User Manual for Pyxis models including: SP-200 Pocket Colorimeter and Fluorometer, SP-200, Pocket Colorimeter and Fluorometer
File Info : application/pdf, 26 Pages, 3.74MB
DocumentDocumentSP-200 Handheld Colorimetric Analyzer User Manual February 8, 2021 Rev. 1.02 Pyxis Lab, Inc. 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com © 2017 Pyxis Lab, Inc. Pyxis Lab Proprietary and Confidential Table of Contents 1 Introduction 3 1.1 Main Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2 Specifications 3 3 Unpacking Instrument 3 3.1 Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 3.2 Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 4 Installation 5 4.1 Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 5 Instrument Overview 6 5.1 Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 5.2 Main Module On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 6 Measurement 7 6.1 Colorimetric Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.2 Select a Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.3 Single-Vial Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 7 Calibration 9 7.1 Set Method Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 7.2 Low Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 7.3 Slope Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 7.4 Restore Factory Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 8 Device Information and Diagnosis 13 8.1 Bluetooth Connection to Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 8.2 Factory Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 9 Use with uPyxis® Mobile App 15 9.1 Download uPyxis® Mobile App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 9.2 Connecting to uPyxis® Mobile App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 9.3 Setting Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 9.4 System Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 9.5 Datalog Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 10 Use with uPyxis® Desktop App 19 10.1 Install uPyxis® Desktop App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 10.2 Connecting to uPyxis® Desktop App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 10.3 System Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 10.4 Setting Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 10.5 Datalog Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 11 Device Maintenance and Precaution 23 11.1 Maintenance Best Practices and Quick Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 11.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 12 Contact Us 23 SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 1 Warranty Information Confidentiality The information contained in this manual may be confidential and proprietary and is the property of Pyxis Lab, Inc. Information disclosed herein shall not be used to manufacture, construct, or otherwise reproduce the goods described. Information disclosed herein shall not be disclosed to others or made public in any manner without the express written consent of Pyxis Lab, Inc. Standard Limited Warranty Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether written or oral, is expressed or implied. Warranty Term The Pyxis warranty term is thirteen (13) months ex-works. In no event shall the standard limited warranty coverage extend beyond thirteen (13) months from original shipment date. Warranty Service Damaged or dysfunctional instruments may be returned to Pyxis for repair or replacement. In some instances, replacement instruments may be available for short duration loan or lease. Pyxis warrants that any labor services provided shall conform to the reasonable standards of technical competency and performance effective at the time of delivery. All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative, or designate. Pyxis warrants these services for 30 days after the authorization and will correct any qualifying deficiency in labor provided that the labor service deficiency is exactly related to the originating event. No other remedy, other than the provision of labor services, may be applicable. Repair components (parts and materials), but not consumables, provided during a repair, or purchased individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the incorporation of a warranted repair component into an instrument extend the whole instrument's warranty beyond its original term. Warranty Shipping A Repair Authorization (RA) Number must be obtained from Pyxis Technical Support before any product can be returned to the factory. Pyxis will pay freight charges to ship replacement or repaired products to the customer. The customer shall pay freight charges for returning products to Pyxis. Any product returned to the factory without an RA number will be returned to the customer. To receive an RMA you can generate a request on our website at https://pyxis-lab.com/request-tech-support/. Pyxis Technical Support Contact Pyxis Technical Support at +1 (866) 203-8397, service@pyxis-lab.com, or by filling out a request for support at https://pyxis-lab.com/request-tech-support/. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 2 1 Introduction The Pyxis SP-200 Handheld Colorimetric Analyzer is a single- and multi-parameter colorimetric testing platform. With the use of auxiliary colorimetric reagent kits, the SP-200 can quickly test a variety of water quality parameters on the spot with high sensitivity. 1.1 Main Features The SP-200 includes the following features: · Innovative technology platform with colorimetric measurement of 6 LED wavelengths · Use of matching reagent kits and prefabricated reagent tubes · Long standby mainframe 10,000+ measurements · Can store up to 8 GB of test data (including test time) · Create user-defined standard curves, view historical data, or analyze data via the uPyxis® App. 2 Specifications Table 1. SP-200 Specifications Item Specification* Part Number (P/N) 50802 Absorbance Linearity Range 02.0 Abs Absorbance Reproducibility ±0.005 Abs (01.0 Abs) Colorimeter Wavelength 365/420/470/525/568/624 nm Wavelength Accuracy ±1 nm Temperature Range 40106 °F (441 °C) Humidity 85% at 106 °F (41 °C) Data Storage 30,000 group data storage, Bluetooth transmission Display 320x240 TFT-LCD, visible under direct sunlight Power Supply 4 AA alkaline batteries Typical Battery Life 10,000 readings Cuvette 24mm diameter cuvette Dimension (L × W × H) Weight 6.69 × 3.15 × 1.77 inch (170 × 80 × 45 mm) 0.88 lbs (400 g) Enclosure Rating IP67 Regulation CE/RoHS * With Pyxis's continuous improvement policy, these specifications are subject to change without notice. Batteries excluded 3 Unpacking Instrument Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipment. Verify that all items listed on the packing slip are included. If any items are missing or damaged, please contact Pyxis Customer Service at service@pyxislab.com. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 3 3.1 Standard Accessories · Quick-Start Guide · Four (4) AA alkaline batteries · Two (2) 10 mL Sample Vials, 24 mm diameter P/N: MA-24 · Bluetooth/USB Adapter for Desktop P/N: MA-NEB · User Manual available online at https://pyxis-lab.com/support/ 3.2 Optional Accessories The following optional accessories can be ordered from Pyxis Customer Service (order@pyxis-lab.com) or Pyxis E-Store at https://pyxis-lab.com/shop/. Table 2. Optional Accessories Accessory Name Part Number (P/N) Pyxis Carrying Case for Handheld Devices 50725 10 mL Sample Vial MA-24 25 mL Sample Vial MA-25 SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 4 4 Installation 4.1 Battery Installation The SP-200 is powered by four alkaline batteries. Typical battery life lasts for 10,000 measurements or 10 months. When the battery capacity is critically low, the SP-200 will display a "LOW BATTERY" warning for 5 seconds and then automatically turn off. *NOTE* Do not use rechargeable nickel cadmium (NiCad) or lithium batteries. Replace the batteries to resume operation of the SP-200 after the battery warning. The SP-200 will automatically turn on in the measurement mode after new batteries are installation. The SP-200 battery compartment, shown in Figure ??, is on the back side of the instrument. Batteries are held in place by a cover secured with two Phillips-head screws. Figure 1. The SP-200 battery compartment Install batteries using the following steps: 1. Remove the battery compartment cover by loosening the two screws. 2. Remove old batteries and dispose of them properly. 3. Following the positive and negative terminal signs in the compartment bottom, snap four new AA alkaline batteries firmly into the battery holder. 4. Replace the battery compartment cover and ensure that the sealing O-ring is lying flat on the battery holder. 5. Fasten the two screws. *NOTE* Failure to properly seat the O-ring may result in water damage to the SP-200. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 5 5 Instrument Overview Figure 2. 5.1 Control Keys The SP-200 has three control keys, as shown in Figure 2. The left ( ), right ( ), and ok ( ) keys are used to launch actions indicated on the LCD display directly above the keys. The labels above the keys indicate the function associated with each key and functions can be changed in different operation modes. 5.2 Main Module On/Off To turn on the SP-200: Press momentarily and release. To turn off the SP-200: Press and hold for about three seconds. Release when the LCD display turns off. The SP-200 turns itself off after 60 seconds without user interaction detected. This is done to conserve battery life. *NOTE* This auto-time off setting may be customized by the user as desired through the uPyxis® Mobile or Desktop App. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 6 6 Measurement 6.1 Colorimetric Methods A wide range of colorimetric methods is supported by the SP-200 analyzer and the number of them keeps increasing with continuous development of Pyxis. The following table is list of Supported Colorimetric Methods: Abbreviation CL-F CL-T Br-T CLO2 PAA H2O2 Bleach-L Bleach-H O3 NH2C CLO2D CLO2H CL2H CL2UH Table 3. Supported Colorimetric Methods Chemical Measured Method Type Free Chlorine DPD Total Chlorine DPD Total Bromine DPD Chlorine Dioxide DPD Peroxyacetic Iodometry Hydrogen Peroxide Iodometry Bleach -- Low Range Direct Read Bleach -- High Range Direct Read Ozone DPD Chloramine, Mono -- Low Range Indophenol Chlorine Dioxide -- Mid Range Direct Read Chlorine Dioxide -- High Range Direct Read Free/Total Chlorine -- High Range DPD Free/Total Chlorine -- Ultra High Range Iodometry Range 2.2 ppm 2.2 ppm 4.5 ppm 5.0 ppm 500 ppm 500 ppm 1.0% 16.0% 1.0 ppm 3.0 ppm 50 ppm 1500 ppm 10 ppm 400 ppm 6.2 Select a Method When powered on, the SP-200 is in the screen of the last selected measurement method, as shown in the Figure 3. Use the following steps to select a different method: 1. Press Color ( ) to launch the Method screen (Figure 4). *NOTE* The top line of the page shows the most frequently selected method, marked by the . 2. Press Up ( ) and Down ( ) to navigate to the desired method. *NOTE* Long press to exit without changing the method. 3. Press to select and start the highlighted method. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 7 Figure 3. Figure 4. 6.3 Single-Vial Measurement 1. Make the Blank sample by filling a sample vial with the water sample and place it in the sample vial compartment of the SP-200. 2. Press Zero ( ). Zero will then appear on the top-left corner of the display (Figure 5). 3. Add the appropriate reagent to the Blank sample. 4. Make the Developed sample by capping the sample vial and inverting several times to fully mix the reagent and sample. 5. Place the Developed sample in the sample vial compartment SP-200. 6. Press TMR1 ( ) to start the method timer (Figure 6). 7. The SP-200 will continuously display the concentration as the timer counts down. Record the final value after timer stops (Figure 7). *NOTE* The rate of the reaction is often faster than the standard preset time. If the concentration remains stable on the concentration-time plot, press Stop ( ) to terminate the timing step early. The last concentration measurement will be displayed. Figure 5. Figure 6. Figure 7. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 8 7 Calibration 7.1 Set Method Parameters Each method has two parameters that can be changed by the user; the units and the chemical formula form. To change these parameters for a given method, use the following steps: 1. Complete a measurement with the desired method. 2. Press Cal/Sys ( ) as needed to highlight either Unit or Form. 3. Press OK ( ) to launch the Choose Unit screen (Figure 8) or the Choose Form screen (Figure 9). 4. Press » ( ) to navigate to the desired parameter option. 5. Press OK ( ) to select the desired parameter option. Figure 8. Figure 9. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 9 7.2 Low Calibration Some methods have a non-zero intercept value in the calibration equation. For these methods, the appropriate non-zero intercept value is shipped pre-loaded in the SP-200. The following steps are used to carry out a low calibration which changes this non-zero intercept value: 1. Complete a measurement with the desired method. 2. Press Cal/Sys ( ) as needed to highlight Low-C. 3. Press OK ( ) to launch the calibration screen (Figure 10). 4. Fill a sample vial with deionized (DI) water or with solution with a known concentration and place it in the sample vial compartment of the SP-200. *NOTE* For best results, the know concentration should be less than half of the max- imum concentration of the method. See the Supported Colorimetric Methods table for concentration ranges for each method. 5. Use « ( ) and ( ) to adjust the concentration value to match the known concentration of the solution or to zero if using deionized (DI) water. 6. Press Calib ( in Figure 11. ) to continue with the entered concentration value. The display updates as shown 7. Press OK ( ) to start the low calibration. 8. The message "Low point calibration succeeded" will appear on the display (Figure 12). 9. Press to exit. Figure 10. Figure 11. Figure 12. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 10 7.3 Slope Calibration Each method has been calibrated prior to shipping, there is no need to calibrate unless a calibration check indicates that the method needs a calibration. The following steps are used to carry out a slope calibration: 1. Complete a measurement with the desired method. 2. Press Cal/Sys ( ) as needed to highlight Slope-C. 3. Press OK ( ) to launch the calibration screen (Figure 13). 4. Fill a sample vial with a solution with known concentration and place it in the sample vial compartment of the SP-200. *NOTE* For best results, the know concentration should be less maximum concentra- tion of the method. See the Supported Colorimetric Methods table for concentration ranges for each method. 