Triplett PCAL300 Loop Calibrator (Source and Measure)

User Manual

Manufacturer: Triplett

Model: PCAL300

Contact: Test Equipment Depot - 800.517.8431 - 5 Commonwealth Ave, MA 01801 - TestEquipmentDepot.com

1 Safety Information

Users should use the calibrator in accordance with the instructions in the Manual, or the protective measures provided by the calibrator may be damaged. The Company is not responsible for any damage caused by failure to follow the safety warning information provided.

⚠️ WARNING

indicates a situation or action that may pose a danger to the user. ? Caution indicates a situation or action that may cause damage to the calibrator or the equipment being tested. Please refer to Table 1 for an explanation of international electrical symbols used in the calibrator and the Manual.

Table 1. International Electrical Symbols

Symbol Description
[Ground Symbol] Earthing
[Warning Symbol] Warning Message

⚠️ WARNING

To avoid electric shock or personal injury:

2 Get to Know the Calibrator

This section provides an overview of the calibrator's components and interface.

2.1 Input and Output Terminals

Figure 2 shows the input and output terminals of the calibrator. Table 2 explains their purpose.

Table 2. Input/output terminals

Terminal Function Description
1 LOOOP terminal: 24V loop power external terminal
2 V terminal: DCV measurement (+) input terminal
3 mA terminal: DCI measurement (+) input terminal
4 Common (-) (return) terminal for all inputs
5 mA- terminal: DCI output (-) terminal
6 mA+ terminal: DCI output (+) terminal

5.2 Buttons

Figure 3 shows the calibrator buttons. Table 3 explains their functions.

(Diagram of buttons would be here if images were allowed)

Table 3. Button Functions

SN Button name Description
1 Power button Power on/off
2 MEASURE/SOURCE Input/output state switching
3 Output waveform switch button Output waveform selection of current output automatic waveform
4 Output START button Start/stop button for automatic waveform output in the output current function
5, 7 Output setting button Output setting position increase/decrease
6, 8 Output setting button Output setting position left/right shift
9 CONFIG button In the output current function, press this button to enter current-related parameter settings
10 ZERO button In the output state, press this button to restore the output value to the default value. Press this button to save the settings in states of factory maintenance setting and parameter setting.
11 FUNC button Press this button to switch functions
12 Backlight/flashlight button Short press to turn on/off the backlight; long press to turn on/off the flashlight

5.3 Display screen

Figure 4 shows a typical display screen. The following elements are indicated:

(Diagram of display screen would be here if images were allowed)

6 Preparations

Operating precautions

Safe use of the calibrator

Malfunctioning Conditions

If the instrument starts to release smoke, emits a strange smell, or shows other abnormal phenomena, immediately turn off the instrument and remove the batteries. Then contact the seller from whom you purchased the instrument.

General operations

Environmental requirements

Use the instrument under the environmental requirements listed below:

Ambient Temperature and Humidity
Do not use the instrument in the following environment

Note:

Install or replace batteries

⚠️ WARNING

? Caution

(Diagram of battery replacement would be here if images were allowed)

Step 1: Before replacing the battery, remove the test lead and turn off the calibrator.

Step 2: Use a slotted screwdriver to rotate the battery door screw by a quarter turn counterclockwise and remove the battery door.

Step 3: Correctly install 3 AA LR6 alkaline batteries into the battery compartment in the direction as it indicates.

Step 4: After replacing the batteries, close the battery door tightly again.

Power Features

Power on/off

Press the Power button to turn on the calibrator when the power is off; press the Power button for 2 seconds to turn off the calibrator when it is on.

Automatic shutdown

The calibrator automatically shuts down when there is no button operation within the factory default of 5 minutes. The automatic shutdown time can be set in the factory settings. Please refer to Chapter 9 "Factory Settings".

Turn the backlight on/off

Press the Backlight button to turn the backlight on, and press it again to turn the backlight off. This makes it easier to see the content on the display screen in dark places or when performing outputs or measurements. Turning on the backlight will reduce the battery life when the calibrator is operating on batteries.

Note: The backlight automatically turns off after a default of approximately 60 seconds. Press the Backlight button to re-illuminate backlight. The backlight illumination time can be set in the factory settings. Please refer to Chapter 9 "Factory Settings".

7 Use the Output Mode

DC signal can be outputted by using the calibrator.

⚠️ Warning

To avoid the electric shock, the nominal voltage that exceeds the value marked on the calibrator should not be exerted between the terminals of the calibrator or between any terminal and ground. The calibrator shall be used in the occasion that the voltage of any terminal over the ground does not exceed 30V peak.

7.1 Output DC current (active)

Step 1: Connect leads to the target equipment

(Diagram of DC current active output would be here if images were allowed)

Step 2: Press the MEASURE/SOURCE key and switch the state into the output state. The characters of ‘SOURCE’, ‘LOOP’, ‘mA’ on display screen will be lighted. By this time, it is the active DC current output function.

