HORIBA Advanced Techno
HF/HCI Concentration Monitor HF-960H
Ideal for measuring high concentrations of hydrofluoric acid, hydrochloric acid, etc.
Product Overview
The HF-960H accurately measures concentrations of hydrofluoric and hydrochloric acid used in semiconductor manufacturing and other processes. Measurement is based on conversion to concentration from four-electrode conductivity measurements, enabling a wide range of measurements from low to high concentrations. The dedicated FES-510 series sensor is made from materials highly resistant to chemicals, making this unit ideal for managing hydrofluoric and hydrochloric acid, even in high concentrations.
Features a four-electrode carbon sensor with outstanding resistance to chemicals.
Image Description: A textual representation of the HORIBA HF-960H Indication Converter, a compact black electronic device with a digital display showing '10.00' and '25.0', and control buttons labeled MEAS, CAL, MNT, A, V, ENT. It is identified as an 'Indication Converter HF-960H'.
Image Description: A textual representation of the HORIBA FES-510 series four-electrode carbon sensor, a flow-type sensor with a black body and a cable. It is labeled 'Sensor FES-510-3/4' and '4-Electrode Carbon Sensor FES-510'.
Key Features
-
Wide range compatibility
Automatic decimal point adjustment enables a wide range of measurement for HF and HCl concentrations, from 0.000 to 10.00%.
-
Automatic display range switching
The display range automatically switches from high to low concentrations. This unit also allows the user to monitor changes while maintaining effective conductivity resolution.
-
Enables four types of measurement
HF concentration, HCl concentration, conductivity, and temperature. HF and HCl concentrations are measured while compensating for temperatures based on relevant characteristic data. This unit can also measure the conductivity of other chemicals.
-
Chemical resistant four-electrode carbon sensor
The sensor is a new FES-510 series four-electrode flow-type sensor made from glass carbon, which is highly resistant to chemicals.
-
4-channel transmission output
Equipped with four output channels, which can be allocated to HF concentration, HCl concentration, conductivity, and temperature. Choose from two types of current output: 4 to 20 mA DC, or 0 to 20 mA DC.
-
24 V DC power supply
-
RoHS compliant
The HF-960H is also an environmentally-friendly product that uses lead-free solder for mounting chips on the PCB.
Specifications (Indication Converter)
Model | HF-960H |
---|---|
Measurement method | 4-electrode carbon flow-type conductivity measurement - concentration conversion |
Sensor input | 1-channel |
Temperature sensor specifications | Platinum resistance 1000Ω/0°C |
Measuring range |
HF concentration: 0 to 10% HCl concentration: 0 to 10% Conductivity: 0 to 1000 mS/cm (without temperature compensation) Temperature: 0 to 100°C (Decimal places selectable: 1 or 2) |
Repeatability |
HF/HCl concentration: Within ±0.5% of full scale (in equivalent input) Conductivity: Within ±1.0% of full scale (in equivalent input) |
Linearity |
HF/HCl concentration: Within ±0.5% of full scale (in equivalent input) Conductivity: Within ±1.0% of full scale (in equivalent input) |
Transmission output |
No. of outputs: 4 Type: 4 to 20 mA DC or 0 to 20 mA DC (I/O insulation type) Maximum load resistance: 900Ω Output range: Custom setting possible within measurement range. (Negative terminals connected internally) |
Contact output |
No. of outputs: 5 (Alarm outputs: R1, R2, R3, R4, Error alarm RF) Contact type: Relay contact, SPST (R1-R4), SPDT (RF) Contact rating: 240 V AC, 1A and 30 V DC, 1A (resistance load) Contact function: Selectable (upper/lower limit operation, ON/OFF control, delay operation for R1-R4; error, trouble warning, maintenance for RF) |
Contact input |
No. of inputs: 1 Contact status: Open collector no-voltage contact Contact function: Transmission output hold |
Communications output | RS-485 input/output |
Calibration function |
Conductivity: Based on specified compensation coefficient for cell constant Temperature: Calibrated by comparison with reference thermometer |
Transmission output Hold function | Selectable (previous value hold, optional value hold; previous value hold only during maintenance mode) |
Self-diagnosis function | Sensor diagnosis (short-circuit, disconnection of temperature sensor), Converter error |
Temperature compensation |
HF and HCl specific temperature compensation Input of optional temperature compensation coefficient (Reference temperature: 25°C, Coefficient: 0 to 3%/°C) Temperature compensation range: HF: 0 to 80°C, HCl: 0 to 80°C, Conductivity: 0 to 100°C. (Out-of-range compensation is extended.) |
Ambient environment | Temperature: -5 to 45°C, Relative humidity: 20 to 85% (without condensation) |
Power supply | 24 V DC ±10%, 10 W (max.) |
Protective structure | Panel: IP65, Rear case: IP20, Terminal: IP00 (indoor-use panel installation type) |
Mass | Approx. 550g |
Conforming standards | CE marking, FCC Part 15 |
Compatible sensors | FES-510 series (flow-type carbon sensor) |
Specifications (4-electrode carbon flow-type conductivity sensor)
Model | FES-510 -1/4 | FES-510 -3/8 | FES-510 -1/2 | FES-510 -3/4 | FES-510 -1 | FES-510 -6_4 | FES-510 -10_8 |
---|---|---|---|---|---|---|---|
Liquid end material | Glass carbon (electrode), PFA (body), Kalrez® (seal) | ||||||
Piping size | 1/4 inch | 3/8 inch | 1/2 inch | 3/4 inch | 1 inch | Ø6/Ø4 mm | Ø10/Ø8 mm (Made by order) |
Sample flow rate | 0 to 2 L/min | 0 to 8 L/min | 0 to 10 L/min | 0 to 15 L/min | 0 to 25 L/min | 0 to 2 L/min | 0 to 8 L/min |
Sample pressure range | 0 to 0.5 MPa for 5 to 50°C, 0 to 0.1 MPa for above 100°C. For 50 to 100°C, pressure range is plotted between 0.5 MPa and 0.1 MPa as shown on the line graph. | ||||||
Sample temperature | 5 to 100°C (Aqueous solution which boils at 100°C) | ||||||
Cable length | 10 m (standard) |
Dimensions and Diagrams
Indication Converter Dimensions
Image Description: Diagram showing the external dimensions of the HORIBA HF-960H Indication Converter, with measurements in millimeters (mm) and inches (in). It includes a 'Panel Cut Size' diagram.
