Volt/mA Calibrator

Instruction Sheet
Introduction
The Fluke 715 Volt/mA Calibrator is a source and measurement tool for 0 to 24 mA current loop testing and dc voltage from 0 to 20/25 V. The calibrator does not source and measure simultaneously.
Your calibrator is supplied with a holster, a set of test leads, an installed 9 V alkaline battery, and this instruction sheet.

Summary of Calibrator Capabilities

Function

Range

Resolution

dc mV input dc mV output dc V input dc V output dc mA input dc mA output Loop power output

0 to 200 mV 0 to 25 V 0 to 20 V 0 to 24 mA
24 V dc output

0.01 mV 0.001 V 0.001 mA N/A

PN 650314 July 1997 Rev. 3, 8/05 1997-2005 Fluke Corporation. All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.

If the calibrator is damaged or something is missing, contact the place of purchase immediately. Contact your Fluke distributor for information about accessories. To order replacement parts or spares, see "Replacement Parts."

To contact Fluke, call one of the following telephone numbers:

USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181
Singapore: +65-*-276-6196
Anywhere in the world: +1-425-356-5500

Address correspondence to:

Fluke Corporation P.O. Box 9090, Everett, WA 98206-9090 U.S.A.

Fluke Europe B.V. P.O. Box 1186, 5602 BD Eindhoven The Netherlands

Visit us on the World Wide Web at: www.fluke.com

Symbol
J I M W
T
$
P

International Symbols Meaning
Earth ground
Fuse
Battery
Refer to this instruction sheet for information about this feature. Double insulated
Conforms to relevant Canadian Standards Association directives. Conforms to European Union directives

Safety Information
Use the calibrator only as specified in this sheet, otherwise the protection provided by the calibrator may be impaired.
A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test.
W Warning
To avoid possible electric shock or personal injury:
· Never apply more than 30 V between any two jacks (terminals), or between any jack and earth ground.
· Make sure the battery door is closed and latched before you operate the calibrator.
· Remove test leads from the calibrator before you open the battery door.
· Do not operate calibrator if it is damaged.
· Do not operate the calibrator around explosive gas, vapor, or dust.
· When servicing the calibrator, use only specified replacement parts.
Caution
Use the proper jacks, function, and range for your measurement or output application.
Turning the Calibrator On
Press the green O pushbutton to turn the calibrator on and off.
For longest battery life:
· For mA output, use current simulate mode instead of source mode when an external 24 to 30 V loop power supply is available. (See Simulating a Transmitter.)
· Turn off the calibrator when not in use.
Auto Shut-Off (Power Saver) Feature
The Calibrator automatically turns off after 30 minutes of inactivity. To reduce this time or disable this feature:

1. With the Calibrator OFF, press the O key. P.S.xx is
displayed, where xx is the turn-off time in minutes. OFF means the power saver is disabled.
2. Press ] and/or } to increase or decrease the turn-
off time in minutes.
3. To disable, press } until the display shows OFF.
HART TM Resistor Mode
The calibrator has a user selectable 250 ohm HARTTM resistor to facilitate use with HARTTM communication devices. The resistor can be switched in/out at any time by simultaneously pressing
the V and A keys. Use a HARTTM communicator when
measuring dc mA with loop power or sourcing mA.
Span Check Feature
The calibrator allows the users to store 0% and 100% setpoints for each output function. Once setpoints are stored, the span check feature allows the user to quickly toggle back and forth from 0% to 100% or to step in 25% increments.
Automatic step and ramp modes can be enabled while in the
span check mode by simultaneously pressing the ] } keys.
First select the desired output mode (V,mV or mA) then preceed to store the setpoints:
1. Storing 0% and 100% Setpoints
2. Using the ] } controls set the output to the desired
output value for 0%.
3. Press the ] } scroll keys simultaneously to store
the 0% value.
4. Now using the ] } controls set the output to the
desired output value for 100%
5. Again, press the ] } scroll keys simultaneously to
store the 100% value.

Measuring dc Volts
INPUT

Press so that INPUT is on the display

V

mA

mV

%

mV or V depending on range
V Press to toggle mV mV and
V ranges

Sourcing dc Volts
OUTPUT

Press so that OUTPUT is on the display

V

mA

mV

%

it01i.eps
mV or V depending on range
V Press to toggle mV mV and V ranges

it02i.eps

Measuring dc mA
INPUT

Press so that INPUT is on the display

V

mA

mV

%

mA %

Press to toggle

mA and % of

4-20 mA scale

it03i.eps

Measuring dc mA with Loop Power

INPUT
Press so that INPUT is on the display

%

V

mA

mV

%

mA %

Press to toggle

mA and % of

4-20 mA scale

+24V output

Two-wire transmitter

HARTTM Communicator

it06i.eps

Using the Current Output Modes
The calibrator provides current output in mA or percent display. Percent is -25.00 to 125.00%, where 0% is 4 mA, and 100% is 20 mA.
In source mode, the calibrator supplies the current. In simulate mode, the calibrator simulates a two-wire transmitter in an externally-powered current loop.
Sourcing mA
Use source mode whenever you need to supply current into a passive circuit such as a current loop with no loop supply. Insert the test leads into the OUTPUT + and - mA jacks as shown below.
Note
A path must exist for current to flow between the OUTPUT + and -mA jacks, otherwise the display will indicate an overload (OL) when you set an output value.

