Portable Charged Plate Monitor

Operation and Maintenance Instructions

Description

The SCS 770720 Portable Charged Plate Monitor (CPM) is a handheld meter designed to check the offset voltage and discharge times of air ionizers per ANSI/ESD SP3.3 and ESD TR53. The meter's integrated timer measures decay rates from ±1000V to ±100V to determine the ionizers' ability to neutralize both positive and negative electrostatic charges within the coverage area. The 180-degree rotary head with a 2.75" x 1.25" plate allows the operator to read the display at various angles. The Portable Charged Plate Monitor is calibrated to NIST standards and includes a certificate.

Manufacturer: Desco Industries Inc. (SCS)

Contact: SCS – 926 JR Industrial Drive, Sanford, NC 27332 | East (919) 718-0000 | West (909) 627-9634 | StaticControl.com

Safety Precautions

Danger: This device does not conform to explosion-proof specifications. Do not install it in locations where flammable gases or solvents are handled, such as painting booths etc. Doing so may result in fire or explosion.

Caution: This device is a precision electrical instrument. Avoid installing it in wet, oily, hot, and humid locations. In particular, avoid locations of high humidity and condensation, as there is a possibility of fire due to breakdown.

Installation

Do not use this device in the following locations, as doing so may cause malfunctions:

Maintenance

Handling

Packaging

Features and Components

The device features a 180-degree rotary head with a 2.75" x 1.25" plate, allowing the operator to read the display at various angles. The sensor head rotates in 45-degree increments, clicking into place at each angle.

Key components include:

Important Points About Taking Measurements

Grounding

This product requires a ground connection. Do not use this product without grounding it. Accurate measurements require a reliable ground connection.

Caution: If the meter contacts or is too close to a large, charged body, there is the danger of an electrical discharge. This may cause a breakdown of the device.

Preparation

  1. Insert the battery: Remove the battery case cover and insert two AA batteries. Replace the cover.

    Note: If you are using the optional AC adapter, remove the rubber cap from the adapter terminal before attaching the AC adapter.

  2. Connect the ground cord to the ground terminal: Connect the opposite end to ground. Connect the ground cord securely to the terminal, and connect its alligator clip to a grounded object. Accurate measurements require a reliable ground connection.

Operation

Measuring Ion Balance

  1. Turn the meter on: Slide the power switch to the ON position. The meter will chirp then activate.
  2. Zero point adjust the meter: Bring the plate into contact with a grounded body, and press the 0ADJ button.

    Note: Zero point adjustment must be reset every time the meter is powered.

  3. Set the desired mode: Press the SELECT button, and the meter will scroll through the modes. To measure balance (offset voltage), use I.B mode. The available modes and their LED displays are:
    ModeLED display
    I.B ModeI.B
    +DECAY Mode+DECAY
    -DECAY Mode-DECAY
  4. Point the plate at the measurement location: The meter will measure the balance.

    Note: If the display panel blinks "+1" or "-1", the charged voltage has exceeded the measurable range (over-range). If this occurs, stop measuring immediately, as this may cause damage to the meter.

  5. Turn the meter off: Slide the power switch to the OFF position.

Measuring Neutralization (Discharge) Times

  1. Turn the meter on: Slide the power switch to the ON position. The meter will chirp then activate.
  2. Zero point adjust the meter: Bring the plate into contact with a grounded body, and press the 0ADJ button.

    Note: Zero point adjustment must be reset every time the meter is powered.

  3. Set the desired decay range: The meter changes modes each time the SELECT button is pressed. Select +Decay to measure positive discharge time. Select -Decay to measure negative discharge time. The default setting measures from ±1,000V to ±100V. The setting may also be changed to ±1,000V to ±50V. To change the range, hold the SELECT button and press the START button three times within one second. The display will indicate the selected decay range for two seconds then return to the standby mode.
  4. Point the plate in the airflow of the ionizer to be measured and press the START button: Voltage will apply onto the plate, and the meter will measure the amount of time needed to neutralize it.

    Note: If the display panel blinks "+1" or "-1", the charged voltage has exceeded the measurable range (over-range). If this occurs, stop measuring immediately, as this may cause damage to the meter.

  5. View neutralization time: The meter will chirp and display the neutralization time when the measurement is completed. The neutralization time in seconds is displayed at the bottom of the display.
  6. Turn the meter off: Slide the power switch to the OFF position.

