9825 Installation & User Manual

26 ago 2024 — The 9825 digital indicator uses panel installation, which requires the thickness of the cabinet's front panel to be no more than 4mm. Before installation ...


File Info : application/pdf, 23 Pages, 3.33MB

PDF preview unavailable. Download the PDF instead.

9825-Installation-User-Manual-Rev-1.7
Installation & User Manual
9825
9825 Installation & User Manual

Info Icons
Note: "Note" means essential information that will help you use the device more effectively. Caution: "Caution" means this may cause damage to your device or data loss if you do not follow the instructions. Warning: "Warning" means potential danger. Example: property damage, personal Injury or even death.
Pre Installation Warnings
Warning: This device must be installed and connected by a professional electrical staff with the power supply disconnected for safe and dependable operation. Warning: This device cannot be used in an unsafe enviornment. Example: Where explosion protection is required.

Table of Contents

Unpacking & Installation

1

Unpacking

1

Storage & Installation

1

Connections

2

Power Connections

2

Load Cell Connections

2

Serial I/O Device Connections

4

USB Port Connections

4

Analog Output Connections

4

Relay Input/Output Control Connections

5

Basic Operation

6

Power On

6

Display Details

6

Keypad Details

7

System Configuration

8

Menu Tree

8

Menu Descriptions

8

Calibration Procedures

10

Industrial Interfaces

12

USB Interface Communication

12

Analog Output Interface

12

I/O Control Applications

13

Setpoint Application

13

Alarm Application

13

Indicator Info

14

Software Version

14

Appendix

15

Appendix 1: Command Output Format 1 - Continuous Mode (ASCII)

15

Appendix 2: Command Output Format 1 - Demand Mode (ASCII)

16

Appendix 3: Condec Format Output (Condec)

17

Appendix 4: FAST MODE

18

Specifications

19

Warranty

20

Unpacking & Installation

Unpacking:

Please follow these inspection procedures after unpacking the product:

- Check the product to ensure that there was no damage in transit.

- Check the following list and confirm that all the items are in the carton:

- 9825 Digital Indicator

- 9825 Installation & User Manual

- External Connecting Terminals - Product Qualification Certificate

- Clamping Strips & Anchor Nuts

- 9825 External Power Supply

- 9825 Grounding Cable Assembly

Storage & Installation:

The 9825 indicator must be stored in a dry, dust-free environment before use. Storage temperature is -20°C to +65°C (-4°F to +149°F), working environment temperature is -10°C to +104°F (+14°F to +104°F), relative humidity no more than 95% (Non-Condensing).
The 9825 digital indicator uses panel installation, which requires the thickness of the cabinet's front panel to be no more than 4mm. Before installation, remove the two mounting screws from the indicator's clamping strips, then remove the clamping strips. Push the indicator into the opening on the cabinet, then reinsert the clamping strips. Gently tignten the two mounting screws.
Indicator structure and physical dimensions (mm)

Interface Inc. www.interfaceforce.com

1

9825 Installation & User Manual 15-386

Connections:
Power Connections:
The 9825 has an input range of 9VDC to 36VDC. The maximum electrical power consumption of the 9825 is 6W (8W Peak).The unit is shipped with an external 24VDC linear power supply and a grounding cable assembly. The GND Terminal should be routed to the grounding lug on the rear of the 9825 housing and then to earth using the provided grounding cable assembly in order to optimize signal stability.
Use the screw down terminals to secure the power supply leads and grounding cable to the 3 position connector in the following configuration:
Pin Assignment: 1 = VDC + 2 = VDC 3 = GND
Warning:
Verify that the power supply connections are correct before powering on.
Note:
Make sure that the power cord does not pose a potential obstacle or tripping hazard. Only use approved accessories and peripherals.
Load Cell Connections:
The 9825 indicator uses a 6-wire load cell signal connection. This indicator provides a 4.5 volt DC excitation voltage to the load cell(s). The voltage difference between +SIG and -SIG is about 0 ~ 9mV when connected to a load cell with a 2mV/V output, and about 0 ~ 13.5mV when connected to a load cell with a 3mV/V output. The 9825 indicator can drive up to six (6) 350-ohm load cells (or the equivalent resistance of all load cells connected in parallel is higher than 87).
If the application requires the 9825 to be connected to multiple load cells, please use a junction box.
Note:
This product does not contain a junction box. If a junction box is necessary for your application, we recommend Interface Model JB104SS as an approved accessory.
The load cell cable requires a shield which must be properly grounded to ensure maximum stability. High quality cable is recommended. Be sure to route the load cell cable away from high voltage / power cables. The maximum length allowed for the load cell or junction box cable is shown in the following table:

