Vabiooth Professional Microscope VB-M006

This operation manual is for the Model VB-M006 microscope. It is recommended to carefully review this manual prior to operating the device to optimize safety and performance. Keep this manual in an easily accessible place for future reference.

WARNING:

To prevent fire and electrical shock, do not expose the unit to rain or moisture.

CAUTION:

No user-serviceable parts inside. Refer servicing to qualified service personnel.

Contents

1. Before Use

1-1. Operation

CAUTION:

It will damage the microscope if you hold the microscope either by the stage, focusing knob, or head.

1-2. Maintenance

CAUTION:

Alcohol is flammable. Do not place this chemical near fire or fire-causing sources. Use this chemical in a ventilated place.

2. Diagram

VB-M006-B Microscope Diagram

This diagram shows the Vabiooth VB-M006-B microscope with the following labeled parts:

VB-M006-C Microscope Diagram

This diagram shows the Vabiooth VB-M006-C microscope, similar to the VB-M006-B, with the addition of a Camera attached to the head.

VB-M006-M Microscope Diagram

This diagram shows the Vabiooth VB-M006-M microscope, similar to the VB-M006-B, with the addition of an LCD Screen attached to the head.

3. Assembly

3-1 Install the Objectives

Install the objectives into the microscope nosepiece, starting from the lowest magnification to the highest, in a clockwise direction.

3-2 Insert the Eyepieces

  1. Remove the eyepiece tube covers.
  2. Insert the eyepieces into the tubes completely.

3-3 Install plug

Insert one side of the power plug into the back socket of the microscope.

4. Operation

4-1 Turn on

  1. Insert the power plug into an outlet and turn on the microscope.
  2. Adjust the light adjustment knob until you achieve the suitable brightness. (Refer to Figure 1, which shows the light control knob).

4-2 Set the specimen slide

  1. Push the arm of the specimen holder back.
  2. Insert the slide with the cover glass facing up, and then allow the arm of the specimen holder to move forward. (Refer to Figure 2, which illustrates placing the specimen slide).
  3. Rotate the X and Y-axis knobs on the mechanical stage to move the specimen to the center, aligning it with the objective.

4-3 Adjust focus

  1. Move the 4X objective into the light path.
  2. Rotate the coarse-focusing knob until an image appears.
  3. Turn the fine-focusing knob to make the specimen clearer. (Refer to Figure 3, which indicates the focus knobs).

Note: When using the 4X and 10X objectives, open the aperture diaphragm and field diaphragm to their widest positions.

4-4 Adjust tension of the focus

If the coarse focus knob is difficult to turn or the stage drifts downwards by itself, adjust the tension adjustment ring. Turn clockwise to tighten the controls or counterclockwise to loosen them. (Refer to Figure 4, showing the tension adjustment ring).

4-5 Adjust the interpupillary distance

When observing with both eyes, hold the bases of the eyepieces and rotate them around their axis until you see a single, unified field of view. (Refer to Figure 5, showing eye distance indication). The numbers on the left and right eyepiece bases should align to indicate the interpupillary distance.

4-6 Adjust diopters

Rotate one eyepiece to the "0" position using the white thread. Focus on a specimen clearly using only that eyepiece (this is for your dominant eye). Then, observe with the other eyepiece and rotate its diopter adjusting ring until the image is clear in that eyepiece as well. (Refer to Figure 6, illustrating diopter adjustment).

4-7 Adjust diaphragm

The diaphragm controls the amount of light entering the condenser, which can strengthen the image. For the clearest image, adjust the diaphragm so its image is just on the edge of the field of view.

4-8 Adjust aperture diaphragm

The aperture diaphragm determines the illumination's numerical aperture (N.A.). Matching the illumination N.A. with the objective's N.A. provides better resolution and contrast. If contrast is low, adjust the condenser's N.A. to 70%-80% of the objective's N.A. Adjust the upper diaphragm ring until the image appears as shown in Figure 7. For example, with a 40X objective (N.A. = 0.65), set the aperture diaphragm to approximately 0.52 (0.65 * 0.8). (Refer to Figure 8 for diaphragm adjustment).

4-9 Use the oil objective

Follow these steps for using the 100X oil objective:

  1. Use the 40X objective to focus on the specimen.
  2. Place a drop of immersion oil on the specimen. (Refer to Figure 9, showing oil position).
  3. Rotate the nosepiece to place the 100X objective (oil) into the light path. Use the fine focusing knob to focus.

Important: Ensure there are no air bubbles in the oil.

To remove air bubbles:

After using the oil objective, clean the 100X lens with clean lens paper and rubbing alcohol if the oil does not come off easily.

Using the Filter

Filters can soften the background. A frosted filter is included. A polarizer can also be placed in this position when using the optional simple polarizing set.

