User Manual for Pro sKit models including: MT-1710N True RMS Auto Range Multimeter, MT-1710N, True RMS Auto Range Multimeter, RMS Auto Range Multimeter, Auto Range Multimeter, Range Multimeter, Multimeter

MT-1710N USERMANUAL

CYWANG

Download

MT-1710N 3-3/4 True-RMS Auto Range Multimeter | Pro'sKit Official Website

MT-1710N 3 3/4 True-RMS Auto Range Multimeter - Pros'Kit

2 6. Don’t input voltage during the resistance measurement. 7. Test lead should be removed from testing point before changing the batteries or fuses and turn off the power.

PDF preview unavailable. Download the PDF instead.

MT 1710N Users Manual
MT-1710N 3 3/4 True-RMS Auto Range Multimeter
User's Manual 1st Edition, 2023
©2023 Copyright by Prokit's Industries Co., Ltd.

I. INTRODUCTION

The instrument is a stable and high-performance digital multimeter driven by

battery. It uses the LCD with 42mm high to make the readings clearly. The instrument has the function of measuring DCV, ACV, DCA, ACA,

Resistance and Capacitance, Frequency, Diode, Triode, Continuity test,

Temperature, Auto power off (can be canceled) and backlight function. The

instrument takes dual-integral A/D converter as key point, is an excellent tool.

It's an ideal tool for lab, factory, and family.

The new generation series of productions add NCV (non-contact ACV

induction measurement) and torch function based on the original function.
II. OPEN THE PACKAGE FOR CHECKING

Open the box, take out the meter, checking the items below if they are

missing or damaging:

K type probe (-20~250)

1pc

hFE test Adapter

1pc

Test leads

1pair

User manual

1pc

Please contact with your supplier if you find out any problems.
III. SAFETY NOTES

This series meter meets the standard of IEC61010 (the safety standards

request issued by IEC or equal GB4793.1 standards). Read it before

operation.

1. Input over range is prohibited in each range during the test.

2. Voltage less than 36V is safety voltage. When measuring voltage higher

than 36V DC, 25V AC, check the connection, insulation of test leads to

avoid electric shock.

3. When changing function and range, test leads should be removed from

testing point.

4. Select correct function and range.

5. Don't operate the meter when the battery case and back cover are not

fixed

1

6. Don't input voltage during the resistance measurement.

7. Test lead should be removed from testing point before changing the

batteries or fuses and turn off the power.

8. Comply with local and national safety regulations, wear personal protective

equipment (approved rubber gloves, masks and flame-retardant clothing,

etc.), to prevent electric shock and arc injury caused by exposure of

dangerous electric conductor.

9. Make measurements using only the correct measurement standard class

(CAT), voltage and current rated probe, test conductor and adapter. 10Safety symbols:

" " exists high voltage " " GND

" " Dual Insulation

" " Must Refer to Manual "

" Low Battery

IV.

SAFETY SYMBOLS

V.

SPECIFICATIONS

GENERAL SPECIFICATIONS

1) Display mode: LCD displaying. 2) Max. indication: 39993 3/4, auto polarity indication

3) Measuring method: Dual integral A/D conversion

4) Sampling rate: Approx. 3times/second

5) Over range indication: "OL" displays

6) Low battery: The " " displays.

7) Operating environment: (0 ~ 40) , humidity<75%RH. 8) Storage environment:-20~60humidity<85%RH

2

9) Power: Four 1.5V AAA batteries, LR03 10) Dimension: 186mm×92mm ×52mm. 11) Weight: approx. 355g (not include batteries).

VI.

EXTERNAL STRUCTUR

1. Sound alarm indicator light

2. LCD display

3. Function / range knob

4. Measuring input terminal

5. Function selection button

6. Auto / manual range

conversion

7. Relative value measurement

/ flashlight control

8. Data hold / backlight control

9. NCV induction position

10. Flashlight window

11. Hang hole

12. Screws for fixing the battery box

13. Bracket

14. Test leads holder

VII. LCD DISPLAY

3

1

Auto range

2

Auto power off

3

DC measurement

4

AC measurement

5

High voltage

6

True RMS

7

Diode/continuity test

8

Data hold

9

Relative value

10

Low battery

11

meaTsruioredme ent

12

Temperature

13

Duty cycle

14 Current / Capacitance

15 Resistance/ Frequency

VIII. KEY DESCRIPTION 1.Data hold: Short press the "HOLD B/L" button to hold the measuring data on the LCD, and the symbol "HOLD" appears. If you press the key again, the data hold function will be closed, and the symbol will disappear.
Warning: to prevent possible electric shock, fire, or personal injury, do not use the data hold function to measure the unknown voltage. When open the HOLD function, the LCD will keep original data when measuring a different voltage.

