OPERATION MANUAL
BX-302
Zero Filter Calibration Kit BX-302-9800 Rev H

BX-302-9800 Manual Rev H.doc

Met One Instruments, Inc. 1600 NW Washington Blvd.
Grants Pass, OR 97526 Telephone: (541) 471-7111 Facsimile: (541) 471-7116
metone.com
Page 2

Met One Instruments, Inc. is now part of the Acoem international group of companies.
Met One Instruments has been designing and manufacturing classleading meteorological, ambient air sensing, and air quality monitoring instrumentation since its inception in 1989. Its line of robust industrial-grade meteorological equipment, air particulate monitoring equipment, and indoor air quality monitoring systems have set the standard for the industry. Headquartered in Grants Pass, OR, Met One Instruments, Inc. is fueled by a dedicated expert team who is diligently working to advance the technology required to ensure continued improvements in human and environmental health now and for generations to come.
Acoem is committed to helping organizations and public authorities find the right balance between progress and preservation -- safeguarding businesses and assets and maximizing opportunities while conserving the planet's resources. Headquartered in Limonest, France, Acoem delivers unrivaled inter-operable AI-powered sensors and ecosystems that empower our customers to make enlightened decisions based on accurate and timely information.
In 2021, Acoem acquired Met One Instruments, marking a pivotal moment when two industry leaders in the air quality monitoring sectors converged -- creating a single, stronger and more future-focused provider of holistic environmental monitoring solutions. Now, Met One Instruments Powered by Acoem has opened new possibilities through an extensive offering of class leading, multi-parameter environmental monitoring and industrial reliability solutions. These integrated measurement systems, technologies, and services deliver comprehensive solutions for a range of applications, including environmental research, regulatory compliance, and industrial safety and hygiene. For more information about Met One Instruments Powered by Acoem, please visit: metone.com
For more information about Acoem, please visit: acoem.com
BX-302 Operation Manual - © Copyright 2024 Met One Instruments, Inc. All Rights Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any form without the express written permission of Met One Instruments, Inc.
BX-302-9800 Manual Rev H.doc

Overview:
This document describes the procedure for using the BX-302 Zero Filter Calibration Kit to audit or adjust the background (BKGD) value on Beta Attenuation Monitors (BAM) particulate monitors. These instructions must be followed carefully to obtain the best accuracy from the unit.

BX-302 Zero Filter Variations

BX-302 with Cover (81418)

About the BAM Background:
The Background (BKGD) value is the zero correction (slope offset) used for all Met One Instruments BAM instruments (e.g. BAM 1020, BAM 1022, E-BAM Plus, and E-BAM).
The BKGD value is factory calibrated for each BAM instrument under laboratory conditions. Units should have this value verified and/or adjusted by the user during initial field deployment of the instrument, using the BX-302 zero filter kit. The BKGD value should also be audited either seasonally or at least annually afterward.
This field zero test corrects the BKGD value to compensate for minor variations caused by local conditions such as inlet heater operation, grounding, and RFI/EMI. It is not uncommon for the initial field zero test to result in a BKGD value that varies from the factory-set value by up to several micrograms.
If possible, the zero test should NOT be performed during a period of rapidly changing barometric pressure or ambient/shelter temperature, because rapid changes in air density can be measured as mass noise and may adversely affect the test results.
NOTE: The included BX-302 cover should always be installed during Background testing to protect against water intrusion and to prevent sunlight from creating false beta counts.
Reference the specific BAM instrument's operation manual for detailed Background test procedures.

BX-302-9800 Manual Rev H.doc