Hach Intellical ISEF121 Fluoride Ion-Selective Electrode User Manual

Document Number: DOC022.53.80028

Edition: 5, May 2021

Section 1: Product Overview

The Intellical ISEF121 series probes are digital, combination ion-selective electrodes (ISE) designed to measure the concentration of fluoride in wastewater, drinking water, and general water samples. These probes feature a built-in temperature sensor, a non-refillable gel reference, and non-replaceable membranes.

Figure 1: Probe Overview

The probe consists of the following components:

Section 2: Specifications

Specifications are subject to change without notice.

Specification Details
Probe typeDigital combination probe with a non-refillable reference junction and a built-in temperature sensor
Measurement range0.01 mg/L (5x10-7 M) to 19,000 mg/L (1 M) F-
Accuracy±0.02 mV or 0.05% (the larger value)
Sample pH rangepH 4 to 8 (must be pH 5.0 to 5.5 after the addition of Fluoride ISA)
Reference typeAg/AgCl
Reference junctionSingle junction (annular porous PTFE)
Slope59 mV/pF (90 to 110% at 25 °C (77 °F) in linear range per Nernstian theoretical value)
Linear region0.1 mg/L to 19,000 mg/L F-
Temperature accuracy±0.3 °C (±0.54 °F)
Temperature sensor type30 kΩ NTC thermistor
Operating temperature5 to 50 °C (41 to 122 °F)
Storage temperature5 to 35 °C (41 to 95 °F)
Response time in linear region< 60 seconds (application dependent)
Minimum sample volume25 mL
Minimum immersion depth25.4 mm (1 in.)
Body material (standard)Epoxy
Sensor materialSolid-state LaF3 crystal
Reference electrolyteNon-refillable Dritek gel reference element
Cable connectionM12 digital output and connector
DimensionsDiameter: 12 mm (0.47 in.)
Length: 175 mm (6.9 in.) total; 103 mm (4.1 in.) below head
Cable length: ISEF12101: 1 m (3.3 ft); ISEF12103: 3 m (9.8 ft)
Warranty1 year on the probe. This warranty covers manufacturing defects, but not improper use or wear.
CertificationsCE, FCC/ISED

Section 3: Safety Information

3.1 Intended Use

The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory. The Intellical probes do not treat or alter water.

3.2 Use of Hazard Information

DANGER: Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

WARNING: Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION: Indicates a potentially hazardous situation that may result in minor or moderate injury.

NOTICE: Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.

3.3 Precautionary Labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.

WEEE Symbol: Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

3.4 Product Hazards

CAUTION: Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.

CAUTION: Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

Section 4: Preparation for Use

Prepare the probe for calibration and measurement as follows:

  1. Remove the sensor protection cap from the probe.
  2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
  3. Condition the probe before use. Soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes. Do not add the ISA to the standard solution.
    Note: Condition the probe each day for 15 to 30 minutes or for 30 to 60 minutes after long-term storage.
  4. Make sure that the meter has the correct date and time settings. The service-life time stamp in the probe comes from the date and time settings in the meter.
    Note: Some meters automatically open the date and time settings when the meter starts for the first time, or after battery replacement.
  5. Connect the probe to the meter.

Section 5: Calibration

The following procedure is applicable to meters that can connect to Intellical ISE probes. Refer to the applicable meter documentation for meter operation and probe-specific settings.

5.1 Calibration Notes

Read the notes that follow before calibration:

5.2 Calibration Procedure

Step 1: Navigate to the calibrate menu. Select the probe, if applicable. The display shows the standard solutions to use for calibration.

Step 2: Add 25 mL of each standard solution to plastic beakers.

Step 3: Add one Fluoride Ionic Strength Adjustor (ISA) Powder Pillow to each 25 mL of standard solution.
Note: As an alternative, add 5 mL of Fluoride ISA Solution.

Step 4: Add a stir bar to the first standard solution. Place the standard solution on an electromagnetic stirrer. Stir at a moderate rate.

Step 5: Rinse the probe with deionized water. Blot dry with a lint-free cloth.

Step 6: Immerse the probe in the standard solution with the sensor fully submerged. Do not place the probe on the bottom or sides of the beaker.