2.Backlight function: Long press the "HOLD B/L" button to turn on the back light of the LCD, and long press this key again to turn off the back light. 3.Function select key (Hz/DUTY SELECT key) At the DC/AC mVuAmAA range, short press the "Hz/DUTY SELECT" key to convert among DCV, ACV, and DCA, ACA for testing; At the diode/continuity test range, short press the "Hz/DUTY SELECT" key to convert the diode and continuity test, the diode range is already set by default. At the ACV range, short press the "Hz/DUTY SELECT" key to cycle switch among the voltage, accompanying frequency, duty cycle. At the temperature range, short press the "Hz/DUTY SELECT" key to convert the Celsius and Fahrenheit; At the
4

Frequency range, short press the "Hz/DUTY SELECT" key to convert the Frequency and duty cycle test. 4.Auto / manual range conversion The auto range is already set by default after the meter is turned on, short press the "RANGE" key, the meter convert to manual range, the "AUTO" symbol on the LCD will disappear. At the manual range mode, the meter starts from the low range, when you press the "RANGE" key, the range change to a higher one, the procedure repeats like this in sequence; long press "RANGE" key to exit the manual range mode, and the LCD displays "AUTO" symbol. 5.Relative measurement At the DC/AC mVVuAmAA range, capacitance, temperature range, short press the "REL" key to enter the relative measurement mode, and the "REL" symbol appears on the LCD. Short press the "REL" key again, the meter exits the relative measurement mode, and the "REL" symbol disappears. 6. Lighting Long press "REL/ " to turn on/off the Flashlight. 7. Cancel the auto power off function Press the "Hz/DUTY SELECT" key to turn on the meter, the "APO" symbol disappears, and then the auto power off function closed.

IX.

MEASUREMENT INSTRUCTIONS

First, please check the battery, and turn the knob to the proper range

that you need. If the battery is out of power, the " " symbol will

appear on the LCD. Pay attention to the symbol next to the jack

for test leads. This is a warning that the voltage and current should not

exceed the indicated value.

1. DCV and ACV measurement

5

1-1. Set the knob to the mV/ACV/DCV range, and connect the test leads across to the circuit under test. The voltage and polarity of the point which the red test lead is connected are display on the LCD.
1-2. Insert the black test lead to "COM" jack, the red one to " " jack.
1-3.The auto range is already set by default after the meter is turned on, LCD displays "AUTO" symbol, press the "RANGE" key to convert to manual range, the range 40mV,400mV (only for DC mV/AC mV range), 4V, 40V, 400V, 1000V/750V (DCV/ACV) are optional.
1-4.You can get the result from LCD. 1-5. At the DC mV/AC mV range, short press the "Hz/DUTY SELECT"
key to convert DC/AC mV measurement. The DC mV range is already set by default after the meter is turn on. Note: (1) Do not input a voltage over DC 1000V or AC 750Votherwise the circuit might be damaged. At the DC/AC mV range, the input voltage should not exceed 250VDC/ AC RMS. (2) Please keep the test leads away from the circuit after finished your testing. (3) When measured voltage is higher than 220V, it's necessary to wear personal protective equipment (such as approved rubber gloves, face masks, and flame-retardant clothing etc.) to prevent the injury from electric shock and arc. 2. DCA and ACA measurement 2-1. Set the knob to a proper DCA/ACA range, and then connect the test leads to the power or circuit under test in series. The polarity and the current value of this point which connects to the red test lead will display on the LCD. 2-2. Insert the black test lead to "COM" jack, the red one to " " (max. 400mA) or "10A" (max. 10A) jack. 2-3. You can get the result from display.
6

