User Manual for Pro sKit models including: MT-1710N True RMS Auto Range Multimeter, MT-1710N, True RMS Auto Range Multimeter, RMS Auto Range Multimeter, Auto Range Multimeter, Range Multimeter, Multimeter
MT-1710N 3-3/4 True-RMS Auto Range Multimeter | Pro'sKit Official Website
2 6. Don’t input voltage during the resistance measurement. 7. Test lead should be removed from testing point before changing the batteries or fuses and turn off the power.
MT-1710N 3 3/4 True-RMS Auto Range Multimeter User's Manual 1st Edition, 2023 ©2023 Copyright by Prokit's Industries Co., Ltd. I. INTRODUCTION The instrument is a stable and high-performance digital multimeter driven by battery. It uses the LCD with 42mm high to make the readings clearly. The instrument has the function of measuring DCV, ACV, DCA, ACA, Resistance and Capacitance, Frequency, Diode, Triode, Continuity test, Temperature, Auto power off (can be canceled) and backlight function. The instrument takes dual-integral A/D converter as key point, is an excellent tool. It's an ideal tool for lab, factory, and family. The new generation series of productions add NCV (non-contact ACV induction measurement) and torch function based on the original function. II. OPEN THE PACKAGE FOR CHECKING Open the box, take out the meter, checking the items below if they are missing or damaging: K type probe (-20~250) 1pc hFE test Adapter 1pc Test leads 1pair User manual 1pc Please contact with your supplier if you find out any problems. III. SAFETY NOTES This series meter meets the standard of IEC61010 (the safety standards request issued by IEC or equal GB4793.1 standards). Read it before operation. 1. Input over range is prohibited in each range during the test. 2. Voltage less than 36V is safety voltage. When measuring voltage higher than 36V DC, 25V AC, check the connection, insulation of test leads to avoid electric shock. 3. When changing function and range, test leads should be removed from testing point. 4. Select correct function and range. 5. Don't operate the meter when the battery case and back cover are not fixed 1 6. Don't input voltage during the resistance measurement. 7. Test lead should be removed from testing point before changing the batteries or fuses and turn off the power. 8. Comply with local and national safety regulations, wear personal protective equipment (approved rubber gloves, masks and flame-retardant clothing, etc.), to prevent electric shock and arc injury caused by exposure of dangerous electric conductor. 9. Make measurements using only the correct measurement standard class (CAT), voltage and current rated probe, test conductor and adapter. 10Safety symbols: " " exists high voltage " " GND " " Dual Insulation " " Must Refer to Manual " " Low Battery IV. SAFETY SYMBOLS V. SPECIFICATIONS GENERAL SPECIFICATIONS 1) Display mode: LCD displaying. 2) Max. indication: 39993 3/4, auto polarity indication 3) Measuring method: Dual integral A/D conversion 4) Sampling rate: Approx. 3times/second 5) Over range indication: "OL" displays 6) Low battery: The " " displays. 7) Operating environment: (0 ~ 40) , humidity<75%RH. 8) Storage environment:-20~60humidity<85%RH 2 9) Power: Four 1.5V AAA batteries, LR03 10) Dimension: 186mm×92mm ×52mm. 11) Weight: approx. 