Step 3: Setting of manual stepping span and current output range:

Note: When the measuring range is 0-20mA, 25% span means 5mA; when the measuring range is 4-20mA, 25% span means 4mA. When the measuring range is 0-20mA, 100% span means 20mA; when the measuring range is 4-20mA, 100% span means 16mA.

Note: After setting the current output range, the set range applies to all current output functions.

Step 4: Set the output value by pressing the output setting key.

7.1.1 Output current auto-stepping output mode

Step 1: [Stepping Symbol] symbol of the display screen will be lighted at this time by pressing the M key in the state of output current function and switching the function to DC step mode.

Step 2: The character 'STEP' at the lower right corner of the display screen indicates the step setting and the main display area in the display screen displays the parameters to be set in the entry of the DC step mode parameter setting interface by pressing the CONFIG key;

Use the /// key to set step length required(1-200S); press ZERO key to keep and set. Then press CONFIG key to exit the setting interface.

Step 3: Press START key to start the automatic step output current, and by this time (RUN) signal on the display screen is lighted.

Step 4: Press START key to stop the automatic step output current when the user is going to end the automatic step output current. By this time, (RUN) sign on the display screen disappears.

7.1.2 Output current automatic sawtooth wave output mode

Step 1: Press the M key and switch the function to DC automatic sawtooth wave mode in the state of output current function. By this time, [Sawtooth Symbol] symbol on display screen is lighted.

Step 2: Press CONFIG key to enter DC sawtooth wave mode parameter setting interface. By this time, the character 'START' at the lower right corner of the display screen indicates the origin current value setting and the parameters to be set in main display area on display screen; The user should utilize / / / key to set the origin current that is required; The user should press ZERO key to keep and set and switch to the next setting item. By this time, the character 'STOP' at the lower right corner of display screen indicates the terminal current value setting and the parameters to be set in the main display area on the display screen; The user should utilize /// key to set the terminal current required; the user should press ZERO key to keep and set and switch to the next setting item and by this time, the character 'CYC' at the lower right corner of display screen indicates the period setting and the parameters to be set in the main display area on display screen; the user should utilize /// key to set the period needed (5-200S) and press ZERO key to keep and set. And then, the user should press CONFIG key to exit the setting interface.

Step 3: Press START key again to start the automatic step output current. By this time, (RUN) sign on the display screen is lighted.

Step 4: Press START key to stop the automatic step output current when the user is going to end the automatic step output current. By this time, (RUN) sign on the display screen disappears.

7.1.3 Output current automatic triangular wave output mode

Step 1: Press the M key and switch the function to DC automatic sawtooth wave mode in the state of output current function. By this time, [Triangular Symbol] symbol on display screen is lighted.

Step 2: Press CONFIG key to enter DC sawtooth wave mode parameter setting interface. By this time, the character 'START' at the lower right corner of display screen indicates origin current value setting and the parameters to be set in main display area on display screen; The user should utilize / / / key to set the origin current that is required; The user should press ZERO key to keep and set and switch to the next setting item. By this time, the character 'STOP' at the lower right corner of display screen indicates the terminal current value setting and the parameters to be set in main display area on display screen; The user should utilize /// key to set the terminal current required; The user should press ZERO key to keep and set and switch to the next setting item and by this time, the character 'CYC' at the lower right corner of display screen indicates the period setting and the parameters to be set in main display area on display screen; The user should utilize /// key to set the period needed (5-200S) and press ZERO key to keep and set. And then, the user should press CONFIG key to exit setting interface.

Step 3: Press START key to start the automatic step output current. By this time, (RUN) sign on display screen is lighted.

Step 4: Press START key to stop automatic step output current when the user is going to end the automatic step output current. By this time, (RUN) sign on display screen disappears.

7.2 Output DC current (passive)

Step 1: Connect the lead to the target equipment

(Diagram of DC current passive output would be here if images were allowed)

Step 2: Press MEASURE/SOURCE key and switch the state into the output state, and the 'SOURCE', 'LOOP', 'mA' characters on display screen will be lighted. By this time, it is the APC DC output function.

Step 3: Press FUNC key and switch the function to passive DC current output. The character ‘SOURCE','mA'on display screen are lighted.

Other operations should be found in the part of active DC function.

Tips: Passive DC current needs an external 5-28V DC power supply

8 Use Measurement Mode

DC and voltage and DC current can be measured by utilizing the calibrator

⚠️ Warning

The instrument might be damaged or staff might be injured if the line is wrongly operated or wrong operation is conducted during the process of measurement. Therefore, you should pay as much attention as possible when you are performing measurement operation.