2-4. Short press the "Hz/DUTY SELECT" key to convert DCA or ACA, the DCA range is already set by default after the meter is turned on. Short press "RANGE" key to convert to manual range, you can choose 400uA, 4000uA at the uA range, and 40mA, 400mA are optional at the mA range, 4A, 10A are optional at the A range. The auto range is already set by default after the meter is turned on.
Note: (1) If the tested current range is unsure beforehand, please start from
the highest range then find the proper range according to the reading. Do not measure the voltage in the current jack. (2) If the reading displays "OL"the value is beyond the present range. Now you need to set the knob to the higher range. (3) Be careful when measuring 10A range. Continuous measurement of high current will heat the circuit, even affect the accuracy and damage the meter. (4) When measuring the large current above 10A, it's necessary to wear personal protective equipment (such as approved rubber gloves, face masks, and flame-retardant clothing etc.) to prevent the injury from electric shock and arc. 3. Resistance measurement 3-1. Set the knob to resistance range, then connecting the test leads to the tested resistance. 3-2. Insert the black test lead to "COM" jack, the red one to " "jack. 3-3. You can read the measurement results from the display. 3-4. Auto range method is already set by default after the meter turned on, short press "RANGE" key to convert to the manual range, The range 400, 4K, 40K, 400K, 4M, 40M are optional. Note: (1) If "OL" displays on LCD, it means over range, now you need to set the range knob to the higher range. When measuring resistance is
7

more than 1M, the reading may take a few seconds to stabilize. This is normal for high resistance measurement. (2) When the input terminal is under open circuit, the figure "OL" will be displayed for the overload condition. (3) When checking in-circuit resistance, be sure the power has been switched off and all capacitors are fully discharged. 4. Capacitance measurement 4-1. Set the range knob to a proper capacitance range, connect the test leads to the capacitor under tested (the polarity of red test lead is "+"). 4-2. Insert the red test lead to " " jack and the black one to "COM" jack. 4-3. You can get the result from LCD. Note: (1) If the LCD displays "OL", it indicates the value is over range. The maximal range is 100mF. (2) When measuring capacitance, because of lead wire and the distributed capacitance, there may be some residual reading without connection to capacitance under measured. It will be clearer when measuring small capacitance range. Measuring result is to subtract the residual reading for getting accurate reading. This will not affect the measuring accuracy. You can short press the "REL" key to clear the residual reading of open circuit, then measuring the relative value. (3) When under a large capacitance range, if capacitor is broken or leakage, the LCD will display some value and it's unstable. (4) Discharge all the capacitance completely before measuring to avoid the damage of the meter. (5) UNIT: 1mF=1000uF 1uF =1000nF 1nF=1000pF 5. TRANSISTOR hFE 5-1. Set the knob to "hFE" range. 5-2. Insert the adapter to " " and "COM" jack, then according to
8

the NPN or PNP, insert emitter, base and collector to proper jack. 5-3. You can get the result from LCD. 6. Diode and continuity test 6-1. Set the range knob to the " " range. The diode range is already set by default after the meter is turned on, short press the "Hz/DUTY SELECT" key to convert the diode and continuity test. Set the black test lead to "COM" jack, and the red one to " "jack (the polarity of red test lead is "+"), meanwhile connect the red test lead to the positive of the diode and the black one to the negative. The reading is the approx. value of the diode positive volt drop. If the test leads connect to the reverse polarity of diode, the LCD will display "OL". 6-2. Connect the test leads to two ends of the circuit under tested, if the resistance is lower than approx.50±20, the LCD displays " " symbol, and the buzzer sounds. 7. Frequency measurement 7-1.Turn the knob to frequency range, and connect the test leads or shielded cable across the signal source or the load under tested. 7-2.Insert the black test lead to the "COM" jack and the red one to the " "jack. 7-3.You can get the result from the LCD. 7-4.Short press the "Hz/DHTY SELECT" key to convert to duty cycle. Note: (1) When it is over 10V RMS, you can get the result, but we cannot promise the accuracy. (2) It is preferable to use shielded cable for measuring small signal in noisy environment. (3) Be careful when measuring high voltage. (4) Do not input a voltage over DC 250V or AC peak value to avoid damage to the meter. (5) It is auto range at the frequency range, you can test 10Hz ~20MHz. 8. Temperature measurement
9

