NARKOVET SAV II VENTILATOR FOR SMALL ANIMALS

QUICK OPERATING GUIDE

1. ABOUT INSTRUMENT COMPONENTS

The NARKOVET SAV II Ventilator consists of several key components. The diagram below illustrates the main parts:

The device features a main unit with a display and controls, and a bellows component for ventilation volume measurement. Various ports are located on the rear and sides for connections.

Instrument Components
SN Part Name Description
1Bellows componentsHouses the bellows for measuring tidal volume.
2DisplayShows system status, parameters, and interface controls.
3External power indicatorIndicates connection to an external power source.
4Power switch buttonTurns the device on/off.
5INSP.HOLD (inspiratory hold) buttonAllows holding the breath at the end of inspiration.
6Shuttle knobUsed for navigating menus and adjusting parameters.
730 mm portConnects to the three-way silicone tube (to ports 16 and 19).
822 mm portConnects to the threaded tube and the anesthesia machine (via port 13).
9USB-B portFor software updates.
10USB portFor data export.
11Pressure sensor portConnected to port 9 (likely for internal sensor connection).
12Battery compartmentHouses the internal battery.
13Exhaust outletConnects to the bellows via the exhaust gas filter tank.
14Drive gas output portConnects the three-way silicone tube to ports 8 and 19.
15Fuse boxContains the fuse for power protection.
16Power cord socketConnects to the external power supply.
1717 mm portConnects to ports 8 and 16 with the three-way silicone tube.

2. MAIN INTERFACE DESCRIPTION

The main interface provides real-time monitoring and control of the ventilator.

Display Overview

The screen displays a pressure-time waveform, real-time parameter values, and mode/setting information.

Key interface elements and their functions:

Interface Parameters and Icons
Parameter/Icon Description
? Alarm SetSet alarm items, lower alarm limit, and upper alarm limit.
? MuteActivates mute for alarms. The mute state lasts for 2 minutes and can be cancelled by pressing the icon again.
⚙️ Self-checkInitiates the device's self-check sequence.
MenuAccesses the main menu with sub-items: parameter list, data record, alarm record, Settings, date and time, and manufacturer maintenance.
Battery StatusDisplays the current battery level.
USB Interface IndicatorDisplayed only when an external device is connected to the USB interface.
Date and TimeDisplays the current date and time, configurable in Settings.

Ventilation Parameters and Modes

The interface shows critical ventilation parameters and allows selection of operating modes.

Ventilation Parameters
Parameter/Icon Description
Pressure-time diagram Visual representation of airway pressure over time during the ventilation cycle.
Real-time parameter display Shows current values for PEAK airway pressure, PEEP, Tidal Volume (VT), Minute Ventilation (MV), and Respiratory Rate (RR).
Mode (VCV, PCV, APNEA) Displays the current ventilation mode. Click to switch between VCV (Volume Control Mode), PCV (Pressure Control Mode), and APNEA mode.
Weight kg The currently set animal weight. The system uses this to suggest recommended parameters.
Tidal volume (VT) Displayed in VCV and APNEA modes. Represents the volume of air inhaled by the animal per breath.
RR (Respiratory Rate) Respiratory rate per minute.
PEAK (Peak Airway Pressure) The target pressure in the inspiratory phase or the peak airway pressure during mechanical ventilation in PCV mode.
I:E (Inspiratory:Expiratory Ratio) The ratio of inspiratory time to expiratory time during mechanical ventilation.
P-Trig (Pressure Trigger) In VCV/PCV modes: triggers the next inspiratory phase when airway pressure reaches the set trigger pressure. In APNEA modes: used to identify spontaneous or mechanical ventilation.
MV (Minute Ventilation) Total volume of air ventilated per minute.
PEEP (Positive End-Expiratory Pressure) A positive pressure maintained in the airway at the end of expiration during mechanical ventilation.
APNEA Time If the animal does not breathe or breathes weakly during this time, APNEA ventilation is initiated.