Adjustment and Maintenance

Rotating the Sensor Head

The 180-degree rotary head with a 2.75" x 1.25" plate allows the operator to read the display at various angles. The sensor head rotates in 45-degree increments. When rotating the head, stop at angles where the sensor head clicks into place.

Battery Indicator

The remaining battery charge is displayed on the lower left of the display panel. Replace the battery when the icon shows only one bar remaining.

Error Display

Troubleshooting

If the device does not operate correctly, it may be the result of one of the following:

The display panel is not active when the power is turned on.

The plate voltage discharges rapidly.

The display does not change when approaching the measurement location.

Specifications

ParameterValue
Signal detection methodOscillating chopper method
Display refresh frequency0.2 seconds
Measurable potential range0 to ±1999 V (resolution: 1 V)
Measurement time range0.0 to 99.9 s
Continuous operating limitApproximately 6 hours (with alkaline battery, I.B mode)
Sensor head rotation angle180° (stops every 45°)
DisplayLCD display
Mode switchingSwitch modes by pressing SELECT button
Battery checkRemaining charge displayed in display panel
BatteryTwo AA batteries, Power adapter (sold separately)
Operating temperature/ humidity range10 to 40° C, 60% RH or less, non-condensing
Measurement accuracy±10% rdg ±2 digit
Dimensions2.8 in. x 1.2 in. x 6.9 in. (70 mm x 30 mm x 174 mm)
Plate dimensions2.8 in. x 1.3 in. (70 mm x 32 mm)
Static electricity capacity20 pF ± 2 pF
Weight (including battery)8 oz (240 g)
Country of OriginJapan

Warranty

See the SCS Warranty: StaticControl.com/Limited-Warranty.aspx

Models: TB-9110 Portable Charged Plate Monitor, TB-9110, Portable Charged Plate Monitor, Charged Plate Monitor

File Info : application/pdf, 10 Pages, 724.09KB

PDF preview unavailable. Download the PDF instead.

770720-Portable-Charged-Plate-Monitor-User-Guide

References

Adobe PDF Library 21.11.71

Related Documents

Preview SCS 963E Benchtop Air Ionizer: Installation, Operation, and Maintenance Guide
Comprehensive user guide for the SCS 963E Benchtop Air Ionizer, covering installation, operation, maintenance, calibration, and technical specifications for effective static charge neutralization in electronic environments.
Preview SCS 770052 Workstation Diagnostic Kit: Operation and Maintenance Guide
Detailed guide for the SCS 770052 Workstation Diagnostic Kit, covering operation, maintenance, components, packaging, specifications, and network setup for effective ESD process control monitoring.
Preview SCS 718 Static Sensor Installation, Operation, and Maintenance Guide
Comprehensive guide for the SCS 718 Static Sensor, detailing installation, operation, maintenance, and specifications. Learn how to accurately measure electrostatic charges and use the device with the SCS Air Ionizer Test Kit 718A for ionizer performance verification.
Preview SCS PTC-IIe Manual: Data Transmission Technology Guide
Comprehensive manual for the SCS PTC-IIe, detailing its features, installation, commands, and operation for advanced data transmission technology, including PACTOR, NAVTEX, FAX, and Packet-Radio.
Preview SCS PTC-II User Manual: Data Transmission Technology
Comprehensive user manual for the SCS PTC-II, detailing its features for data transmission technology, PACTOR, AMTOR, Packet-Radio, and more. Includes installation, operation, and command references.
Preview SCS PRO 500 Automatic Gate Opener Installation and User Manual
Comprehensive guide for installing and using the SCS PRO 500 automatic gate opener system, covering mechanical and electrical setup, programming, and troubleshooting.
Preview SCS P4dragon DR-7800 / DR-7400 Installation Guide
This installation guide provides instructions for setting up the SCS P4dragon DR-7800 and DR-7400 high-performance HF radio modems. It covers package contents, system requirements, and an introduction to the PACTOR communication protocols.
Preview SG5 Sleep Trainer Clock User Manual
User manual for the SG5 Sleep Trainer Clock, detailing its features, operation, safety information, and specifications. Learn how to set alarms, use the sleep training functions, and customize settings for optimal sleep routines for children.