Instructions for ARROW models including: CDA-45703-P1A Pressure Injectable Arrowg ard Blue Plus Three-Lumen CVC, CDA-45703-P1A, Pressure Injectable Arrowg ard Blue Plus Three-Lumen CVC

linux, amd64

CDA-45703-P1A - Teleflex Incorporated - Vascular Access Product Catalog

PDF preview unavailable. Download the PDF instead.

LBL053595
10

5

5

CDA-45703-P1A

7 3 Fr.

Lumen

20cm catheter length

.032 inch dia. spring-wire guide

Pressure Injectable Arrowg+ard Blue Plus® Three-Lumen CVC

Contents:
1: Arrowg+ard Blue Plus®1 Three-Lumen Catheter: 7 Fr. (2.5 mm OD) x 20 cm, Pressure Injectable
1: Spring-Wire Guide, Marked: .032" (0.81 mm) dia. x 23-5/8" (60 cm) (Straight Soft Tip on One End - "J" Tip on Other) with Arrow Advancer
1: Catheter: 18 Ga. x 2-1/2" (6.35 cm) Radiopaque over 20 Ga. RW Introducer Needle
1: Pressure Transduction Probe 1: Introducer Needle: Echogenic 18 Ga. x 2-1/2" (6.35 cm) XTW and 5 mL Arrow®
Raulerson Spring-Wire Introduction Syringe 1: Injection Needle: EclipseTM2 22 Ga. x 1-1/2" (3.81 cm) RW and 5 mL Luer-Slip
Syringe 1: Injection Needle: EclipseTM2 25 Ga. x 1" (2.54 cm) RW and 3 mL Luer-Lock
Syringe 1: Syringe: 10 mL Luer-Lock 1: Tissue Dilator: 8.5 Fr. (2.8 mm) x 10.2 cm 1: SecondSiteTM Adjustable Hub: Fastener 1: SecondSiteTM Adjustable Hub: Catheter Clamp 1: SharpsAway® Disposal Cup 1: SharpsAway® II Locking Disposal Cup 1: Maximal Barrier DrapeTM with 4" fenestration 1: Towel 1: Needle Holder 1: Forceps 1: Safety Scalpel: #11 1: Checklist/CLIP Sheet 1: Sterile Procedure Sign 2: Gauze Pad: 2" x 2" (5 cm x 5 cm)

5: Gauze Pad: 4" x 4" (10 cm x 10 cm) 1: Dressing: Tegaderm®3 10 cm x 12 cm 1: Surgical Apparel: Impervious Gown 1: Surgical Apparel: Bouffant Cap 1: Surgical Apparel: Mask with Eye Shield 1: Suture: 3-0 Silk with Curved Needle 1: HemoHopper® Fluid Receptacle
1Licensed under US Patent No. 7,329,412. 2A trademark of Becton, Dickinson and Company. 3A registered trademark of 3M Company.

Not made with natural rubber latex. Store below 25°C (77°F). Avoid excessive heat above 40°C (104°F).
Contraindications: The Arrowg+ard Blue Plus antimicrobial catheter is contraindicated for patients with known hypersensitivity to chlorhexidine, silver sulfadiazine and/or sulfa drugs.

Lumen

Priming Volume*
(mL)

Distal (16 Ga.)

0.39

Medial (18 Ga.)

0.39

Proximal (18 Ga.) 0.40

Gravity Flow Rate
(mL/hr) 2283 913 996

MAX Pressure Injection
Flow Rate** (mL/sec) 10 5 5

* Priming volumes are approximate and are done without accessories.  Flow rate values are approximate and are determined using deionized water at 100 cm head height. ** Pressure injection flow rates are determined using media of 11.8 centipoise (cP) viscosity, with 152 cm administration set tubing.