OUTPUT
Press to step up/down 4 mA (25%)

715 VOLT/mA CALIBRATOR

V

mA

mV

%

mA %

Press to toggle

mA and % of

4-20 mA scale

Press so that OUTPUT is on the display.

INPUT

+ LOOP 30V mA 30V COM 30V V

MAX

MAX

MAX

- mA SIMULATE, +

- mA +

VOLTS

OUTPUT (30V MAX)

Press to scroll up/down 0.001 mA. Hold down to scroll faster.

HARTTM Communicator

it04i.eps

Simulating a Transmitter
Use simulate mode when an external 24 to 30 V loop power supply is available.
Insert the test leads into the mA SIMULATE - and + jacks as shown below.

OUTPUT
Press to step up/down 4 mA (25%)
Press so that OUTPUT is on the display

V

mA

mV

%

COM

mA %

Press to toggle

mA and % of

4-20 mA scale

Press to scroll up/down 0.001 mA. Hold down to scroll faster.

30V dc Power max. supply

it05i.eps

Maintenance
For maintenance procedures not described in this sheet, contact a Fluke Service Center.
In Case of Difficulty
· Check the battery and test leads. Replace as necessary.
· Review this sheet to make sure you are using the correct jacks and pushbuttons.
If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement below for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.

Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center.
Replacing the Battery W Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator (B) appears.
it07i.eps

Replacement Parts and Accessories

Item

Replacement Parts

Description

PN or

Model

Qty.

no.

BT1

9V battery, ANSI/NEDA 1604A or 614487

1

IEC 6LR61

H80M

Holster, Yellow, hanging

H80M

1

MP85

Case top

620200

1

MP86

Case bottom

2397526

1

H2, 3, 4 Case screw

832246

3

MP89, 90 Non-skid foot

824466

2

MP8

O-ring for input/output receptacle 831933

1

MP92

Battery door

619947

1

H5, 6

Battery door fasteners

948609

2

S1

Keypad

687100

1

TL75

Test lead set

TL75

1

-

Instruction sheet

650314

1

AC72

Alligator clips

AC72

1

TL20

Industrial test lead set

TL20

Option

-

71X Calibration Manual

686540 Option

MP85
MP8 BT1 MP86

S1
mV V % mA

H2, 3, 4

H5, 6

MP92

Instruction Sheet

AC72 Alligator Clips

TL75 Test Lead Set

MP89, 90

Holster

it10c.eps

Specifications
Specifications are based on a one year calibration cycle and apply from +18°C to +28°C unless stated otherwise. "Counts" means number of increments or decrements of the least significant digit.
DC V Input and Output

Range

Resolution

Accuracy, ±(% of Reading + Counts)

200 mV

0.01 mV

0.015% + 2

20 V output 25 V input

0.001 V

0.01% + 2

Input impedance: 1 M (nominal), < 100 pF Overvoltage protection: Fuseless Voltage drive capability: 1 mA

DC mA Input

Range

Resolution

24 mA

0.001 mA

Overload protection: Fuseless

Accuracy, ±(% of Reading + Counts)
0.01% + 2

DC mA Output
Range: 0 mA to 24 mA Percent display: 0% = 4 mA, 100% = 20 mA Accuracy: ±(0.01% of reading + 2 counts) Source mode:
Compliance: 1000  at 20 mA for battery voltage 6.8 V (700  at 20 mA for battery voltage 5.8 to 6.8 V)
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power
24 V ±10%

General Specifications
Maximum voltage applied between any jack and earth ground or between any two jacks: 30 V
Storage temperature: -40°C to 60°C
Operating temperature: -10°C to 55°C
Operating altitude: 3000 meters maximum
Temperature coefficient: ±0.005% of range per °C for the temperature ranges -10 to 18°C and 28 to 55°C
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 to 500 Hz
Shock: 1 meter drop test
Safety: Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L);
With holster: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight: 349 g (12.3 oz); With holster: 601 g (21.2 oz)

LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, send your defective calibrator to the nearest Fluke Authorized Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.