Interface Inc. www.interfaceforce.com

2

9825 Installation & User Manual 15-386

Sensor Input Terminal Pin Assignment: Four-Wire Analog (Load Cell) or (Junction Box) Connection: Six-Wire Analog (Load Cell) or (Junction Box) Connection:

Interface Inc. www.interfaceforce.com

3

9825 Installation & User Manual 15-386

Serial I/O Device Connections:
The 9825 indicator comes standard with one USB port.
USB Port Connections:
The 9825 indicator comes standard with a MINI-USB port that can be connected to a PC. This USB port is designed for data communication and firmware upgrades.
Analog Output Connections:
Use the JP1 pin header on the internal analog option board to configure the analog output for current output (4-20mA, 0-24mA) or voltage output (0-10V, 0-5V). Please note that voltage and current outputs cannot be used at the same time. We suggest using a PLC or PC to monitor the analog output calibration.
Configure the voltage or current output as follows. Output type is selected in the Analog Out Setup menu, under the Output Type submenu.
Voltage output: Select either 0-5V or 0-10V. Use the Analog + & Analog - terminals.
Current output: Select either 0-24mA or 4-20mA. Use the Analog + & Analog - terminals.

Interface Inc. www.interfaceforce.com

4

9825 Installation & User Manual 15-386

Relay Input/Output Control Connections: The Output Connections: The 9825 optional I/O control port is relay-based and can be used with an AC or DC
power supply. The DC power supply range is 24VDC to 100VDC. The AC power supply range is up to 220VAC.
The COM terminal can be connected to the positive or negative of the power supply. The maximum power output of each relay is 90W / 5A.
Output control interfaces and load connection diagram:
Output control interfaces and PLC connection diagram:
The Input Connections: The input interfaces are isolated, passive inputs. The interfaces can be connected to many control control keys (Buttons), and the wiring is as follows.

Interface Inc. www.interfaceforce.com

5

9825 Installation & User Manual 15-386

Basic Operation

Power On:
The display will show the Interface logo followed by the Device Mode and Firmware Version. After that, the current force value will be displayed.

Display Details:

The 9825 utilizes a 128 x 32 dot OLED display with adjustable LED back-light. The table below summarizes the display annunciators.

Basic Force Display:

Fast Analog Enabled - Display Locked:

Force Display with Tare Function Active:

Force Display with Tare and Relay Function Active:

Interface Inc. www.interfaceforce.com

6

9825 Installation & User Manual 15-386

Keypad Details:

Keypad Functions:
Tare (Exit,  ):
- When in display mode (Tare Function) - Pressing this key sets force value to zero (sets tare). - If tare is already set, pressing this key removes tare.
- When in the setup menu (Exit Function) - Return to a previous menu.
- Increase the value when used as directional key (  ).
- Hold to exit setup menu.
PK/Val (  ):
- When in display mode (PK/Val Function) - Cycle among real time, peak and valley display modes.
- When in the setup menu (  Function)
- Enter sub-menu.
- Decrease a value when used as a directional key (  ).
Reset (  ):
- When in display mode (Reset Function) - Resets peak and valley values.
- When in setup menu (  Function)
- Moves left when used as directional key. - Used to toggle FAST ANALOG mode. Menu (Enter):
- When in display mode (Menu Function) - Hold this key until buzzer sounds to enter the setup menu.
- When in setup menu (Enter Function) - Saves current setting.

Interface Inc. www.interfaceforce.com

7

9825 Installation & User Manual 15-386

System Configuration
Menu Tree:

Warning: Do not access the Advanced Menu unless instructed to do so by a qualified technician.