355g (not include batteries). VI. EXTERNAL STRUCTUR 1. Sound alarm indicator light 2. LCD display 3. Function / range knob 4. Measuring input terminal 5. Function selection button 6. Auto / manual range conversion 7. Relative value measurement / flashlight control 8. Data hold / backlight control 9. NCV induction position 10. Flashlight window 11. Hang hole 12. Screws for fixing the battery box 13. Bracket 14. Test leads holder VII. LCD DISPLAY 3 1 Auto range 2 Auto power off 3 DC measurement 4 AC measurement 5 High voltage 6 True RMS 7 Diode/continuity test 8 Data hold 9 Relative value 10 Low battery 11 meaTsruioredme ent 12 Temperature 13 Duty cycle 14 Current / Capacitance 15 Resistance/ Frequency VIII. KEY DESCRIPTION 1.Data hold: Short press the "HOLD B/L" button to hold the measuring data on the LCD, and the symbol "HOLD" appears. If you press the key again, the data hold function will be closed, and the symbol will disappear. Warning: to prevent possible electric shock, fire, or personal injury, do not use the data hold function to measure the unknown voltage. When open the HOLD function, the LCD will keep original data when measuring a different voltage. 2.Backlight function: Long press the "HOLD B/L" button to turn on the back light of the LCD, and long press this key again to turn off the back light. 3.Function select key (Hz/DUTY SELECT key) At the DC/AC mVuAmAA range, short press the "Hz/DUTY SELECT" key to convert among DCV, ACV, and DCA, ACA for testing; At the diode/continuity test range, short press the "Hz/DUTY SELECT" key to convert the diode and continuity test, the diode range is already set by default. At the ACV range, short press the "Hz/DUTY SELECT" key to cycle switch among the voltage, accompanying frequency, duty cycle. At the temperature range, short press the "Hz/DUTY SELECT" key to convert the Celsius and Fahrenheit; At the 4 Frequency range, short press the "Hz/DUTY SELECT" key to convert the Frequency and duty cycle test. 4.Auto / manual range conversion The auto range is already set by default after the meter is turned on, short press the "RANGE" key, the meter convert to manual range, the "AUTO" symbol on the LCD will disappear. At the manual range mode, the meter starts from the low range, when you press the "RANGE" key, the range change to a higher one, the procedure repeats like this in sequence; long press "RANGE" key to exit the manual range mode, and the LCD displays "AUTO" symbol. 5.Relative measurement At the DC/AC mVVuAmAA range, capacitance, temperature range, short press the "REL" key to enter the relative measurement mode, and the "REL" symbol appears on the LCD. Short press the "REL" key again, the meter exits the relative measurement mode, and the "REL" symbol disappears. 6. Lighting Long press "REL/ " to turn on/off the Flashlight. 7. Cancel the auto power off function Press the "Hz/DUTY SELECT" key to turn on the meter, the "APO" symbol disappears, and then the auto power off function closed. IX. MEASUREMENT INSTRUCTIONS First, please check the battery, and turn the knob to the proper range that you need. If the battery is out of power, the " " symbol will appear on the LCD. Pay attention to the symbol next to the jack for test leads. This is a warning that the voltage and current should not exceed the indicated value. 