Glucose analysis comparability evaluation of the YSI 2900C Biochemistry Analyzer to the legacy 2300 STAT Plus™

White Paper

Abstract

Background

The YSI 2300 STAT Plus™ (YSI 2300) D-glucose (glucose) and Lactate Analyzer is a Class II in-vitro diagnostics (IVD) medical device, widely accepted for reference measurements and system calibration by most manufacturers of blood glucose (BG) monitoring systems. YSI has developed a new generation analyzer, the YSI 2900C Biochemistry Analyzer (YSI 2900C), employing the same biosensor technology as the YSI 2300, but with updated usability features and form factor. A comparative study was performed to evaluate the accuracy and precision of the new YSI 2900C platform for whole blood (WB), plasma, and serum.

Introduction

The YSI 2300 is a Class II IVD medical device designed for clinical diagnostics and sports physiology applications. Today, the YSI 2300 analyzer is widely accepted as a method for reference measurements and system calibration by most manufacturers of BG monitoring systems.

The YSI 2900D is a laboratory instrument intended for use in research, biotechnology and food-processing applications. The YSI 2900C is an In Vitro Diagnostic Device that is a direct replacement for the YSI 2300. The aim of this study was to demonstrate equivalence of the YSI 2300 and YSI 2900C analyzers concerning glucose analytical performance using human WB, plasma, and serum samples.

Materials and Methods

YSI 2300 and YSI 2900C analyzers

Both YSI 2300 and YSI 2900C analyzers (YSI, Inc., Yellow Springs, OH, USA) employ an enzyme-based, amperometric biosensor for measuring glucose concentrations. The biosensor uses a glucose oxidase-containing membrane for oxidizing glucose to gluconolactone and hydrogen peroxide. The hydrogen peroxide is in turn oxidized at the platinum anode, producing electrons. The electron flow (current) is linearly proportional to the steady state hydrogen peroxide concentration and, therefore, to the concentration of glucose in the sample. Analysis time is approximately 60 seconds/sample.

Output from each YSI 2300 probe was aligned with a paired probe on a YSI 2900C of the same membrane lot to create the data pairs.

Whole Blood (WB), plasma and serum analysis

After three instrument pairs were qualified by daily checks, donor subject blood draws initiated. Subjects had been previously recruited and the specimens collected under external IRB constraints and oversight to protect their rights, safety, and well-being. The agency providing this ethical oversight was Advarra IRB, Columbia, MD. All specimens were collected, handled, and stored according to the CLSI guidelines GP41, 7th Edition, “Collection of Diagnostic Venous Blood Specimens”. All samples were immediately stored in ice water or refrigerated at 4°C when not in direct use. All testing occurred within four hours of sample collection to minimize preanalytical variability caused by glycolysis during storage. Each WB, PL, or serum sample was tested on the analyzers as close in time as possible. During venous WB testing the test matrix mode selection was set to “Whole Blood” in the instrument set-up control panel. During testing of plasma and serum the mode selection was set to “Plasma”. Instrument calibration was performed immediately prior to sampling. Each analyzer has two probes and both probes have glucose sensing membranes. One calibrator lot was used for all instruments and samples in this report. At the conclusion of the study all bloodrelated samples were disposed of by an authorized medical biohazard waste service.

Results and discussion

Daily membrane & linearity checks

All instruments utilized for the study were required to pass daily performance checks prior to analyzing samples. For acceptable glucose oxidase membrane integrity, a Ferrocyanide (FCN) test value of ≤ 5 mg/dL apparent glucose is required. FCN values for all membranes were well below 5 mg/dL, indicating structurally intact membranes. The acceptable range for the 900 mg/dL glucose linearity standard is 855 - 945 mg/dL, which is ± 5% of the linearity check point. All instruments met the linearity specifications for each day of testing.

Data

The data presented here is a glucose assay from whole blood. Similar data was collected for plasma, serum, and fingerstick capillary whole blood. These were assayed in both the YSI 2300 and the YSI 2900C. In both simulated and actual data sets, the x-axis is always considered the reference assay (output from the predicate device YSI 2300) and the y-axis is always considered the blood glucose monitoring system (output from the YSI 2900C).

Plot 1. Data pairs over the whole range of D-glucose values tested in whole blood.

A (1/X²) Weighted Deming fit was used as the regression model. The weighting was used to compensate for leveraged bias due to the increasing variance (heteroscedasticity) over the range of the data. The shaded area around the regression line shows the 95% confidence interval (CI) around the slope and intercept of the model. The bootstrap technique was used to calculate the CI. Data are from 111 unique clinical samples.