When measuring temperature, insert the negative of the thermocouple sensor into the "COM" jack, and insert the positive into the " " jack; put the sensing part of the temperature probe on or inside the tested object. Read the value from the LCD which shows as Celsius. Short press the "Hz/DHTY SELECT" key can convert the Celsius or Fahrenheit unit. 9. Non-contact voltage induction measurement NCV 9-1. Turn the knob to "NCV" range. 9-2. NCV induction voltage range is 48V~220V, let the upper position of the meter close to the measured electric AC power line, when the ACV is sensed, the upper red indicator lights of the meter flashing and at the same time a buzzer alarm on, the more close to the AC power line, the more stronger the ACV sensed, the corresponding flashing lights, and the buzzer alarm sounds faster. 10. Automatic power on/off To save power consumption and prolong battery life, the APO automatic shutdown function will be turned on by default after the meter is turned on. If the user does not operate the meter within 14 minutes, the meter will be prompted with 3sounds. If there is still no operation, the meter will automatically turn off the power after a long sound one minute later. Short press "SELLECT" can turn on the multimeter. To cancel APO function, please refer to the instructions in section 8, "key function".

X.

Technical characteristics

Accuracy: ± (a% × reading + digit)

Environment temperature: (23±5) 

Relative humidity: <75%RH.

One year guarantee since production date.

10-1. DC Voltage
10

Range
4V 40V 400V 1000V

Accuracy

Resolution

±(0.5%+4d) ±(1.0%+6d)

1mV 10mV 100mV
1V

Input impedance

Overload protection

About 10M

1000V DC / AC RMS

10-2. DC mV

Range Accuracy

Resolution

40mV ±(0.5%+5d)
400mV

0.01mV 0.1mV

Input impedance
>40M

Overload protection 250VDC / AC RMS

10-3. AC mV (True RMS)

Range

Accuracy

Resolution

40mV ±(1.0%+6d) 400mV

0.01mV 0.1mV

Input impedance
About 10M

Overload protection
250V DC/AC RMS

10-4. ACV (True RMS)

Range

Accuracy

Resolution

Input impedance

Overload protection

4V 40V 400V
750V

± (0.8%+10d)
± (1.2%+10d)

1mV 10mV 100mV

About 10M

1V

1000V DC / 750AC RMS

Accuracy measurement range: 10%-100% of the range.

Frequency response: 40Hz-1 kHz.

Measurement mode (sine wave): true RMS.

11

Crest factor: CF3, adding an additional error of 1% to the reading at CF2.
AC frequency measurement error: 0.2%+0.02Hz. AC frequency measurement range: 40Hz~1k Hz. AC frequency input sensitivity: 80V~600V

10-5. DCA

Range

Accuracy Resolution

Load voltage

Overload protection

400uA 4000uA 40mA 400mA

±(1.0%+10d) ±(1.2%+8d)

0.1uA 1uA 10uA 100uA

0.1mV/ mA 0.1mV/ mA 1.552mV/ mA 1.637mV / mA

FUSE 400mA/250V

4A

1mA

±(1.2%+10d)

10A

10mA

31.789mV/ A 31.789mV/ A

FUSE 10A/250V

10A (test within 10 seconds); Recovery time is 15 minutes.

10-6. ACA (True RMS)

Range

Accuracy Resolution Load voltage

Overload protection

400uA

0.1uA 0.1mV/ mA

4000uA ±(1.5%+10d) 40mA

1uA 10uA

0.1mV/ mA 1.552mV/ mA

FUSE 400mA/250V

400mA

100uA 1.637mV / mA

4A

±(2.0%+5d)

1mA

31.789mV/ A

10A

10mA 31.789mV/ A

FUSE 10A/250V

Measuring range of accuracy value: 10% - 100% of the range.

Frequency response: 40Hz - 1k Hz

Measuring way (sine wave): True RMS

Crest factor: CF3, when CF2, add an additional error of 1% of the

12

reading. 10A (The measurement should not be more than 10 seconds); Recovery time is 15 minutes.

10-7. Resistance

Range

Accuracy

Short-circuit Resolution
current

400 ± (0.8%+5d) 0.1 About 0.4mA

4k

1

About 100uA

40k ± (0.8%+4d)
400k

10 100

About 10uA About 1uA

4M

1k

About 0.2uA

40M ± (1.2%+10d) 10k About 0.2uA

Measuring error does not include lead resistance.

Overload protection: 250VDC/AC RMS

Open-circuit voltage
About 1V
About 0.5V

10-8. Capacitance

Range 6nF

Accuracy ±(5.0%+40d)

Resolution 0.001nF

Overload protection

60nF

0.01nF

600nF 6uF

±(3.5%+20d)

0.1nF 0.001uF

60uF

0.01uF

250VDC/AC RMS

600uF

0.1uF

6mF

±(5.0%+10d) 0.001mF

60mF 100mF

±10%

0.01mF 0.1mF

Measuring range of accuracy: 10% - 100% of the range

Large capacitance response time: 1uF about 8s

Measurement error does not include lead distribution capacitance.