1. DCV and ACV measurement 5 1-1. Set the knob to the mV/ACV/DCV range, and connect the test leads across to the circuit under test. The voltage and polarity of the point which the red test lead is connected are display on the LCD. 1-2. Insert the black test lead to "COM" jack, the red one to " " jack. 1-3.The auto range is already set by default after the meter is turned on, LCD displays "AUTO" symbol, press the "RANGE" key to convert to manual range, the range 40mV,400mV (only for DC mV/AC mV range), 4V, 40V, 400V, 1000V/750V (DCV/ACV) are optional. 1-4.You can get the result from LCD. 1-5. At the DC mV/AC mV range, short press the "Hz/DUTY SELECT" key to convert DC/AC mV measurement. The DC mV range is already set by default after the meter is turn on. Note: (1) Do not input a voltage over DC 1000V or AC 750Votherwise the circuit might be damaged. At the DC/AC mV range, the input voltage should not exceed 250VDC/ AC RMS. (2) Please keep the test leads away from the circuit after finished your testing. (3) When measured voltage is higher than 220V, it's necessary to wear personal protective equipment (such as approved rubber gloves, face masks, and flame-retardant clothing etc.) to prevent the injury from electric shock and arc. 2. DCA and ACA measurement 2-1. Set the knob to a proper DCA/ACA range, and then connect the test leads to the power or circuit under test in series. The polarity and the current value of this point which connects to the red test lead will display on the LCD. 2-2. Insert the black test lead to "COM" jack, the red one to " " (max. 400mA) or "10A" (max. 10A) jack. 2-3. You can get the result from display. 6 2-4. Short press the "Hz/DUTY SELECT" key to convert DCA or ACA, the DCA range is already set by default after the meter is turned on. Short press "RANGE" key to convert to manual range, you can choose 400uA, 4000uA at the uA range, and 40mA, 400mA are optional at the mA range, 4A, 10A are optional at the A range. The auto range is already set by default after the meter is turned on. Note: (1) If the tested current range is unsure beforehand, please start from the highest range then find the proper range according to the reading. Do not measure the voltage in the current jack. (2) If the reading displays "OL"the value is beyond the present range. Now you need to set the knob to the higher range. (3) Be careful when measuring 10A range. Continuous measurement of high current will heat the circuit, even affect the accuracy and damage the meter. (4) When measuring the large current above 10A, it's necessary to wear personal protective equipment (such as approved rubber gloves, face masks, and flame-retardant clothing etc.) to prevent the injury from electric shock and arc. 3. Resistance measurement 3-1. Set the knob to resistance range, then connecting the test leads to the tested resistance. 3-2. Insert the black test lead to "COM" jack, the red one to " "jack. 3-3. You can read the measurement results from the display. 3-4. Auto range method is already set by default after the meter turned on, short press "RANGE" key to convert to the manual range, The range 400, 4K, 40K, 400K, 4M, 40M are optional. Note: (1) If "OL" displays on LCD, it means over range, now you need to set the range knob to the higher range. When measuring resistance is 7 more than 1M, the reading may take a few seconds to stabilize. This is normal for high resistance measurement. (2) When the input terminal is under open circuit, the figure "OL" will be displayed for the overload condition. (3) When checking in-circuit resistance, be sure the power has been switched off and all capacitors are fully discharged. 