Menu Description:

Menu Data Capture

Sub-Menu Sampling Rate
FIR Filter (Finite Impulse Response)

SMA Filter (Simple Moving Average)
Select Unit Type

Units

Cal Basic Unit Nominal Capacity Disp Resolution

Description Number of Samples per second.
Reduces the influence of nearby electrical or mechanical noise sources.
Smooths signal by averaging samples over a given span.
Force, Torque, Electric, Distance, NONE.

Default 30Hz
On
1
Force Torque Electric Distance

Select engineering units displayed
Sets display output range
Set decimal placement and increments

Force 100,000 1:100,000

Options 30, 40, 50, 60, 75, 80, 100, 120, 150, 170, 200, 240, 300, 400, 600, 1200Hz Off, On
Integer values from 1 to 100
LB, MT, KLB, ozf, KN, N, t, g, KG oz-in, kg.m, kg.cm, kg.mm, N.m, cN.m, mN.m, lb-ft, lb-in mV/V, V in, mm, defines what units are avalable in "select unit type" Integer values from 1 to 100,000
Menu options are based on Nominal Capacity value

Interface Inc. www.interfaceforce.com

8

9825 Installation & User Manual 15-386

Calibration Live or Key-In Set Pos Span Set Neg Span Set Zero Point Cal Stability
USB Setup Baud Rate
Bit / Parity Port Mode Protocol
Analog Out Output Type Scale Output

Set calibration type Set span from zero to positive capacity Set span from zero to negative capacity Set zero A greater value can produce higher accuracy calibration points but will require a more stable mV/V input signal during calibration as well. Serial communication rate in bits per second Set Binary format and check bit Set port mode Set Protocol (see protocol descriptions in the appendix) Set Analog output type

Live
1
9600 8-bit None Demand Condec 0-10V

Fine Tune

Relay

IO Application Input Point n (Setpoint) Input Hysteresis n (Setpoint) Input High Point (Alarm) Input Low Point (Alarm) Custom In-1

None 4000 200 5000 3000 None

System

Version / Date
Unique ID Power-On Tare Sys Reset
Advanced Menu

Display firmware version and date Display Unique ID
Reset to default settings. Password Required

Disable

Live, Key-In
Press  and enter to start
sequence Fixture
Press  to start sequence
Integer values from 0 to 320 represent the number of samples averaged when capturing a calibration point. Larger values = greater stability required
2400, 4800, 9600, 19200, 38400, 57600,115200
8-bit None, 8-bit Even, 7-bit Even, 7-bit Odd Demand, Continuous Condec, ASCII
4-20mA, 0-10V, 0-5V, 0-24mA
Adjust Low and High points using keypad Adjust 0%, 50%, 100% output points using keypad None, Setpoint, Alarm Adjust values using keypad
None, Reset Key, Tare Key, Print Key
Press  to view
Off, On
Press  to execute
Enter Password 336699 to access the Advanced Menu

Interface Inc. www.interfaceforce.com

9

9825 Installation & User Manual 15-386

Calibration Procedures:
Calibration Overview:
The 9825 Indicator can be calibrated by using either a Live calibration method or a Key-In calibration method. It is important to set the Nominal Capacity Value before beginning calibration.
Live Calibration:
The Live calibration method produces the best possible system accuracy. This method requires one of the following:
- The load cell being paired with the 9825 Indicator will be connected to the instrument while a series of nominal force loads are applied to calibrate the instrument.
- A load simulator will be connected to the 9825 Indicator while a series of simulated mV/V loads are applied to calibrate the instrument.
Live calibration is accomplished by setting the Positive Span, Negative Span and Zero. To execute a Live calibration, follow the steps below:
1. Press and hold the Menu button to enter the setup menu. A beep will sound as the setup menu is activated.
2. Using the  (Menu) button, scroll until Calibration is displayed on the screen. Press the  (Pk/Val) button to enter the calibration sub-menus.
3. Using the  (Menu) button, scroll until Set Pos (or Neg) Span is displayed on the screen. Press the  (Pk/Val) button to initiate the calibration process.
4. The term Fixture will show on the screen. At this point the load cell should be set into its fixturing, but with no additional calibration loads applied. If a simulator is being used for the Live calibration, connect the simulator, but set its value to 0mV/V. Press the Menu (Enter) button to save this point.
5. After the Fixture value has been set, the term C1 (calibration point #1) will appear on the screen. The user should set the numerical field so that it displays the nominal force load that is about to be applied. Once this value has been input and the applied force load has stablized, pressing the Menu (Enter) button will capture this point.
6. The C2 term will then appear. If the user would like to add another calibration point (up to six are possible) they can repeat the actions in Step 5. If the user would like to end the calibration, they should leave the numerical field as 0 and press the Menu (Enter) button.
Note: if the calibration was unsuccessful, an error message will appear:
- "Err2" : There is not enough signal from the load cell. This is most commonly caused by incorrect wiring or a damaged load cell.
Repeat this process in the opposing polarity, than proceed to Zero calibration.

