User Guide for LIFESCAN models including: DCO 06078902A, OUS QRG US, OneTouch UltraSmart Blood Glucose Meter, OneTouch, UltraSmart, Blood Glucose Meter, UltraSmart Blood Glucose Meter, OneTouch UltraSmart Blood Glucose Meter

DCO 06078902A OUS QRG US

macuser

DCO 06078902A OUS QRG US - northcoastmed.com

Enter FastFacts . Press FastFacts SmartButton. Select “Logbook” from Menu. Press . Press SmartButton for type of electronic logbook entry (exercise, health, medication, or food). Enter Set-Up Mode.


File Info : application/pdf, 2 Pages, 172.56KB

PDF preview unavailable. Download the PDF instead.

OneTouch-UltraSmart-Quick-Reference-Guide-English
Coding the Meter Code the meter before using it for the first time and every time you change to another vial of OneTouch® Ultra® Test Strips.
CAUTION: Matching the code on the meter and the code on the test strip vial is essential to obtain accurate results. Each time you test, check to make sure the code numbers match.
STEP 1 Enter Code Mode.
Insert test strip to turn on meter. "OneTouch UltraSmart" will appear, followed by code number.
STEP 2 See Code Number.
Code number will appear on display.

(Example) (Example)

STEP 3 Match Code Numbers.
Compare code number on meter display with code number on test strip vial. If matching, begin testing. If not matching, see Step 4.

STEP 4 Code Meter. Scroll up or down with Arrow button . Meter is ready for testing.
(Example)
Test Summary Note: Before testing, wash hands with warm, soapy water. Rinse and dry thoroughly.
STEP 1 Insert Test Strip. Push it all the way in until it will go no further. Meter turns on automatically.

Apply Blood Alternate Site

STEP 2 Select Test Mode.
Use Arrow button to select "Apply Blood" for fingertip sample or "Alternate Site" for forearm and palm samples. Do Not select Control Solution mode when testing blood. Failure to select correct test mode may cause blood glucose averages to be incorrect.

STEP 3 Apply Sample.
Touch and hold drop of blood to the narrow channel in the top edge of the test strip.

STEP 4 Fill Confirmation Window.
Be sure confirmation window is full before meter counts down. If window is not full before meter begins countdown, repeat test with new test strip. Do Not add more blood to the test strip; discard the test strip and retest.

Sep 8-02 10:30 am
104 mg/dL
(Example)

STEP 5
Accurate Results in Just 5 Seconds.
WARNING: If mg/dL does not appear with the test result, call LifeScan Customer Service at 1 800 227-8862.

Display Symbols and Messages

Code
---
Check Code

1) New meter not coded. Code meter; or
2) Meter lost power briefly and entered Set-up mode. Reset code number.

Sep 14-02 9:45 am
414 mg/dL Check Ketones Comment? Press OK
(Example)
Sep 14-02 9:30 am
35 mg/dL Do you need a snack? Comment? Press OK
(Example)

Test result with suggestion to check ketone levels.
Test result is from 20 through 59 mg/dL with meal or snack reminder.

WARNING LOW
GLUCOSE
Below 20 mg/dL

Blood glucose level lower than 20 mg/dL.

WARNING HIGH
GLUCOSE
Above 600 mg/dL
WARNING Error 1
Call customer service See Owner's Booklet

Very high blood glucose level exceeding 600 mg/dL.
There may be a problem with meter. 1) Turn off meter. Enter Set-up mode.
Reset meter before testing; or 2) Turn off meter. Enter Set-up mode.
Reset language, date, time and meter code. If Error 1 message continues, Do Not use the meter. Call LifeScan Customer Service at 1 800 227-8862 for a replacement meter.

WARNING Error 2
Meter or strip problem Please repeat test
See Owner's Booklet

Used test strip inserted or problem with meter. Repeat test with new test strip.

WARNING Error 3
Meter not ready Please repeat test
See Owner's Booklet

Blood or control solution sample applied before "Apply Blood", "Alternate Site", or "Control Solution" appeared on display. Repeat test with new test strip.

