CEB09BV Transmitter User Manual
USER MANUAL CEB09BV
TRANSMITTER
GUADARRAMA FLOW
Flowmeters and flow measurement technologies. Excellence in precision and repeatability. Made in Spain since 1972.
1

CEB09BV Transmitter User Manual
Index
1. General information ........................................................................................................................ 3 2. CEB09BV Transmitters..................................................................................................................... 4 3. Installation and connections of the CEB09BV transmitters ............................................................ 5
3.1. CEB09BV Transmitters......................................................................................................... 5 3.2. CEB09BVSI Transmitters ...................................................................................................... 6 3.3. CEB09BVSIA Transmitters.................................................................................................... 7 4. CEB09BV Transmitter operating modes.......................................................................................... 8 4.1. Measurement mode................................................................................................................ 8 4.1.1. Display description .......................................................................................................... 8 4.1.2. Display modes ................................................................................................................. 9 4.2. Configuration mode .............................................................................................................. 11 4.2.1. Access and Navigation in Configuration Mode. Modification of Parameters ............... 11 4.2.2. Description of Parameters............................................................................................. 17 5. Basic maintenance operations ...................................................................................................... 26 5.1. Battery change. ..................................................................................................................... 26 5.2. Electronic board change........................................................................................................ 33 6. Contact .......................................................................................................................................... 34
2

CEB09BV Transmitter User Manual
1. General information
The CEB09BV Transmitter is an electronic device that is used to display, locally or remotely, the flow and volume of liquid that passes through the Turbine or Oval Wheel flow meter. This Transmitter transforms the received pulses and, using the specific pulse/liter factor of each flow meter, calculates and displays:
· The Total Volume of liters since the start of operation of the equipment, with a resolution of eight digits. It can be reseted by entering the Configuration Mode (see point 4.2).
· The Partial Volume of liters with easy zeroing. · The Instantaneous Flow, with a resolution of five digits. There are models of CEB09BVTransmitter that have: · Signal outputs:
o Pulses o Analog (4-20 mA.) · Protective box for electronics: o Aluminum and glass. · Positioning relative to the flowmeter: o Attached to the flowmeter.
3

CEB09BV Transmitter User Manual

2. CEB09BV Transmitters

There are different models of CEB09BV Transmitters to better adapt to the needs of each client and application. The following table details the different models of Transmitters that are manufactured and their main characteristics.

Model
CEB09BV CEB09BVSI CEB09BVSIA

Power supply
Internal Internal 24 Vdc

Output Material
Indication Pulses 4 ­ 20 mA

Description

X

Aluminum

Local Indication, without outputs

X

X

Aluminum

NPN ­ PNP transistor output and local indication

X

X

X

Aluminum NPN ­ PNP transistor and 4 ­ 20 mA output and local indication

4

CEB09BV Transmitter User Manual
3. Installation and connections of the CEB09BV transmitters
These transmitters are attached to the Turbine Flowmeter from which they receive the pulses, so they do not require specific mechanical installation. You only have to take into account the mechanical installation of the Flowmeter that is explained in the corresponding manual. Models that have pulse and/or analog outputs require specific electrical connections. VERY IMPORTANT: The external screw of the transmitter must be grounded using a cable with a minimum section of 4 mm.
3.1. CEB09BV Transmitters The CEB09BV transmitters have a protective aluminum casing, resistant to shocks and humidity. They do not have an electrical output, so they DO NOT require specific mechanical installation or electrical connection.
5

CEB09BV Transmitter User Manual
3.2. CEB09BVSI Transmitters The CEB09BVSI transmitters have a protective aluminum casing and a pulse output through a 4-pin M12 connector as seen in the following image.

The pulse output can be connected PNP or NPN according to the following diagrams:

M12 Gray Connector

PNP pulse connection

M12 Gray Connector

NPN pulse connection

+24 Vdc
Pulse IN
0 Vdc PLC
Pulse Input
CONNECTION CABLES · 1 Screened two wire hose. · Ground only one end of the screen. · Do not take the hoses through the power pipes.

+24 Vdc
Pulse IN
0 Vdc
PLC Pulse Input
PULSE OUTPUT · Opto-coupled pulse output. · 32ms configurable pulse width. · Maximum voltage: 40 V · Maximum current: 30 mA

6

CEB09BV Transmitter User Manual
3.3. CEB09BVSIA Transmitters The CEB09BVSIA transmitters have a protective aluminum casing, it has a pulse and analog output through a 4-pin M12 connector as seen in the following image.

