helix+ User Manual

HIS CAPTURE KIT 2

for ligand strand 2 with green dye Ga - only for Y-structure

Dynamic Biosensors GmbH | Order Number: HK-NTA-2 v1.0

Key Features

  • This kit is designed for capture of histidine-tagged proteins (His6 or His10) using Tris-NTA.
  • Includes Ligand strand 2 with Tris-NTA for 20 regenerations.
  • For functionalization of ligand strand 2 (green arm) of the Y-structure.
  • Used in combination with HK-NTA-1 for the red arm of the Y-structure, it is ideal for dimerization projects.
  • Homo-/hetero-proteins can be coupled easily to the arms via his-tag capture.
  • Compatible with heliX® Adapter Chip, both Spot 1 and Spot 2

heliX® Adapter Chip Overview

The heliX® Adapter Chip features 2 spots with 2 different anchor sequences for DNA-encoded addressing. Spot 1 is functionalized with the capture molecule while Spot 2 is used as real-time reference.

Single Capture

The 'Single Capture' diagram illustrates the functionalization of a single spot on the heliX® Adapter Chip. Spot 1 is shown with 'Anchor strand 1' and is functionalized with 'Ligand strand 2 (L2) - NTA', indicated by a green dye. A captured histidine-tagged protein is shown bound to the NTA ligand. Spot 2, equipped with 'c-Anchor strand 2', is designated as a real-time reference. The diagram indicates the 'chip surface'.

Double Capture

The 'Double Capture' diagram depicts a Y-structure setup on the heliX® Adapter Chip. Spot 1 features 'Anchor strand 1' and 'c-Anchor strand 2', functionalized with 'Ligand strand - NTA' (red dye) on one arm and 'Ligand strand 2 (L2) - NTA' (green dye) on the other. Two histidine-tagged proteins are shown captured, one on each arm of the Y-structure. Spot 2, with 'c-Anchor strand 2', serves as a reference. The diagram indicates the 'chip surface'.

Product Description

Order Number: HK-NTA-2

Contents and Storage Information

MaterialCapConcentrationAmountBufferStorage
NTA - Ligand strand 2Purple500 nM2 x 100 μLTE40 [1]-20°C
Loading Solution (NiCl2)Transparent10 mM5 x 1500 μLTE40 [1]-20°C
Imidazole SolutionTransparent250 mM10 x 2000 μLTE140 [2]-20°C

For research use only.

This product has a limited shelf life; please see expiry date on label. After preparation of ready-to-use solution, the expiry date is 6 months.

Preparation

  1. Step 1: Hybridization of Y-structure strands

    For surface functionalization, the Y-structure Red Adapter strand (harboring the red dye Ra) and the Y-structure Green Adapter strand (harboring the green dye Ga) need to be pre-hybridized with the NTA - Ligand strand and NTA - Ligand strand 2 to build the Y-structure. In solution hybridization:

    1. Mix Y-structure Red Adapter strand with Ra (400 nM), Y-structure Green Adapter strand with Ga (400 nM), NTA Ligand strand (500 nM), and NTA - Ligand strand 2 (500 nM) at a 1:1 ratio (v/v).
    2. Incubate the solution from step i) at room temperature (RT) for at least 2 hours to ensure complete hybridization. Overnight incubation at 4°C is also possible.
  2. Step 2: Mix with cAnchor strand 2

    Mix the solution from step ii) with cAnchor strand 2 (100 nM) at a 1:1 ratio (v/v).

  3. Step 3: Ready for functionalization

    The solution is now ready to use for heliX® Adapter Chip functionalization.

Example

Required volume for one functionalization for double capture on Y-structure (Spot 1): 35 μL with a final concentration of 50 nM.

Red Adapter strand with Ra (400 nM)Green Adapter strand with Ga (400 nM)NTA - Ligand strand (500 nM)NTA - Ligand strand 2 (500 nM)cAnchor strand 2 (100 nM)
4.5 μL4.5 μL4.5 μL4.5 μL18 μL

Assay Setup in heliOS

This specific kit requires a customized method for His-tag capture plus Y-Structure FRET Kinetics. This method is not currently provided among the verified assays. An advanced heliOS user can create it by applying default parameters from existing workflows. For assistance in creating the new method, please contact the support team at support@dynamic-biosensors.com.

TIP: As the stability of his capture is affected by the protein, in case of long dissociations, consider using the classic conjugation approach.

Useful Order Numbers

Product NameCommentOrder No
heliX® Adapter ChipChip with 2 detection spotsADP-48-2-0
Y-Structure Kit 1: for proximity binding assay on Spot 1For studying ternary complex formationHK-YS-1
Y-Structure Kit 2: for proximity binding assay on Spot 2For studying ternary complex formationHK-YS-2
His Capture Kit: with red dye (Ra)For capturing of histidine-tagged proteins using Tris-NTA on the red arm of the Y-structure.HK-NTA-1

Contact

Dynamic Biosensors GmbH
Perchtinger Str. 8/10
81379 Munich
Germany

Bruker Scientific LLC
40 Manning Road, Manning Park
Billerica, MA 01821
USA

Order Information: order@dynamic-biosensors.com

Technical Support: support@dynamic-biosensors.com

Website: www.dynamic-biosensors.com

Instruments and chips are engineered and manufactured in Germany.

©2025 Dynamic Biosensors GmbH

For Research Use Only. Not for use in clinical diagnostic procedures.

[1] TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA

[2] TE140: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA

Models: HK-NTA-2 HIS Capture Kit 2, HK-NTA-2, HIS Capture Kit 2, Capture Kit 2, Kit 2

File Info : application/pdf, 5 Pages, 1.53MB

PDF preview unavailable. Download the PDF instead.

HK-NTA-2

References

HexaPDF version 1.4.0

Related Documents

Preview Bruker Biosensors heliX+ New Y-Structure His Capture Kit User Manual
Comprehensive user manual for the Bruker Biosensors heliX+ New Y-Structure His Capture Kit (HK-NYS-NTA), detailing key features, preparation, assay setup, and order information for histidine-tagged protein capture.
Preview Bruker Biosensors helix+ New Y-Structure Kit 2 User Manual for Proximity Binding Assays
User manual for the Bruker Biosensors helix+ New Y-Structure Kit 2, detailing key features, product contents, preparation steps, adapter chip overview, and assay setup for proximity binding assays. Covers ternary complex formation and binary binding using FRET and Fluorescence Proximity Sensing (FPS).
Preview Bruker Biosensors helix+ Y-Structure Kit 1: User Manual for Proximity Binding Assays
Comprehensive user manual for the Bruker Biosensors helix+ Y-Structure Kit 1 (HK-NYS-1). Learn about key features, contents, preparation steps, and assay setup for studying ternary complex formation and binary binding using FRET and Fluorescence Proximity Sensing (FPS).
Preview Bruker Biosensors Amine Coupling Kit 2 User Manual
User manual for the Bruker Biosensors Amine Coupling Kit 2, detailing the process for coupling molecules with primary amines to the ligand strand using spin column purification.
Preview Bruker Biosensors heliX cyto Labeling Kit Green Dye User Manual
User manual for the Bruker Biosensors heliX cyto Labeling Kit Green Dye, detailing its features, a 3-step labeling workflow, product contents, required materials, troubleshooting, and order information for labeling protein analytes with a green fluorescent dye.
Preview Bruker Bioscience Helix Cyto Normalization Solution (Red Dye) User Manual
User manual for the Bruker Bioscience Helix Cyto Normalization Solution (Red Dye), detailing its features, product description, preparation, and application notes for scIC measurements.