User Guide for senseca models including: ECO 130.2 Response 2 Channel Alarm Thermometer, ECO 130.2, Response 2 Channel Alarm Thermometer, Channel Alarm Thermometer, Alarm Thermometer, Thermometer

x x2 x3 ECO 130.2 - Quick response 2 channel-alarm-thermometer

Operating manual

Senseca,

Munta Anca / Deml Klaus Senseca Germany GmbH

B-H89.0.1X.DK2-2.0 ECO130 2 - Operating Manual EN[A5]

ECO 130.2 - quick response 2 channel-alarm-thermometer (formerly G 1202) - Portable | Senseca Germany GmbH

Thiết bị đo nhiệt độ Senseca ECO 130.2 (-65.0 ... 1200.0 ° C / -85.0 ... 2192 ° F) | SIEUTHITHIETBI.COM.VN

ECO 130.2 - Quick response 2 channel-alarm-thermometer


File Info : application/pdf, 22 Pages, 592.83KB

PDF preview unavailable. Download the PDF instead.

B-H89.0.1X.DK2-2.0 ECO130 2 - Operating Manual EN[A5]
OPERATING MANUAL
ECO 130.2
Quick response 2 channel-alarm-thermometer
for exchangeable Type K thermocouple sensors
B-H89.0.1X.DK2-2.0

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Table of contents

Table of contents

1

About this documentation .......................................................................................................4

1.1 Foreword ................................................................................................................................................ 4

1.2 Legal notices.......................................................................................................................................... 4

1.3 Further information ........................................................................................................................... 4

2

Safety ..............................................................................................................................................5

2.1 Explanation of safety symbols ....................................................................................................... 5

2.2 -Foreseeable misuse .......................................................................................................................... 5

2.3 Safety instructions.............................................................................................................................. 6

2.4 Intended use ......................................................................................................................................... 6

2.5 Qualified personnel............................................................................................................................ 6

3

The device at a glance ................................................................................................................7

3.1 Display elements ................................................................................................................................. 7

3.2 Connections .......................................................................................................................................... 7

3.3 Operating elements ........................................................................................................................... 8

4

Operation ......................................................................................................................................9

4.1 Opening the configuration menu .................................................................................................. 9

4.2 Adjustment of the measuring input .......................................................................................... 11

5

Measurement Basics ...............................................................................................................13

5.1 Sensor/device accuracy ................................................................................................................. 13

5.2 Possible measuring errors ............................................................................................................ 13 5.2.1 Immersion depth............................................................................................................. 13 5.2.2 Surface effects and poor heat transfer .................................................................. 13 5.2.3 Cooling / evaporation ................................................................................................... 13 5.2.4 Response time ................................................................................................................. 14

6

Operation and maintenance.................................................................................................15

6.1 Operating and maintenance notices ........................................................................................ 15

6.2 Battery ................................................................................................................................................. 15 6.2.1 Battery indicator ............................................................................................................ 15 6.2.2 Changing battery............................................................................................................ 15

B-H89.0.1X.DK2-2.0

Page 2 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Table of contents

7

Error and system messages...................................................................................................17

8

Technical data ...........................................................................................................................18

9

Disposal .......................................................................................................................................19

10 10.1 10.2 10.3

Service .........................................................................................................................................20
Manufacturer .................................................................................................................................... 20
Calibration and adjustment service .......................................................................................... 20
Accessories......................................................................................................................................... 21 10.3.1 Spare parts: ...................................................................................................................... 21 10.3.2 Suitable sensors: ............................................................................................................ 21 10.3.3 Further accessories: ..................................................................................................... 21

B-H89.0.1X.DK2-2.0

Page 3 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
1 About this documentation

About this documentation

1.1 Foreword
Read this document carefully and familiarize yourself with the operation of the device before you use it. Keep this document ready to hand and in the immediate vicinity of the device so that it is available to the personnel/user for reference at all times in case of doubt. The user must have carefully read and understood the operating manual before beginning any work.

1.2 Legal notices
The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse, disregarding this document, disregarding safety notices, assignment of inadequately qualified technical personnel and arbitrary modifications of the device. This document is entrusted to the recipient for personal use only. Any transmission, duplication, translation into other languages or excerpts from this operating manual require the consent of the manufacturer. The manufacturer assumes no liability for print errors.

1.3 Further information
Software version of the device: ­ V1.0 or later For the exact product name, refer to the type plate on the rear side of the device.
Note For information about the software version, press and hold the ON button to switch on for longer than 5 seconds. The series is shown in the main display and the software version of the device is shown in the secondary display.