3. OPERATING PROCEDURES

3.1 Preparation before use

  1. Check that the anesthesia machine is functioning correctly, connected to a stable oxygen source, and has sufficient anesthetic agent and soda lime.
  2. Ensure the anesthesia machine is set to the rebreathing circuit and the APL valve is closed.
  3. Connect the ventilator to the anesthesia machine: Attach one end of the threaded tube to the anesthesia machine's breathing bag port and the other end to the 22 mm port on the ventilator.
  4. Connect the ventilator to the exhaust gas filter: Attach one end of the bellows to the ventilator's EXHAUST port and the other to the exhaust gas filter.
  5. Connect the ventilator to the power cord. While the device has a built-in battery, using external power is recommended for continuous operation. Check battery level and recharge as needed.
  6. Assess bellows replacement: For animals weighing approximately 20 kg, consider replacing the bellows with a 1,500 mL capacity. Estimate if the tidal volume needs to exceed 300 mL.

Replacing the bellows: Rotate the bellows cover counterclockwise to remove it. Take off the black folding pouch, fit the replacement pouch onto the ring, straighten its ripples, and turn the bellows cover clockwise until it snaps into place, ensuring the scale-marked side faces forward.

3.2 Power-on and Self-test

Press and hold the power button for 1 second to turn on the device and initiate the self-test program. Follow on-screen prompts and click [Start Self-Test]. If the self-test passes, a confirmation message will appear, and the system will automatically proceed to the home page in 3 seconds, ready for connection to the animal. If the self-test fails, refer to the user manual for troubleshooting.

Self-test steps typically involve:

  1. Adjusting the flowmeter to 0.
  2. Using the circulation circuit and closing the APL valve.
  3. Sealing the animal port of the circuit.
  4. Connecting the threaded pipe to the breathing bag port of the circuit.
  5. Pressing the O2 flush button to fill the bellows completely.

After successful self-test, the system displays a "Home" screen, and a "Start" button is available.

3.3 Initiating mechanical ventilation

  1. Enter weight: Select the "Weight" parameter on the screen and enter the animal's ideal weight using the shuttle knob. The ventilator will suggest appropriate respiratory parameters for healthy animals (dogs and cats).
  2. Select breathing mode: Choose between "VCV" (Volume Control Mode), "PCV" (Pressure Control Mode), or "APNEA" mode by tapping the screen or using the shuttle knob.
    • VCV: Tidal volume (VT) is the reference. Inspiration ends when the set VT is delivered.
    • PCV: Peak inspiratory pressure (PIP) is the reference. Inspiration ends when the set PIP is reached. Tidal volume may vary.
    • APNEA: Based on VCV, it adds start/stop control for when the animal breathes spontaneously.
  3. Adjust breathing parameters: For specific cases (young, elderly, respiratory diseases, non-mammals), adjust parameters based on intelligent suggestions or manual input.
  4. Anesthesia and Intubation: Select an appropriately sized tracheal cannula. After intubation, inflate the cuff to ensure a seal and prevent air leakage.
  5. Connect to breathing circuit: Set the anesthesia machine's oxygen flow meter to above 0.5 L/min and adjust anesthetic gas concentration as needed.
  6. Start ventilation: Once the animal has reached the desired anesthesia level and spontaneous breathing is weak, click "Start ventilation" on the screen.
  7. Intraoperative monitoring: Observe the airway pressure curve and real-time breathing parameters displayed on the screen to monitor the animal's condition.
  8. Detaching the animal: During suturing, reduce anesthetic gas, turn off the vaporizer, lower respiratory rate, and set pressure adjustment to "-1" (PEEP off) to encourage spontaneous breathing. If irregular pressure or significant spontaneous breathing occurs, mechanical ventilation may need to be maintained temporarily.
  9. Device reset: After detaching the animal, close the breathing circuit. Long-press the quick oxygenation button on the anesthesia machine for 20-30 seconds to clear residual gas. Turn off the anesthesia machine's flow meter. If the ventilator is not used for immediate subsequent surgery, open the APL valve and reconnect tubing.

4. CLEANING AND MAINTENANCE

4.1 Cleaning the product

Always disconnect the device from the power supply before cleaning. Reconnect only after all parts are completely dry.

4.2 Battery performance maintenance

To maintain battery life, it is recommended to service the battery every 3 months:

  1. Remove the device and let it rest for at least 2 hours at room temperature (5°C - 30°C).
  2. Connect to external power and recharge the battery for 5 - 7 hours.
  3. Reposition the device.

4.3 Consumable replacement cycle

Turbo air filter: Replace the turbine filter annually. Check and clean the primary effect cotton every three months.

For detailed information, please refer to the full manual available on the Eickemeyer website.