LifeScan Customer Service toll-free numbers: English 1 800 227-8862 Español 1 800 381-7226 Visit the LifeScan Web site at: www.LifeScan.com
©2006 LifeScan,Inc. Milpitas,CA 95035 Rev. date: 4/2006 AW 060-789-02A
AW 06078902A

WARNING Error 4
Strip problem Please repeat test
See Owner's Booklet

Three possible reasons:
1) Possible high glucose in a test environment near low end of operating temperature range (43­111°F/6­44°C). Retest in a warmer environment. or
2) Sample improperly applied. Review blood application and repeat test with new test strip. or
3) Test strip damaged or moved during testing. Repeat test with new test strip.

WARNING Error 5
Invalid test Please repeat test
See Owner's Booklet

1) Test strip damaged or confirmation window not full. Retest with new test strip.
2) There may be a problem with the meter. If the error message appears again, call LifeScan Customer Service at 1 800 277-8862.

WARNING Low Batteries!
+ You should replace the batteries soon See Owner's Booklet

Battery power getting low. Replace batteries as soon as possible.

WARNING Batteries!
+ You MUST replace the batteries now See Owner's Booklet

Battery power too low for testing. Replace batteries immediately.

WARNING Meter is outside of
operating temperature range
Test not possible See Owner's Booklet

Temperature outside system operating range. Do Not perform a test until the meter and test strips reach a temperature within the operating range (43­111°F/6­44°C).

Adding Comment to Test Result

Sep 8-02 10:30 am
135 mg/dL
Comment? Press OK
(Example)

STEP 1 Complete Blood Test. Start with test result on display.

STEP 2 Press OK Button.
You may add food, health, or exercise comment.

Sep 8-02 10:30 am Gluc. 135 Comment

Food Aft Brkft

Health --------

Exer

--------

Save

STEP 3 Select Comment Type.
Use to highlight type and press .

Sep 8-02 10:30 am Gluc. 135 Comment

Food Health Exer Save

-------Stress
Feel Hypo Illness Menses

STEP 4 Add Comment.
Select comment from menu. Press with comment highlighted.

Sep 8-02 10:30 am Gluc. 135 Comment

Food Health 1 Health 2 Exer Save

Bef Lunch Stress ---------------

STEP 5 Save Comment. Scroll to "Save". Press .

Making Logbook Entry
Press SmartButton® for type of electronic logbook entry (exercise, health, medication, or food).

Exercise SmartButton®
Health SmartButton®

Food SmartButton®
Medication SmartButton®

Basic Meter Set-Up

STEP 1 Enter Set-Up Mode.
Press and hold down BOTH and buttons together for at least two
seconds.

Set up in?
English Español Francais Deutsch Italiano Português Nederlands

STEP 2 Select Display Language.
To use English, press . To use another language, scroll to it with . Press .

Basic Set-up

Time

12:00 am

Date

Sep 15-02

STEP 3 Set Date and Time.

Scroll with and press each is correct.

when

Do you want to customize your meter now?
Yes No

STEP 4 Complete Basic Meter Set-Up.
Select "No" and press . Basic meter Set-up is complete. Select "Yes" and press to customize other meter settings. See Owner's Booklet and next panel for details.*

Accessing Your Logbook To review test results in logbook memory, use FastFacts®.

FastFacts®

STEP 1 Enter FastFacts®. Press FastFacts® SmartButton®.

FastFacts
Logbook Glucose by Meals Glucose Analysis
Insulin Intake Hypo Info
Food Averages Health Checks

STEP 2 Select "Logbook" from Menu. Press .

Logbook
Sep 8-02 10:30 am Glucose 125 mg/dL Bef Lunch Before Exer

STEP 3 Scroll Back through Results. Use to scroll.

Logbook

Sep 15-02 Exercise
Hard

3:30 pm 00:10

*Note: It is important that the time of day is correct and that your medication set-up and mealtimes are set according to your personal daily schedule. Otherwise, some FastFacts® information will not be correct.

Meter and Electronic Logbook In One
Quick Reference
Guide
See inside for instructions on using OneTouch® UltraSmart®. Important: Before testing, read detailed instructions in your OneTouch® UltraSmart® Owner's Booklet.



References

GPL Ghostscript 9.50