Data pairs are (YSI 2300, YSI 2900C) in mg/dL Glucose.

Experimental Data-Whole Blood

Plots 2 & 3. Error grid analyses. Upper left – Parkes* (Consensus) EGA. Upper right – Surveillance EGA**. In the PEG all points are within risk zone A. Bottom – Table of data per risk grade in SEG. The Parkes EGA plot is shown because it covers the range of measured data whereas the SEG is bounded at 600 mg/dL. All Parkes EGAs shown are for Type 1 Diabetes.

*Parkes, Joan L., Pardo, Scott, Slatin, Stephen L., and Ginsberg, Barry H., "A New Consensus Error Grid to Evaluate the Clinical Significance of Inaccuracies in the Measurement of Blood Glucose", Diabetes Care, Vol. 23, No. 8, pp. 1143-1148, August 2000.

**Klonoff, David C. et al., “The Surveillance Error Grid”, Journal of Diabetes Science and Technology 2014, Vol. 8(4), 658-672.

Summary and Conclusion - Comparability

Summary

This study was done to gain a comparative performance description between the predicate instrument, YSI 2300 Biochemistry Analyzer and the YSI 2900C Biochemistry Analyzer. Data pairings were plotted, and the comparison described by a Deming weighted regression model, allowing for error on both axes. More than 99% of all the data pairs compared were within zone A of clinical risk for all bloodrelated matrices tested, for the whole range of glucose concentrations.

Conclusion

The device, the YSI 2900C Biochemistry Analyzer, performance was equivalent or better when compared to the predicate device, the YSI 2300 Biochemistry Analyzer for all four sample matrices tested. In conclusion, this study demonstrates the YSI 2900C Biochemistry Analyzer's ability to be an acceptable replacement for the YSI 2300 Biochemistry Analyzer as an In Vitro Diagnostic Device.

PDF preview unavailable. Download the PDF instead.

Glucose-Analysis-Comparability-Evaluation-YSI-2900C-Biochemistry-Analyzer-White-Paper ?srsltid=AfmBOopd2 bUhIHOCAZvhyHGOVYD2e Adobe PDF Library 17.0 Adobe InDesign 20.4 (Macintosh)

Related Documents

Preview YSI Model 32 Conductance Meter Instruction Manual
Instruction manual for the YSI Model 32 Conductance Meter, detailing its general description, specifications, operating procedures, maintenance, and calibration.
Preview YSI Certificate of Analysis for Bioanalytical Standard 2367
Certificate of Analysis from YSI Incorporated for the Bioanalytical Standard, Model 2367, detailing the testing results for GLUCOSE, including determination, tolerance, and actual results, verified by QA.
Preview YSI 1531C Bioanalytical Standard Certificate of Analysis
Certificate of Analysis for YSI 1531C Bioanalytical Standard, detailing test results for GLUCOSE, including average production vs. metrology values and coefficient of variation, verified by QA.
Preview YSI 2801 Certificate of Analysis for Bioanalytical Standard
Certificate of Analysis for YSI Model 2801 Bioanalytical Standard, detailing test results for GLUCOSE, including average production, metrology, tolerance, and coefficient of variation. Verified by QA and completed by Amanda Fenner on 7/23/2025.
Preview YSI Bioanalytical Standard Certificate of Analysis - Model 2776
Certificate of Analysis for YSI Bioanalytical Standard, Model 2776 (Lot 25G100401), detailing test results for Glucose and Lactate with specified tolerances.
Preview YSI 1531 Bioanalytical Standard Certificate of Analysis - Glucose Test Results
Certificate of Analysis from YSI Incorporated detailing the test results, determination, and tolerance for the 1531 Bioanalytical Standard, specifically for Glucose.
Preview YSI 2900 Biochemistry Analyzer: Measuring Glucose and Sucrose in Potatoes - Application Note 218LS
This application note details the use of the YSI 2900 Series Biochemistry Analyzer for the rapid and accurate measurement of glucose and sucrose concentrations in potatoes, utilizing enzyme technology. It covers materials, setup, methodology, calculations, and ordering information.
Preview Certificate of Analysis - YSI Bioanalytical Standard 2772
YSI Incorporated Certificate of Analysis for Bioanalytical Standard, Model 2772. Details lot number 25G100268 and testing results for CHOLINE, including determination, tolerance, and result.