Step 7: Shake the probe from side to side to remove air bubbles.

Step 8: Stir for 30 to 60 seconds, then read the fluoride concentration of the standard solution.

Step 9: Repeat steps 3 through 8 for the remaining standard solutions.

Step 10: Save the calibration.

Section 6: Sample Measurement

The following procedure is applicable to meters that can connect to Intellical ISE probes. Refer to the applicable meter documentation for meter operation and probe-specific settings.

6.1 Sample Measurement Notes

Read the notes that follow before sample measurements:

6.2 Sample Measurement Procedure

Step 1: Pour 25 mL of fresh sample into a 50-mL plastic beaker.

Step 2: Add one Fluoride Ionic Strength Adjustor (ISA) Powder Pillow.
Note: As an alternative, add 5 mL of Fluoride ISA Solution.

Step 3: Add a stir bar. Place the beaker on an electromagnetic stirrer. Stir at a moderate rate.

Step 4: Rinse the probe with deionized water. Blot dry with a lint-free cloth.

Step 5: Immerse the probe in the sample with the sensor fully in the sample. Do not place the probe on the bottom or sides of the beaker.

Step 6: Shake the probe from side to side to refresh the reference junction and remove air bubbles.

Step 7: Stir for 30 to 60 seconds, then read the fluoride concentration of the sample. The display shows the value when the reading is stable.

6.3 Low-Level Measurements

Use the guidelines that follow for measurements at low concentrations (<1 mg/L F-):

6.4 Interferences

The sensing element will measure some other ions that are known to interfere with the method. The probe response to other ions usually increases the mV potential and causes a positive error. The response to other ions can be semi-quantitatively calculated through the Nikolsky equation, an extended Nernst equation:

E = E° + (RT/(zF))ln[aNa + KNax × ax]

Where:

Cations and most anions do not interfere with the fluoride measurements. Anions commonly associated with fluoride (chloride (Cl-), bromide (Br-), sulfate (SO42-), bicarbonate (HCO3-), phosphate (PO43-) and acetate (C2H3O2-)) do not interfere with the probe operation.

Hydroxide ions (OH-) interfere with the probe response when the pH is higher than pH 8. Some ions, such as carbonate (CO32-) or phosphate (PO43-), increase the pH, which increases the hydroxide (OH-) interference, but the ions do not directly interfere. Hydroxide ions (OH-) interfere with the probe response when the level of hydroxide is more than 10% of the level of fluoride. The error increases with an increase in pH and a decrease in fluoride.

When the Fluoride ISA is added to standards and samples, the pH is buffered between 5.0 and 5.5, which prevents interference from hydroxide ion.

The selectivity coefficient is the approximate apparent increase in the measured concentration that is caused by one unit of the interfering ion (e.g., 1 unit of OH- increases the fluoride concentration by 0.1). The approximate selectivity coefficients for some ions with the Intellical Fluoride ISE are shown in Table 1.

Interference Selectivity coefficient
Hydroxide (OH-)0.1 (below pH 8)

Section 7: Verify the Calibration

Measure the value of a fresh standard solution at regular intervals to make sure the result is accurate. The meter compares the expected standard solution value to the measured value and accepts or rejects the measurement. The user can change the standard solution and acceptance criteria for verification in the probe-specific settings.

Note: Password protection may prevent access to the acceptance criteria.

7.1 Verification Procedure

Step 1: Go to the verification menu. The display shows the standard solution to use for verification.
Note: Menu name for HQd meters: Run check standard.

Step 2: Pour 25 mL of the standard solution into a 50-mL plastic beaker.

Step 3: Add one Fluoride Ionic Strength Adjustor (ISA) Powder Pillow.
Note: As an alternative, add 5 mL of Fluoride ISA Solution.

Step 4: Add a stir bar. Place the beaker on an electromagnetic stirrer. Stir at a moderate rate.

Step 5: Rinse the probe with deionized water. Blot dry with a lint-free cloth.

Step 6: Immerse the probe in the standard solution with the sensor fully in the solution. Do not place the probe on the bottom or sides of the beaker.

Step 7: Shake the probe from side to side to remove air bubbles.

Step 8: Stir for 30 to 60 seconds, then read the value of the standard solution. The meter accepts or rejects the result.