B-H89.0.1X.DK2-2.0

Page 4 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
2 Safety

Safety

2.1 Explanation of safety symbols
Danger! This symbol warns of imminent danger, which can result in death, severe bodily injury, or severe property damage in case of non-observance.
Caution! This symbol warns of potential dangers or harmful situations, which can cause damage to the device or to the environment in case of non-observance. Note Blue underlining indicates processes, which can have a direct influence on operation or can trigger an unforeseen reaction in case of non-observance.

2.2 -Foreseeable misuse
The fault-free function and operational safety of the device can only be guaran-teed if applicable safety precautions and the device-specific safety instructions for this document are observed. If these notices are disregarded, personal injury or death, as well as property damage can occur.
Danger! Incorrect area of application!
In order to prevent erratic behavior of the device, personal injury and property damage, the device must be used exclusively as described in the chapter Descrip-tion in the operating manual.  The device is not suitable for use in explosion-prone areas!  The device must not be used for diagnostic or other medical purposes on patients!  Not suitable for use with requirements on functional safety, e.g. SIL!  Do not use in safety / emergency stop devices!

B-H89.0.1X.DK2-2.0

Page 5 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Safety

2.3 Safety instructions

Caution! Risk of injury at the temperature sensor!
 Devices with insertion probes entail the risk of stab injuries due to the pointed probe design.
 There is a risk of burns when measuring in high (and very low) temperatures. Use gloves if necessary.
 Empty batteries and batteries of inferior quality can leak more easily, which can destroy the device. Please also observe the instructions in the chapter "Operation and maintenance".
Note This device does not belong in children's hands! The device housing is not designed for continuous contact with foods. For measurements in food in accordance with EC Regulation 1935 / 2004, suitable temperature probes must be used:

2.4 Intended use
The device is thermometer for type K interchangeable probes. It is designed, with the appropriate probe, for precise and instantaneous temperature measurements in the following media: · Liquids, gases, soft plastic materials, bulk material · Food
2.5 Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have adequate knowledge of the measuring process and the significance of the measurements. The instructions in this document must be understood, observed and followed. In order to avoid any risks arising from interpretation of the measurements in the concrete application, the user must have additional expertise. The user is solely liable for damages/danger resulting from misinterpretation due to inadequate expertise.

B-H89.0.1X.DK2-2.0

Page 6 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
3 The device at a glance

The device at a glance

LCD Display

Top view

Front view

3.1 Display elements Battery indicator Unit display Main display
Auxiliary display

Evaluation of the battery status
Display of units or min/max/hold/diff
Measurement of the current temperature of the selected channel or value for min/max/hold With INP = both or diFF: Display of the current temperature of channel 2 or value for min/max/hold With INP = CH1: Measured value of the current temperature in min/max/hold mode

3.2 Connections Thermocouple plugs

Connection for temperature sensor CH1 and CH2

Caution!
The device ensures extensive protection against splashing water and rain. However, immersion in water is not possible due to the socket opening.  Protect the connector opening / contacts from soiling and moisture!  Dry damp device and plug connectors as quickly as possible!

B-H89.0.1X.DK2-2.0

Page 7 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

The device at a glance

3.3 Operating elements On / Off button Press briefly Long press
Up / Down button

Switch on the device Activate / deactivate lighting
Switch off the device Reject changes in a menu

Press briefly
Long press Both simultaneously Function button

Display of the min/max value Change value of the selected parameter
Reset the min/max value of the current measurement
Rotate display, overhead display

Press briefly Long press, 2s

Freeze measurement (Hold) Call up next parameter
Start menu "configuration", (ONF appears in the display

Operating status

device is in measured value display device is in a menu

B-H89.0.1X.DK2-2.0

Page 8 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
4 Operation

4.1 Opening the configuration menu

1 Press the Function button for 2 seconds to open the Configuration menu. 2 (ONF appears in the display. Release the Function button.