The connections of the pulse and analog outputs are detailed in the following diagrams:

M12 Gray Connector

PNP pulse connection
+24 Vdc Analog IN 0 Vdc
PLC Analog Input
+24 Vdc Pulse IN 0 Vdc
PLC Pulse Input

M12 Gray Connector

NPN pulse connection

+24 Vdc

Analog IN

PLC

0 Vdc

Analog Input

+24 Vdc Pulse IN
0 Vdc PLC
Pulse Input

CONNECTION CABLES

PULSE OUTPUT

· 1 Screened two wire hose.

· Opto-coupled pulse output.

· Ground only one end of the screen.

· 32ms configurable pulse width.

· Do not take the hoses through the power pipes. · Maximum voltage: 40 V.

· Maximum current: 30 mA.

ANALOG OUTPUT · By control loop. · Maximum current: 30 mA. · Voltage: 24 Vac · Maximum load: 1.000 Ohm

Note: If you want to use the CEB09BVSIA transmitter without the analog output (such as if you want a transmitter with reading and pulse output without using a battery), the electrical connection is the same in both cases (PNP and NPN), only you must set PIN 3 to 0 Vdc instead of IN Analog.
7

CEB09BV Transmitter User Manual
4. CEB09BV Transmitter operating modes
The CEB09BV Transmitter, like all the display flow meter transmitters that exist on the market, have 2 modes of operation:
· MEASUREMENT MODE. It is the normal operation mode of the instrument. · CONFIGURATION MODE. It is the one used to modify the instrument's operating parameters.
4.1. Measurement mode 4.1.1. Display description
The main elements of the CEB09BVtransmitter display are shown below:
1. Top line. Depending on the "Display Mode", it can indicate Instantaneous Flow or Partial Volume (or Mass).
2. Lower Line. Depending on the "Display Mode", you can indicate the Total Volume (or Mass) or the Partial Volume (or Mass).
3. This label indicates that the top line shows Instantaneous Flow. 4. This label indicates that the Lower line (and sometimes the upper line) shows the Volume
(Partial or Total) of liters. If, in the configuration of the equipment, a different density than 1000 gr/liter is selected, this label disappears. 5. MODE key. This key has several uses depending on the "Display Mode" and is activated with a magnetic key that is supplied with the equipment. 6. ZERO key. If the Partial Volume is displayed on the Upper or Lower line, this key sets it to zero using the magnetic key supplied with the equipment. 7. Arrows indicating:
· Instantaneous Flow Units (l / m, l / h, m3 / h, kg / m, kg / h, t / h). · If the data outputs (Pulse and Analog) are active.
8

4.1.2. Display modes

CEB09BV Transmitter User Manual

The CEB09BV transmitter display can show three types of data:
· Instantaneous Volumetric Flow (volume per unit of time) · Partial Volume with reset · Total volume without resetting
NOTE: If a liquid density other than 1,000 gr/liter is selected in the equipment configuration (see section 4.2), the CEB09BV transmitter would indicate the Flow, Partial or Total mass.
These 3 values can be displayed on the 2 lines of the screen, in different ways depending on the Display Mode. The CEB09BV transmitter has 3 different Display Modes:

MODE
MODE 1 MODE 2 MODE 3

DISPLAY MODES

TOP LINE

LOWER LINE

FLOW FLOW FLOW / PARTIAL

PARTIAL / TOTAL PARTIAL TOTAL

The user can select the "Display Mode" that best suits his needs, accessing the equipment configuration (see section 4.2). Modes 1 and 3 are typically used in fixed installations and Mode 2 in portable installations.
Display Mode 1 The "Display Mode 1" shows the Flow on the upper line, and the Partial or Total on the lower line. When the bottom line shows the Partial, a P appears on the left. To see alternately the Partial or the Total, simply position the magnetic key on the MODE key.

