Concentrate Flow Adjustments
Revision 3
Applicable Models: Barnstead Pacific TII, Barnstead Pacific RO, Barnstead Smart2Pure, Barnstead LabTower EDI, RO & TII
Provided by Unity Lab Services, Part of Thermo Fisher Scientific
Definitions & Tools
- Concentrate: Water that is rejected by the membrane to Drain.
- Permeate: Water that has been purified - Product Water which flows to the tank.
- CCW: Counter Clock Wise (direction of concentrate pressure valve adjustment).
- CW: Clock Wise (direction of concentrate pressure valve adjustment).
To adjust the Concentrate Flow, you will need the following:
- Phillips Screwdrivers
- Volumetric Flask or Cylinder
- Stopwatch
Concentrate Flow Importance
- After installation of any unit with a Reverse Osmosis Membrane, the Concentrate Flow MUST be adjusted.
- Due to possible changes in feed water temperature and pressure, the concentrate flow must be checked and readjusted (if necessary) every 3 months to ensure proper water production and to maximize membrane life.
- Failure to adjust concentrate flow can result in membrane damage or failure.
Measuring Concentrate Flow
All Flows (concentrate & permeate) are listed in Liters Per Hour (LPH).
The "36 Second" Rule:
To save time, you do not want to measure the concentrate (reject) flow rate for one hour. The 36-second rule is based on a simple mathematical formula:
60 seconds x 60 minutes / 100 = 36 seconds
When units are in production mode (making RO water), fill a Volumetric flask or cylinder with the concentrate for 36 seconds (using a stopwatch).
Measure the volume of water in the flask or cylinder and multiply by 100. Your results will be the concentrate flow in LPH.
Example: 320ml of concentrate x 100 = 32 LPH Concentrate Flow.
Models Needing Concentrate Flow Adjustment
The following models require Concentrate Flow adjustment during installation and approximately every 3 months:
- Barnstead Smart2Pure
- Barnstead Pacific TII
- Barnstead LabTower RO
- Barnstead Pacific RO
- Barnstead LabTower EDI
- Barnstead LabTower TII
Smart2Pure Concentrate Adjustment
The unit must be producing water to measure the concentrate flow.
If the storage tank is full, drain enough water from the tank to lower the float. After a 3-minute delay, the unit will start producing water again.
Locate the Concentrate Tubing
The tubing is approximately 1/4" going to the drain.
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) |
---|---|---|
Smart2Pure 3 | 3 | 50* |
Smart2Pure 6 | 6 | 50** |
Smart2Pure 12 | 12 | 60 |
Check and readjust every 3 months.
Based on 17% recovery, the concentrate flow can be adjusted to produce more water, especially if there are problems filling the tank. More water going to concentrate lowers back pressure on the membrane to produce water.
- Adjust concentrate flow down to 15 LPH is acceptable if necessary for permeate production.
- Change to 30 LPH is acceptable if necessary for permeate production.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the blue cover and locate the Pressure Hold Valve (V2).
- Turn the concentrate valve CCW (Counter Clock Wise) to increase or CW (Clock Wise) to decrease concentrate flow in order to obtain the correct concentrate flow.
Image Description: A close-up view of a valve assembly, with a red circle highlighting a brass fitting, identified as the Pressure Hold Valve (V2). The image also shows labels "Konzentrat Concentrate".
Pacific RO Concentrate Adjustment
Locate the Concentrate tubing to measure the concentrate flow.
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) |
---|---|---|
Pacific RO 3 | 3 | 9 |
Pacific RO 7 | 7 | 21 |
Pacific RO 12 | 12 | 36 |
Pacific RO 20 | 20 | 60 |
Pacific RO 40 | 40 | 120 |
Check & readjust every 6 months.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the back panel and locate the Pressure Hold Valve (V2).
- The valve is located at the top of the Distribution Block.
- Turn the concentrate valve CCW to increase the concentrate flow or CW to decrease the concentrate flow.
Image Description: A diagram of connection ports on a unit, showing labels for "Niveau-schalter level switch", "Druckeranschluß Printer connection", "Ko Concentrate outlet", "Ro Raw water inlet", and "High pu".
Pacific TII Concentrate Adjustment
Locate the Concentrate tubing to measure the concentrate flow.
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) |
---|---|---|
Pacific TII/TII UV 3 | 3 | 9 |
Pacific TII/TII UV 7 | 7 | 21 |
Pacific TII/TII UV 12 | 12 | 36 |
Pacific TII/TII UV 20 | 20 | 60 |
Pacific TII/TII UV 40 | 40 | 120 |
Check & readjust every 6 months.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the back panel and locate the Pressure Hold Valve (V2).