Parameter Values

Meaning

Operation

INP
AL.
AL.LO AL.XI POFF

Measuring channel

(H.1

Main display = channel 1

BOTH

Main display = channel 1 Secondary display = channel 2

D,FF

Main display = difference channel 1 ­ channel 2

Secondary display = channel 2

Alarm *1

OFF

No alarm active

ON

Alarm via text insertion, acoustic signal and flashing

of the background lighting

BEEP

Alarm via text insertion and acoustic signal

L,TE

Alarm via text insertion and flashing of the back-

ground lighting

Min. alarm limit (only available if AL <> off)

-65.0 .. AL.X,

a min. alert is triggered if the value falls below this value. (for °F: -85.0 .. AL.Hi)

Max. alarm limit (only available if AL <> off)

AL.LO .. 1200.0

If the value is exceeded, a max. alarm is triggered. (for °F: AL.Hi .. 2192)

Shut-off time

OFF

No automatic shut-off

0:15, 0:30, 1:00, 4:00, Automatic shut-off after a selected time in

12:00

hours:minutes, during which no buttons have been

pressed

B-H89.0.1X.DK2-2.0

Page 9 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Operation

Parameter Values

Meaning

L,TE

Backlight

OFF

Backlight deactivated

0:15, 0:30, 1:00, 2:00, Automatic shut-off of the backlight after a select-ed

4:00

time in minutes:seconds, during which no buttons

have been pressed

ON

No automatic shut off of the backlight

UN,T

Display unit

°(

Temperature display in °C

°F

Temperature display in °F

IN,T

Factory settings

NO

Use current configuration

YES

Reset device to factory settings.

After confirming with the function-button, the

display shows: IN,T DONE

*1 = The alarm function works depending on the set measuring channel:  with INP = ,,(H.1" on channel 1  with INP = ,,BOTH" on channel 1 and channel 2  with INP = ,,D,FF" on the difference temperature

Note to D,FF The adjustable alarm limits continue to refer to the measuring range limits of the device, which means: negative values can only be set to a very limited extent! if necessary, swapping channel 1 and channel 2 could help here

B-H89.0.1X.DK2-2.0

Page 10 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Operation

4.2 Adjustment of the measuring input

The temperature input can be adjusted with the zero point correction and the gradient correction. If an adjustment is made, you change the pre-adjusted factory settings. This is signaled with the display text T.OF or T.SL when switching on.
1 Switch the device off. 2 Hold the down button and press the On/Off button briefly to switch on the device and
open the Adjustment menu. 3 The display shows the first parameter. Release the down button.

Parameter Values

Meaning

T.OF

Zero point correction of measuring channel 1

0.0

No zero point correction

-5.0 ... 5.0

Zero point correction in °C. (at °F -9.0 .. 9.0)

T.SL

Gradient correction of the temperature of measuring channel 1

0.00

No gradient correction of the temperature

-5.00 ... 5.00

Gradient correction in %

T2.OF

Zero point correction of measuring channel 2

0.0

No zero point correction

-5.0 ... 5.0

Zero point correction in °C. (at °F -9.0 .. 9.0)

T2.SL

Gradient correction of the temperature of measuring channel 2

0.00

No gradient correction of the temperature

-5.00 ... 5.00

Gradient correction in %

Formula used by device: Temperature = °C: Display = (measured value - T.OF) * (1 + T.SL / 100) Temperature = °F: Display = (meas. value - 32 °F - T.OF) * (1 + T.SL / 100) + 32 °F Adjustment example: Temperature unit= °C, the adjustment is carried out in 2 separate steps at 0 °C (e.g. ice water) and at a reference temperature (e.g. clinical thermometer with water bath 37 °C).  First set the values of t.oF and t.SL in the adjustment menu to 0

B-H89.0.1X.DK2-2.0

Page 11 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Operation

 Zero point:  Slope:

 Set the temperature probe to 0 °C and let it adjust.  Start the adjustment menu and enter the display value at 0 °C for
t.oF
 After leaving the menu, the device should now display 0.0 °C.
 Set the temperature sensor to the reference temperature and let it adjust.
 Calculate slope correction: .  =  - 1  100  Start adjustment menu and enter the calculated value at t.SL
 Exit menu, the device should now display the reference temperature.

B-H89.0.1X.DK2-2.0

Page 12 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
5 Measurement Basics

Measurement Basics

5.1 Sensor/device accuracy

The device can be equipped with different interchangeable sensors. The temperature sensors are divided into the following classes according to EN 60584-2.

class

deviation

Temperature range

1

±1.5 °C or ±0.4 % of measured value * -40 ... 1000 °C

2

±2.5 °C or ±0.75 % of measured value * -40 ... 1200 °C

3

±2.5 °C or ±1.5 % of measured value * -200 ... 40 °C

* = larger value counts

5.2 Possible measuring errors

5.2.1 Immersion depth

 Liquid:  Gas:

 Immerse to a depth of at least 20 mm and then stir. Otherwise, measuring errors can occur due to the heat transmission of the sensor tube if the immersion depth is too shallow.
 Immerse as far as possible into the gas to be measured so that the measuring sensor is subjected to a heavy flow.