Section 8: Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time, and life of the probe. Keep the probe clean between measurements.

8.1 Clean the Probe

Clean the probe regularly to remove mineral or sample buildup on the sensing element. Symptoms of contamination:

Steps:

  1. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
  2. Clean the sensing element as follows:
    1. Lightly polish the sensing element with a small quantity of plain fluoride toothpaste (no teeth whiteners or abrasives). Use a circular movement and polish until the sensing element is clean.
    2. Fully rinse with deionized water.
    3. Soak the probe in isopropyl alcohol for 30 minutes.
    4. Rinse the probe with deionized water.
    5. Condition the probe. Refer to Preparation for use on page 5.
  3. Soak the probe for 30 minutes in a 1-mg/L fluoride standard solution.

8.2 Storage

Keep the sensor protection cap on the probe during storage to prevent damage to the sensing element.

Note: Make sure to condition the probe after long-term storage. Refer to Preparation for use on page 5.

  1. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
  2. Make sure that the probe sensor is dry, then install the sensor protection cap. Refer to Product overview on page 3.

Section 9: Troubleshooting

Refer to Table 2 for general troubleshooting information. To check the probe performance, refer to Slope check on page 12. To check the accuracy of sample measurements, refer to Standard additions check on page 12.

Problem Possible Cause Solution
Decreased probe performance causes slow stabilization and prevents accurate calibrations or measurements. The probe has contamination on the sensing element. Clean and condition the probe. Refer to Clean the probe on page 10.
The reference junction is clogged. Fully rinse the reference junction with deionized water. Shake the probe down to remove air bubbles.
The probe is not conditioned to the sample sufficiently. Condition the probe. Refer to Preparation for use on page 5.
Sample properties cause slow stabilization or inaccurate measurements. The sample pH with ISA is incorrect. Make sure that the sample pH after the ISA is added is between 5.0 and 5.5. Make sure to add one ISA pillow to each 25 mL of sample.
The sample temperature is low, or there is a large temperature difference between samples. Increase the sample temperature or adjust the temperature of different samples to be the same (within 2 °C (3.6 °F)).
Air bubbles are around or below the probe tip. Carefully tap or shake the probe to remove air bubbles.
Procedure problem causes slow stabilization and prevents accurate calibrations or measurements. The ISA was not added. Add one ISA powder pillow to each 25 mL of sample and standard solution.
The stir speed is too slow or too fast. Try a different stir speed.
Magnetic stirrers can become warm and increase the solution temperature. Put a piece of insulating material between the stirrer and beaker.
An incorrect standard solution was used or the standard solution has contamination. Use the specified standard solution of good quality.
Too much time passed after the ISA was added. Measure the sample or standard solutions immediately after the ISA is added.

9.1 Slope Check

Use the mV value of two standard solutions to make sure the probe gives the correct slope.

  1. Prepare two standard solutions that are ten times apart in concentration (e.g., 1 mg/L and 10 mg/L F-). Select standard solutions with a concentration above and below the typical sample concentration. Use a minimum concentration of 1 mg/L.
  2. Use the measurement procedure to add the ISA and measure the mV value of each standard solution.
  3. Calculate the difference in the mV value of the two standard solutions to find the slope. If the probe is in good condition, the slope will be 59 mV (within the ± slope limits of the method) at 25 °C (77 °F).

9.2 Standard Additions Check

To make sure that the sample measurement is accurate, add a small volume of a standard solution to the sample and calculate the percent recovery. The sample with the known volume of standard solution is known as a spiked sample.

  1. Use the measurement procedure to measure the concentration of a 25-mL sample.
  2. Use a pipet to add the applicable volume of standard solution to the sample. Refer to Table 3.
Measured sample concentration Volume of standard to add Concentration of standard solution
1 to 2 mg/L0.5 mL100 mg/L F-
3 to 6 mg/L1.0 mL100 mg/L F-
7 to 15 mg/L0.3 mL1000 mg/L F-
15 to 30 mg/L0.5 mL1000 mg/L F-
30 to 60 mg/L1.0 mL1000 mg/L F-

3. Measure the concentration of the spiked sample.

4. Calculate the expected (theoretical) concentration of the spiked sample:

CE = (CS × VS/VT) + (CSS × VSS/VT)

Where:

5. Calculate the percent recovery of the standard addition. A percent recovery of 100 (±5)% is an indication that the sample measurements are accurate.

