Ocean Next™ for Mako, Cobia, Piranha, and RTI Scatter Probe – v4.1.1.0 - RTI
File Info : application/pdf, 1 Pages, 78.20KB
DocumentDocumentPN_Ocean_Next-v4.1.1.0_Cobia.docx Product Note Cobia News in Ocean Next Release v4.1.1.0 Release information Release date Release name Cobia System version Cobia Firmware version 2024-05-20 Ocean Next v4.1.1.0 2019.12A 5.13C New Customer Values No new Cobia firmware with this release. Known Limitations The table below describes limitations that have been observed and not been fixed in the release. Description Using RTI Chamber Adapter with multiple ion chambers, the Cobia cannot handle chambers with identical serial numbers. The result is that even if selecting the second chamber, Cobia will assume that the first one is in use. This will cause measuring error if present. However normal use of RTI chamber adapter will not include ion chambers with identical serial numbers. *** END *** RTI Group Headquarters Phone: +46 (0) 31 746 36 00 Flöjelbergsgatan 8C info@rtigroup.com Company registration number: 556230-2462 SE-431 37 Mölndal, SWEDEN www.rtigroup.com VAT number: SE556230246201 1(1)