5.2.2 Surface effects and poor heat transfer
Special measuring sensors are required for this purpose. Surface characteristics, de-sign of the measuring sensor, heat transfer and environmental temperature influence the measurement result.
Note Thermally conductive paste between the measuring sensor and surface can also in-crease measurement accuracy in some cases.

5.2.3 Cooling / evaporation
When measuring the air temperature, the probe should be dry, otherwise the temperature measured will be too low.

B-H89.0.1X.DK2-2.0

Page 13 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Measurement Basics

5.2.4 Response time
An adequate wait time must be observed for the measuring process before reading the measured value. The response time T90 describes the time in which the displayed measured value reached 90% of the end value.

B-H89.0.1X.DK2-2.0

Page 14 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
6 Operation and maintenance

Operation and maintenance

6.1 Operating and maintenance notices
Note The device and temperature probe must be handled with care and used in accordance with
the technical data. Do not throw or strike. Plugs and sockets must be protected from soiling. If the not used for an extended period of time, the batteries must be removed. Leaks from
the batteries are avoided as a result.

6.2 Battery
6.2.1 Battery indicator If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time. If the BAT display text appears in the main display, the battery voltage is no longer adequate for operation of the device. The battery is fully depleted.
6.2.2 Changing battery
Danger! Danger of explosion! Using damaged or unsuitable batteries can generate heat, which can cause the batter-ies to crack and possibly explode!  Only use high-quality and suitable alkaline batteries!
Caution! Damage! If the batteries have different charge levels, leaks and thus damage to the device can occur.  Only use high-quality and suitable alkaline batteries!  Do not use different types of batteries!  Remove depleted batteries immediately and dispose of them at a suitable collection
point.

B-H89.0.1X.DK2-2.0

Page 15 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Operation and maintenance

Note
 Unnecessary unscrewing endangers the protection against moisture and should therefore be avoided.
 Read the following handling instructions before replacing batteries and follow them step by step.
 If disregarded, the device could be damaged or the protection from moisture could be diminished.

1 Unscrews the Phillips screws (A)and remove the cover. 2 Carefully replace the two Mignon AA batteries (B). Ensure
that the polarity is correct! It must be possible to insert the batteries in the correct position without using force. 3 The O-ring (C) must be undamaged, clean and positioned at the intended depth. 4 Fit the cover (D) on evenly. The O-ring must remain at the intended depth! 5 Tighten the Phillips screws (A).

B-H89.0.1X.DK2-2.0

Page 16 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Error and system messages

7 Error and system messages

Display ----

Meaning
No suitable measuring probe connected Measurement far outside of the measuring range

No display, Battery depleted

unclear

System error

characters or Device is defective no response

when buttons

are pressed

BAT

Battery depleted

ERR.1

Measuring range exceeded

ERR.2

Measuring range is undercut

ERR.7

System error

ERR.1 1 SYS ERR

value could not be calculated
System error

Possible causes

Remedy

· Incorrect measuring probe
· No 2 probes connected for diff. measurement
· Measuring probe or device defect

 Connect suitable measuring probe(s)
 Measurement leaves the permissible range
 Send in for repair

· Battery depleted  Replace battery · Error in the device  Send in for repair

· Battery depleted  Replace battery

· Measurement too high
· Incorrect measuring probe connected
· Measuring probe or device defect

 Stay within allowable measurement range
 Check measuring probe  Send in for repair

· Measurement too low
· Incorrect measuring probe connected
· Measuring probe or device defect

 Stay within allowable measurement range
 Check measuring probe  Send in for repair

· Cold junction outside operating range or defect

 Stay within allowable operating temperature
 Send in for repair

· Min. 1 measured value invalid

 Operate within the permissible measuring range

· Error in the device

 Switch device on/off  Replace batteries  Send in for repair

B-H89.0.1X.DK2-2.0

Page 17 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Technical data

8 Technical data

Sensor connection Supported sensor type Measuring ranges Accuracy (at nominal temperature)
Device Cold junction Temperature drift Measuring cycle Display
Additional functions
Adjustment Housing
Protection rating
Dimensions L*W*H Weight Nominal temperature Environmental conditions Current supply Current requirement battery life
Battery indicator
Auto-power-OFF function