5. Specification

5-1 VB-M006 Microscope General Specifications

VB-M006 FeatureDetails
EyepiecesWF10X, WF16X, and WF25X eyepieces
4 Objective lenses4X, 10X, 40X (Spring), 100X (Spring, Oil)
Magnification levels40X, 64X, 100X, 160X, 250X, 400X, 640X, 1000X, 1600X, 2500X
Interpupillary distance55-75mm
Adjustable ocular diopter+/-5dp
Head30 degree inclined, 360 degree rotatable trinocular head
Mechanical stageMechanical stage with rack stop, 110x120mm
Mechanical stage travel range70x30mm
Abbe condenserNA 1.25 Abbe condenser with iris diaphragm
KnobsCoaxial coarse and fine focus knobs
LED lightTransmitted illumination with 1W LED bulb, adjustable brightness
Power supplyType-C 5V/1A 100-240V power input by AC adaptor or 3AA power supply, low voltage input, safe and power-saving, portable

6. Troubleshooting

Optics

SymptomCauseRemedy
(1) The lamp is bright, but the light in the field of view is not bright enough.Field diaphragm is not large enough.
Condenser is too low.
Condenser is not centered.
Enlarge the field diaphragm.
Adjust the condenser.
Center the condenser.
(2) The side of the field of view is dark or not even.The nosepiece is not in the right position.
Stain or dust has accumulated on the condenser, objective, eyepieces, and base lens.
Condenser adjustment needed (see step 3-7)
Turn the nosepiece into the right position.
Clean the lens.
Center the condenser.
(3) Stain or dust is observed in the field of view.Stains have accumulated on the specimen.
Stains have accumulated on the lens.
Clean the specimen.
Clean the lens.
(4) Unclear imageNo cover glass on the specimen slide.
The cover glass is not standard.
The cover glass faces down.
The immersion oil has accumulated on the dry objective.
The immersion oil is not used for oil objective 100XR.
Air bubble in the immersion.
Use wrong immersion oil.
Add the cover glass.
Use a standard cover glass with thickness 0.17mm.
Put the cover glass to face up.
Clean thoroughly.
Use immersion oil.
Get rid of the air bubble.
Use correct one.
(5) One side of the field of view is dark or the image moves while focusing.The aperture is not opened correctly.
Stain or dust has accumulated on the lens in the inlet of the head.
The condenser is not in the right position.
The specimen slide is not fixed.
The nosepiece is not in the right position.
Condenser is centered incorrectly.
Adjust the iris diaphragm.
Clean the lens.
Adjust the condenser.
Fix with clips.
Turn the nosepiece into the right position.
Center the condenser.
(6) The eyes feel tired easily. The right field of view doesn't superpose with the left.Interpupillary distance is wrong.
Diopter adjustment is wrong.
The eyepiece for the right is different from the left.
Adjust the interpupillary distance.
Adjust the diopter.
Use the same eyepieces.

Mechanics & Electrical

SymptomCauseRemedy
(1) Can not get the objective focused.The cover glass faces down.
The cover glass is not standard.
Put the cover glass to face up.
Use a standard cover glass with thickness 0.17mm.
(2) The objective touches the cover glass while turning the nosepiece.The cover glass faces down.
The cover glass is not standard.
Put the cover glass to face up.
Use a standard cover glass with thickness 0.17mm.
(3) Coarse focusing knob is too tight.Tension knob is too tight.Loosen a little.
(4) Stage declines itself and can't stay in the focus plane.Tension knob is too loose.Tighten a little.
(5) Coarse focusing knob can't rise.The limit stop knob is locked.Loosen the knob.
(6) Coarse focusing knob can't decline.The base of the condenser is too low.Raise the base.
(7) Can not move the slide smoothly.The slide is not fixed correctly.
The movable specimen holder is not fixed properly.
Adjust it correctly.
Adjust it correctly.
(8) The image moves obviously when touching the stage.The stage is fastened incorrectly.Fasten the stage correctly.
Electrical:
(1) The bulb does not work.
No power supply.
The bulb is not inserted correctly.
Check the connection of the power cable.
Insert it correctly.
(2) The bulb burnt out usually.The bulb burnt out.Replace it.
(3) The field of view is not bright enough.Use a wrong bulb.
The use of light adjustment knob is wrong.
Replace with a correct one.
Adjust correctly.
(4) The bulb flickers or the brightness is not stable.The bulb will burn out soon.
The wire doesn't connect all right.
Replace with a new one.
Connect correctly.

7. How to install the monitor or camera?

To install a monitor or camera:

  1. Rotate the head of the tube counterclockwise; the bottom of the tube is twisted onto the machine.
  2. The tube will then divide into two parts: one for the 23.2mm ocular port or tube, and the other for the C-Mount.

The monitor fits for the 23.2mm tube, while the camera is typically C-Mount.

Monitor Operation:

For instructional videos, visit the "Vabiooth" YouTube channel.

8. Microscope Camera (Only for 5MP Camera Version)

Features:

This camera is widely used in pattern recognition, industrial testing, medical microscopy, traffic management, machine vision, and various other fields requiring image acquisition.

It features a C-Mount eyepiece interface (adapter) designed for the microscope.

Camera Specification:

The kit includes a software CD with imaging application and calibration slide for measurement accuracy. If your computer lacks a CD drive, the software can be downloaded from: https://bit.ly/3XEZvxX

9. Dual Power Supply & Power cord hanger design

The microscope offers dual power options:

10. Warranty

Warranty information is not detailed in the provided document pages.

Models: VB-M006, VB-M006 Compound Binocular Microscope, Compound Binocular Microscope, Binocular Microscope, Microscope

File Info : application/pdf, 16 Pages, 741.55KB

PDF preview unavailable. Download the PDF instead.

917 k3HqagL

References

Microsoft Word 2016