Laboratory pH Electrode Selection Guide

YSI, a Xylem Brand W75-04

Introduction

pH is one of the most fundamental parameters measured in nearly every application. Different electrode requirements exist for different applications. Due to these various requirements, YSI offers a suite of laboratory combination pH electrodes to ensure selection of an electrode that meets application demands. This guide provides a description of each component of a pH electrode, an overview of each YSI lab electrode family, and specifications for each YSI electrode. For additional assistance in selecting a pH electrode, consult the YSI Laboratory pH Electrode Application Guide, and for practical guidance regarding pH measurement, refer to the YSI pH Handbook.

pH Electrode Design

Choosing the right electrode for your needs begins with an understanding of the different structures that make up a pH electrode.

Temperature sensor (optional)

Some electrodes feature an integrated temperature sensor. pH values are dependent on temperature, so pH measurements should always be completed with an accurate temperature sensor.

Reference electrolyte

The electrolyte has connection to the sample through the junction. Potassium chloride (KCl) is the most common electrolyte used and can be in the form of liquid or gel. Liquid electrolyte provides faster measurement results and decreases the possibility of measurement errors due to diffusion potentials.

Electrode body

Plastic body electrodes are more rugged, less likely to crack, and commonly feature gel electrolyte. Glass body electrodes have a wider range of operating temperatures and many are refillable.

Glass membrane

Specialized glass that is sensitive to hydrogen ion activity. Filled with buffer of known pH, creating an environment of constant binding of hydrogen ions on the inside while the sample outside has a variable amount of hydrogen ions. This difference in hydrogen ions creates an electrical potential. The shape of the membrane can vary in order to ensure optimal moistening. Unique applications may require a specialized membrane shape.

Refill opening (optional)

Since electrolyte escapes through the reference junction, electrodes that can be refilled with liquid electrolyte tend to have a longer life than non-refillable electrodes.

Reference electrode

Designed to maintain a constant electrical potential. The difference in electrical potential between the reference and sensing electrode results in a voltage that is used to calculate a pH value. The silver/silver chloride reference system is the most common, although the iodine/iodide system has many advantages.

Reference junction

Allows for electrical contact between the reference electrode and the solution. Slightly permeable to prevent electrolyte from escaping too quickly. Different junctions are available, each with unique characteristics.

YSI pH Electrode Families

YSI has three different families of combination pH electrodes from which to choose, each with unique characteristics. The IDS electrodes are designed for use with the YSI MultiLab, a multiparameter digital lab instrument. TruLine and Science electrodes feature BNC connection and are designed for use with the YSI TruLab, although they can be connected to the MultiLab via a BNC adapter.

MultiLab IDS

  • Intelligent digital sensors (IDS) for the YSI MultiLab
  • Store calibration status and serial number, even if disconnected from the meter
  • Double-junction with Ag/AgCl reference system
  • All electrodes have an integrated temperature sensor
  • Available in gel-filled or liquid electrolyte versions; glass or plastic body

Science

  • Designed for use in unique applications that require a particular junction
  • Features the Silamid reference system, a unique construction of the Ag/AgCl reference system resulting in a more stable and longer lasting electrode with a fast response
  • Double-junction reference and integrated temperature sensor
  • Ground-joint junction with a fast electrolyte outflow is specifically designed for samples high in solids or with low ionic strength
  • Platinum junction for applications that do not require a rate of electrolyte outflow as fast as the ground-joint
  • BNC connector; compatible with any pH meter that has a BNC input

TruLine

  • Designed for use in typical laboratory applications
  • Single-junction electrodes with Ag/AgCl reference system
  • Integrated temperature sensor option
  • Durable electrodes with gel electrolyte for general use; liquid electrolyte sensors for more critical measurements
  • Electrodes with special membranes for unique applications
  • BNC connector; compatible with any pH meter that has a BNC input

YSI pH Electrode Comparison

The following comparison is designed to assist in selecting the electrode that best fits your needs. For additional information, the manual for each electrode can be found on YSI.com.