13

10-9. Frequency

Range

Accuracy

Resolution Overload protection

10Hz

0.01Hz

100Hz

0.1Hz

1kHz 10kHz

±(0.5%+10d)

1Hz 10Hz

250VDC/AC RMS

100kHz

100Hz

1MHz

1kHz

20MHz

10kHz

Note: The reading will be zero if the signal is below 3Hz.

Input sensitivity: 1V

10-10. Diode and continuity Test

Range Displaying value

Test Condition

Error

The positive DC Current

Positive voltage

is approx. 1.5mA.

drops of diode

Negative voltage is

approx. 3.2V.

5%

Buzzer sounds, the resistance is less
than 50±20

The test current is approx. 0.4 mA

Warning: Do not input voltage at this range.

Overload protection: 250V DC/AC RMS

10-11. Temperature

Range

Accuracy

Resolutio n

Overload protection

(-20-1000) <400 ± (1.0%+5d)

1

400 ± (1.5%+15d)

250V DC/AC

(-4~1832) <752 ± (1.0%+5d)

1

752 ± (1.5%+15d)

RMS

Sensor: Type-K thermocouple (Ni-Cr ­ Ni-Si) banana probe
14

10-12. Transistor hFE DATA TEST

Range

Displaying range

hFE NPN or PNP

0 ~ 1000

Test condition Basic current is approx. 10uA 
Vce is about 1.5V

XI.

Replace the battery or fuse.

1. Move away the test leads from the circuit under test, pull out the test lead

from the input jack, turn the range knob to the "OFF" range to turn off the

power.

2. Use a screwdriver to twist off the screws on the battery cover and remove

the battery cover and bracket.

3. Take out the old battery or the broken fuse, then replace with a new alkaline

battery 9V or a new fuse.

4. Close the battery cover and use a screwdriver to tighten the screws on the

battery cover.

5. Battery specifications:4 * 1.5V AAA alkaline batteries

6. Fuse specifications:

mA input fuse "FS1": 5 * 20mm 400mA 250V

10A input fuse "FS2": 5 * 20mm 10A 250V

Note: When the low voltage " " symbol displays on the LCD, the battery

should be replaced immediately, otherwise the measuring accuracy will be

affected.

XII.

MAINTENANCE AND CARE

It is an accurate meter. Do not try to modify the electric circuit.

1. Pay attention to the waterproof, dustproof and break proof of the meter.

2. Please do not store or use it in environment of high temperature, high

humidity, high flammability or strong magnetic.

3. Please wipe the meter with a damp cloth and soft detergent, and abrasive

and drastic solvent such as alcohol are forbidden.

15

4. If do not operate for a long time, should take out the battery to avoid leakage. 5. When replacing fuse, please use another same type and specification fuse.

13. Trouble Shooting

If the meter does not work properly, check the meter as following:

Fault

Solution

Power off--- Pls turn on the

power

No reading on LCD

Holding key-- Pls set a correct mode

Replace battery

The signal appears

Replace battery

No current or temperature input Replace fuse

Error Value

Replace battery

This user's manual is subject to any change without further notice. The content in this user's manual is deemed correct; if you find any mistake, omission, etc., please contact the manufacturer. We will not be held liable for any accidents or harms caused due to your wrong operations. The functions set forth in this user's manual shall not be regarded as reasons for applying this product for special purposes.

16

3-3/4 

 
MT-1710N3 3/4  42mm  LCD  
 DCVACVDCAACA NCV LED   A / D  
 

 x1 K (-20250)x1/ x1  x1
 

 IEC61010  1.  2.  36V  36V DC/25V AC  3.  4.   5.  6.  7.  8.   9. CAT  10. 

" "

" "

" " 

,

" " "

"

 

!







17











 CE 

 



39993 3/4

 A/D 



 3 

 "OL"

"

"

 0 40< 75 RH

-20 60< 85 RH

1.5V  AAAx4LR03

186× 92× 52 

 355 

  1

1.  2. LCD  3. / 4.  5.  6. / 7. // 8. // 9. NCV  10. 11.  12.  13.  14. 
 

