplied. Other side of the Power Cable can be connected to the instrument, if probe power connector is equipped. Please refer page 6 for connecting external power supply, if the instrument has no probe power connector.

+15V(RED) GND(BLACK) -15V(BLUE)
. Panel cutout size is as following to mount the Auxiliary Power Connecter on a panel, if necessary. Maximum allowable thickness of the panel is 5mm.

. Turn on the power switch.

. Connect fiber connector to the O/E converter and adjust voltage and sweep time of the instrument observe the light signal. Also make sure 50[] impedance is selected on the instrument.

Page 5

Specifications

Item

Test Condition

SPD-3

SPD-4

OE Device

Si PIN PD

Si PIN PD

Peak Sensitivity Wavelength

700nm

700nm

Wavelength Range

More than 1/5

380nm 

380nm 

of peak sens.

950nm

950nm

Conversion Gain

50GI Fiber

500mV/mW

300mV/mW

@850nm

@850nm

Acceptable Core Diameter

Less than

Less than

0.5mm

0.5mm

Acceptable Fiber NA

Less than

Less than

0.25

0.25

Frequency Bandwidth

-3dB electrical

DC2.0GHz

DC3.0GHz

Maximum Optical Power for

400mV Out

-1dBm

+1dBm

Proper Signal Output

@850nm

@850nm

Wideband Noise Voltage

Up to 12.4GHz

Less than

Less than

1.5mVrms

1.8mVrms

Noise Equivalent Power

850nm

Less than

Less than

-25.2dBm

-22.4dBm

Output Offset Voltage

No Input Signal

Within

Within

+/- 0.5mV

+/- 0.5mV

Output Impedance

50ohms

50ohms

Input Optical Connector

FC

FC

Output Coax Connector

BNC Plug

BNC Plug

Supply Voltage

+/-15V DC

+/-15V DC

Supply Current

+150mA/-50mA

+150mA/-50mA

Physical Dimension[mm]

103x44x21

103x44x21

Weight[g]

130

130

Operational Temperature +5[]+35[]

Storage temperature -20[]+50[]

This specifications are subject to change without notice.

Page 7

Page 6

The Spectral Sensitivity curve of -  -
Reference point is 1 at 850nm

1.6

1.4

1.2

1

0.8

0.6

0.4

300

400

500

600

700

800

900 1000

Wave Lengthnm

Contact to : Graviton Inc. 1-8-1 Kasuga-cho, IRUMA-SHI, SAITAMA, 358-0006 JAPAN TEL+81-4-2966-0816 FAX+81-4-2966-0817 "GRAVITON" is trade mark of Graviton lnc. in Japan. Copyright 2006 by Graviton Inc. All right reserved. Never reproduce or republish without written permission.

Page 8

No. 20220831-01



References

M.NAGATOMO Foxit PDF Creator Version 9.7.0.2220