Prompt: The "OL" symbol is displayed in the main display area of the display screen when the measured value exceeds the measuring range of this range.

8.1 Measure DC voltage

Step 1: Ensure that the measurement lead is disconnected from the equipment under test.

Step 2: Press the MEASURE/SOURCE button to switch the state to measurement, and the display screen (MEASURE) will light up at this moment.

Step 3: Press the FUNC button to switch to the DC voltage function (the default function of the measurement state is DCV), and the V character of the display screen is lighted at this moment.

Step 4: Connect the measurement lead to the signal end of the equipment under test, and the real-time measurement value is displayed in the main display area of the display screen at this moment.

(Diagram for measuring DC voltage would be here if images were allowed)

8.2 Measure DC current

Step 1: Ensure that the measurement lead is disconnected from the equipment under test.

Step 2: Press the MEASURE/SOURCE button to switch the state to measurement, and the display screen (MEASURE) will light up at this moment.

Step 3: Press the FUNC button to switch to the DC current function, and the mA character of the display screen is lighted at this moment.

Step 4: Connect the measurement lead to the signal end of the equipment under test, and the real-time measurement value is displayed in the main display area of the display screen at this moment.

(Diagram for measuring DC current would be here if images were allowed)

8.3. Measure current with external 24V supply

Step 1: Ensure that the measurement lead is disconnected from the equipment under test.

Step 2: Press the MEASURE/SOURCE button to switch the state to measurement, and the display screen (MEASURE) will light up at this moment.

Step 3: Press the FUNC button to switch to the DC current function, and the mA and the mA and LOOP characters of the display screen are lighted at the same time.

Step 4: Connect the measurement lead to the measurement end of the equipment under test, and the real-time measurement value is displayed in the main display area of display screen at this moment.

Prompt: This function provides 24V power for the external equipment loop and measures the current value in the loop.

(Diagram for measuring current with external 24V supply would be here if images were allowed)

9 Factory Settings

The default factory settings can be changed by the calibrator.

Method of entering: Hold the backlight button, and then press the power button to start, and release the backlight button after the instrument enters the setting interface.

9.1 Automatic shutdown time setting

Step 1: "APOF" is displayed in the display screen, which indicates the automatic shutdown setting after entering the setting interface.

Step 2: Use the setting buttons such as /// to set necessary parameters. The displayed value unit of the automatic shutdown time is minute.

Setting range: 0-60 minutes; 0 indicates canceling the automatic shutdown, and other values indicate shutting down the instrument after the corresponding time.

Step 3: Press ZERO button and save the setting when there is the sign of "SAVE" in the display screen.

9.2 Backlight time setting

Step 1: Press the MEASURE/SOURCE and the "BLOF" is displayed, indicating the backlight time setting.

Step 2: Use // / to set necessary parameters. The displayed value unit of the backlight time is second.

Setting range: 0-3600 seconds; 0 indicates that the automatic backlight turning off is cancelled. Other values indicate that the instrument backlight is turned off after the corresponding time.

Step 3: Press the ZERO button and save the setting when there is the sign of "SAVE" in the display screen.

9.3 Flashlight time setting

Step 1: Press the MEASURE/SOURCE and the "LTOF" is displayed, indicating the flashlight time setting.

Step 2: Use // / to set necessary parameters. The displayed value unit of the flashlight time is minute.

Setting range: 0-30 minutes; 0 indicates that the automatic backlight turning off is cancelled. Other values indicate that the instrument backlight is turned off after the corresponding time.

Step 3: Press the ZERO button and save the setting when there is the sign of "SAVE" in the display screen.

9.4 Default factory settings

Step 1: Press the MEASURE/SOURCE and the "FACT" is displayed indicating the factory default setting.

Step 2: Use / to set necessary parameters;

NO indicates that all settings are not restored to factory defaults. YES indicates that all settings are restored to factory defaults.

Step 3: Press the ZERO button and save the setting when there is the sign of "SAVE" in the display screen.

All setting factory parameters are as follows:

Prompt: ZERO button needs to be pressed to save the setting as long as the setting of any item is changed. Only the latest setting value is saved when ZERO button is pressed randomly.

10 Replace Battery or Fuse

⚠️ Warning

The test lead wire must be removed from the calibrator before opening the battery door in order to avoid electric shock. The battery door must be closed tightly before the calibrator is utilized.

Be cautious

(Diagram of battery and fuse replacement would be here if images were allowed)

Step 1: Remove the test wire and charger and turn off the calibrator, before batteries or fuses are to be replaced.

Step 2: Use a flat-head screwdriver to turn the battery door screw one quarter of a circle counterclockwise and remove the battery door.

Step 3: Install 3 identical AA LR6 batteries correctly into the battery compartment in line with the directions indicated by the battery compartment. Or replace the same type of fuse (100mA/250V).

