MEAN WELL GSM120A Series: 120W AC-DC Reliable Green Medical Adaptor
Features
- 3 pole AC inlet IEC320-C14, Class I power unit
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Extremely low leakage current
- No load power consumption <0.15W
- Energy efficiency level VI and meet CoC Version 5
- -30~+70°C wide range working temperature
- Protections: Short circuit / Overload / Over voltage / Over temperature
- LED indicator for power on ?
- Lifetime > 95 K hours
- Various DC plug quick adapter accessory available (Plug kit sold separately, please refer to: https://www.meanwell.com/upload/pdf/DC_plug.pdf)
- 3 years warranty
Applications
- Mobile clinical workstation
- Oral irrigator
- Portable hemodialysis machine
- Breath Machine
- Medical computer monitor
Description
The GSM120A series is a highly reliable, 120W desktop style single-output green medical adaptor. It features a 3-pin (with FG) standard IEC320-C14 power plug and accepts an input range from 80VAC to 264VAC. The series offers various output voltages from 12VDC to 48VDC, catering to diverse medical electrical devices. Its circuitry adheres to international medical standards (2*MOPP) with an ultra-low leakage current (<100μA), making it suitable for medical devices in direct electrical contact with patients. With up to 91.5% efficiency and less than 0.15W no-load power consumption, the GSM120A complies with energy standards like USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP, and CoC Version 5. It utilizes a 94V-0 flame retardant plastic case and has obtained international medical safety certifications.
Model Encoding
The model number is structured as GSM120A [Output Voltage]-[DC Plug Type]. For example, GSM120A12-R7B indicates a 12V output with the R7B DC plug type. R7B signifies a standard model with a Power DIN 4Pin with lock type. Other DC plug options are available upon customer request.
Technical Specifications
Parameter | Specification |
---|---|
Order No. | GSM120A12-R7B, GSM120A15-R7B, GSM120A20-R7B, GSM120A24-R7B, GSM120A48-R7B |
Safety Model No. | GSM120A12, GSM120A15, GSM120A20, GSM120A24, GSM120A48 |
DC Voltage | 12V, 15V, 20V, 24V, 48V |
Rated Current | 8.5A, 7A, 6A, 5A, 2.5A |
Rated Power (max.) | 102W, 105W, 120W |
Voltage Tolerance (max.) | ±5.0%, ±3.0%, ±2.5% |
Line Regulation | ±1.0% |
Load Regulation | ±5.0%, ±4.0%, ±3.0%, ±2.5% |
Setup, Rise Time | 1500ms, 30ms / 230VAC; 2000ms, 30ms/115VAC at full load |
Hold Up Time (Typ.) | 40ms/230VAC; 24ms/115VAC at full load |
Voltage Range | 80~264VAC |
Frequency Range | 47~63Hz |
Power Factor (Typ.) | PF>0.93/230VAC; PF>0.97/115VAC at full load |
Efficiency (Typ.) | 88% to 91.5% |
AC Current (Typ.) | 1.4A/115VAC; 0.7A/230VAC |
Inrush Current (Typ.) | 35A/115VAC; 70A/230VAC (Cold start) |
Leakage Current (max.) | Earth leakage < 115 μA/264VAC, Touch current <100 µA/264VAC |
Overload Protection | 105~160% rated output power; Hiccup mode, recovers automatically |
Over Voltage Protection | 105~135% rated output voltage; Shut down o/p voltage, re-power on to recover |
Over Temperature Protection | Shut down o/p voltage, re-power on to recover |
Working Temp. | -30~+70°C |
Working Humidity | 20% ~ 90% RH non-condensing |
Storage Temp., Humidity | -40~+85°C, 10~95% RH non-condensing |
Temp. Coefficient | ±0.03% / °C (0~40°C) |
Vibration | 10~500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes |
Operating Altitude | 3000 meters |
Safety Standards | IEC 60601-1, BS EN/EN 60601-1, ANSI/AAMI ES60601-1, CAN/CSA C22.2 No. 60601-1, EAC TP TC 004 |
Isolation Level | Primary-Secondary: 2xMOPP, Primary-Earth: 1xMOPP |
Withstand Voltage | I/P-O/P:4KVAC, I/P-FG:2KVAC, O/P-FG:0.5KVAC |
Isolation Resistance | >100M Ohms / 500VDC / 25°C/70% RH |
EMC Emission | Conducted emission: BS EN/EN55011 (CISPR11), FCC PART 15/CISPR22, CAN ICES-3(B)/NMB-3(B) Class B; Radiated emission: BS EN/EN55011 (CISPR11), FCC PART 15/CISPR22, CAN ICES-3(B)/NMB-3(B) Class B |
EMC Immunity | ESD: BS EN/EN61000-4-2 Level 4; RF field susceptibility: BS EN/EN61000-4-3 Level 3; EFT bursts: BS EN/EN61000-4-4 Level 3; Surge susceptibility: BS EN/EN61000-4-5 Level 3; Conducted susceptibility: BS EN/EN61000-4-6 Level 3; Magnetic field immunity: BS EN/EN61000-4-8 Level 4; Voltage dip, interruption: BS EN/EN61000-4-11 |
MTBF | 2413.9K hrs min. (Telcordia SR-332); 388.5K hrs min. (MIL-HDBK-217F) |
Dimension | 167*67*35mm (L*W*H) |
Packing | 0.6Kg; 20pcs/13.0Kg/0.89CUFT |
Connector/Plug | Various DC plug options available (see page 4-5) |
Derating Curve
A derating curve is provided showing the relationship between ambient temperature and load percentage. The unit can operate at 100% load up to 50°C, after which the load capacity decreases linearly to 50% at 70°C.
Static Characteristics
A graph illustrates the static characteristics, showing the load percentage versus input voltage. The adaptor maintains full load capacity across the input voltage range of 90VAC to 264VAC.
Mechanical Specification
The unit dimensions are 167mm (Length) x 67mm (Width) x 35mm (Height), with a tolerance of ±1mm. It features an IEC320-C14 AC inlet and a standard R7B DC output plug (4-pin DIN with lock). The cable length for 12V models is 1000±50mm (UL2464 16AWGx4C), and for 15-48V models is 1200±50mm (UL2464 18AWGx4C).
DC Output Plug Options
MEAN WELL offers customizable DC plug options and quick adapter accessories. Various connector types are available, including Tuning Fork Style (P1J, P1M), DIN 4 Pin with Lock (R7BF), DIN 5 Pin (R1B), Neutrik XLR (MIC4), Molex (C6P), AMP (C4P), and stripped and tinned leads. Detailed pin assignments and output configurations are provided for each type. More information can be found at: https://www.meanwell.com/upload/pdf/DC_plug.pdf
Installation Manual
For installation guidance, please refer to: http://www.meanwell.com/manual.html