18

1 

2 

3 

4 

5 

6 

7 /

8 

9 

10 

11 

12 /

13 

nF/uF/mF; 14 mV/V
uA/mA/A

15 :/K/M; :Hz KHz MHz

 
1. "HOLD B/L" LCD "HOLD"
"HOLD"
 HOLD   HOLD  2.
"HOLD B/L" LCD  3."Hz/DUTY SELECT":
 mVuAmAA "Hz/DUTY SELECT" /   Hz   4./RANGE :
19

"AUTO", "RANGE" 

"AUTO"

"RANGE""RANGE"

"AUTO"

5.REL/

:

 mVVAmAA "REL/ "

"REL"

"REL"

6.REL/

:

"REL/

"/ LED /

7. APO 

"Hz/DUTY SELECT""APO"



 



"

"

"

""

"



1. 

1.1"

" COM 

1.2 mV  ACV/DCV

1.3

"-""+"



1.4"AUTO""RANGE"

 40mV400mV mV 4V40 V400 V1000

/750V

1.5 mV "Hz/DUTY SELECT"



1.6"Hz/DUTY SELECT"



  mV  V 

"V""COM"

20

  . DC1000V  AC750V mV
 250V DC/AC RMS.    220V 


2. 

2.1"

" 400mA 10A 10A

 COM 

2.2A/mA/A

2.3



2.4

2.5"Hz/DUTY SELECT"

"RANGE"A 

 400A4000A mA  40mA 400mA  A  4A

10A 

  
     OL,   10A  10     10A 

21



3. 

3.1. 

3.2. "

" COM 



3.3. 

3.4. "RANGE"

 4004K 40K400K4M40M

              ,  
"OL",  1M ,  "OL"  

4. 

4.1. 

4.2. "

" COM 

4.3. 

    OL 
 100mF.  
   "REL"        1Mf=1000uF 1uF=1000nF 1nF=1000pF
22

5. hFE 

5.1.  hFE 

5.2. hFE "

""COM" NPN  PNP



5.3

6. /

6.1."

"

6.2."

""COM"

6.3.

"OL".

6.4."Hz/DUTY"

6.5. 50±20



7. /

7.1 "Hz" Hz "Hz/DUTY

SELECT"

7.2 "

""COM"



7.3 

  10V rms 
  
    250V 
              
10Hz~20MHz.

8. 

8.1."/"""

"Hz/DUTY SELECT"""

8.2."

""COM"

23

 8.3.
9. NCV
9.1."NCV". 9.2. LED   LED   LED 
  
    

10. 
APO  14  3 1  "SELECT"APO, 

 
+   18 28 75

 (DCV)





4V

40V

±(0.5%+4 )

400V

1000V

±(1.0%+6 )

 1mV 10mV 100mV 1V





 10M 1000V DC/AC RMS

24

 (DCmV)





40mV 400mV

±(0.5%+5 )

 0.01mV 0.1mV

 40M

 250V DC/AC RMS

 mV  (ACmV True RMS)









40mV 400mV

±(1.0%+6 )

0.01mV 0.1mV

 10M

 V  (ACV True RMS)









4V

1mV

40V 400V

±(0.8%+10 )

10mV 100mV

 10M

1000V

±(1.2%+10 ) 1V

 250V DC/AC RMS
 1000V DC 750V AC RMS

 10%100%  40Hz- 1kHz    CP3 CP2  1% 0.2%+0.02Hz; 40Hz- 1kHz80V600V

 (DCA)





400uA 4000uA

±(1.0%+10 )

40mA 400mA

±(1.2%+8 )

4A

10A

±(1.2%+10d)

 0.1uA
1uA 10uA 100uA 1mA 10mA

 0.1mV/mA 0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A

10A  10  15 


FUSE 400mA/250V
FUSE 10A/250V

 (ACA True RMS)







400uA ±(1.5%+10 )

0.1uA

25

 0.1mV/mA

 FUSE

4000uA 40mA 400mA 4A 10A

±(2.0%+5 )

1uA 10uA 100uA 1mA 10mA

0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A

 10%~100%  40Hz- 1kHz    CP3 CP2  1% 10A  10  15 

400mA/250V
FUSE 10A/250V

 ()





400 ±(0.8%+5 )

4k

40k 400k ±(0.8%+4 )

4M

40M ±(1.2%+10 )