- The valve is located at the top of the Distribution Block.
- Turn the concentrate valve CCW to increase the concentrate flow or CW to decrease the concentrate flow.
Image Description: A close-up view of a valve assembly, with a red circle highlighting a brass fitting, identified as the Pressure Hold Valve (V2).
LabTower EDI Concentrate Adjustment
The LabTower EDI requires its RO Concentrate and EDI Product & Concentrate Flows to be adjusted. Adjust the RO Concentrate Flow first.
Locate the Concentrate Tubing
The Concentrate Outlet is located in the front of the unit, below the blue cover.
You will also need to remove the front panel of the LabTower base and adjust the EDI Flow rates via adjustable flow meters (V8 & V9).
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) | Concentrate Flow EDI (Liters/Hour) |
---|---|---|---|
LabTower EDI 15 | 15 | 60 | 2 |
LabTower EDI 30 | 30 | 90 | 4 |
LabTower EDI 45 | 45 | 135 | 5 |
Check & readjust every month.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the back panel and locate the Pressure Hold Valve (V2).
- The valve is located at the top of the Distribution Block towards the front of the unit.
- Turn the concentrate valve to increase or decrease concentrate flow in order to obtain the correct concentrate flow.
Image Description: A close-up view of a valve assembly with a red arrow pointing to a fitting, identified as part of the adjustment mechanism.
LabTower EDI Product & Concentrate Adjustment
After adjusting the concentrate flow, the EDI Product & Concentrate Flows must be adjusted while the unit is running. Start with the Product Flow.
The Product Flow should be adjusted to the model by the flow meter:
- LabTower EDI 15 = 15 LPH
- LabTower EDI 30 = 30 LPH
- LabTower EDI 45 = 45 LPH
The Concentrate Flow should be adjusted by the flow meter to approximately 10% of the Product Flow:
- 15 LPH – Concentrate setting: 1.5-2 LPH
- 30 LPH – Concentrate setting: 3 LPH
- 45 LPH – Concentrate setting: 4.5-5 LPH
Image Description: Two flow meters are shown, one labeled "Product" and the other "Concentrate".
LabTower RO Concentrate Adjustment
Locate the Concentrate tubing to measure the concentrate flow.
The Concentrate Outlet is located on the front of the unit, below the blue cover.
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) |
---|---|---|
LabTower RO 20 | 20 | 60 |
LabTower RO 40 | 40 | 120 |
LabTower RO 60 | 60 | 180 |
Check & readjust every month.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the back panel and locate the Pressure Hold Valve (V2).
- The valve is located at the top of the Distribution Block.
- Turn the concentrate valve CCW to increase the concentrate flow or CW to decrease the concentrate flow.
Image Description: A close-up view of a valve assembly, with a red circle highlighting a brass fitting, identified as the Pressure Hold Valve (V2).
LabTower TII Concentrate Adjustment
Locate the Concentrate tubing to measure the concentrate flow.
The Concentrate Outlet is located on the front of the unit, below the blue cover.
Version | Permeate Flow (Liters/Hour) | Concentrate Flow (Liters/Hour) |
---|---|---|
LabTower TII/TII UV 20 | 20 | 60 |
LabTower TII/TII UV 40 | 40 | 120 |
LabTower TII/TII UV 60 | 60 | 180 |
Check & readjust every month.
Following the 36-second rule, compare the concentrate flow against the chart above. If the flows are not correct, adjustment is necessary.
Adjustment Procedure:
- Remove the back panel and locate the Pressure Hold Valve (V2).
- The valve is located at the top of the Distribution Block.
- Turn the concentrate valve CCW to increase the concentrate flow or CW to decrease the concentrate flow.
Image Description: A close-up view of a valve assembly, with a red circle highlighting a brass fitting, identified as the Pressure Hold Valve (V2).
Revisions
Revision | Change | Date |
---|---|---|
1 | Initial Release | March-2012 |
2 | Page 10, 12, 15 & 17 concentrate flow correction to 110 LPH. Page 15 & 17 permeate flow correction to 60 LPH. | |
3 | Changed concentrate flow adjustments based on updates in manuals for LabTower (50153117 Rev A May 2020) and Pacific units (50152835 Rev A May 2020). This is based on 25% concentrate flow recovery. | August 2020 |
3 | Suggestions for S2P concentrate adjustment could have more room for adjustment, based on 17% recovery per Charlie Serpa. Formula: Permeate/(Permeate+ Concentrate) = % concentrate recovery. | August 2020 |
Unity Lab Services supports your laboratory so you can focus on the science. We implement and manage service programs daily for our customers. Let our expertise complement yours.