5. Use « ( ) and ( ) to adjust the concentration value to match the known concentration of the solution. 6. Press Calib ( in Figure 14. ) to continue with the entered concentration value. The display updates as shown 7. Press OK ( ) to start the slope calibration. 8. The message "High point calibration succeeded" will appear on the display (Figure 15). 9. Press to exit. Figure 13. Figure 14. Figure 15. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 11 7.4 Restore Factory Default Values Each method can be restored to the factory default intercept and slope values, individually. The following steps are used to carry out a factory default of the desired method's intercept and slope values: 1. Complete a measurement with the desired method. 2. Press Cal/Sys ( ) as needed to highlight Default. 3. Press OK ( ). The display updates as shown in Figure 16. 4. Press OK ( ) to start the factory default of the desired method. 5. The message "Restore data succeeded" will appear on the display (Figure 17). Figure 16. Figure 17. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 12 8 Device Information and Diagnosis The Information screen is launched when Cal/Sys ( ) is pressed while not in an ongoing measurement procedure. This screen contains the device serial number, software version, and hardware version (Figure 18). The battery life as a percentage and the MAC addresses for main module also shown. Press Diag ( ) to launch the Diagnosis screen where raw measurement data are displayed (Figure 19). The information has no use for normal operation, but instead is used for device troubleshooting. Provide an image of both the Information screen and the Diagnosis screen when you contact Pyxis (service@pyxislab.com) for troubleshooting your device or call +1 (866) 203-8397. Figure 18. Figure 19. 8.1 Bluetooth Connection to Devices The SP-200 uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone via the uPyxis® Mobile App or to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the uPyxis® Desktop App. To allow the SP-200 to connect via Bluetooth with other devices follow the steps below: 1. Power on the SP-200 by pressing . 2. Press Cal/Sys ( ) to launch the Information screen. 3. Press Diag ( ) to launch the Diagnosis screen. 4. Allow 510 seconds for the message in the top-right corner of the display change from Starting BTLE... to BTLE Started (Figure 19). 5. Choose to connect via one of two options: (a) The uPyxis® Mobile App (see the Use with uPyxis® Mobile App section), or (b) The uPyxis® Desktop App (see the Use with uPyxis® Desktop App section). SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 13 8.2 Factory Reset Use the following steps to restore all device parameters to factory default: 1. Power on the SP-200 by pressing . 2. Press Cal/Sys ( ) to launch the Information screen. 3. Press Diag ( ) to launch the Diagnosis screen. 4. Allow 510 seconds for the message in the top-right corner of the display change from Starting BTLE... to BTLE Started . 5. Press Factory ( ). The display updates as shown in Figure 20. 6. Press OK ( ) to start the factory reset. 7. The message "Data recovery is successful!" will appear on the display (Figure 21). Figure 20. Figure 21. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 14 9 Use with uPyxis® Mobile App 9.1 Download uPyxis® Mobile App Download uPyxis® Mobile App from Apple App Store or Google Play. Figure 22. uPyxis® Mobile App installation SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 15 9.2 Connecting to uPyxis® Mobile App Connect the SP-200 sensor to a mobile smart phone according to the following steps: 1. Follow the steps in the Bluetooth Connection to Devices section to make the SP-200 discoverable. 2. Open uPyxis® Mobile App. 3. On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices. 4. If the connection is successful, the SP-200 and its Serial Number (SN) will be displayed (Figure 26). 5. Press on the SP-200 image. Figure 23. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 16 9.3 Setting Screen From the Setting screen, the user can set the Power off time and Screen off time in seconds. Figure 24. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 17 9.4 System Screen When connected, the uPyxis® Mobile App will default to the System screen. From the System screen, users can change the Device Name, find the Serial Number, Hardware Version, and Firmware Version, as well as update the firmware of the SP-200 by pressing Check Update. If a firmware update is available, press Get Firmware. Once the new firmware is downloaded, press Upgrade Firmware. *NOTE* The firmware update process takes some time and will require the SP-200 to stay within range (approximately 10 ft without obstructions) for the entire duration of the update. Once the update is complete, the SP-200 will reboot which will disconnect the SP-200 from the uPyxis® Mobile App. Figure 25. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 18 9.5 Datalog Screen From the Datalog screen, the user can view and export the internal log files of the SP-200 by pressing Read Datalogs and selecting the desired datalog (these are separated by month). The SP-200 will then populate any relevant log event from the selected datalog which can be viewed in more detail by pressing Read Record or exported as a CSV document by pressing Export/Share. Figure 26. 10 Use with uPyxis® Desktop App 10.1 Install uPyxis® Desktop App Download the latest version of uPyxis® Desktop software package from: https://pyxis-lab.com/upyxis/ this setup package will download and install the Microsoft.Net Framework 4.5 (if not previously installed on the PC), the USB driver for the USB-Bluetooth adapter (MA-NEB), the USB-RS485 adapter (MA-485), and the main uPyxis® Desktop application. Double click the uPyxis.Setup.exe file to install. Figure 27. uPyxis® Desktop App installation Click Install to start the installation process. Follow the screen instructions to complete the USB driver and uPyxis installation. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 19 10.2 Connecting to uPyxis® Desktop App Connect the SP-200 to a Windows computer using a Bluetooth/USB adapter (P/N: MA-NEB) according to the following steps: 1. Follow the steps in the Bluetooth Connection to Devices section to make the SP-200 discoverable. 2. Plug the Bluetooth/USB adapter into a USB port in the computer. 3. Launch uPyxis® Desktop App. 4. On uPyxis® Desktop App, click Device Connect via USB-Bluetooth (Figure 30). 5. If the connection is successful, the SP-200 and its Serial Number (SN) will be displayed in the left pane of the uPyxis® window. Figure 28. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 20 10.3 System Screen Once connected to the device, a picture of the device will appear on the top-left corner of the window and the uPyxis® Desktop App will default to the System screen. From the System screen, users can upgrade the firmware by selecting an appropriate firmware file (contact service@pyxis-lab.com for these firmware files) and clicking Upgrade Firmware. *NOTE* The firmware update process takes some time and will require the SP-200 to stay within range (approximately 10 ft without obstructions) for the entire duration of the update. Once the update is complete, the SP-200 will reboot which will disconnect the SP-200 from the uPyxis® Mobile App. Figure 29. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 21 10.4 Setting Screen From the Setting screen, the user can set the Power off time and Screen off time in seconds. Figure 30. 10.5 Datalog Screen From the Datalog screen, the user can view, delete, and export the internal log files of the SP-200 by clicking Read Datalog List and selecting the desired datalog (these are separated by month). The SP-200 will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Read Datalog, deleted by clicking Delete, or exported by clicking Export as .CSV File. Figure 31. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 22 11 Device Maintenance and Precaution 11.1 Maintenance Best Practices and Quick Tips For greatly increased working life and ease of use of the SP-200 follow the list of maintenance best practices and quick tips below: · Use a soft cloth or lint free paper tissue to clean the sample vial compartment periodically. Remove debris, scale, and deposit promptly. · Although the SP-200 is protected from water damage, it is a good practice to avoid water entering the sample vial compartment. Deposits left behind when the water is evaporated could affect performance. · During storage and transportation, do not leave a sample vial in the sample vial compartment. · Replace batteries when the SP-200 displays a warning message indicating LOW BATTERY voltage. Remove batteries from the SP-200 battery compartment if the SP-200 is going to be placed in storage for a long period time. · When the SP-200 is shipped, a desiccant pack is included in the desiccant compartment underneath the cover of the battery compartment. It is recommended that a new desiccant pack is replaced each time the batteries are replaced. 11.2 Storage The SP-200 should be stored in the temperature range of 0140 °F (-1860 °C) and relative humidity less than 85% at 106 °F (41 °C). Do not leave the SP-200 in a parked vehicle as the temperature inside can reach above 150 °F in summer and -20 °F in winter. Remove batteries from the SP-200 battery compartment when in storage for a long period time. Do not leave a sample vial in the sample vial compartment. 12 Contact Us Pyxis Lab, Inc 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com Phone: +1 (866) 203-8397 Email: service@pyxis-lab.com SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397 23