Step 4: Close the battery door again and lock the screw latch after batteries are replaced.

11 Maintenance

11.1 Cleaning the calibrator

Warning

The specified replacement parts should be used and water is not allowed to enter the shell for the purpose of avoiding personal injury or damage to calibrator.

Be cautious

Solvents or abrasive cleaners are not allowed in order to avoid damaging plastic lenses and shells. Clean the calibrator with a soft cloth dipped in water or mild soapy water.

11.2 Calibration or reparation service center

Calibration, reparation or maintenance of the instrument should only be carried out by experienced service staff. If the calibrator is not functioning properly, batteries should be checked first and they should be replaced if necessary. It should be guaranteed that the calibrator is operated in line with the instructions in this manual. Please return the calibrator back with a fault description if the calibrator does not work properly. Please securely package the calibrator and send it to your nearest service center (please pay postage and insurance) if the original packing box is still kept. Our company is not liable for damage in transit. Calibrators covered by our warranty can be repaired or replaced quickly (at our discretion) and returned free of charge. Please refer to the warranty terms of this manual. There is certain charge for the repair of the calibrator if the warranty period has expired. Please contact our authorized service center to inquire about reparation and fees if the calibrator is not covered by the warranty. Please refer to the "Contact Us" section in the previous section of the Manual if you want to find an authorized service center.

12 Index

Input measurement function

[Used within one year after calibration, 23 °C± 5 °C, 20-70% RH, accuracy within the range of ± (%setting value + character)]

Measurement function Measurement range Measurement scope Resolution Accuracy Remarks
DCV 30V -30.000V~30.000V 0.001V 0.02%+2mV Input resistance: about 1MΩ
DCI 30mA -30.000mA~30.000mA 0.001mA 0.02%+4uA Shunt resistance: about 10Ω
Input resistance: about 20Ω
LOOP 24 V 10%

Other characteristics:

Analog output function

[Used within one year after calibration, 23 °C± 5 °C, 20-70% RH, accuracy within the range of ± (%setting value + character)]

Output function Measurement range Output scope Resolution Accuracy Remarks
DCI 30mA 0.000mA~30.000mA 0.001mA 0.05%+4uA In the case of 20 mA, maximum load is 1000Ωresistance,
When the transmitter is simulated, the external loop provides electricity within the range of 5~28V

Other characteristics:

Models: PCAL300 Loop Process Calibrator, PCAL300, Loop Process Calibrator, Process Calibrator, Calibrator

File Info : application/pdf, 25 Pages, 1.38MB

PDF preview unavailable. Download the PDF instead.

Triplett pcal300 manual f808 Microsoft Word for Microsoft 365

Related Documents

Preview Triplett VCT1000 Voltage and Current Tester User Manual
User manual for the Triplett VCT1000 Voltage and Current Tester. This guide details features, safety, operation for AC/DC voltage, current, phase rotation, continuity, resistance, and NCV testing, along with specifications and warranty.
Preview Triplett CM1000 1000A TRMS AC Clamp Meter User Manual
Comprehensive user manual for the Triplett CM1000 1000A True RMS AC Clamp Meter, covering features, operation, safety, specifications, and maintenance for accurate electrical measurements.
Preview Triplett VDL48 DC Voltage Datalogger User Manual
Comprehensive user manual for the Triplett VDL48 DC Voltage Datalogger, detailing its features, operation, LCD indicators, specifications, firmware updates, and warranty information.
Preview Triplett CTT42 Coating Thickness Tester User Manual
Comprehensive user manual for the Triplett CTT42 Coating Thickness Tester, detailing its features, operation, calibration, and specifications for measuring coating thickness on ferrous and non-ferrous substrates using magnetic induction and eddy current technologies.
Preview Triplett UTG300 Ultrasonic Thickness Gauge User Manual | Accurate Material Measurement
Explore the Triplett UTG300 Ultrasonic Thickness Gauge. This comprehensive user manual provides detailed instructions on operation, specifications, calibration, and maintenance for precise material thickness measurements.
Preview Triplett RHT415 Hygro-Thermometer with Remote Probe User Manual
User manual for the Triplett RHT415 Hygro-Thermometer with Remote Probe, detailing its features, operation, specifications, and warranty. This device measures temperature and humidity with high precision.
Preview Tritech Test Equipment LLC (Triplett) ISO 9001:2015 Certification by NQA
Certificate of Registration for Tritech Test Equipment LLC, DBA Triplett, confirming their Quality Management System meets ISO 9001:2015 standards, certified by NQA.
Preview Triplett GSM120 Carbon Monoxide Meter User Manual
User manual for the Triplett GSM120 Carbon Monoxide Meter, detailing its operation, safety guidelines, specifications, and warranty information provided by Triplett and Test Equipment Depot.