4. Capacitance measurement 4-1. Set the range knob to a proper capacitance range, connect the test leads to the capacitor under tested (the polarity of red test lead is "+"). 4-2. Insert the red test lead to " " jack and the black one to "COM" jack. 4-3. You can get the result from LCD. Note: (1) If the LCD displays "OL", it indicates the value is over range. The maximal range is 100mF. (2) When measuring capacitance, because of lead wire and the distributed capacitance, there may be some residual reading without connection to capacitance under measured. It will be clearer when measuring small capacitance range. Measuring result is to subtract the residual reading for getting accurate reading. This will not affect the measuring accuracy. You can short press the "REL" key to clear the residual reading of open circuit, then measuring the relative value. (3) When under a large capacitance range, if capacitor is broken or leakage, the LCD will display some value and it's unstable. (4) Discharge all the capacitance completely before measuring to avoid the damage of the meter. (5) UNIT: 1mF=1000uF 1uF =1000nF 1nF=1000pF 5. TRANSISTOR hFE 5-1. Set the knob to "hFE" range. 5-2. Insert the adapter to " " and "COM" jack, then according to 8 the NPN or PNP, insert emitter, base and collector to proper jack. 5-3. You can get the result from LCD. 6. Diode and continuity test 6-1. Set the range knob to the " " range. The diode range is already set by default after the meter is turned on, short press the "Hz/DUTY SELECT" key to convert the diode and continuity test. Set the black test lead to "COM" jack, and the red one to " "jack (the polarity of red test lead is "+"), meanwhile connect the red test lead to the positive of the diode and the black one to the negative. The reading is the approx. value of the diode positive volt drop. If the test leads connect to the reverse polarity of diode, the LCD will display "OL". 6-2. Connect the test leads to two ends of the circuit under tested, if the resistance is lower than approx.50±20, the LCD displays " " symbol, and the buzzer sounds. 7. Frequency measurement 7-1.Turn the knob to frequency range, and connect the test leads or shielded cable across the signal source or the load under tested. 7-2.Insert the black test lead to the "COM" jack and the red one to the " "jack. 7-3.You can get the result from the LCD. 7-4.Short press the "Hz/DHTY SELECT" key to convert to duty cycle. Note: (1) When it is over 10V RMS, you can get the result, but we cannot promise the accuracy. (2) It is preferable to use shielded cable for measuring small signal in noisy environment. (3) Be careful when measuring high voltage. (4) Do not input a voltage over DC 250V or AC peak value to avoid damage to the meter. (5) It is auto range at the frequency range, you can test 10Hz ~20MHz. 8. Temperature measurement 9 When measuring temperature, insert the negative of the thermocouple sensor into the "COM" jack, and insert the positive into the " " jack; put the sensing part of the temperature probe on or inside the tested object. Read the value from the LCD which shows as Celsius. Short press the "Hz/DHTY SELECT" key can convert the Celsius or Fahrenheit unit. 9. Non-contact voltage induction measurement NCV 9-1. Turn the knob to "NCV" range. 