MODE

9

CEB09BV Transmitter User Manual If you want to reset the Partial Volume (or Mass), just position the magnetic key on the ZERO key.
ZERO
Display Mode 2 The "Display Mode 2" shows the Flow in the Upper line and the Partial in the lower line. This Mode is widely used in installations where the flowmeter is portable and is installed between hoses. To reset the Partial Volume (or Mass), simply position the magnetic key on the ZERO key.
ZERO
Display Mode 3 The "Display Mode 3" shows the Flow or Partial on the top line, and Total on the bottom line. When the upper line shows Partial, the "FLOW" label and the arrow indicating the flow unit disappear. To see alternately the Partial or the Flow, simply position the magnetic key on the MODE key.
MODE
If you want to reset the Partial Volume (or Mass), just position the magnetic key on the ZERO key.
ZERO
Note: If a liquid density other than 1,000 gr/liter is selected in the equipment configuration (see section 4.2), the CEB09BV Transmitter would indicate the Flow, Partial or Total mass.
10

4.2. Configuration mode

CEB09BV Transmitter User Manual

Access, navigation and modification of the equipment parameters is done from outside the CEB09BV transmitter. You don't need to physically access the electronics like you did in previous versions.

Access and modification of internal parameters must always be carried out by qualified personnel. Any incorrect alteration can cause the Transmitter to malfunction.

4.2.1. Access and Navigation in Configuration Mode. Modification of Parameters a. Keys and their meaning The CEB09BV transmitter has only 2 keys:
1. MODE located in the lower left part of the Transmitter and 2. ZERO located lower right

To activate each of these keys, just move the magnetic key supplied with each device. However, these two keys become 4 depending on how long we keep the magnetic key on each key:

MODE ZERO

SHORT KEY Approach and quickly remove the magnetic key.

MODE ZERO

LONG KEY
Approach and hold the magnetic key for 4 seconds until the message ESPERE (wait) appears. A decreasing counter
is displayed on the top line until the message SUELTE (release) is displayed.

11

CEB09BV Transmitter User Manual Note: In the different diagrams of this manual these symbols are used so that the user knows at all times if he has to apply a LONG KEY (more alive symbol) or SHORT (dimmer symbol). Screens that appear when the LONG KEY is activated: When the key is held over each key, the LONG KEY is activated and the following screens appear to guide the user:
· WAIT screen. At the bottom appears the word ESPERE (wait) and at the top the number 4 that decreases until it reaches zero. The magnetic key must remain next to the key since, if it is separated from the key before reaching zero, the spindle returns to the previous screen without having performed any function.
· RELEASE screen. When the countdown reaches zero, the SUELTE (release) screen appears. This message tells the user that they can remove the magnetic key from the key. In this way, you will go to the next screen, which will be different depending on each situation.
Note: This operation has been developed so that the user knows at all times what he is doing and has been the result of communication with our clients.
12

CEB09BV Transmitter User Manual
b. Access to Configuration Mode To access the equipment configuration parameters, proceed as follows:
1. Position the magnetic key for 4 seconds on the key MODE. The screen to enter the PASSWORD will appear, in which the first digit from the left flashes:

2. Enter the password 0300 using the keys:
· Key MODE: increases the flashing digit by one unit. · Key ZERO: Shifts the blinking digit to the right.
3. Confirm the KEY by positioning the magnetic key for 4 seconds on the Key ZERO and the first configuration parameter of the equipment is accessed (0-ADJUSTMENT)
c. Navigating the Configuration Mode
· Description of the parameter screens. When positioned on a parameter, a screen appears in which:
o The top line shows:  One or 2 digits to identify each parameter.  A blank space or a comma (in case of 2 digits)  3 letters to describe each parameter
o The bottom line shows the current value of that parameter.

1 or 2 digits to identify the parameter
A blank space or a comma

3 letters to describe the parameter
Current value of the Parameter

· Navigation. Once you have accessed the Configuration Mode, you have the following Navigation options:
o Move to the next Parameter. Press the Key MODE. If the last parameter is reached and the key is pressed again MODE, returns to the first parameter.
o Exit Configuration Mode. Press the Key MODE for 4 seconds.
o Access Parameter Modification. Press the Key ZERO.

13

CEB09BV Transmitter User Manual
d. Modification of Parameters As explained in the previous section, once positioned on the parameter to be modified, press the Key ZERO and the edition of the Parameter is accessed. There are two ways to modify a parameter, depending on the type of Parameter to enter:
· Options Parameter. The value can only be chosen from several options. When editing a parameter with Options, the entire bottom line of the display flashes, and the following actions can be performed: o Select the next Option: Press the Key MODE or the Key ZERO. o Save changes and Exit Configuration Mode: Press the Key ZERO for 4 seconds.
PARAMETER MODIFICATION SCHEME

ZERO

MODE ZERO

ZERO
4 seconds

Position yourself in the Parameter

Access Edition

The first Option flashes

Use the ZERO or MODE keys to select Record

the desired Option

and Exit

· Numerical Parameter. The value can be any number, between a minimum and a maximum. When this type of parameter is edited, the left digit flashes, indicating that it is the one that can be modified. The possible actions are:

o Increase the value of the flashing digit: Press the Key MODE

o Edit the next digit to the right of the current one: Press the Key ZERO.

o Save changes and Exit Configuration Mode: Press the Key ZERO for 4 seconds.