Sensor Dimensions
Image Description: Diagrams illustrating the dimensions of the HORIBA FES-510 series four-electrode carbon sensors, with measurements in millimeters (mm). Two variants, FES-510-3/4 and FES-510-1/2, are shown with detailed views of their construction and connection points.
Pressure vs. Temperature Graph
Diagram Description: A line graph titled 'FES-510 Pressure resistance type' plotting 'Pressure range MPa' on the Y-axis against 'Liquid temperature °C' on the X-axis. The graph shows how pressure resistance changes with liquid temperature, with specific data points and ranges indicated.
Contact Information
Please read the operation manual before using this product to assure safe and proper handling of the product.
The contents of this catalog are subject to change without prior notice, and without any subsequent liability to this company. It is strictly forbidden to copy the content of this catalog in part or in full.
Website: http://www.horiba-adt.jp/index_e.htm
HORIBA Advanced Techno, Co., Ltd.
-
Head Office
31 Miyanonishicho, Kisshoin, Minami-ku, Kyoto, Japan
Phone: 81-75-321-7184
Fax: 81-75-321-7291 -
Tokyo Sales Office
Arute-Bldg. HigashiKanda, 4th Fl, 1-7-8 Higashi-Kanda, Chiyoda-ku, Tokyo, Japan
Phone: 81-3-3851-3150
Fax: 81-3-3851-3140
Other HORIBA Offices:
- HORIBA KOREA Ltd.
112-6 Sogong-Dong, Choong-ku, Seoul, Korea
Phone: 82-2-753-7911
Fax: 82-2-756-4972 - HORIBA INSTRUMENTS Pte. Ltd. SINGAPORE
10 Ubi Crescent, #05-11/12 Ubi Techpark, Singapore 408564
Phone: 65-745-8300
Fax: 65-745-8155 - HORIBA / STEC INCORPORATED
Santa Clara Head Office: 3265 Scott Boulevard, Santa Clara, CA 95054, U.S.A.
Phone: 1-408-730-4772
Fax: 1-408-730-8975
Austine Office: 9701 Dessau Road, Suite 605, Austin, TX 78754, U.S.A.
Phone: 1-512-836-9560
Fax: 1-512-836-8054 - HORIBA TRADING (SHANGHAI) Co., Ltd.
Shanghai Office: Room 1701, United Plaza, 1468 Nanjing Rd. West, Shanghai 200040, China
Phone: 21-6289-6060
Fax: 21-6289-5553
Beijing Office: Room 1801, Capital Tower, Beijing, Tower 1 No. 6Jin, Jianguomenwai Ave., Chaoyang District, Beijing, 100022 China
Phone: 10-8567-9966
Fax: 10-8567-9066 - HORIBA INSTRUMENTS LIMITED
Kyoto Close, Summerhouse Road, Moulton Park, Northampton NN3 6FL, England
Phone: 44-1604-542600
Fax: 44-1604-542696
e-mail: hil.semicon@horiba.co.jp - HORIBA EUROPE GmbH
Head Office: Hans-Mess-Str.6, D-61440 Oberursel/Ts. Germany
Phone: 49-6172-1396-0
Fax: 49-6172-137385 - HORIBA Ltd. Taiwan Representative Office
3F NO.18 Lane 676, Chung Hua Rd, Chupei City, Hsinchu Hsien, 302, Taiwan
Phone: 886-3-656-1012
Fax: 886-3-656-8231 - HORIBA FRANCE
Rue L. et A. Lumiere, Technoparc F-01630 St-Genis-Pouilly France
Phone: 33-4-50-42-27-63
Fax: 33-4-50-42-07-74
Bulletin: HAE-T0152B
[Recycled Paper] Printed in Japan
0903SK13