9-2. NCV induction voltage range is 48V~220V, let the upper position of the meter close to the measured electric AC power line, when the ACV is sensed, the upper red indicator lights of the meter flashing and at the same time a buzzer alarm on, the more close to the AC power line, the more stronger the ACV sensed, the corresponding flashing lights, and the buzzer alarm sounds faster. 10. Automatic power on/off To save power consumption and prolong battery life, the APO automatic shutdown function will be turned on by default after the meter is turned on. If the user does not operate the meter within 14 minutes, the meter will be prompted with 3sounds. If there is still no operation, the meter will automatically turn off the power after a long sound one minute later. Short press "SELLECT" can turn on the multimeter. To cancel APO function, please refer to the instructions in section 8, "key function". X. Technical characteristics Accuracy: ± (a% × reading + digit) Environment temperature: (23±5) Relative humidity: <75%RH. One year guarantee since production date. 10-1. DC Voltage 10 Range 4V 40V 400V 1000V Accuracy Resolution ±(0.5%+4d) ±(1.0%+6d) 1mV 10mV 100mV 1V Input impedance Overload protection About 10M 1000V DC / AC RMS 10-2. DC mV Range Accuracy Resolution 40mV ±(0.5%+5d) 400mV 0.01mV 0.1mV Input impedance >40M Overload protection 250VDC / AC RMS 10-3. AC mV (True RMS) Range Accuracy Resolution 40mV ±(1.0%+6d) 400mV 0.01mV 0.1mV Input impedance About 10M Overload protection 250V DC/AC RMS 10-4. ACV (True RMS) Range Accuracy Resolution Input impedance Overload protection 4V 40V 400V 750V ± (0.8%+10d) ± (1.2%+10d) 1mV 10mV 100mV About 10M 1V 1000V DC / 750AC RMS Accuracy measurement range: 10%-100% of the range. Frequency response: 40Hz-1 kHz. Measurement mode (sine wave): true RMS. 11 Crest factor: CF3, adding an additional error of 1% to the reading at CF2. AC frequency measurement error: 0.2%+0.02Hz. AC frequency measurement range: 40Hz~1k Hz. AC frequency input sensitivity: 80V~600V 10-5. DCA Range Accuracy Resolution Load voltage Overload protection 400uA 4000uA 40mA 400mA ±(1.0%+10d) ±(1.2%+8d) 0.1uA 1uA 10uA 100uA 0.1mV/ mA 0.1mV/ mA 1.552mV/ mA 1.637mV / mA FUSE 400mA/250V 4A 1mA ±(1.2%+10d) 10A 10mA 31.789mV/ A 31.789mV/ A FUSE 10A/250V 10A (test within 10 seconds); Recovery time is 15 minutes. 10-6. ACA (True RMS) Range Accuracy Resolution Load voltage Overload protection 400uA 0.1uA 0.1mV/ mA 4000uA ±(1.5%+10d) 40mA 1uA 10uA 0.1mV/ mA 1.552mV/ mA FUSE 400mA/250V 400mA 100uA 1.637mV / mA 4A ±(2.0%+5d) 1mA 31.789mV/ A 10A 10mA 31.789mV/ A FUSE 10A/250V Measuring range of accuracy value: 10% - 100% of the range. Frequency response: 40Hz - 1k Hz Measuring way (sine wave): True RMS Crest factor: CF3, when CF2, add an additional error of 1% of the 12 reading. 10A (The measurement should not be more than 10 seconds); Recovery time is 15 minutes. 