Electrode Characteristics

Electrode Item Number Connector Shaft Refillable Temp Sensor Reference System Junction Type Number of Junctions Membrane Shape Reference Electrolyte
4110 103740Y Digital Plastic (PPE/PS) No Yes Ag/AgCl Fiber Double Cylindrical Gel, KCI
4110-3 103741Y Digital Plastic (PPE/PS) No Yes Ag/AgCl Fiber Double Cylindrical Gel, KCI
4120 103750Y Digital Plastic (Polyamide) Yes Yes Ag/AgCl Ceramic Double Cylindrical Liquid, 3 M KCI
4130 103780Y Digital Glass Yes Yes Ag/AgCl Platinum Double Conical Liquid, 3 M KCI
pH 15 400350 BNC Glass Yes Yes Ag/AgCl Platinum Single Conical Liquid, 3 M KCI
pH 17 400351 BNC Glass Yes No Ag/AgCl Platinum Single Conical Liquid, 3 M KCI
pH 25 400353 BNC Plastic (PPE/PS) No No Ag/AgCl Fiber Single Cylindrical Gel, KCI
pH 26 400352 BNC Plastic (PPE/PS) No Yes Ag/AgCl Fiber Single Cylindrical Gel, KCI
pH 21 400357 BNC Plastic (PPE/PS) No No Ag/AgCl Hole Single Spear tip Polymer
pH 27 400354 BNC Glass No No Ag/AgCl KPG annular gap Single Flat Polymer
pHT-Pt 400360 BNC Glass Yes Yes Silamid®* Platinum Double Sphere (bulb) Liquid, 3 M KCI
pHT-G 400361 BNC Glass Yes Yes Silamid®* Ground Joint Double Sphere (bulb) Liquid, 3 M KCI
pHT-Micro 400362 BNC Glass Yes Yes Silamid®* Platinum Double Cylindrical Liquid, 3 M KCI

Membrane Shape Properties

Shape Property / Application
sphere (bulb) constant quality, low resistance because of large surface area, for most applications
cone robust, smooth, easy to clean, universally applicable
cylinder shock-proof, for general applications, especially useful for fermenter electrodes
spear high resistance, suitable for penetrating semi-solid media
flat high resistance, shock-proof, easy to clean, suitable for surface measurements

Detailed Electrode Specifications

Electrode Item Number Connector Temp. Sensor pH Range Temp. Range Length Diameter Minimum Immersion Depth* Cable Length (meter) Warranty (months)
4110 103740Y Digital Included in digital connection 0 ... 14 0 ... 80 °C 120 mm 12 mm 12-13 mm 1.5 12
4110-3 103741Y Digital Included in digital connection 0 ... 14 0 ... 80 °C 120 mm 12 mm 12-13 mm 3 12
4120 103750Y Digital Included in digital connection 0 ... 14 0 ... 80 °C 120 mm 12 mm 12-13 mm 1.5 12
4130 103780Y Digital Included in digital connection 0 ... 14 0 ... 100 °C 120 mm 12 mm 16-18 mm 1.5 12
pH 15 400350 BNC Banana plug 0 ... 14 0 ... 100 °C 120 mm 12 mm 16-18 mm 1 12
pH 17 400351 BNC N/A (no temp. sensor) 0 ... 14 0 ... 100 °C 120 mm 12 mm 16-18 mm 1 12
pH 25 400353 BNC N/A (no temp. sensor) 0 ... 14 0 ... 80 °C 120 mm 12 mm 12-13 mm 1 12
pH 26 400352 BNC Banana plug 0 ... 14 0 ... 80 °C 120 mm 12 mm 12-13 mm 1 12
pH 21 400357 BNC N/A (no temp. sensor) 2 ... 13 0 ... 80 °C top: 65 mm
bottom: 25 mm
top: 15 mm
bottom: 5 mm
11-12 mm 1 12
pH 27 400354 BNC N/A (no temp. sensor) 2 ... 13 0 ... 50 °C 120 mm 12 mm N/A (flat membrane) 1 12
pHT-Pt 400360 BNC Banana plug 0 ... 14 -5 ... 100 °C 170 mm 12 mm 17-19 mm 1 12
pHT-G 400361 BNC Banana plug 0 ... 14 -5 ... 100 °C 170 mm 12 mm 12-14 mm 1 12
pHT-Micro 400362 BNC Banana plug 0 ... 14 -5 ... 100 °C top: 70 mm
bottom: 130 mm
top: 12 mm
bottom: 5 mm
10-11 mm 1 12

* Minimum immersion depth is the distance from the tip of the membrane to the outer junction. The junction must always be completely immersed when taking a measurement.

Calibration Buffers and Solutions

YSI calibration buffers are available in pH values of 4, 7, and 10. The following are available:

  • 003821: pH 4 Buffer (box of 6 pints)
  • 003822: pH 7 Buffer (box of 6 pints)
  • 003822: pH 10 Buffer (box of 6 pints)
  • 603824: Assorted Case (2 pints each of pH 4, 7, and 10 buffers)

Potassium chloride (3 M KCl) solution is available (109 705Y) for refillable electrodes. 3 M KCl is also the recommended storage solution for all YSI lab pH electrodes.