Percent recovery = CM/CE × 100

Where:

Section 10: Consumables

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.

Description Quantity Item no.
Fluoride Ionic Strength Adjustor Powder Pillows100/pkg258999
Fluoride Ionic Strength Adjustor Solution3.78 L2829017
Fluoride Standard Solution, 0.5 mg/L as F500 mL40505
Fluoride Standard Solution, 1 mg/L as F500 mL29149
Fluoride Standard Solution, 2 mg/L as F500 mL40520
Fluoride Standard Solution, 5 mg/L as F500 mL2797149
Fluoride Standard Solution, 10 mg/L as F500 mL35949
Fluoride Standard Solution, 100 mg/L as F500 mL23249
Wash bottle, polyethylene, 500 mL162011
Disposable wipes, 11 x 22 cm280/pkg2097000
Beaker, 30 mL, plastic, colorless80/pkgSM5010
Beaker, 100 mL, polypropylene1108042
Probe stand for standard Intellical probes18508850

10.1 Accessories

Description Quantity Item no.
Beaker, polypropylene, 50 mL, low form1108041
Disposable wipes, 11 x 22 cm280/pkg2097000
Wash bottle, polyethylene, 500 mL162011
Probe stand for standard Intellical probes18508850
Models: ISEF121, Fluoride Ion Selective Electrode

File Info : application/pdf, 18 Pages, 364.58KB

PDF preview unavailable. Download the PDF instead.

User Manual HQd ISEF121 Probe DOC0225380028 AH XSL Formatter V6.4 R1 for Windows (x64) : 6.4.2.26942 (2016/12/07 15:30JST) Antenna House PDF Output Library 6.4.928 (Windows (x64))

Related Documents

Preview Hach Intellical ISENO3181 Nitrate ISE Probe User Manual
User manual for the Hach Intellical ISENO3181 series digital combination ion-selective electrode (ISE) probes, detailing specifications, calibration, sample measurement, maintenance, and troubleshooting for nitrate analysis in water samples.
Preview HACH LBOD101 Intellical Luminescent Dissolved Oxygen Sensor User Manual
User manual for the HACH Intellical LBOD101 probe, a digital luminescent dissolved oxygen sensor designed for measuring dissolved oxygen concentration in BOD samples. This guide covers product overview, specifications, safety information, preparation, calibration procedures (100% air-saturated and zero-point), sample measurement, maintenance, troubleshooting, and consumables.
Preview Hach Ultra Low Range CL17sc Colorimetric Chlorine Analyser: Technical Overview
Comprehensive overview of the Hach Ultra Low Range CL17sc online colorimetric chlorine analyser. Details its applications in water quality monitoring, technical specifications, principle of operation, dimensions, and ordering information for industrial and utility use.
Preview Hach EZ3007 Fluoride Analyser: Specifications and Features
Detailed overview of the Hach EZ3007 Fluoride Analyser, highlighting its ISE technology for online fluoride measurements, key specifications, and included components for industrial and environmental applications.
Preview Hach SC4500 Digital Controller: Advanced Water Quality Monitoring
Explore the Hach SC4500 Digital Controller, an advanced solution for water quality monitoring. Learn about its features, technical specifications, applications in wastewater, drinking water, and industrial sectors, and its seamless integration capabilities with Hach's Claros system.
Preview HACH DR300 User Manual
Comprehensive user manual for the HACH DR300 portable filter photometer, detailing operation, testing procedures, calibration, maintenance, and troubleshooting for water quality analysis.
Preview HACH SL1000 Portable Parallel Analyzer: Basic User Manual
Get started with the HACH SL1000 Portable Parallel Analyzer. This basic user manual covers essential information on specifications, installation, operation, maintenance, and troubleshooting for accurate water quality analysis.
Preview HACH TL2360 Basic User Manual - Turbidity Measurement Guide
User manual for the HACH TL2360 laboratory turbidimeter. Learn about installation, operation, calibration, and maintenance for accurate water turbidity measurements.