PARAMETER MODIFICATION SCHEME

Position yourself in the Parameter

ZERO
Access Edition

The first digit from the left

MODE ZERO

Increase the value of the digit
The next digit to the right blinks

ZERO
4 seconds

Carry out these operations until the desired value is entered

Record and Exit

14

CEB09BV Transmitter User Manual SUMMARY SCHEME: On the next page you can see a summary diagram so that the user can navigate in an easier way through the Configuration Mode.
15

CEB09BV Transmitter User Manual 16

CEB09BV Transmitter User Manual
4.2.2. Description of Parameters
The characteristics of each of the CEB09BV Transmitter Configuration Parameters and how to edit and modify them are detailed below.
a. - Calibration Adjustment (0 AJT) · This parameter is used to adjust the flowmeter measurement to the customer's installation.
Flowmeters leave the factory calibrated in our Approved Laboratory so, in principle, this parameter should always be zero. However, due to the characteristics of the installation and the type of flowmeter, it may be necessary to perform an Adjustment or Recalibration of the equipment on-site. To make an accurate adjustment you need a reliable and precise standard or reference (Volume or Weight) (Scale or Calibrated Vessel, etc.).
· The user can choose a value between + 99.99% and -99.99% to adjust the flowmeter value to the Standard or Reference.
PARAMETER MODIFICATION SCHEME

1
Position yourself in the Parameter

ZERO
2
Access Edition

The first digit from the left flashes

MODE

Change the value of the digit

ZERO

The next digit to the

right blinks
3

Carry out these operations until the desired value is entered

ZERO
4 seconds
4
Record and Exit

Note: A variation of the value of this parameter alters the flowmeter measurement

17

CEB09BV Transmitter User Manual b. - Liquid Density (1 DEN) · This parameter is used to enter the density of the liquid that passes through the flowmeter. · The user can choose a value between 0 and 2000 gr / liter.
PARAMETER MODIFICATION SCHEME

1
Position yourself in the Parameter

ZERO
2
Access Edition

The first digit from the left flashes

MODE

Increase the value of the digit

ZERO

The next digit to the

right blinks
3

Carry out these operations until the desired value is entered

ZERO
4 seconds
4
Record and Exit

Note: When this parameter has a value of 1000 gr / liter, the CEB09BV Transmitter displays all units in volume (Liters, m3) or volumetric flow (L / h, L / min, m3 / h). However, if this parameter is different from 1000 gr / liter, all units will be displayed in mass (Kg, Tm) or mass flow (Kg / h, K / min or Tm / h).
Note: This parameter can be interesting to use when the density of the liquid that passes through the flowmeter has a small variation.

18

CEB09BV Transmitter User Manual
c. - Calibration Constant (2 CTE) · This parameter shows the constant of the flow meter that has been obtained when calibrating the
equipment in our Approved Laboratory. For this reason, it is important not to modify this parameter unless you do a Recalibration in a Laboratory. If you need to adjust the flowmeter measurement, the above 2 parameters can be used. · The user can select a value between 1 and 99999999 (8 digits) pulses per 1,000 liters.
PARAMETER MODIFICATION SCHEME

1
Position yourself in the
Parameter

ZERO
2
Access Edition

The first digit from the left flashes

MODE

Increase the value of the digit

ZERO

The next digit to the

right blinks

3

Carry out these operations until the desired value is entered

ZERO
4 seconds
4
Record and Exit

Note: A variation of the value of this parameter alters the flowmeter measurement
d. - Reset Totalizer (3 TOT) · This parameter is used to reset the General Totalizer that appears on the bottom line of the work
mode screen. · The user can choose between:
o "Reset YES". Resets the totalizer o "Reset NO". Doesn't zero it
PARAMETER MODIFICATION SCHEME

1
Position yourself in the
Parameter

ZERO
2
Access Edition

The first Option flashes

3
MODE
ZERO
Use the ZERO or MODE keys to select the desired Option

ZERO
4 seconds
4
Record and Exit

19

CEB09BV Transmitter User Manual
e. - Display Modes (4 VIS) · This parameter is used to select the Mode in which you want the main quantities to be seen (Flow,
Partial and Total) on the main screen of the Transmitter (Work mode). · The user can choose between:

MODE

TOP LINE

LOWER LINE

MODE 1

FLOW

PARTIAL / TOTAL

MODE 2

FLOW

PARTIAL

MODE 3

FLOW / PARTIAL

TOTAL

1
Position yourself in the Parameter

ZERO
2
Access Edition

PARAMETER MODIFICATION SCHEME 3
MODE

ZERO

Current option flashes

Use the ZERO or MODE keys to select the desired Option

ZERO
4 seconds
4
Record and Exit

20

CEB09BV Transmitter User Manual

f. - Number of Decimals (5 DEC)
· This parameter is used to select the number of decimal places shown by the 2 totalizers (Partial and Total) and the Flow on the main screen.
· The user can select from the following options: o No decimals o 1 decimal o 2 decimals
PARAMETER MODIFICATION SCHEME

3

ZERO

1
Position yourself in the Parameter

2
Access Edition

Current option flashes

MODE
ZERO
Use the ZERO or MODE keys to select the desired Option

ZERO
4 seconds
4
Record and Exit

21

CEB09BV Transmitter User Manual
g. - Units of Measurement (6 UNL) · This parameter is used to select the flow unit shown on the Transmitter display. · The user can choose between the following options:
o Liters / minute o Liters / hour o m3 / hour
PARAMETER MODIFICATION SCHEME

ZERO

1
Position yourself in the Parameter

2
Access Edition

Current option flashes

3
MODE
ZERO
Use the ZERO or MODE keys to select the desired Option

ZERO
4 seconds
4
Record and Exit

Notes:
· When the flow rate is displayed in Work Mode, a small arrow appears on the right side of the display indicating the selected flow unit.
· If m3 / hour is selected, Totalizers (Partial and Total) are shown in m3

h. ­ Flow Display Filter (7 FDC) · This parameter is a time filter and serves to stabilize the flow reading. Sets a time interval in which
it receives the pulses from the flowmeter and does the instantaneous flow calculations. · User can enter a value between 1 and 60 seconds
PARAMETER MODIFICATION SCHEME

MODE

Increase the value of the digit

ZERO

ZERO

ZERO

The next digit to the 4 seconds

right blinks

Position yourself in the Parameter

Access Edition

The first digit from the left flashes Carry out these operations until the desired value is entered

Record and Exit

Note: it is useful when the pulse frequency (Pulses / Second) is very low or not very stable due to, for example, the use of a pulsating pump

22

CEB09BV Transmitter User Manual
i. - Minimum Flow Alarm (8 QMI) · This parameter is used to indicate on the main screen that the flow rate at which the equipment
works is lower than this value. · The user can choose a value between 0 and 9999999. The flow units used are those defined in
parameter 6.
PARAMETER MODIFICATION SCHEME

1
Position yourself in the Parameter

ZERO
2
Access Edition

The first digit from the left flashes

MODE

Increase the value of the digit

ZERO

The next digit to the

right blinks
3

Carry out these operations until the desired value is entered

ZERO
4 seconds
4
Record and Exit

j. - Maximum Flow Alarm (9 QMA)
· This parameter is used to indicate on the main screen that the flow rate at which the equipment works exceeds this value.
· The user can choose a value between 0 and 9999999. The flow units used are those defined in parameter 6.

PARAMETER MODIFICATION SCHEME

1
Position yourself in the Parameter

ZERO
2
Access Edition

The first digit from the left flashes

MODE

Increase the value of the digit

ZERO

The next digit to the

right blinks
3

Carry out these operations until the desired value is entered

ZERO
4 seconds
4
Record and Exit

23

CEB09BV Transmitter User Manual
The following parameters will only be active for versions of the Transmitter with:
· Pulse Output "SI" (Parameters 10 and 11), or · Pulse and Analog Output "SIA" (Parameters 10, 11 and 12)
k. - Pulses per unit of Volume (10 PHL)
· This parameter is used to adjust the number of pulses per unit volume that the Transmitter pulse output generates.
· The user can choose between the following values: o 100 Pulses / Liter o 10 Pulses / Liter o 1 Pulse / Liter o 1 Pulse / 10 Liters o 1 Pulse / 100 Liters o 1 Pulse / 1000 Liters
PARAMETER MODIFICATION SCHEME

ZERO

MODE ZERO

ZERO
4 seconds

Position yourself in the Parameter

Access Edition

The first digit from the left flashes

Carry out these operations until the desired value is entered

Record and Exit

Note: The electronics of the Transmitter is a pulse divider, therefore, the selected value cannot be higher than that supplied by the Turbine or Oval Wheel flowmeter.