10-7. Resistance Range Accuracy Short-circuit Resolution current 400 ± (0.8%+5d) 0.1 About 0.4mA 4k 1 About 100uA 40k ± (0.8%+4d) 400k 10 100 About 10uA About 1uA 4M 1k About 0.2uA 40M ± (1.2%+10d) 10k About 0.2uA Measuring error does not include lead resistance. Overload protection: 250VDC/AC RMS Open-circuit voltage About 1V About 0.5V 10-8. Capacitance Range 6nF Accuracy ±(5.0%+40d) Resolution 0.001nF Overload protection 60nF 0.01nF 600nF 6uF ±(3.5%+20d) 0.1nF 0.001uF 60uF 0.01uF 250VDC/AC RMS 600uF 0.1uF 6mF ±(5.0%+10d) 0.001mF 60mF 100mF ±10% 0.01mF 0.1mF Measuring range of accuracy: 10% - 100% of the range Large capacitance response time: 1uF about 8s Measurement error does not include lead distribution capacitance. 13 10-9. Frequency Range Accuracy Resolution Overload protection 10Hz 0.01Hz 100Hz 0.1Hz 1kHz 10kHz ±(0.5%+10d) 1Hz 10Hz 250VDC/AC RMS 100kHz 100Hz 1MHz 1kHz 20MHz 10kHz Note: The reading will be zero if the signal is below 3Hz. Input sensitivity: 1V 10-10. Diode and continuity Test Range Displaying value Test Condition Error The positive DC Current Positive voltage is approx. 1.5mA. drops of diode Negative voltage is approx. 3.2V. 5% Buzzer sounds, the resistance is less than 50±20 The test current is approx. 0.4 mA Warning: Do not input voltage at this range. Overload protection: 250V DC/AC RMS 10-11. Temperature Range Accuracy Resolutio n Overload protection (-20-1000) <400 ± (1.0%+5d) 1 400 ± (1.5%+15d) 250V DC/AC (-4~1832) <752 ± (1.0%+5d) 1 752 ± (1.5%+15d) RMS Sensor: Type-K thermocouple (Ni-Cr Ni-Si) banana probe 14 10-12. Transistor hFE DATA TEST Range Displaying range hFE NPN or PNP 0 ~ 1000 Test condition Basic current is approx. 10uA Vce is about 1.5V XI. Replace the battery or fuse. 1. Move away the test leads from the circuit under test, pull out the test lead from the input jack, turn the range knob to the "OFF" range to turn off the power. 2. Use a screwdriver to twist off the screws on the battery cover and remove the battery cover and bracket. 3. Take out the old battery or the broken fuse, then replace with a new alkaline battery 9V or a new fuse. 4. Close the battery cover and use a screwdriver to tighten the screws on the battery cover. 5. Battery specifications:4 * 1.5V AAA alkaline batteries 6. Fuse specifications: mA input fuse "FS1": 5 * 20mm 400mA 250V 10A input fuse "FS2": 5 * 20mm 10A 250V Note: When the low voltage " " symbol displays on the LCD, the battery should be replaced immediately, otherwise the measuring accuracy will be affected. XII. MAINTENANCE AND CARE It is an accurate meter. Do not try to modify the electric circuit. 1. Pay attention to the waterproof, dustproof and break proof of the meter. 2. Please do not store or use it in environment of high temperature, high humidity, high flammability or strong magnetic. 3. Please wipe the meter with a damp cloth and soft detergent, and abrasive and drastic solvent such as alcohol are forbidden. 15 4. If do not operate for a long time, should take out the battery to avoid leakage. 5. When replacing fuse, please use another same type and specification fuse. 13. Trouble Shooting If the meter does not work properly, check the meter as following: Fault Solution Power off--- Pls turn on the power No reading on LCD Holding key-- Pls set a correct mode Replace battery The signal appears Replace battery No current or temperature input Replace fuse Error Value Replace battery This user's manual is subject to any change without further notice. The content in this user's manual is deemed correct; if you find any mistake, omission, etc., please contact the manufacturer. We will not be held liable for any accidents or harms caused due to your wrong operations. The functions set forth in this user's manual shall not be regarded as reasons for applying this product for special purposes. 