 0.1 1 10 100 1k 10k



  0.4mA  100uA  10uA
 1uA  0.2uA  0.2uA

  1V
 0.5V


250V DC/AC RMS

 (F)
 6nF 60nF 600nF 6uF 60uF 600uF 6mF 60mF 100 mF

 ±(5.0%+40 )
±(3.5%+20 )
±(5.0%+10 ) ±10%

 0.001nF 0.01nF 0. 1nF 0.001uF 0.01uF 0.1uF 0.001mF 0.01mF 0.1mF


250V DC/AC RMS

 10%~100% 1mF  8 

26

 (Hz)





10Hz

100Hz

1KHz

10kHz

±(0.5%+10 )

100kHz

1MHz

20MHz

 0.01Hz 0.1Hz 0.001KHz 0.01KHz 0.1KHz 0.001MHz 0.01MHz


250V DC/AC RMS

3Hz  01.0V 











 1.5mA.  3.2V.

,   50±20.

 0.4mA



 5%


250V DC/AC RMS

 (/)





(-20-1000)

<400 ±(1.0%+5 ) 400 ±(1.5%+15 )

(-4-1832)

<752 ±(1.0%+5 ) 752 ±(1.5%+15 )

K -

 1 1


250V DC/AC RMS

 hFE 







NPN or PNP

0~1000

 10uA, Vce .1.5V



27

 
1."OFF" 2. PH0 3. 1.5V  4.  5. 
mA "FS1":5x20mm 400mA 250V 10A "FS2":5x20mm 10A 250V  LCD " " 

 
 1. 2. 3. 4. 5.

 









1. 2 

28

" 

" 

 

















29

3-3/4 

 

MT-1710N (3 3/4  42mm  LCD  
 DCVACVDCAACA NCV LED   A / D  
 
 x1 K  x1 / x1   x1 
 
 IEC61010  1.  2.  36V  36V DC/25V AC  3.  4.   5.  6.  7.  8.   9. CAT  10. 

" ", " "

" " 

" " "

"

 

!

















 CE 

30

 



39993 3/4

 A/D 



 3 

 "OL"

"

"

 0 40< 75 RH

-20 60< 85 RH 1.5V  AAAx4LR03 186× 92× 52   355 
 

1.  2. LCD  3. / 4.  5.  6. / 7. // 8. // 9. NCV  10. 11.  12.   13.  14. 
 

31

1 

2 

3 

4 

5 

6 

7 /

8 

9 

10 

11 

12 /

13 

nF/uF/mF; 14 mV/V
uA/mA/A

15 :/K/M; :Hz KHz MHz

 
1. "HOLD B/L" LCD "HOLD"
"HOLD"

 HOLD 

 HOLD 

2.

"HOLD B/L" LCD 

3."Hz/DUTY SELECT":

 mVuAmAA "Hz/DUTY SELECT"

/



 Hz 



4./RANGE :

"AUTO", "RANGE" 

"AUTO"

"RANGE""RANGE"

"AUTO"

5.REL/

:

 mVVAmAA "REL/ "

"REL"

"REL"

6.REL/

:

"REL/

"/ LED /

7. APO 

32

"Hz/DUTY SELECT""APO" 

 



"

"

"

""

"



1. 

1.1"

" COM 

1.2 mV  ACV/DCV

1.3

"-""+"



1.4"AUTO""RANGE"

 40mV400mV mV 4V40 V400 V1000

/750V

1.5 mV "Hz/DUTY SELECT"



1.6"Hz/DUTY SELECT"    mV  V  "V""COM"

 33

 DC1000V  AC750V  mV  250V DC/AC RMS.
  220V 


2. 

2.1"

" 400mA 10A 10A

 COM 

2.2A/mA/A

2.3



2.4

2.5"Hz/DUTY SELECT"

 "RANGE"A 

 400A4000A mA  40mA 400mA  A  4A

10A 

  
     OL,   10A  10     10A   

3. 

3.1. 

3.2. "

" COM 

34


3.3. 
3.4.  "RANGE"
 4004K 40K400K4M40M
  ,"OL",
  1M,   "OL"   
4. 
4.1. 
4.2. " " COM 
4.3. 
    OL 
 100mF. 
   "REL"        1Mf=1000uF 1uF=1000nF 1nF=1000pF
5. hFE 
5.1.  hFE 
5.2. hFE " ""COM" NPN  PNP
 35

5.3

6. /

6.1." "

6.2."