24

CEB09BV Transmitter User Manual
l. - Duration or pulse width (11 ANP) · This parameter is used to adjust the width value of each spindle output pulse. · The user can choose a value between 1 and 999 milliseconds.
PARAMETER MODIFICATION SCHEME

MODE ZERO

Increase the value of the digit
The next digit to the right blinks

ZERO
4 seconds

Position yourself in the Parameter

Access Edition

The first digit from

the left flashes

Carry out these operations until the desired value is entered

Record and Exit

m. ­ 20 mA Flow value - Full Scale (12 FES)

· This parameter is used to adjust the full-scale value of the Analog Output, that is, the flow rate produced by a 20-mA output.
· The user can choose a value between 0.01 and 999999.99 Liters / hour.

PARAMETER MODIFICATION SCHEME

,
Position yourself in the Parameter

ZERO
Access Edition

MODE

Increase the value of the digit ,

,
The first digit from the left flashes

ZERO

The next digit to the

right blinks

,

Carry out these operations until the desired value is entered

4 seconds
Record and Exit

25

CEB09BV Transmitter User Manual
5. Basic maintenance operations
5.1. Battery change.
The CEB09BV and CEB09BVSI models are self-powered by a 3.6V Lithium AA format battery. The duration of this battery varies depending on the use that is given to the equipment, but normally they have a work life of 4 to 7 years. A few weeks before the battery is completely discharged, a flashing battery symbol appears on the screen.
Low battery symbol
This can also occur due to a shock, vibration, rough jolt, etc., which causes the battery to be removed in place and causes a small loss of voltage. To change the battery, follow these steps:
26

Steps 1

CEB09BV Transmitter User Manual
Description Remove the screw cap from the Transmitter

Remove the 4 screws that hold the Transmitter cover and remove the Transmitter cover 2
Raise the electric bridge. Wait 5 seconds and put it back on, if the battery icon persists, follow 3
the procedure, if it disappears, go to step 10.

27

CEB09BV Transmitter User Manual

Remove the 2 screws that hold the aluminum bracket to the Transmitter body 4

5

Remove the 4 screws that hold the plate to the aluminum bracket

28

CEB09BV Transmitter User Manual

Remove the depleted 3.6V Lithium battery 6

7

Put a new 3.6V Lithium battery

29

CEB09BV Transmitter User Manual

Screw in the 4 screws. Accommodate wiring and silica bags inside Transmitter body 8

9

Screw in the 2 screws of the aluminum bracket.

30

CEB09BV Transmitter User Manual

Put the electric jumper on the top two pins 10

eleven

Put the Transmitter cover and screw the 4 screws

31

CEB09BV Transmitter User Manual Put the screw cap on the Transmitter 12
32

CEB09BV Transmitter User Manual
5.2. Electronic board change
The customer can replace, by himself, the electronics of the CEB09BV Transmitters, in those cases that it is necessary, following the following steps:
1. Remove from damaged electronics.
a. Open the front cover b. Disconnect the connection bridge if the board is live. c. Loosen the 4 screws that hold the electronics d. Loosen the terminal block and loosen the cables. This terminal block connects the
electronics with the pulse sensor. Jumper
disconnected
Jumper disconnected
2. Fit the new electronics. The electronics are shipped with the supply voltage disconnected and
must remain so until installed. a. Connect the terminal block to the cables coming from the pulse sensor. It does not matter the order in which the cables are placed, since they do not have polarity. b. Put the board in its place and put the screws that hold it. c. Give voltage to the board by placing the jumper between the top 2 pins as seen in the image. d. Place the front cover of the Transmitter.
Jumper Connected
3. Adjust the flow meter constant. Check that the constant (Pulses / Liter) on the plate coincides
with that of the flow meter, for which the procedure is as indicated in section 4.2.
33

CEB09BV Transmitter User Manual
6. Contact
For any problem you may encounter or service, you need, do not hesitate to contact the G - Flow offices.

Telephone:

+34 916378174 / +34 916378175

E-mail:

serviciotecnico@g-flow.com

Address:

Office Laboratory Factory

Calle Justina Velasco Martín 2, Pol. Ind. Los Llanos 28260 ­ Galapagar ­ Madrid.

34



References

Microsoft Word for Microsoft 365