Temperature Sensors

External temperature sensors are available for electrodes that do not have an integrated temperature sensor. Examples include ScienceLine Temp 135 and ScienceLine Temp 136 Sensors.

YSI Instruments

YSI MultiLab

IDS (Intelligent Digital Sensor) pH electrodes feature a digital connector and are designed for use with the YSI MultiLab, a multiparameter digital instrument. A BNC adapter (108132Y) also allows these electrodes to be connected to the MultiLab 4010-2W and 4010-3W.

Instrument Model Item Number Description
4010-1W 1FD350Y One channel multiparameter digital instrument with memory and USB connectivity
4010P-1W 1FD350PY One channel multiparameter digital instrument with memory, USB connectivity, and integrated printer
4010-2W 1FD560Y Two channel multiparameter digital instrument with memory and USB connectivity
4010-3W 1FD570Y Three channel multiparameter digital instrument with memory and USB connectivity

YSI TruLab

The TruLine and Science pH electrodes feature BNC connection and are compatible with any pH meter that has a BNC input, such as YSI TruLab meters. The YSI TruLab can also make use of analogue ISE (Ion Selective Electrode) probes with a BNC connector.

Instrument Model Item Number Description
1110 1AA120Y One channel pH/mV instrument
1310 1AA320Y One channel pH/mV instrument with memory and USB connectivity
1310P 1AA320PY One channel pH/mV instrument with memory, USB connectivity, and integrated printer
1320 1GA340Y Two channel pH/mV/ISE instrument with memory and USB connectivity
1320P 1GA340PY Two channel pH/mV/ISE instrument with memory, USB connectivity, and integrated printer

*There is an integrated printer option for the MultiLab 4010-1, TruLab 1310, and TruLab 1320.

Models: 1110 Benchtop pH Meter For Oxygen, 1110, Benchtop pH Meter For Oxygen, pH Meter For Oxygen, Meter For Oxygen, For Oxygen

File Info : application/pdf, 6 Pages, 985.17KB

PDF preview unavailable. Download the PDF instead.

W75-04-YSI-Lab-pH-Electrode-Selection-Guide---Web

References

Adobe PDF Library 17.0 Adobe InDesign 18.3 (Macintosh)

Related Documents

Preview YSI 3821 Buffer Solution pH 4.00 Safety Data Sheet (SDS)
Comprehensive Safety Data Sheet (SDS) for YSI 3821 Buffer Solution pH 4.00, detailing product identification, hazards, composition, first aid, handling, storage, and regulatory information.
Preview YSI Certificate of Analysis for Buffer and Solid Products - Model 2357
Official Certificate of Analysis from YSI Incorporated for Buffer and Solid Products, Model 2357 (Lot 250703GFS098005), confirming average mass and pH specifications.
Preview YSI 1579 Certificate of Analysis for Buffer and Solid Products
Certificate of Analysis for YSI Model 1579 Buffer and solid products, detailing average mass and pH verification results. Includes lot number, method, and compliance information.
Preview YSI 1579 Certificate of Analysis for Buffer and Solid Products
Certificate of Analysis for YSI Model 1579 buffer and solid products, detailing mass and pH verification results. Includes lot number 25G100490 and completion date 7/28/2025.
Preview YSI Certificate of Analysis - Buffer and Solid Products (Model 1579)
Official Certificate of Analysis from YSI Incorporated for Buffer and Solid Products, Model 1579. Includes lot number 25H100065, average mass, and average pH test results verified by QA.
Preview ProDSS Factory Service Plans for YSI Meters
Optimize the performance and lifetime of your ProDSS meter with YSI's comprehensive Factory Service Plans. Learn about preventative maintenance, regular check-ups, discounted replacement parts, and priority treatment.
Preview YSI Water Monitoring Panel: Advanced Water Quality Measurement and Control
Discover the YSI Water Monitoring Panel (WMP), a comprehensive solution for precise measurement of pH, ORP, conductivity, chlorine, and dissolved oxygen. Ideal for municipal water treatment, aquaculture, commercial, and industrial applications, the WMP integrates YSI's Intelligent Digital Sensor (IDS) technology with the IQ SensorNet platform for reliable real-time data acquisition and seamless system integration.
Preview YSI 1579 Certificate of Analysis for Buffer and Solid Products
Certificate of Analysis for YSI Model 1579 buffer and solid products, detailing average mass and pH verification results. Issued by YSI Incorporated.