Interface Inc. www.interfaceforce.com

10

9825 Installation & User Manual 15-386

Zero Calibration:
1. Press and hold the Menu button to enter the setup menu. A beep will sound as the setup menu is activated.
2. Using the  (Menu) button, scroll until Calibration is displayed on the screen. Press the  (Pk/Val) button to enter the calibration sub-menus.
3. Using the  (Menu) button, scroll until Set Zero Point is displayed on the screen.
4. At this point, the Zero Calibration is ready to begin. Be sure the load cell is connected and in an unloaded state. If using a simulator, be sure that the simulator is set to 0mV/V. Press the  (Pk/Val) button to initiate the Zero Calibration. Dashed lines will be displayed at the bottom right of the screen to indicate that the 9825 is capturing the zero point.
Key-In Calibration:
The Key-In calibration method is typically only used in emergency cases when the indicator cannot receive a Live calibration. The Key-In method uses a single point to establish the span of the load cell. It ignores load cell nonlinearity and any asymmetry between the opposing loading modes.
To execute a Key-In calibration, follow the steps below:
1. Press and hold the Menu button to enter the setup menu. A beep will sound as the setup menu is activated.
2. Using the  (Menu) button, scroll until Calibration is displayed on the screen. Press the  (Pk/Val) button to enter the calibration sub-menus.
3. The Live or Key-In sub-menu is the first Calibration sub-menu and should be displayed on the screen. Press the  (Reset) button to change the flashing value from Live to Key-In. Press the Menu (Enter) button to save this setting.
4. Press the  (Menu) button to change the sub-menu to Rated Output. Press the  (Pk/Val) button to enter the Rated Output sub-menu.
5. Enter the sensitivity of the load cell into the numerical field. This is typically the mV/V output of the load cell at its rated capacity. Press the Menu (Enter) button to save this value.
6. Press the  (Menu) button to change the sub-menu to Sensor Capacity. Press the  (Pk/Val) button to enter the Sensor Capacity sub-menu.
7. Enter the rated capacity of the load cell into the numerical field. Press the Menu (Enter) button to save this value.
8. Press the  (Menu) button to change the sub-menu to Set Zero Point. User should preform a Zero Calibration as detailed above.

Interface Inc. www.interfaceforce.com

11

9825 Installation & User Manual 15-386

Industrial Interfaces
USB Interface Communication:
The 9825 indicator can connect to a PC via USB cable. First a USB driver must be installed on the PC to access the 9825. Measurement data can be accessed by using a terminal emulation application such as HyperTerminal. The USB Port output has Two fixed strings: ASCII and Condec.

Analog Output Interface:

Analog Output Calibration:
The mode of the Analog Output can be selected from its Output Type sub-menu. There are four modes of analog output: 4-20mA, 0-24mA, 0-5V and 0-10V. Please refer to the wiring section for the proper jumper setting of the optional analog output board. To calibrate the Analog Output, follow the steps below:
Scale Output:
1. While within the Analog Output menu, scroll to Scale Output and press the  (Pk/Val) button to initiate the Scale Output sequence.
2. Scale Output is set by inputting a Low and High force value. To set a given value, use the numerical field on the screen to input the desired force. The first character can be used to switch the sign convention from + to - and back. Press the  (Pk/Val) button to save the setting.
Fine Tune:
Before performing this part of the device setup, the Analog Output of the 9825 should be connected to whatever instrument will be accepting and measuring the analog signal.
1. In the Analog Output setup menu, scroll to Fine Tune and press the  (Pk/Val) button to initiate the Fine Tune sequence.
2. The screen will display "0%", which indicates the lowest point of the analog scale. For voltage outputs, this is 0VDC. For current outputs this is either 0mA (0-24mA) or 4mA (4-20mA).
3. By adjusting the numerical value on the screen, the Analog output will be Fine Tuned. The digit furthest to the left creates the greatest change in output, while the digit furthest to the right creates the smallest change in output. Adjust this number until the measured value on the connected meter or PLC shows the minimum point on the analog scale. Press Menu (Enter) to save this value and proceed.
4. Repeat this process for the 50% point. For a 0-5V setting the output will be 2.5V. For a 4-20mA setting the output will be 12mA and so on.
5. Repeat this process for the 100% point.
Notes: - The analog output mode setup to 4mA-20mA: If the load 0kg, the voltage output is 0. If the load is the full range of the scale, than the voltage output is 24mA.
- The analog output mode setup to 0-10V: If the load 0kg, the voltage output is 0. If the load is the full range of the scale, than the voltage output is 10.8V.

Interface Inc. www.interfaceforce.com

12

9825 Installation & User Manual 15-386

I/O Control Applications:
SetPoint Application:
The following conditions should occur when you use SetPoint application:
1. When the load is less than the value of "Input Point 1": - The symbol will show on the display. - The OUT-1 relay will close.
Otherwise, the symbol will show on the display and the OUT-1 relay will open.
2. When the load is less than the value of "Input Point 2", But greater than the value of "Input Point1": - The symbol will show on the display. - The OUT-2 relay will close. Otherwise, the symbol will show on the display and the OUT-2 relay will open.
3. When the load is less than the value of "Input Point 3", but greater than the value of "Input Point2": - The symbol will show on the display. - The OUT-3 relay will close. Otherwise, the symbol will show on the display and the OUT-3 relay will open.
4. When the load is less than the value of "Input Point 4", but greater than the value of "Input Point 3": - The symbol will show on the display. - The OUT-4 relay will close. Otherwise, the symbol will show on the display and the OUT-4 relay will open.
Alarm Application:
The loads of the four configurable Alarm points must follow this formula:
Input ExtraHigh > Input HighPoint > Input LowPoint > Input ExtraLow
1. When the load is less than the value of "Input ExtraHigh": - The symbol will show on the display - The alarm will sound - The OUT-1 relay will close - The display will send a warning message
Otherwise, the symbol will show on the display and the OUT-1 relay will open.
2. When the load is less than the value of "Input ExtraHigh", but greater than the value of "Input HighPoint": - The symbol will show on the display - The alarm will sound - The OUT-2 relay will close - The display will send a warning message Otherwise, the symbol will show on the display and the OUT-2 relay will open.

Interface Inc. www.interfaceforce.com

13

9825 Installation & User Manual 15-386

3. When the load is less than the value of "Input LowPoint", but greater than the value of "Input ExtraLow": - The symbol will show on the display - The alarm will sound - The OUT-3 relay will close - The display will send a warning message Otherwise, the symbol will show on the display and the OUT-3 relay will open.
4. When the load is less than the value of "Input ExtraLow": - The symbol will show on the display - The alarm will sound - The OUT-4 relay will close - The display will send a warning message
Otherwise, the symbol will show on the display and the OUT-4 relay will open.
Indicator Info
Software Version:
This Information can be accessed from the main menu under Menu_System_Version/Date.
- Software Version:
- Last Update:

Interface Inc. www.interfaceforce.com

14

9825 Installation & User Manual 15-386

Appendix
Appendix 1: Command Output Format 1 - Continuous Mode (ASCII)
In this mode of communication, the indicator transmits the data frame continuously. The load value in the frame is expressed in ASCII.