""COM"

6.3.

"OL".

6.4."Hz/DUTY"

6.5. 50±20



7. /
7.1 "Hz" Hz "Hz/DUTY SELECT"

7.2 "

""COM"


7.3 
  10V rms 
  
    250V 
               
10Hz~20MHz.

8. 

8.1."/"""

"Hz/DUTY SELECT"""

8.2."

""COM"



8.3.

9. NCV
9.1."NCV".
36

9.2. LED   LED   LED 
  
    

10. 
APO  14  3 1   "SELECT"APO, 

 
+   18 28 75

 (DCV)





4V

40V

±(0.5%+4 )

400V

1000V

±(1.0%+6 )


1mV 10mV 100mV 1V





 10M 1000V DC/AC RMS

 (DC mV)





40mV 400mV

±(0.5%+5 )

 0.01mV 0.1mV

 40M

 250V DC/AC RMS

 mV  (AC mV True RMS)









40mV 400mV

±(1.0%+6 )

0.01mV 0.1mV

 10M

 V  (ACV True RMS)

37

 250V DC/AC RMS

 4V 40V 400V 1000V

 ±(0.8%+10 ) ±(1.2%+10 )

 1mV 10mV 100mV 1V

  10M


1000V DC 750V AC RMS

 10%~100%  40Hz- 1kHz    CP3 CP2  1% 0.2%+0.02Hz; 40Hz- 1kHz80V~600V

 (DCA)





400uA 4000uA

±(1.0%+10 )

40mA 400mA

±(1.2%+8 )

4A ±(1.2%+10 )
10A

 0.1uA
1uA 10uA 100uA 1mA 10mA

 0.1mV/mA 0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A

10A  10  15 


FUSE 400mA/250V
FUSE 10A/250V

 (ACA True RMS)





 

400uA

0.1uA

0.1mV/mA

4000uA 40mA

±(1.5%+10 )

1uA 10uA

0.1mV/mA 1.552mV/mA

400mA

100uA 1.637mV/mA

4A

1mA

31.789mV/A

±(2.0%+5 )

10A

10mA 31.789mV/A

 10%~100%  40Hz- 1kHz    CP3 CP2  1% 10A  10  15 


FUSE 400mA/250V
FUSE 10A/250V

38

 ()





400 ±(0.8%+5 )

4k

40k 400k

±(0.8%+4 )

4M

40M ±(1.2%+10 )

 0.1
1 10 100 1k 10k



  0.4mA  100uA  10uA
 1uA  0.2uA  0.2uA

  1V
 0.5V


250V DC/AC RMS

 (F)
 6nF 60nF 600nF 6uF 60uF 600uF 6mF 60mF 100 mF

 ±(5.0%+40 )
±(3.5%+20 )
±(5.0%+10 ) ±(10%+0 )

 0.001nF 0.01nF 0.1nF 0.001uF 0.01uF 0.1uF 0.001mF 0.01mF 0.1mF


250V DC/AC RMS

 10%~100% 1mF  8 

 (Hz)
 10Hz 100Hz 1KHz 10kHz 100kHz 1MHz 20MHz

 ±(0.5%+10 )

 0.01Hz 0.1Hz 0.001KHz 0.01KHz 0.1KHz 0.001MHz 0.01MHz


250V DC/AC RMS

3Hz  01.0V 

39











 1.5mA.  3.2V.

,   50±20.

 0.4mA



 5%


250V DC/AC RMS

 (/)
 (-20-1000)
(-4-1832)


<400 ±(1.0%+5 ) 400 ±(1.5%+15 )
<752 ±(1.0%+5 ) 752 ±(1.5%+15 )

K -

 1 1


250V DC/AC RMS

 hFE 







NPN or PNP

0~1000

 10uA, Vce .1.5V



 
1."OFF" 2.PH0   3. 1.5V  4.  5. 
mA "FS1":5x20mm 400mA 250V 10A "FS2":5x20mm 10A 250V  LCD " " 

40

 
 1. 2. 3. 4. 5.

 









1. 2 

" 

" 

 

 

 

   

41

















 MT-1710N-C

  12 ( )
     
  · (
) ·  · 
 ·  ·  · ... ·  · ,


   130  7  5   886-2-22183233 E-mail: pk@mail.prokits.com.tw

/  1365  25  021-68183050   400 1699 629

42


Microsoft Word 适用于 Microsoft 365