16 3-3/4 MT-1710N3 3/4 42mm LCD DCVACVDCAACA NCV LED A / D x1 K (-20250)x1/ x1 x1 IEC61010 1. 2. 36V 36V DC/25V AC 3. 4. 5. 6. 7. 8. 9. CAT 10. " " " " " " , " " " " ! 17 CE 39993 3/4 A/D 3 "OL" " " 0 40< 75 RH -20 60< 85 RH 1.5V AAAx4LR03 186× 92× 52 355 1 1. 2. LCD 3. / 4. 5. 6. / 7. // 8. // 9. NCV 10. 11. 12. 13. 14. 18 1 2 3 4 5 6 7 / 8 9 10 11 12 / 13 nF/uF/mF; 14 mV/V uA/mA/A 15 :/K/M; :Hz KHz MHz 1. "HOLD B/L" LCD "HOLD" "HOLD" HOLD HOLD 2. "HOLD B/L" LCD 3."Hz/DUTY SELECT": mVuAmAA "Hz/DUTY SELECT" / Hz 4./RANGE : 19 "AUTO", "RANGE" "AUTO" "RANGE""RANGE" "AUTO" 5.REL/ : mVVAmAA "REL/ " "REL" "REL" 6.REL/ : "REL/ "/ LED / 7. APO "Hz/DUTY SELECT""APO" " " " "" " 1. 1.1" " COM 1.2 mV ACV/DCV 1.3 "-""+" 1.4"AUTO""RANGE" 40mV400mV mV 4V40 V400 V1000 /750V 1.5 mV "Hz/DUTY SELECT" 1.6"Hz/DUTY SELECT" mV V "V""COM" 20 . DC1000V AC750V mV 250V DC/AC RMS. 220V 2. 2.1" " 400mA 10A 10A COM 2.2A/mA/A 2.3 2.4 2.5"Hz/DUTY SELECT" "RANGE"A 400A4000A mA 40mA 400mA A 4A 10A OL, 10A 10 10A 21 3. 3.1. 3.2. " " COM 3.3. 3.4. "RANGE" 4004K 40K400K4M40M , "OL", 1M , "OL" 4. 4.1. 4.2. " " COM 4.3. OL 100mF. "REL" 1Mf=1000uF 1uF=1000nF 1nF=1000pF 22 5. hFE 5.1. hFE 5.2. hFE " ""COM" NPN PNP 5.3 6. / 6.1." " 6.2." ""COM" 6.3. "OL". 6.4."Hz/DUTY" 6.5. 50±20 7. / 7.1 "Hz" Hz "Hz/DUTY SELECT" 7.2 " ""COM" 7.3 10V rms 250V 10Hz~20MHz. 8. 8.1."/""" "Hz/DUTY SELECT""" 8.2." ""COM" 23 8.3. 9. NCV 9.1."NCV". 9.2. LED LED LED 10. APO 14 3 1 "SELECT"APO, + 18 28 75 (DCV) 4V 40V ±(0.5%+4 ) 400V 1000V ±(1.0%+6 ) 1mV 10mV 100mV 1V 10M 1000V DC/AC RMS 24 (DCmV) 40mV 400mV ±(0.5%+5 ) 0.01mV 0.1mV 40M 250V DC/AC RMS mV (ACmV True RMS) 40mV 400mV ±(1.0%+6 ) 0.01mV 0.1mV 10M V (ACV True RMS) 4V 1mV 40V 400V ±(0.8%+10 ) 10mV 100mV 10M 1000V ±(1.2%+10 ) 1V 250V DC/AC RMS 1000V DC 750V AC RMS 10%100% 40Hz- 1kHz CP3 CP2 1% 0.2%+0.02Hz; 40Hz- 1kHz80V600V (DCA) 400uA 4000uA ±(1.0%+10 ) 40mA 400mA ±(1.2%+8 ) 4A 10A ±(1.2%+10d) 0.1uA 1uA 10uA 100uA 1mA 10mA 0.1mV/mA 0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A 10A 10 15 FUSE 400mA/250V FUSE 10A/250V (ACA True RMS) 400uA ±(1.5%+10 ) 0.1uA 25 0.1mV/mA FUSE 4000uA 40mA 400mA 4A 10A ±(2.0%+5 ) 1uA 10uA 100uA 1mA 10mA 0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A 10%~100% 40Hz- 1kHz CP3 CP2 1% 10A 10 15 400mA/250V FUSE 10A/250V () 400 ±(0.8%+5 ) 4k 40k 400k ±(0.8%+4 ) 4M 40M ±(1.2%+10 ) 0.1 1 10 100 1k 10k 0.4mA 100uA 10uA 1uA 0.2uA 0.2uA 1V 0.5V 250V DC/AC RMS (F) 6nF 60nF 600nF 6uF 60uF 600uF 6mF 60mF 100 mF ±(5.0%+40 ) ±(3.5%+20 ) ±(5.0%+10 ) ±10% 0.001nF 0.01nF 0. 1nF 0.001uF 0.01uF 0.1uF 0.001mF 0.01mF 0.1mF 250V DC/AC RMS 10%~100% 1mF 8 26 (Hz) 10Hz 100Hz 1KHz 10kHz ±(0.5%+10 ) 100kHz 1MHz 20MHz 0.01Hz 0.1Hz 0.001KHz 0.01KHz 0.1KHz 0.001MHz 0.01MHz 250V DC/AC RMS 3Hz 01.0V 1.5mA. 3.2V. , 50±20. 0.4mA 5% 250V DC/AC RMS (/) (-20-1000) <400 ±(1.0%+5 ) 400 ±(1.5%+15 ) (-4-1832) <752 ±(1.0%+5 ) 752 ±(1.5%+15 ) K - 1 1 250V DC/AC RMS hFE NPN or PNP 0~1000 10uA, Vce .1.5V 27 1."OFF" 2. PH0 3. 