Contact Information

Eickemeyer KG (Germany Headquarters)
Eltastraße 8, 78532 Tuttlingen
T: +49 7461 96 580 0 | F: +49 7461 96 580 90
info@eickemeyer.de | www.eickemeyer.de

Eickemeyer S.R.L. (Italy)
Via G. Verdi 8, 65015 Montesilvano (PE)
T: +39 085 935 4078 | F: +39 085 935 9471
info@eickemeyer.it | www.eickemeyer.it

Eickemeyer Ltd. (United Kingdom)
3 Windmill Business Village, Brooklands Close, Sunbury-on-Thames, Surrey, TW16 7DY
T: +44 20 8891 2007
info@eickemeyer.co.uk | www.eickemeyer.co.uk

Eickemeyer AG (Switzerland)
Sandgrube 29, 9050 Appenzell
T: +41 71 788 23 13 | F: +41 71 788 23 14
info@eickemeyer.ch | www.eickemeyer.ch

Eickemeyer ApS (Denmark)
Solbakken 26, Hammelev, 6500 Vojens
T: +45 7020 5019
info@eickemeyer.dk | www.eickemeyer.dk

Eickemeyer Inc. (Canada)
617 Douro Street, Suite 205, Stratford, Ont. Canada N5A 0B5
T: +1 519 273 5558
info@eickemeyervet.ca | www.eickemeyercanada.ca

Eickemeyer Sp. z o.o. (Poland)
Al. Jana Pawła II 27, 00-867 Warszawa
T: +48 22 185 55 76 | F: +48 22 185 59 40
info@eickemeyer.pl | www.eickemeyer.pl

Eickemeyer B.V. (Netherlands)
Doejenburg 203, 4021 HR Maurik
T: +31 345 58 9400
info@eickemeyer.nl | www.eickemeyer.nl

Item No. 213020

Models: 213020 SAV II Ventilator FOR Small Animals, 213020, SAV II Ventilator FOR Small Animals, Ventilator FOR Small Animals, FOR Small Animals, Small Animals, Animals

File Info : application/pdf, 8 Pages, 2.22MB

NarkoVet SAV II Operationsguide (Engelsk)

References

Adobe PDF Library 17.0 Adobe InDesign 19.4 (Windows)

Related Documents

Preview Eickemeyer LED 4 Untersuchungsleuchte Benutzerhandbuch
Umfassende Anleitung für die Eickemeyer LED 4 Untersuchungsleuchte, einschließlich Montage, Bedienung, Reinigung, Desinfektion und technischen Daten für den Einsatz in der tierärztlichen Praxis.
Preview Eickemeyer Stainless Steel Table Cleaning and Disinfection Guide
Official cleaning and disinfection instructions for Eickemeyer stainless steel veterinary tables. Learn how to maintain your durable quality product for longevity.
Preview EickWarm Air Veterinary Warming System User Manual
Comprehensive user manual for the EickWarm Air Veterinary Warming System by Eickemeyer. Learn about its features, operation, installation, safety, and maintenance for optimal patient temperature management in veterinary care.
Preview Eickemeyer Laboratory Centrifuge User Manual - Model 718200
Comprehensive user manual for the Eickemeyer Laboratory Centrifuge (Model 718200), covering operation, safety, maintenance, troubleshooting, and specifications. This guide provides essential information for laboratory professionals.
Preview OXYVET IV Oxygen Concentrator User Manual - Eickemeyer
Comprehensive user manual for the Eickemeyer OXYVET IV Oxygen Concentrator (Model 213180), detailing setup, operation, maintenance, and troubleshooting for veterinary use.
Preview Eickemeyer Electric Lift Operating Table User Manual
User manual for the Eickemeyer electric lift operating table, providing operating instructions, safety guidelines, troubleshooting, cleaning, and maintenance information for veterinary professionals.
Preview OrthoVet PRO Battery Powered Drill System User Manual
Comprehensive user manual for the Eickemeyer OrthoVet PRO Battery Powered Drill System, detailing its features, operation, safety instructions, care and maintenance, sterilization procedures, troubleshooting guide, system specifications, and ordering information for veterinary surgical applications.
Preview Social Media Course for Veterinary Clinics: Success with SoMe
Eickemeyer offers a two-module course 'Success with SoMe' for veterinary clinics, focusing on social media strategy, platform selection, audience targeting, content creation, and practical application. Learn from expert Rikke Hansen.