Interface Inc. www.interfaceforce.com

15

9825 Installation & User Manual 15-386

Appendix 2: Command Output Format 1 - Demand Mode (ASCII) This host device (PC) will output demand command through the serial ports when the scale is in the normal loading status. The demand command format is shown below:
The serial output data format is as follows:

Interface Inc. www.interfaceforce.com

16

9825 Installation & User Manual 15-386

Appendix 3: Condec Format Output (Condec) Condec Demand Output:
Demand Commands: "P" > Print "T" > Tare "Z" > Zero "G" > Gross "N" > Net Condec Continuous Output:

Note: The starting address 40001 of MODBUS is not suitable for SIEMENS soft.

Interface Inc. www.interfaceforce.com

17

9825 Installation & User Manual 15-386

APPENDIX 4: FAST MODE The 9825 features a fast analog to digital converter. However, the OLED display update rate limits the effective bandwidth unless display is set to FAST MODE.
· When active, the display updates for 5 times per second and these update takes 20 ms. · During the 20 ms updates, the analog output freezes at current value. Also, the peak/valley
does not update during that time. · To allow faster analog update and peak valley response, the FAST MODE should be enabled. EXAMPLE DATA TRACE WITH FAST MODE = OFF. NOTE 20 ms flat spots in data occur every 200 ms.
EXAMPLE DATA TRACE WITH FAST MODE = ON Note smooth response with no flat spots.
18

Specifications

EXCITATION

Excitation Voltage ­ VDC

4.5

Current ­ mA

100

PERFORMANCE

Maximum Display Counts Internal Resolution Counts Signal Input Range ­ mV/V Sensitivity ­ V/count Readings Per Second - MAX

±999,999 1,000,000
±4.5 0.03 1000

Latency Filter Settings

Variable up to 20ms (Affects Analog out & Peak/Valley) Off, Static, Dynamic FIR and/or Moving Average

Serial Interfaces

USB 2.0 standard

Operating Temperature

ENVIRONMENTAL
°C °F

-10 to +45 +14 to 113

Relative Humidity ­ %

at °C

MAX

at °F

10% to 90%, non-condensing 10% to 90%, non-condensing

POWER

Supply

VDC

24 VDC with supplied 120V 60Hz, AC/ DC adapter or 9-36 VDC external supply

Power Consumption

W

Switching frequency of internal PSU

Provides isolation

6 RMS, 8 Peak 300kHz 6kV

Dimensions - W x H x D Weight

MECHANICAL
mm in g lbs

106 x 66 x 150 4.17 x 2.6 x 5.91
68 1.5

Display ­ mm(in) Panel Cutout - W x H

128 x 32 OLED dot matrix display. Font size is 9.5 (0.37) H and 6.5 (0.26) W

mm

91 x 46

in

3.58 x 1.81

FAST ANALOG OUTPUT ­ kHz

VDC 0-5, 0-10, 2.5+/-2.5, 5+/-5 mA 4-20, 0-24, 12+/-8, 12 +/-12

Interface Inc. www.interfaceforce.com

19

9825 Installation & User Manual 15-386

Warranty
Warranty
All indicator products from Interface Inc., ('Interface') are warranted against defective material and workmanship for a period of (1) one year from the date of dispatch. If the 'Interface' product you purchase appears to have a defect in material or workmanship or fails during normal use within the period, please contact your Distributor, who will assist you in resolving the problem. If it is necessary to return the product to 'Interface' please include a note stating name, company, address, phone number and a detailed description of the problem. Also, please indicate if it is a warranty repair. The sender is responsible for shipping charges, freight insurance and proper packaging to prevent breakage in transit. 'Interface' warranty does not apply to defects resulting from action of the buyer such as mishandling, improper interfacing, operation outside of design limits, improper repair or unauthorised modification. No other warranties are expressed or implied. 'Interface' specifically disclaims any implied warranties of merchantability or fitness for a specific purpose. The remedies outlined above are the buyer's only remedies. 'Interface' will not be liable for direct, indirect, special, incidental or consequential damages whether based on the contract, tort or other legal theory. Any corrective maintenance required after the warranty period should be performed by 'Interface' approved personnel only
www.interfaceforce.com

Interface Inc. www.interfaceforce.com

20

9825 Installation & User Manual 15-386