1.5V 4. 5. mA "FS1":5x20mm 400mA 250V 10A "FS2":5x20mm 10A 250V LCD " " 1. 2. 3. 4. 5. 1. 2 28 " " 29 3-3/4 MT-1710N (3 3/4 42mm LCD DCVACVDCAACA NCV LED A / D x1 K x1 / x1 x1 IEC61010 1. 2. 36V 36V DC/25V AC 3. 4. 5. 6. 7. 8. 9. CAT 10. " ", " " " " " " " " ! CE 30 39993 3/4 A/D 3 "OL" " " 0 40< 75 RH -20 60< 85 RH 1.5V AAAx4LR03 186× 92× 52 355 1. 2. LCD 3. / 4. 5. 6. / 7. // 8. // 9. NCV 10. 11. 12. 13. 14. 31 1 2 3 4 5 6 7 / 8 9 10 11 12 / 13 nF/uF/mF; 14 mV/V uA/mA/A 15 :/K/M; :Hz KHz MHz 1. "HOLD B/L" LCD "HOLD" "HOLD" HOLD HOLD 2. "HOLD B/L" LCD 3."Hz/DUTY SELECT": mVuAmAA "Hz/DUTY SELECT" / Hz 4./RANGE : "AUTO", "RANGE" "AUTO" "RANGE""RANGE" "AUTO" 5.REL/ : mVVAmAA "REL/ " "REL" "REL" 6.REL/ : "REL/ "/ LED / 7. APO 32 "Hz/DUTY SELECT""APO" " " " "" " 1. 1.1" " COM 1.2 mV ACV/DCV 1.3 "-""+" 1.4"AUTO""RANGE" 40mV400mV mV 4V40 V400 V1000 /750V 1.5 mV "Hz/DUTY SELECT" 1.6"Hz/DUTY SELECT" mV V "V""COM" 33 DC1000V AC750V mV 250V DC/AC RMS. 220V 2. 2.1" " 400mA 10A 10A COM 2.2A/mA/A 2.3 2.4 2.5"Hz/DUTY SELECT" "RANGE"A 400A4000A mA 40mA 400mA A 4A 10A OL, 10A 10 10A 3. 3.1. 3.2. " " COM 34 3.3. 3.4. "RANGE" 4004K 40K400K4M40M ,"OL", 1M, "OL" 4. 4.1. 4.2. " " COM 4.3. OL 100mF. "REL" 1Mf=1000uF 1uF=1000nF 1nF=1000pF 5. hFE 5.1. hFE 5.2. hFE " ""COM" NPN PNP 35 5.3 6. / 6.1." " 6.2." ""COM" 6.3. "OL". 6.4."Hz/DUTY" 6.5. 50±20 7. / 7.1 "Hz" Hz "Hz/DUTY SELECT" 7.2 " ""COM" 7.3 10V rms 250V 10Hz~20MHz. 8. 8.1."/""" "Hz/DUTY SELECT""" 8.2." ""COM" 8.3. 9. NCV 9.1."NCV". 36 9.2. LED LED LED 10. APO 14 3 1 "SELECT"APO, + 18 28 75 (DCV) 4V 40V ±(0.5%+4 ) 400V 1000V ±(1.0%+6 ) 1mV 10mV 100mV 1V 10M 1000V DC/AC RMS (DC mV) 40mV 400mV ±(0.5%+5 ) 0.01mV 0.1mV 40M 250V DC/AC RMS mV (AC mV True RMS) 40mV 400mV ±(1.0%+6 ) 0.01mV 0.1mV 10M V (ACV True RMS) 37 250V DC/AC RMS 4V 40V 400V 1000V ±(0.8%+10 ) ±(1.2%+10 ) 1mV 10mV 100mV 1V 10M 1000V DC 750V AC RMS 10%~100% 40Hz- 1kHz CP3 CP2 1% 0.2%+0.02Hz; 40Hz- 1kHz80V~600V (DCA) 400uA 4000uA ±(1.0%+10 ) 40mA 400mA ±(1.2%+8 ) 4A ±(1.2%+10 ) 10A 0.1uA 1uA 10uA 100uA 1mA 10mA 0.1mV/mA 0.1mV/mA 1.552mV/mA 1.637mV/mA 31.789mV/A 31.789mV/A 10A 10 15 FUSE 400mA/250V FUSE 10A/250V (ACA True RMS) 400uA 0.1uA 0.1mV/mA 4000uA 40mA ±(1.5%+10 ) 1uA 10uA 0.1mV/mA 1.552mV/mA 400mA 100uA 1.637mV/mA 4A 1mA 31.789mV/A ±(2.0%+5 ) 10A 10mA 31.789mV/A 10%~100% 40Hz- 1kHz CP3 CP2 1% 10A 10 15 FUSE 400mA/250V FUSE 10A/250V 38 () 400 ±(0.8%+5 ) 4k 40k 400k ±(0.8%+4 ) 4M 40M ±(1.2%+10 ) 0.1 1 10 100 1k 10k 0.4mA 100uA 10uA 1uA 0.2uA 0.2uA 1V 0.5V 250V DC/AC RMS (F) 6nF 60nF 600nF 6uF 60uF 600uF 6mF 60mF 100 mF ±(5.0%+40 ) ±(3.5%+20 ) ±(5.0%+10 ) ±(10%+0 ) 0.001nF 0.01nF 0.1nF 0.001uF 0.01uF 0.1uF 0.001mF 0.01mF 0.1mF 250V DC/AC RMS 10%~100% 1mF 8 (Hz) 10Hz 100Hz 1KHz 10kHz 100kHz 1MHz 20MHz ±(0.5%+10 ) 0.01Hz 0.1Hz 0.001KHz 0.01KHz 0.1KHz 0.001MHz 0.01MHz 250V DC/AC RMS 3Hz 01.0V 39 1.5mA. 3.2V. , 50±20. 0.4mA 5% 250V DC/AC RMS (/) (-20-1000) (-4-1832) <400 ±(1.0%+5 ) 400 ±(1.5%+15 ) <752 ±(1.0%+5 ) 752 ±(1.5%+15 ) K - 1 1 250V DC/AC RMS hFE NPN or PNP 0~1000 10uA, Vce .1.5V 1."OFF" 2.PH0 3. 1.5V 4. 5. mA "FS1":5x20mm 400mA 250V 10A "FS2":5x20mm 10A 250V LCD " " 40 1. 2. 3. 4. 5. 1. 2 " " 41 MT-1710N-C 12 ( ) · ( ) · · · · · ... · · , 130 7 5 886-2-22183233 E-mail: pk@mail.prokits.com.tw / 1365 25 021-68183050 400 1699 629 42Microsoft Word 适用于 Microsoft 365