Medtronic Communicator 4NR016

Getting Started Guide

CAUTION - Investigational device. Limited by Federal (or United States) law to investigational use.

Recharge Components

This document describes the recharge components for the Medtronic HAPTIX Implantable Neurostimulator Controller (INC). The Model 4NR016 Communicator is intended for use in conjunction with the MDT 97745 Recharge Controller, 97755 Recharger, and 97745AC adapter to recharge the HAPTIX 4NR015 INC.

Communicator Components and Accessories

Model 4NR016 Communicator Package:

Additional Recharge Components Supplied Separately by Medtronic:

Instructions for Use

Notes:

The Bluetooth®, Bluetooth Low Energy®, and BLE® trademarks and logos are registered trademarks owned by Bluetooth SIG, Inc. Any use of such marks by Medtronic is under license.

Storage and Operating Conditions

ParameterStorageOperating
Minimum Temperature-35 °C (-31 °F)10 °C (50 °F)
Maximum Temperature58 °C (136 °F)40 °C (104 °F)

Installing Batteries in the Communicator

  1. To open the battery compartment, slide the lid downward (as shown in Figure 1).
  2. Lift the battery compartment lid (as shown in Figure 2). The compartment is designed to hold batteries.
  3. Insert 2 new AAA alkaline batteries according to the orientation indicated on the inside of the battery compartment.
  4. Close the battery compartment lid.
  5. Secure the battery compartment lid by sliding it upward.
  6. Remove batteries when the device is not in use.

Turning the Communicator On and Off

The power switch, marked with a power symbol (⏻), is located on the side of the communicator (as shown in Figure 3).

To turn the communicator on:

Slide the power switch as illustrated and then release it. Confirm the Battery Icon is illuminated green. The communicator will start its communication activity. Further information on communicator activity during recharge is contained in the Communicator Technical Manual.

To turn the communicator off:

Slide the power switch, hold it for 2 seconds, then release it. Confirm the Battery Icon is not illuminated or flashing.

Connecting the 97745AC Power Supply Cable to the Communicator

The 97745AC Power Supply can be used to power the communicator if desired (e.g., when fresh batteries are unavailable).

  1. Connect the power supply cable to the communicator:
    • Insert the custom plug of the power supply cable into the custom port on the communicator (as shown in Figure 4).
  2. Plug the 97745AC Power Supply into a US main 60Hz 110-120VAC power source.
  3. Power the communicator on and off as directed above.

Disconnecting the 97745AC Power Supply Cable from the Communicator

The 97745AC Power Supply can be disconnected when external power is no longer required for the communicator (e.g., recharge session complete, desire to switch to battery power).

  1. Power the communicator off as described above.
  2. Unplug the 97745AC Power Supply from the communicator.

PDF preview unavailable. Download the PDF instead.

GetApplicationAttachment.html?id=4469273 Adobe PDF Library 19.10.96

Related Documents

Preview Medtronic Intellis 97715 Neurostimulator Implant Manual
Comprehensive implant manual for the Medtronic Intellis 97715 Rechargeable Neurostimulator, detailing device specifications, operating values, implantation procedures, and safety guidelines.
Preview Medtronic InterStim Systems MRI Guidelines
Comprehensive guidelines for performing Magnetic Resonance Imaging (MRI) scans on patients with Medtronic InterStim neurostimulation systems. This document details MR Conditional status, potential risks, eligibility determination, and procedural steps for both full-body and head-only MRI scans.
Preview Medtronic InterStim MRI Guidelines: Ensuring Patient Safety
Comprehensive guidelines for conducting MRI scans on patients with Medtronic InterStim neurostimulation systems. Covers eligibility, equipment, scan conditions, and safety precautions for both full-body and head-only scans.
Preview Medtronic Sacral Neuromodulation Systems: Product Catalog and Technical Overview
Comprehensive overview of Medtronic's InterStim and Verify systems for sacral neuromodulation, including neurostimulators, programmers, leads, test electrodes, cables, and accessories. Details technical specifications and key benefits for managing pelvic floor dysfunction.
Preview Medtronic Micra AV Transcatheter Pacing System Specifications and Information
Comprehensive product specifications, indications, contraindications, warnings, precautions, and potential complications for the Medtronic Micra AV Transcatheter Pacing System (MC1AVR1). Includes details on physical characteristics, battery, pacing parameters, MRI safety, and delivery system.
Preview Medtronic Azure XT DR MRI SureScan W1DR01 Device Manual
Comprehensive guide for the Medtronic Azure XT DR MRI SureScan Model W1DR01 dual chamber implantable pulse generator (IPG), detailing system overview, indications, contraindications, warnings, implant procedures, device parameters, and technical specifications.
Preview Medtronic Azure S DR MRI SureScan W3DR01 Device Manual
Comprehensive device manual for the Medtronic Azure S DR MRI SureScan W3DR01, an implantable pulse generator (IPG) featuring MRI SureScan technology and Bluetooth wireless telemetry, detailing system overview, warnings, implant procedures, and product specifications for healthcare professionals.
Preview LINQ II™ Insertable Cardiac Monitor (LNQ22) - Medtronic Product Overview and Safety Information
Comprehensive overview of the Medtronic LINQ II™ Insertable Cardiac Monitor (Model LNQ22), detailing its technical and battery specifications, indications for use, contraindications, warnings, precautions, potential adverse events, and contact information.