Connection sockets for miniature flat plug Thermocouple type K (NiCr-Ni) -65.0 ... +1200.0 °C (-85.0 ... +2192 °F)
Note The system accuracy depends on the accuracy of the used temperature sensor!
±0.1 % of m.v. ±1 °C ±0.3 °C ±0.01 % of m.v. / K ±0.025 °C / K approx. 3 measurements per second 3-line segment LCD, additional symbols, illuminated (white, luminous duration adjustable), Display orientation 180° rotatable (overhead display) Min/max/hold, alarm (optical and acoustic) Offset and gradient correction Break-proof ABS housing IP40 (sensor socket) IP65 / IP67 (rest of the housing) 108 * 54 * 28 mm approx.. 125 g, incl. batteries 25 °C -20 to 50 °C; 0 to 85 %RH (non-condensing) 2 * AA batteries (mignon) approx. 1 mA, approx. 3 mA with backlight Service life > 2500 hours with alkaline batteries (without backlighting) 4-stage battery status indicator, Replacement indicator for depleted batteries: "BAT" The device switches off automatically if this is activated

B-H89.0.1X.DK2-2.0

Page 18 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer

Disposal

Directives and standards

The devices conform to the following Directives of the Council for the harmonization of legal regulations of the Member States:
· 2014/30/EU EMC Directive
· 2011/65/EU RoHS
Applied harmonized standards:
· EN 61326-1:2013 Emission limits: Class B Immunity according to Table 1 Additional errors: < 1 % FS
· EN IEC 63000:2018
The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.

9 Disposal
Separation by material and recycling of device components and packaging must take place at the time of disposal. The valid regional statutory regulations and directives applicable at the time must be observed.
Note The device must not be disposed of with household waste. Return it to us, freight prepaid. We will then arrange for the proper and environmentally-friendly disposal.
 Private end users in Germany have the possibility of dropping off the device at the municipal collection center. Batteries must be removed beforehand!
 Please dispose of empty batteries at the collection points intended for this purpose

B-H89.0.1X.DK2-2.0

Page 19 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
10 Service

Service

10.1 Manufacturer If you have any questions, please do not hesitate to contact us.

10.2 Calibration and adjustment service
The purpose of the calibration is to verify the precision of the measuring device by comparing it with a traceable reference. Both ISO calibration certificates and DAkkS calibration certificates are available from Senseca.
Note · The ISO standard 9001 is applied for the iso-calibration certificates. · These certificates area affordable alternative to the DAkkS calibration certificates and
provide information of the traceable reference, a list of individual values and documentation. · The DAkkS calibration is based on DIN EN ISO/17025, the accreditation basis is recognized worldwide. These certificates offer high-quality calibration and consistently high quality. The DAkkS calibration includes any necessary adjustment with the purpose of minimizing a deviation of the measuring device. · The device is delivered with a test report. · This confirms that the measuring device has been adjusted and tested, without making any statement about the accuracy of a temperature sensor. · Only the manufacturer can check the basic settings and make corrections if necessary.

B-H89.0.1X.DK2-2.0

Page 20 of 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
10.3 Accessories

Service

10.3.1 Spare parts: GB-AA-2

art. no. 479249

Spare batteries AA (2 pcs.)

10.3.2 Suitable sensors:

GTE 130 OK

art. no. 601483

GF 1TK-T3

art. no. 609695

GF 1TK-E3

art. no. 609697

GF 1TK-E1.5

art. no. 609699

GF 1TK-L3

art. no. 611299

GTF 300

art. no. 600072

GTF 40K-620

art. no. 610829

Insertion probe without cable, Ø1.5 mm
Compact immersion probe with silicone handle, Ø3 mm
Compact insertion probe with silicone handle, Ø3 mm
Compact insertion probe with silicone handle, Ø1.5 mm
Compact air probe with silicone handle, Ø3 mm
Wire probe for measurements in fractions of a second
T-handle insertion probe for bulk material, soil, compost

10.3.3 Further accessories:

ST-G1000 GCLIP 1000 G1000_BASE GKK 1000

art. no. 611373 art. no. 475820 art. no. 481885 art. no. 611603

GKK 252

art. no. 601056

Protective bag with belt clip
Metal belt clip, self-adhesive
Table stand, wall holder
Case (235 x 185 x 48 mm), with punched lining for 1 device of the G1xxx-series
Case (235 x 185 x 48 mm), with foam lining for universal use

B-H89.0.1X.DK2-2.0

Page 21 of 22

senseca.com
Senseca Germany GmbH Hans-Sachs-Straße 26
93128 Regenstauf GERMANY
INFO@SENSECA.COM WEEE reg. no.: DE 93889386



References

Munta Anca / Deml Klaus Adobe PDF Library 24.2.121